Retention System and Splinting on Morse Taper Implants in the Posterior Maxilla by 3D Finite Element Analysis

Abstract

The purpose of this study was to evaluate different retention systems (cement- or screw-retained) and crown designs (non-splinted or splinted) of fixed implant-supported restorations, in terms of stress distributions in implants/components and bone tissue, by 3-dimensional (3D) finite element analysis. Four 3D models were simulated with the InVesalius, Rhinoceros 3D, and SolidWorks programs. Models were made of type III bone from the posterior maxillary area. Models included three 4.0-mm-diameter Morse taper (MT) implants with different lengths, which supported metal-ceramic crowns. Models were processed by the Femap and NeiNastran programs, using an axial force of 400 N and oblique force of 200 N. Results were visualized as the von Mises stress and maximum principal stress (σmax). Under axial loading, there was no difference in the distribution of stress in implants/components between retention systems and splinted crowns; however, in oblique loading, cemented prostheses showed better stress distribution than screwed prostheses, whereas splinted crowns tended to reduce stress in the implant of the first molar. In the bone tissue cemented prostheses showed better stress distribution in bone tissue than screwed prostheses under axial and oblique loading. The splinted design only had an effect in the screwed prosthesis, with no influence in the cemented prosthesis. Cemented prostheses on MT implants showed more favorable stress distributions in implants/components and bone tissue. Splinting was favorable for stress distribution only for screwed prostheses under oblique loading.

Key Words:
dental implants; cement-retained; screw-retained; splinted; finite element analysis

Resumo

O objetivo deste estudo foi avaliar diferentes sistemas de retenção (cimentada x parafusada) e configuração da coroas (unitárias x esplintadas) de próteses fixas implantossuportadas em relação a distribuição de tensões nos implantes/componentes e tecido ósseo pela análise de elementos finitos 3D. Quatro modelos 3D foram simulados com auxílio dos programas Invesalius, e Rhinoceros 3D, e SolidWorks. Os modelos foram confeccionados simulando bloco ósseo de região posterior da maxila (tipo ósseo III), com 3 implantes cone Morse com 4,0 mm de diâmetro e diferentes comprimentos, suportando prótese metalocerâmica de 3 elementos. Os modelos foram processados pelos programas FEMAP e NEiNastran sob força axial de 400 N e oblíqua de 200N. Os resultados foram plotados através de mapas de tensão de von Mises (vM) (implantes e componentes) e tensão máxima principal (TMP) (tecido ósseo). Sobre o carregamento axial, não foi observada diferenças entre os diferentes sistemas de retenção e tipo de prótese na distribuição das tensões nos implantes/componentes, porém, sobre o carregamento oblíquo as próteses cimentadas apresentaram melhor distribuição de tensões em comparação com as próteses parafusadas, enquanto que as próteses esplintadas apresentou uma tendência de redução das tensões no implante do primeiro molar. No tecido ósseo as próteses cimentadas apresentaram melhor distribuição das tensões em comparação com as próteses parafusadas, independente do carregamento. A esplintagem foi favorável somente para as próteses parafusadas, não havendo influência sobre as próteses cimentadas. As próteses cimentadas sobre implantes cone Morse apresentam melhor comportamento biomecânico nos implantes/componentes e tecido ósseo. A esplintagem foi efetiva somente nas próteses parafusadas sob carregamento oblíquo.

Introduction

Rehabilitation with implant-supported fixed prostheses is a predictable modality to restore lost function and esthetics in partially or totally edentulous patients. Treatment longevity depends on the preservation of bone around the implant. Morse taper (MT) implants have shown high success rates with good bone preservation in implant-supported restorations 11 Mangano, C; Iaculli, F; Piattelli, A; Mangano, F. Fixed restorations supported by Morse-taper connection implants: a retrospective clinical study with 10-20 years of follow-up. Clin Oral Implants Res 2015;26:1229-1236.. This clinical performance can be explained by the reduced incidence of microgap and bacterial contamination 22 Alves, DC; Carvalho, PS; Martinez, EF. In vitro microbiological analysis of bacterial seal at the implant-abutment interface using two Morse taper implant models. Braz Dent J2014;25:48-53.. Furthermore, compared to other connections, the MT connection permits superior bone crest stability under axial and lateral loads 33 Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300..

