Study on implant stability in cementless total knee arthroplasty

Abstracts

Objetives: Determine the stability of tibial and femoral components of 20 cementless knee arthroplasties with rotating platform. METHODS: The 20 patients (20 knees) underwent an analysis of dynamic radiographs with an image amplifier and maneuvers of varus and valgus which were compared to static frontal and lateral radiographs of the knees and analyzed by two experienced surgeons in a double-blind way. RESULTS: We could observe in this study that both methods showed very similar results for the stability of the tibial and femoral components (p<0,001) using the Kappa method for comparison. CONCLUSION: The tibial component was more unstable in relation to the femoral component in both static and dynamic studies. Level of evidence IV, Case Series.

Knee arthroplasty; Knee prosthesis; Knee osteoarthritis; Biomechanics


OBJETIVO: Comparar dois métodos de avaliação da estabilidade dos componentes tibial e femoral nas artroplastias de joelho não cimentadas com plataforma rotatória. MÉTODOS: Para isso foram avaliados 20 pacientes (20 joelhos) através de uma análise de radiografias dinâmicas com intensificador de imagem e manobras de estresse em varo e valgo, que foram comparadas com radiografias estáticas em frente e perfil dos joelhos, analisadas por dois cirurgiões experientes, cegos um em relação ao outro. RESULTADOS: Os resultados das análises estáticas e dinâmicas foram comparados e demonstraram forte correlação estatística (p<0,001), utilizando-se o método Kappa de comparação. CONCLUSÃO: O componente tibial mostrou-se mais instável quando comparado com o componente femoral, tanto na análise estática, quanto na dinâmica. Nível de evidência IV, Série de Casos.

Artroplastia do joelho; Prótese do joelho; Osteoartrite do joelho; Biomecânica


ORIGINAL ARTICLE

Study on implant stability in cementless total knee arthroplasty

Luís Eduardo Passarelli Tírico; Thiago Pasqualin; José Otávio Pécora; Riccardo Gomes Gobbi; José Ricardo Pécora; Marco Kawamura Demange

Institute of Orthopedics and Traumatology of Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo

Mailling Address

ABSTRACT

OBJETIVES: Determine the stability of tibial and femoral components of 20 cementless knee arthroplasties with rotating platform.

METHODS: The 20 patients (20 knees) underwent an analysis of dynamic radiographs with an image amplifier and maneuvers of varus and valgus which were compared to static frontal and lateral radiographs of the knees and analyzed by two experienced surgeons in a double-blind way.

RESULTS: We could observe in this study that both methods showed very similar results for the stability of the tibial and femoral components (p<0.001) using the Kappa method for comparison.

CONCLUSION: The tibial component was more unstable in relation to the femoral component in both static and dynamic studies. Level of Evidence IV, Case Series.

Keywords: Knee arthroplasty. Knee prosthesis. Knee osteoarthritis. Biomechanics.

INTRODUCTION

Total knee arthroplasty (TKA) is one of the more effective, reliable procedures with reproducible results in orthopedic surgery. It presents exceptional results in pain relief and functional restoration in patients with inflammatory arthritis and osteoarthrosis, with low rates of aseptic loosening at 10 years.1-3

Nowadays, the gold standard in knee arthroplasty is the use of metallic components cemented into the femur and tibia.4-9 However, countless surveys have been carried out in an attempt for us to increase the quality of component fixation in the bone and to reduce the long-term loosening rate.

One of the causes of early loosening of TKA components is the fixed tibial baseplate, where the polyethylene is rigidly secured to the component, disallowing movement. This rigid fixation leads to greater friction between and among the arthroplasty components, producing greater wear and tear on the polyethylene and the release of debris, which contributes to the aseptic loosening of the implant. A tibial component with rotating platform was created in an attempt to minimize this friction on the implant-bone interface, where the polyethylene can move rotationally on the tibial component, thus reducing the friction.10-12

