Acessibilidade / Reportar erro

Accuracy of Transthoracic Echocardiogram as a Screening Method in the Clinical Practice of Pulmonary Hypertension Investigation

Abstract

Background

The transthoracic echocardiogram (TTE) plays a screening role in the diagnostic algorithm of pulmonary hypertension (PH). Studies have shown a significant disagreement between TTE measurements of the systolic pulmonary artery pressure (sPAP) and right atrial pressure (RAP) and those obtained by right heart catheterization (RHC).

Objective

To compare TTE measurements of sPAP and RAP with those obtained by RHC in patients being investigated for PH.

Methods

Patients referred to a PH reference center with a high or intermediate TTE probability of PH upon admission were submitted to RHC. The agreement between sPAP and RAP from both procedures was assessed through the Bland-Altman test. Differences of up to 10 mmHg for sPAP and 5 mmHg for RAP were considered within the variability of the test. Receiver Operating Characteristic (ROC) curve was constructed to determine the most accurate sPAP and Tricuspid regurgitation maximal velocity (TRV)values associated with the diagnosis of PH by RHC. The adopted level of statistical significance was 5%.

Results

Ninety-five patients were included. The Bland-Altman analysis showed a bias of 8.03 mmHg (95% CI:-34.9-50.9) for sPAP and -3.30 mmHg (95% CI:-15.9-9.3) for RAP. AUC for sPAP and TRV measured by TTE for discrimination of probable PH were 0.936 (95% CI: 0.836-1.0) and 0.919 (95% CI: 0.837-1.0), respectively. However, only 33.4% of the echocardiographic estimate of sPAP and 55.1% of RAP were accurate, as compared to the measurements obtained by RHC.

Conclusion

TTE has a high discriminatory power as a screening diagnostic method for PH despite presenting disagreements between sPAP and RAP absolute values when compared to RHC measurements.

Echocardiography; Hypertension, Pulmonary; Data Accuracy

Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br