The make decision about the use or not of splinted crowns is based on different factors such as: implant length, occlusion, hygiene, connection type and difficulty achieving a passively fitting framework 44 Clelland, N; Chaudhry, J; Rashid, RG; McGlumphy, E. Split-mouth comparison of splinted and non splinted prostheses on short implants: 3-year results. Int J Oral Maxillofac Implants. 2016;31:1135-1141.. The maxillary posterior area after tooth loss often can limits the use of the longer implants. In these situations, short implants may be recommended with the aim of avoiding auxiliary surgical procedures 55 Mendonca, JA; Francischone, CE; Senna, PM; Matos de Oliveira, AE; Sotto-Maior, BS. A retrospective evaluation of the survival rates of splinted and non-splinted short dental implants in posterior partially edentulous jaws. J Periodontol 2014;85:787-794.. However, short implants may present greater risks when compared to implants with conventional lengths, especially when placed in the posterior region 66 Lemos, CA; Ferro-Alves, ML; Okamoto, R; Mendonca, MR; Pellizzer, EP. Short dental implants versus standard dental implants placed in the posterior jaws: A systematic review and meta-analysis. J Dent 2016; 47:8-17.. Thus, the splinted crowns of short implants can be indicated in atrophic maxilla because transmit more uniform stress in structures with higher reduction of stress in the peri-implant region 44 Clelland, N; Chaudhry, J; Rashid, RG; McGlumphy, E. Split-mouth comparison of splinted and non splinted prostheses on short implants: 3-year results. Int J Oral Maxillofac Implants. 2016;31:1135-1141..

Different retention systems, such as screw- or cement-retained systems, are available for implant-supported restorations. The retention system is chosen on the basis of the advantages and disadvantages of each system, as well as the clinician’s preference or clinical experience 77 Lemos, CA; de Souza Batista, VE; Almeida, DA; Santiago, JFJunior; Verri, FR; Pellizzer, EP. Evaluation of cement-retained versus screw-retained implant-supported restorations for marginal bone loss: A systematic review and meta-analysis. J Prosthet Dent 2015;115:419-427.. Screw-retained prostheses have advantage of reversibility allowing maintenance mainly in multiple prostheses, predictable retention, lower biological risks, but it presents an unfavorable aesthetic, more difficulty in achieving stable occlusal contacts, as well as greater risks of mechanical complications. Cement-retained prosthesis has as main characteristic aesthetics, flexibility to correct implant position, occlusal table integrates, more likely to achieve passive fit, and ease fabricate and cost; however, there is no predictability regarding reversibility, show greater biological risks due to excess cement, need for a minimum of inter occlusal space for the abutment to allow retention of the crown 88 Shadid, R; Sadaqa, N. A comparison between screw- and cement-retained implant prostheses. A literature review. J Oral Implantol 2012;38:298-307.. However, there is currently no consensus about the use of splinting or the best type of retention system for fixed implant-supported restorations with MT connections.

The aim of this study was to evaluate the influence of splinting and different retention systems in the MT connection. The null hypotheses were as follows: 11 Mangano, C; Iaculli, F; Piattelli, A; Mangano, F. Fixed restorations supported by Morse-taper connection implants: a retrospective clinical study with 10-20 years of follow-up. Clin Oral Implants Res 2015;26:1229-1236. there would be no difference in stress distribution between cement- and screw-retained implant-supported fixed restorations; and 22 Alves, DC; Carvalho, PS; Martinez, EF. In vitro microbiological analysis of bacterial seal at the implant-abutment interface using two Morse taper implant models. Braz Dent J2014;25:48-53. there would be no difference in stress distribution between single-unit and splinted prostheses.

