Compression or distraction forces applied on a pedicular fixation system: an experimental study

An experimental study was conducted with the objective of evaluating the effects of compression and distraction (approach and deviation) forces applied on implants placed at vertebral pedicles. The vertebral segments T8-T9, T10-T11, T12-L1, L2-L3 and L4-L5 of Landsark pigs were used in the study. Each vertebral segment was fixed with pedicular implants, with compression or distraction forces being applied with a displacement that ranged from 5.1 mm to 9.35 mm. Displacement caused by forces application was measured in four points of the vertebral segment: A (anterior portion of the vertebral segment), B (medial portion of the vertebral segment), C (posterior portion of the vertebral segment) and D (superior joint process). Compression forces caused the deviation of points located at regions A and B, and the approach of points at region D. Distraction forces caused the approach of points at regions A and B and deviation of the points at region D.

Spine fusion; Spine fixation