Another alteration introduced over the years was the substitution of cemented component fixation by cementless arthroplasty. Cementless TKA was initially idealized for young individuals, in the hope of reducing bone loss during the reviews that would take place in the future, and of reducing the surgical time and the systemic complications of cementation. For this purpose implants were created with rough surfaces of hydroxyapatite, which would theoretically stimulate bone growth and implant osteointegration, increasing the duration of the effects of surgery.13-15

Cementless knee arthroplasties have been in use for more than 20 years with good outcomes in our field.16 However, articles are still controversial in relation to the results of this technique, when compared with cemented arthroplasties.11,12,17-20 The main problem pointed out by critics of the cementless technique is early tibial loosening. The main parameter for evaluating the loosening of arthroplasty components is the presence of radiolucent lines in postoperative radiographs, situated on the fringes of the components. Radiolucent lines can be present right from the immediate postoperative period and will only become a concern to the surgeon when their pattern changes over the course of the follow-up period, which can mean the loosening of the components and the need to review the arthroplasty.21

There are numerous parameters of evaluation and quantification of radiolucent lines and of the loosening of the arthroplasty components. The parameter most commonly used today is the classification of the American Knee Society based on anterior and lateral knee radiographs,22 where radiolucent lines are quantified according to their location around the tibial and femoral components. This evaluation is usually carried out by experienced surgeons who determine whether the components are stable or unstable. Another method for evaluating component stability is the use of dynamic images with an image intensifier (radioscopy). This method involves the performance of knee maneuvers with varus and valgus stress, which are simultaneously documented by the image intensifier, with the ability to determine the movement of the components and its stability.23

Due to the reduced number of studies conducted in the investigation of cementless knee arthroplasty stability and in the comparison of evaluation methods, we carried out this study to compare two femoral and tibial component stability evaluation methods in cementless total knee arthroplasty with rotating platform.

MATERIALS AND METHODS

We evaluated 20 cementless arthroplasties (20 patients) with rotating platform with at least one year of follow-up. The patients were operated in 2007 and 2008, averaging 17.6 months of postoperative time (12-31 months), being chosen at random for performance of the cementless arthroplasty. The patients' average age was 62 years (ranging between 52 and 75 years) at the time of surgery, with 16 women and four men. All the patients were informed of the surgical procedure and signed a consent form to take part in the study.

All the operations were performed using a medial parapatellar approach, sacrificing the posterior cruciate ligament. The bone cuts were made with standard guides and the surgeons used cementless total knee arthroplasty with a New Wave rotating platform from the Lepine group.

The femoral and tibial components were of chromium-cobalt-molybdenum with porous coating of 120 micrometers. The patellar component was fixed directly on the patella through cementation and two fixator pins. Vacuum drainage was used in all the cases, and was removed when the debit was below 100ml at 24 hours and in the other cases, at 48 hours.

The patients were discharged on the fourth postoperative day allowing partial weight bearing with walking frame. Weight bearing was progressively increased, making the transition to a contralateral cane at three to four weeks and to total weight bearing at six to eight weeks.

The patients were assessed at 1, 3, 6 and 12 and 24 weeks. After this initial phase, the patients made a return outpatient visit every six months up to two years after the surgery and annually after this date.

In the first stage of the study radiographs were taken of the operated knee in orthostatic frontal view with monopodal support and in absolute lateral view of the knee with 30 degrees of flexion in each patient. The images were analyzed, on different occasions, by two surgeons with experience in knee arthroplasty, to check the pattern of the radiolucent lines in the femur and in the tibia, verifying whether the femoral and tibial components were stable or unstable, analyzing each component separately. The responses were noted down in different places for each surgeon (double-blind procedure).

The second phase of the study consisted of the dynamic analysis of the femoral and tibial components, recording real-time images with an image intensifier, when we performed varus and valgus stress maneuvers on the operated knees.

The patients were positioned in dorsal decubitus on the radiography table and an initial static image of the knee operated in extension was documented. After the image was captured and photographed, we performed a varus and valgus stress maneuver of the knee in extension, where both images were documented and analyzed in real time. At this time we were able to analyze whether one of the components, which appeared stable in the static image, would present instability in the dynamic examination.