Materials and Methods

Experimental Design

This research was designed to consider three variation factors: the retention system (screw- or cement-retained), crown design (non-splinted or splinted) and loading condition (axial and oblique). Four models were created to test these factors (Table 1).

Table 1
Specifications of the models of Morse taper connection implants with lengths of 10 mm (1st premolar), 8.5 mm (2nd premolar) and 7 mm (1st molar).

3D Finite Element Modeling

The 3D finite element analysis (FEA) methodology used here follows that of previous studies 33 Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300.,99 de Moraes, SL; Verri, FR; Santiago, JF Jr; Almeida, DA; de Mello, CC; Pellizzer, EP. A 3-D finite element study of the influence of crown-implant ratio on stress distribution. Braz Dent J 2013;24:635-341.. Each model was composed of a bone block of the maxillary area of the first premolar to first molar, with trabecular bone in the center surrounded by 1 mm of cortical bone. Information for the bone block was obtained by computerized tomography (sagittal section) of the region. Modeling was performed with InVesalius software (CTI Renato Archer, Campinas, SP, Brazil) and surface simplification with Rhinoceros 4.0 software (NURBS Modeling for Windows, Seattle, WA, USA).

The implant design was a simplified version of the original 1-connection MT design (Morse taper - Torq, Conexão Sistemas de Prótese Ltda, Aruja, SP, Brazil). The MT implant measured 4 mm in diameter and 10, 8.5, and 7 mm in length for the first premolar, second premolar and first molar, respectively. A prosthetic UCLA-indexed abutment of Co-Cr was used for both retention systems. MT implants were simulated with 1-mm subcrestal placement.

The crown design was obtained from an artificial mandibular molar tooth (Odontofix Indústria e Comércio de Material Odontológico Ltda., Ribeirão Preto, SP, Brazil) and digitized by using a 3D scanner (MDX-20; Roland DG, São Paulo SP, Brazil). Images were exported to Rhinoceros 4.0 CAD software for modeling. Occlusal surface details were added with the SolidWorks CAD software package (SolidWorks Corp, Waltham, MA, USA). Implant-supported crowns were simulated with a screw or cement connection to the MT implant. The cement layer was 50-µm-thick zinc phosphate cement. Crowns were designed as single-unit (non-splinted) or splinted crowns. All implant and abutment designs were simplified by using SolidWorks and Rhinoceros 4.0 software packages. Simplified designs maintained similarity to the real models (Fig. 1).

Figure 1
Schematic view of simulated solid and meshes of structures with analyzed structures, load conditions and restrictions.

After modeling the solids, geometries were exported to FEA software for pre- and post-processing (FEMAP 11.2, Siemens PLM, Santa Ana, CA, USA), to obtain tetrahedral parabolic solid elements with 10 nodes used for mesh generation for all involved structures. Mechanical properties of each simulated material were attributed to the meshes by using literature values 1010 Spazzin, AO; Dos Santos, MB; S Sobrinho, LC; Consani, RL; Mesquita, MF. Effects of horizontal misfit and bar framework material on the stress distribution of an overdenture-retaining bar system: a 3D finite element analysis. J Prosthodont 2011;20:517-522.,1111 Sertgöz, A. Finite element analysis study of the effect of superstructure material on stress distribution in an implant-supported fixed prosthesis. Int J Prosthodont. 1997;10:19-27.,1212 Anusavice, KJ; Hojjatie, B. Stress distribution in metal-ceramic crowns with a facial porcelain margin. J Dent Res 1987;66:1493-1498. (Table 2). All materials were assumed to be homogeneous, isotropic and linearly elastic.

Table 2
Mechanical properties of the simulated materials

Interface, Boundary Conditions and Loadings

All contacts were simulated as symmetric welds, except for the abutment/implant and single-unit crown contacts that were simulated by symmetric contacts. Axes (x, y and z) were fixed in both bone block sections (cortical and trabecular). All other model surfaces were unrestricted. The non-linear applied axial force was 400 N (100 N for both premolars and 200 N for the first molar) at a point determined on the tip of each cusp. The non-linear oblique load was 200 N (50 N for both premolars and 100 N for first molar), suppressing the lingual cusps (Fig. 1).

Criteria for Stress Analyses

Models generated in FEMAP 11.2 were exported to the NeiNastran 11.0 software package (Noran Engineering, Inc., Westminster, CA, USA). After the analytical problems were solved, the models were imported to the FEA software to plot maps of the stresses in units of mega-Pascal (MPa). Von Mises stress analysis was used for implant and prosthetic components (ductile materials). Maximum principal stress (σmax) was used for bone tissue (friable materials), and are distinguished between tensile stresses (positive values) and compressive stresses (negative values) 33 Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300..

Results

Von Mises Stress Analysis of Implants/Abutments

Under axial loading, no differences in the stress distributions of implants/components were observed. Under oblique loading, greater stress concentration was observed compared to axial loading. Compared to screw-retained prostheses (M1, M2), cement-retained prostheses (M3, M4) showed better stress distribution in implants/components, regardless of the type of crown. Splinted crowns (M2, M4) favored stress distribution, especially in the first molar region (Fig. 2).

Figure 2
Von Mises maps to stress distribution on implant/components.

Maximum Principal Stress Analysis of Bone Tissue

Axial loading, similar levels of tensile and compressive stresses in cortical bone were obtained for the first and second premolars. The first molar showed higher tensile stress in all models, with greater concentration of stress for screw-retained prostheses (M1, M2) compared to cemented prostheses (M3, M4). No difference was found for the splinted crown design for both retention systems (Figs. 3 and 4). Under oblique loading, cemented prostheses (M3, M4) showed better stress distribution than screwed prostheses (M1, M2). The splinted crown was favorable to stress distribution only for the screwed prosthesis, with no difference for the cemented prosthesis (Figs. 3 and 4).

Figure 3
Maximum principal stress maps of distribution on cortical and trabecular bone tissue.

Figure 4
Stress values (MPa) in maximum principal stress (σmax) on cortical bone tissue under axial and oblique loading.

Discussion

MT implants present better biomechanical properties than other connections 33 Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300., favoring long-term bone preservation 11 Mangano, C; Iaculli, F; Piattelli, A; Mangano, F. Fixed restorations supported by Morse-taper connection implants: a retrospective clinical study with 10-20 years of follow-up. Clin Oral Implants Res 2015;26:1229-1236.. The internal conical connection provides better dissipation of tension, due to the high stability of the MT connection 1313 Minatel, L; Verri, FR; Kudo, GAH; Almeida, DAF; de Souza Batista, VE; Lemos, CAA; et al.. Effect of different types of prosthetic platforms on stress-distribution in dental implant-supported prostheses. Mater Sci Eng C Mater Biol Appl 2017;71:35-42.. In this report, the first null hypothesis was rejected. Differences were found between retention systems, with cemented prostheses showing better stress distribution. These results corroborate those of other studies reporting the superiority of cemented systems compared to screwed prostheses 1414 de Vasconcellos, DK; Kojima, AN; Mesquita, AM; Bottino, MA; Ozcan, M. A microstrain comparison of passively fitting screw-retained and cemented titanium frameworks. J Prosthet Dent 2014;112:834-838.,1515 Silva, GC; Cornacchia, TM; de Magalhaes, CS; Bueno, AC; Moreira, AN. Biomechanical evaluation of screw- and cement-retained implant-supported prostheses: a nonlinear finite element analysis. J Prosthet Dent 2014;112:1479-1488..