The images of the static and dynamic examinations were documented and kept for subsequent comparison with the radiographic examination.

For the statistical analysis of results and for the evaluation of the agreement of results between the different examiners and the radioscopy, we used the Kappa test at a significance level of 5%, which evaluates the concordance between responses. The tested hypothesis is whether the Kappa index is equal to 0, which would indicate null concordance, or if higher than zero, which means concordance is greater than chance. Upon finding a p-value <0.05, this indicates that the measure of concordance is significantly greater than zero, which would indicate the existence of some concordance. This does not necessarily mean that the concordance is high.

To supplement the analysis we observe the Kappa index that points out the degree of concordance: the closer to 0 the higher the concordance and the closer to 1, the better the concordance. Thus the maximum value of the Kappa measure is 1 (total concordance) and values close to or below 0 (indicating no concordance).

The interpretation of the concordance values followed the methodology proposed by Landis and Koch.24 (Table 1)

RESULTS

The results of the static and dynamic analysis by the specialists and of the radioscopy, respectively, were compared separately for the tibia and femur, for us to test two methods for analysis of the stability of cementless knee arthroplasties. (Table 2)

The absolute frequencies (n) and relative frequencies (%) were presented for the qualitative variables. The mean and median and standard deviation, minimum and maximum were used as summary measures to indicate variability for the quantitative variables. (Tables 3 to 8)

Comparison of results for the tibia

To compare the results obtained by the first examiner in relation to the radioscopy, note that there is perfect concordance (p<0.05) between the two results, i.e., the result was identical (kappa=1). (Table 9)

To compare the results obtained by the second surgeon in relation to the radioscopy, note that there is almost perfect concordance (p<0.05) between the two results, presenting a Kappa equal to 0.828. (Table 10)

After this there was a comparison of the results obtained by the two surgeons. Based on the results it can be seen that there is almost perfect concordance (p<0.05) between the two results, presenting a Kappa equal to 0.828. (Table 11) and (Figures 1-3)


Comparison of results for the femur

The results obtained by the two surgeons and by the radioscopy were stable for all the patients, indicating 100% of concordance of results. In this case (total concordance) we do not calculate the Kappa value.

DISCUSSION

Beaupré et al.25 presented a comparative study between cemented and cementless knee prostheses evaluating clinical aspects and did not find any difference between the two groups. Most authors consider that the absence of pain after the use of cementless prostheses consists of the clinical method for evaluating implant integration. In this study, besides the clinical evaluation, we opted to add methodology with dynamic and static radiographic examination in the implant analysis.

In our study the methodology executed to evaluate implant stability presented good inter-observer correlation. The comparative analysis of the femoral components between the two examiners and the radioscopy presented concordance in all the cases with a Kappa index of 1.

After analyzing the comparative data between the static and dynamic techniques for evaluating stability of the tibial and femoral components in cementless knee arthroplasty with rotating platform, we can conclude that the two methods produce similar results in terms of stability of the components. This was demonstrated by the positive statistical correlation (p<0.001) in the comparison between the two methods.

Hildebrand et al.26 conducted a study analyzing the migration of cementless implants and demonstrated that there is some degree of migration of the components until implant osteointegration occurs. Regner et al.27 demonstrated that cemented implants migrate less initially, yet constantly over a five-year period, and that cemented implants only migrate in the first six months. This may be due to osteointegration of the implant. The presence of radiolucency around the cementless implant does not necessarily mean the loosening of the component, but may result from its initial migration. In our study three patients exhibited signs of radiolucency around the implant, but did not present movement of the implants in the analysis performed or any other signs of loosening. These patients will be followed up for a longer period of time with the intention of observing the evolution of the radiolucent lines.

When we compare the stability of the femoral components from the present study with the data presented in literature, we can conclude that the pattern of stability of the components followed most studies, demonstrating that cementless total knee arthroplasties with rotating platform can show signs of instability in the tibial component, while the femoral component is stable in the majority of cases.