There are different theories to explain the better biomechanical behavior of cemented prostheses. For example, it has been proposed that the screw access hole helps to transfer stress off the long axis of the implant, and that different restorative materials can transfer occlusal loads laterally to the implant instead of axially 77 Lemos, CA; de Souza Batista, VE; Almeida, DA; Santiago, JFJunior; Verri, FR; Pellizzer, EP. Evaluation of cement-retained versus screw-retained implant-supported restorations for marginal bone loss: A systematic review and meta-analysis. J Prosthet Dent 2015;115:419-427.. The cement also may be better at filling discrepancies, absorbing and equalizing the stress of deformations caused by contact between the abutment and implant in the prosthesis structure 1616 Guichet, DL; Caputo, AA; Choi, H; Sorensen, JA. Passivity of fit and marginal opening in screw- or cement-retained implant fixed partial denture designs. Int J Oral Maxillofac Implants 2000;15:239-246.. These factors can contribute to fixation screw of screwed prosthesis present stress overload in the structure, which contributes to greater complication rates can explain the frequent failures (e.g. screw loosening or fracture) that are observed in screwed prostheses 77 Lemos, CA; de Souza Batista, VE; Almeida, DA; Santiago, JFJunior; Verri, FR; Pellizzer, EP. Evaluation of cement-retained versus screw-retained implant-supported restorations for marginal bone loss: A systematic review and meta-analysis. J Prosthet Dent 2015;115:419-427.,1515 Silva, GC; Cornacchia, TM; de Magalhaes, CS; Bueno, AC; Moreira, AN. Biomechanical evaluation of screw- and cement-retained implant-supported prostheses: a nonlinear finite element analysis. J Prosthet Dent 2014;112:1479-1488.,1717 Nissan, J; Narobai, D; Gross, O; Ghelfan, O; Chaushu, G. Long-term outcome of cemented versus screw-retained implant-supported partial restorations. Int J Oral Maxillofac Implants 2011;26:1102-1107.. The high risk in the screw-retained prosthesis can be related to lower preload in the screw with displacement (penetration and gaps) higher and concentrated on the threads of the screw/or abutment when compared to cement-retained 1515 Silva, GC; Cornacchia, TM; de Magalhaes, CS; Bueno, AC; Moreira, AN. Biomechanical evaluation of screw- and cement-retained implant-supported prostheses: a nonlinear finite element analysis. J Prosthet Dent 2014;112:1479-1488.. This difference can be even greater when use a one-piece abutment for cemented prostheses 33 Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300., since the type of abutment can influence biomechanical behavior 1818 Camargos, G de V; Sotto-Maior, BS; Silva, WJ; Lazari, PC; Del Bel Cury, AA. Prosthetic abutment influences bone biomechanical behavior of immediately loaded implants. Braz Oral Res 2016;30. pii: S1806-83242016000100901. doi: 10.1590/1807-3107BOR-2016.vol30.0065.
https://doi.org/10.1590/1807-3107BOR-201...
. Thus, to eliminate the possibility of bias, the same abutment (UCLA casting in Co-Cr) was considered for both retention systems. The results of this study reinforce the notion that cemented prostheses on MT implants may be clinically indicated, as they have a greater chance of bone preservation when compared to screwed prostheses.

The second null hypothesis of this study was also rejected. Different stress distributions were observed in the implants/components for the splinted crowns on both retention systems. In the bone tissue was observed difference for splinted compared to non-splinted crowns for the screwed prostheses under oblique loading. Similarly, other studies reported that splinting favors a reduction in stress in the implants/components and, consequently, bone tissue, especially for screwed prostheses under oblique loading 1919 Yilmaz, B; Seidt, JD; McGlumphy, EA; Clelland, NL. Comparison of strains for splinted and nonsplinted screw-retained prostheses on short implants. Int J Oral Maxillofac Implants 2011;26:1176-1182.,2020 Pellizzer, EP; de Mello, CC; Santiago, JF Junior; de Souza Batista, VE; de Faria Almeida, DA; Verri, FR. Analysis of the biomechanical behavior of short implants: The photo-elasticity method. Mater Sci Eng C Mater Biol Appl 2015;55:187-192.. Splinting favors the sharing of stress between implants, because a rigid structure acts to unite the crowns 2020 Pellizzer, EP; de Mello, CC; Santiago, JF Junior; de Souza Batista, VE; de Faria Almeida, DA; Verri, FR. Analysis of the biomechanical behavior of short implants: The photo-elasticity method. Mater Sci Eng C Mater Biol Appl 2015;55:187-192.. The larger load in the molar area is due to a greater occlusal table over a short implant. Short implants are recommended for maxillary posterior region (molar) due to poor availability of bone tissue; however, short implants are more susceptible to failures rates than conventional implants 66 Lemos, CA; Ferro-Alves, ML; Okamoto, R; Mendonca, MR; Pellizzer, EP. Short dental implants versus standard dental implants placed in the posterior jaws: A systematic review and meta-analysis. J Dent 2016; 47:8-17.. Thus, the splinting of short implants with implants of longer length is indicated in these situations 44 Clelland, N; Chaudhry, J; Rashid, RG; McGlumphy, E. Split-mouth comparison of splinted and non splinted prostheses on short implants: 3-year results. Int J Oral Maxillofac Implants. 2016;31:1135-1141.. Splinted crowns help to distribute the stress between the splinted implants. In this study the increase in stress in implants placed in the premolar area after splinting prostheses is not harmful, since these implants show greater length, and the increased length of the implant favors to stress distribution on the bone tissue 2121 Anitua, E; Orive, G. Short implants in maxillae and mandibles: a retrospective study with 1 to 8 years of follow-up. J Periodontol 2010;81:819-826.. So, the splinting of short implants with implants of longer length is indicated in these situations 2020 Pellizzer, EP; de Mello, CC; Santiago, JF Junior; de Souza Batista, VE; de Faria Almeida, DA; Verri, FR. Analysis of the biomechanical behavior of short implants: The photo-elasticity method. Mater Sci Eng C Mater Biol Appl 2015;55:187-192..

The difference in stress with splinting was observed only for the screwed prostheses. This finding is consistent with the study of Yilmaz et al. 2222 Yilmaz, B; Mess, J; Seidt, J; Clelland, NL. Strain comparisons for splinted and non-splinted cement-retained implant crowns. Int J Prosthodont 2013;26:235-238. which found no influence of splinting for cemented prostheses. Thus, in cases where a cemented prosthesis is used, a single-unit prosthesis should be preferred. This non-splinted crowns allows ease of access in the interproximal region, favoring patient hygiene 2323 Lindhe, J; Meyle, J. Peri-implant diseases: Consensus Report of the Sixth European Workshop on Periodontology. J Clin Periodontol 2008;35:282-285. and adaptation to the prosthesis, with increased quality of life 2425 Goiato, MC; Torcato, LB; Dos Santos, DM; Moreno, A; Antenucci, RM; de Carvalho Dekon, SF. Quality of life and satisfaction of patients wearing implant-supported fixed partial denture: a cross-sectional survey of patients from Aracatuba city, Brazil. Clin Oral Implants Res 2015;26:701-708..