Going from the methodology applied, we can claim that static radiographic analysis performed by experienced surgeons and dynamic analysis through radioscopy produce similar results in the evaluation of the stability of tibial and femoral components in cementless total knee arthroplasties with rotating platform. This result leads us to believe that the cementless arthroplasties evaluated present osteointegration.

We consider that the current generation of cementless implants is different to that presented in the past. The first generation of cementless knee prostheses presented poor and discouraging results. This is possibly due to the inappropriate design of the implants and other factors such as the presence of metalback in the patellar component and the use of polyethylene that may have increased debris production.28 The current generation of cementless implants presents positive results in relation to osteointegration, and it is worthwhile conducting further studies on this line of arthroplasties that may be a trend in knee prostheses as has been observed historically in hip prostheses.

As a study limitation, we considered the non-inclusion in this analysis of other methods for evaluating the stability of tibial and femoral components, such as bone scintigraphy, and we did not define the gold standard in the evaluation of the stability of components in a knee arthroplasty. The study conducted out here is not intended to provide a comparative analysis between the functional or durability result of cementless knee prostheses and cemented knee prostheses. Further studies are necessary for better definition of these standards.

CONCLUSION

Based on this study, we can conclude that the static method of evaluation of the stability of cementless knee arthroplasties by experienced surgeons is just as effective as the dynamic evaluation by radioscopy, whereas both can be used with accuracy in evaluating patients.