All of the models analyzed in this study showed peri-implant areas of compressive and tensile stresses that were within physiological limits described in the literature of ultimate strength of cortical bone (7276 MPa for tensile stress, 140-170 MPa for compressive stress) 2525 Papavasiliou, G; Kamposiora, P; Bayne, SC; Felton, DA. Three-dimensional finite element analysis of stress-distribution around single tooth implants as a function of bony support, prosthesis type, and loading during function. J Prosthet Dent 1996;76:633-640.. FEA is a useful tool for evaluating preclinical situations and determining the stress distribution in individualized structures 33 Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300.,99 de Moraes, SL; Verri, FR; Santiago, JF Jr; Almeida, DA; de Mello, CC; Pellizzer, EP. A 3-D finite element study of the influence of crown-implant ratio on stress distribution. Braz Dent J 2013;24:635-341.,1111 Sertgöz, A. Finite element analysis study of the effect of superstructure material on stress distribution in an implant-supported fixed prosthesis. Int J Prosthodont. 1997;10:19-27.. One of the limitations present in this study was the use of symmetric contacts instead of frictional contacts for the analyzed structures, because this contact demanding on the computational capacity of high-performance in these complex models. Thus, the results from FEA studies permit a better understanding of biomechanical behavior; however, these inherent limitations of FEA must be kept in mind and should be carefully extrapolated to the clinic. Thus, when combined with well-designed clinical studies, the results in this paper could serve as a guideline for implant dentistry, but controlled clinical studies should be conducted to verify the biomechanical effects clinically. Within the limitations of this study, cemented prostheses on MT implants had more favorable stress distributions in the implants/components and bone tissue. However, splinting improved the stress distribution only in screwed prostheses on MT implants under oblique loading.

Acknowledgements

This work was supported by The São Paulo State Research Foundation (FAPESP) in the form of a scholarship grant #2014/02490-8.

References

  • 1
    Mangano, C; Iaculli, F; Piattelli, A; Mangano, F. Fixed restorations supported by Morse-taper connection implants: a retrospective clinical study with 10-20 years of follow-up. Clin Oral Implants Res 2015;26:1229-1236.
  • 2
    Alves, DC; Carvalho, PS; Martinez, EF. In vitro microbiological analysis of bacterial seal at the implant-abutment interface using two Morse taper implant models. Braz Dent J2014;25:48-53.
  • 3
    Santiago, JF Junior; Verri, FR; Almeida, DA; de Souza Batista, VE; Lemos, CA; Pellizzer, EP. Finite element analysis on influence of implant surface treatments; connection and bone types. Mater Sci Eng C Mater Biol Appl 2016;63:292-300.
  • 4
    Clelland, N; Chaudhry, J; Rashid, RG; McGlumphy, E. Split-mouth comparison of splinted and non splinted prostheses on short implants: 3-year results. Int J Oral Maxillofac Implants. 2016;31:1135-1141.
  • 5
    Mendonca, JA; Francischone, CE; Senna, PM; Matos de Oliveira, AE; Sotto-Maior, BS. A retrospective evaluation of the survival rates of splinted and non-splinted short dental implants in posterior partially edentulous jaws. J Periodontol 2014;85:787-794.
  • 6
    Lemos, CA; Ferro-Alves, ML; Okamoto, R; Mendonca, MR; Pellizzer, EP. Short dental implants versus standard dental implants placed in the posterior jaws: A systematic review and meta-analysis. J Dent 2016; 47:8-17.
  • 7
    Lemos, CA; de Souza Batista, VE; Almeida, DA; Santiago, JFJunior; Verri, FR; Pellizzer, EP. Evaluation of cement-retained versus screw-retained implant-supported restorations for marginal bone loss: A systematic review and meta-analysis. J Prosthet Dent 2015;115:419-427.
  • 8
    Shadid, R; Sadaqa, N. A comparison between screw- and cement-retained implant prostheses. A literature review. J Oral Implantol 2012;38:298-307.
  • 9
    de Moraes, SL; Verri, FR; Santiago, JF Jr; Almeida, DA; de Mello, CC; Pellizzer, EP. A 3-D finite element study of the influence of crown-implant ratio on stress distribution. Braz Dent J 2013;24:635-341.
  • 10
    Spazzin, AO; Dos Santos, MB; S Sobrinho, LC; Consani, RL; Mesquita, MF. Effects of horizontal misfit and bar framework material on the stress distribution of an overdenture-retaining bar system: a 3D finite element analysis. J Prosthodont 2011;20:517-522.
  • 11
    Sertgöz, A. Finite element analysis study of the effect of superstructure material on stress distribution in an implant-supported fixed prosthesis. Int J Prosthodont. 1997;10:19-27.
  • 12
    Anusavice, KJ; Hojjatie, B. Stress distribution in metal-ceramic crowns with a facial porcelain margin. J Dent Res 1987;66:1493-1498.
  • 13
    Minatel, L; Verri, FR; Kudo, GAH; Almeida, DAF; de Souza Batista, VE; Lemos, CAA; et al.. Effect of different types of prosthetic platforms on stress-distribution in dental implant-supported prostheses. Mater Sci Eng C Mater Biol Appl 2017;71:35-42.
  • 14
    de Vasconcellos, DK; Kojima, AN; Mesquita, AM; Bottino, MA; Ozcan, M. A microstrain comparison of passively fitting screw-retained and cemented titanium frameworks. J Prosthet Dent 2014;112:834-838.
  • 15
    Silva, GC; Cornacchia, TM; de Magalhaes, CS; Bueno, AC; Moreira, AN. Biomechanical evaluation of screw- and cement-retained implant-supported prostheses: a nonlinear finite element analysis. J Prosthet Dent 2014;112:1479-1488.
  • 16
    Guichet, DL; Caputo, AA; Choi, H; Sorensen, JA. Passivity of fit and marginal opening in screw- or cement-retained implant fixed partial denture designs. Int J Oral Maxillofac Implants 2000;15:239-246.
  • 17
    Nissan, J; Narobai, D; Gross, O; Ghelfan, O; Chaushu, G. Long-term outcome of cemented versus screw-retained implant-supported partial restorations. Int J Oral Maxillofac Implants 2011;26:1102-1107.
  • 18
    Camargos, G de V; Sotto-Maior, BS; Silva, WJ; Lazari, PC; Del Bel Cury, AA. Prosthetic abutment influences bone biomechanical behavior of immediately loaded implants. Braz Oral Res 2016;30. pii: S1806-83242016000100901. doi: 10.1590/1807-3107BOR-2016.vol30.0065.
    » https://doi.org/10.1590/1807-3107BOR-2016.vol30.0065.
  • 19
    Yilmaz, B; Seidt, JD; McGlumphy, EA; Clelland, NL. Comparison of strains for splinted and nonsplinted screw-retained prostheses on short implants. Int J Oral Maxillofac Implants 2011;26:1176-1182.
  • 20
    Pellizzer, EP; de Mello, CC; Santiago, JF Junior; de Souza Batista, VE; de Faria Almeida, DA; Verri, FR. Analysis of the biomechanical behavior of short implants: The photo-elasticity method. Mater Sci Eng C Mater Biol Appl 2015;55:187-192.
  • 21
    Anitua, E; Orive, G. Short implants in maxillae and mandibles: a retrospective study with 1 to 8 years of follow-up. J Periodontol 2010;81:819-826.
  • 22
    Yilmaz, B; Mess, J; Seidt, J; Clelland, NL. Strain comparisons for splinted and non-splinted cement-retained implant crowns. Int J Prosthodont 2013;26:235-238.
  • 23
    Lindhe, J; Meyle, J. Peri-implant diseases: Consensus Report of the Sixth European Workshop on Periodontology. J Clin Periodontol 2008;35:282-285.
  • 25
    Goiato, MC; Torcato, LB; Dos Santos, DM; Moreno, A; Antenucci, RM; de Carvalho Dekon, SF. Quality of life and satisfaction of patients wearing implant-supported fixed partial denture: a cross-sectional survey of patients from Aracatuba city, Brazil. Clin Oral Implants Res 2015;26:701-708.
  • 25
    Papavasiliou, G; Kamposiora, P; Bayne, SC; Felton, DA. Three-dimensional finite element analysis of stress-distribution around single tooth implants as a function of bony support, prosthesis type, and loading during function. J Prosthet Dent 1996;76:633-640.

Publication Dates

  • Publication in this collection
    Jan-Feb 2018

History

  • Received
    11 Jan 2017
  • Accepted
    11 Sept 2017