REFERENCES

  • 1. Keating EM, Meding JB, Faris PM, Ritter MA. Long-term follow-up of nonmodular total knee replacements. Clin Orthop Relat Res. 2002;(404):34-9.
  • 2. Kelly MA, Clarke HD. Long-term results of posterior cruciate-substituting total knee arthroplasty. Clin Orthop Relat Res. 2002;(404):51-7.
  • 3. Ma HM, Lu YC, Ho FY, Huang CH. Long-term results of total condylar knee arthroplasty. J Arthroplasty. 2005;20(5):580-4.
  • 4. Font-Rodriguez DE, Scuderi GR, Insall JN. Survivorship of cemented total knee arthroplasty. Clin Orthop Relat Res. 1997;(345):79-86.
  • 5. Ranawat CS, Flynn WF Jr, Saddler S, Hansraj KK, Maynard MJ. Long-term results of the total condylar knee arthroplasty. A 15-year survivorship study. Clin Orthop Relat Res. 1993;(286):94-102.
  • 6. Rasquinha VJ, Ranawat CS, Cervieri CL, Rodriguez JA. The press-fit condylar modular total knee system with a posterior cruciate-substituting design. A concise follow-up of a previous report. J Bone Joint Surg Am. 2006;88(5):1006-10.
  • 7. Ritter MA, Berend ME, Meding JB, Keating EM, Faris PM, Crites BM. Long-term followup of anatomic graduated components posterior cruciate-retaining total knee replacement. Clin Orthop Relat Res. 2001;(388):51-7.
  • 8. Rodriguez JA, Bhende H, Ranawat CS. Total condylar knee replacement: a 20-year followup study. Clin Orthop Relat Res. 2001;(388):10-7.
  • 9. Berger RA, Rosenberg AG, Barden RM, Sheinkop MB, Jacobs JJ, Galante JO. Long-term followup of the Miller-Galante total knee replacement. Clin Orthop Relat Res. 2001;(388):58-67.
  • 10. Callaghan JJ, Insall JN, Greenwald AS, Dennis DA, Komistek RD, Murray DW, et al. Mobile-bearing knee replacement: concepts and results. Instr Course Lect. 2001;50:431-49.
  • 11. Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM. Insall Award paper. Why are total knee arthroplasties failing today? Clin Orthop Relat Res. 2002;(404):7-13.
  • 12. Sorrells RB. The rotating platform mobile bearing TKA. Orthopedics. 1996;19(9):793-6.
  • 13. Nilsson KG, Kärrholm J, Carlsson L, Dalén T. Hydroxyapatite coating versus cemented fixation of the tibial component in total knee arthroplasty: prospective randomized comparison of hydroxyapatite-coated and cemented tibial components with 5-year follow-up using radiostereometry. J Arthroplasty. 1999;14(1):9-20.
  • 14. Nilsson KG, Henricson A, Norgren B, Dalén T. Uncemented HA-coated implant is the optimum fixation for TKA in the young patient. Clin Orthop Relat Res. 2006;448:129-39.
  • 15. Onsten I, Nordqvist A, Carlsson AS, Besjakov J, Shott S. Hydroxyapatite augmentation of the porous coating improves fixation of tibial components. A randomised RSA study in 116 patients. J Bone Joint Surg Br. 1998;80(3):417-25.
  • 16. Ritter MA, Meneghini RM. Twenty-year survivorship of cementless anatomic graduated component total knee arthroplasty. J Arthroplasty. 2010;25(4):507- 13.
  • 17. Stiehl JB, Voorhorst PE. Total knee arthroplasty with a mobile-bearing prosthesis: comparison of retention and sacrifice of the posterior cruciate ligament in cementless implants. Am J Orthop (Belle Mead NJ). 1999;28(4):223-8.
  • 18. Cook SD, Barrack RL, Thomas KA, Haddad RJ Jr. Quantitative histologic analysis of tissue growth into porous total knee components. J Arthroplasty. 1989;4 Suppl:S33-43.
  • 19. Barrack RL, Nakamura SJ, Hopkins SG, Rosenzweig S. Winner of the 2003 James A. Rand Young Investigator's Award. Early failure of cementless mobile-bearing total knee arthroplasty. J Arthroplasty. 2004;19(7 Suppl 2):101-6.
  • 20. Tai CC, Cross MJ. Five- to 12-year follow-up of a hydroxyapatite-coated, cementless total knee replacement in young, active patients. J Bone Joint Surg Br. 2006;88(9):1158-63.
  • 21. Aebli N, Krebs J, Schwenke D, Hii T, Wehrli U. Progression of radiolucent lines in cementless twin-bearing low-contact-stress knee prostheses: a retrospective study. J Arthroplasty. 2004;19(6):783-9.
  • 22. Ewald FC. The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res. 1989;(248):9-12.
  • 23. Stähelin T, Kessler O, Pfirrmann C, Jacob HA, Romero J. Fluoroscopically assisted stress radiography for varus-valgus stability assessment in flexion after total knee arthroplasty. J Arthroplasty. 2003;18(4):513-5.
  • 24. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159-74.
  • 25. Beaupré LA, al-Yamani M, Huckell JR, Johnston DW. Hydroxyapatite-coated tibial implants compared with cemented tibial fixation in primary total knee arthroplasty. A randomized trial of outcomes at five years. J Bone Joint Surg Am. 2007;89(10):2204-11.
  • 26. Hildebrand R, Trappmann D, Georg C, Müller HH, Koller M, Klose KJ, et al. [What effect does the hydroxyapatite coating have in cementless knee arthroplasty?]. Orthopade. 2003;32(4):323-30.
  • 27. Regnér L, Carlsson L, Kärrholm J, Herberts P. Tibial component fixation in porous- and hydroxyapatite-coated total knee arthroplasty: a radiostereo metric evaluation of migration and inducible displacement after 5 years. J Arthroplasty. 2000;15(6):681-9.
  • 28. Regnér L, Carlsson L, Kärrholm J, Herberts P. Ceramic coating improves tibial component fixation in total knee arthroplasty. J Arthroplasty. 1998;13(8):882-9.
  • Correspondência:
    Secretaria da Graduação
    R. Dr. Ovídio Pires de Campos, 333 - 3º andar
    Cerqueira César - 05403-010 - São Paulo - SP. Brasil
    E-mail:
  • Publication Dates

    • Publication in this collection
      01 Aug 2012
    • Date of issue
      2012

    History

    • Received
      22 Jan 2010
    • Accepted
      25 May 2010
    ATHA EDITORA Rua: Machado Bittencourt, 190, 4º andar - Vila Mariana - São Paulo Capital - CEP 04044-000, Telefone: 55-11-5087-9502 - São Paulo - SP - Brazil
    E-mail: actaortopedicabrasileira@uol.com.br