First results on the fast neutral lithium beam diagnostics probing the edge plasma of the ETE tokamak

Abstract

One of the main objectives of the research program at ETE is devoted to the study of confinement properties in a low aspect ratio tokamak plasma configuration, and to attain this purpose the determination of plasma parameters of the edge region plays an essential role. A Fast Neutral Lithium Beam (FNLB) diagnostics was developed to perform the measurement of the density profile and its fluctuations at the edge of the plasma in ETE. The first light signals were collected after the filling of ETE chamber with He gas (10-4 Torr), where the optimization of this diagnostics was done, to adjust the better point for focalization of the beam and for the optical detection system setup. An extensive study of the variation of the intensity of the light signal with the variation of the gas pressure was carried out. This calibration will be used during shots in ETE to measure the rapidly varying pressure of the pulsed discharge by FNLB technique, a requirement imposed by the use of puff valves for injection of the gas. This paper will also describe the several phases of the development of this diagnostics, discuss the problems occurred and the adopted solutions and show the first measurements of the probing of the edge plasma of ETE with FNLB.


First results on the fast neutral lithium beam diagnostics probing the edge plasma of the ETE tokamak

Rogério Moraes Oliveira; Mário Ueda; Luiz Ângelo Berni

Instituto Nacional de Pesquisas Espaciais - INPE. Laboratório Associado de Plasmas, 12227-010, São José dos Campos, SP, Brazil

ABSTRACT

One of the main objectives of the research program at ETE is devoted to the study of confinement properties in a low aspect ratio tokamak plasma configuration, and to attain this purpose the determination of plasma parameters of the edge region plays an essential role. A Fast Neutral Lithium Beam (FNLB) diagnostics was developed to perform the measurement of the density profile and its fluctuations at the edge of the plasma in ETE. The first light signals were collected after the filling of ETE chamber with He gas (10-4 Torr), where the optimization of this diagnostics was done, to adjust the better point for focalization of the beam and for the optical detection system setup. An extensive study of the variation of the intensity of the light signal with the variation of the gas pressure was carried out. This calibration will be used during shots in ETE to measure the rapidly varying pressure of the pulsed discharge by FNLB technique, a requirement imposed by the use of puff valves for injection of the gas. This paper will also describe the several phases of the development of this diagnostics, discuss the problems occurred and the adopted solutions and show the first measurements of the probing of the edge plasma of ETE with FNLB.

1 Introduction

Plasma boundary study in tokamaks became an important issue recently due to the consensus that many phenomena related to the plasma core are correlated with the conditions of the plasma edge [1-3]. However, this region is characterized by steep gradients of the plasma parameters [4] and by high contamination due to plasma-wall interaction [5], being difficult to be diagnosed. Hence, to conveniently probe the edge plasma, the chosen diagnostics system should fulfill the requirements of allowing high spatial and temporal resolution during measurements, no contamination of the plasma and low disturbance on the discharge. These characteristics, that are inherent to the Fast Neutral Lithium Beam Probes (FNLBP), disseminated the use of this diagnostics to the edge density measurement purposes in some tokamaks and helical systems with various device sizes [6-11] around the world.

A 60keV mono-energetic beam of neutral Li atoms was used to probe the top edge region of the JET plasma. There, the radiation emitted from a 17cm long section of the beam path is sampled employing an array of 50 optical fibers and fast CCD cameras. Plasma background radiation is separated from the resonance beam emission line by a low resolution spectrometer. Recently, a 30keV lithium beam probe system was used to measure the edge density and its fluctuation in discharges of the LHD device. For the photon detection from the Li atoms, a thermoelectric cooled CCD camera is used, and a 1nm band-pass filter is installed between the camera lens and the CCD chip. In this experiment, time resolution, i.e., exposure time of CCD is 10-100ms. Innovative two-dimensional neutral lithium beam probe was designed and installed on the CHS device [1]. In this case the injected beam line is mechanically tilted to cover the edge region of the plasma two-dimensionally. Light emission from the 10-20keV injected beam is collected by an eight-channel detector system using optical band-pass filters and avalanche photodiodes. High energy Li beam (60keV-100keV) was used to probe the plasma of the ASDEX tokamak. Interference filter and photomultiplier were used for the collection of light emitted from the lithium resonance line. At TEXTOR, thermal Li and He beams and fast neutral lithium beam probed the edge plasma region. Both, interference filter - photomultiplier and/or interference filter - CCD camera combinations, were used to monitoring the line emission intensity along the injected beam. Separation of Li beam induced line emission from the plasma background radiation is made via beam modulation.

In the particular case of the ETE tokamak [12] (major radius R=0.30m, aspect ratio A = 1.5, designed plasma current of 200KA with a toroidal magnetic field up to 0.4T), a FNLB was designed to probe its edge region. The beam is operated at an energy of 15keV with a neutralization efficiency of 80%. The lithium ion source is a vitreous b-eucryptite [13,14]. The principle of the determination of the desnsity lies in the fact that the intensity of the Li resonance line (6708 Å) emitted from the probing beam due to the interaction with plasma electrons and ions is proportional to the local electron density [15]. The one-channel optical detection system being used is composed of camera lens, optical fiber, interference filter (FW H M = 100Å) and photomultiplier tube.

In a comparison to other FNLB devices previously cited, main advantages in favor of the FNLB device installed in ETE are the good spatial resolution, of less than 1cm, due to the relatively low energy of the acceleration voltage of the injected beam, and excellent time resolution during measurements, of the order of 1ms, due to the use of photomultiplier tube. When high energy beams are used, as is the case in JET, ASDEX, LHD, the plasma can be probed more deeply but with the disadvantage of worst spatial resolution. For the case of a 60keV acceleration voltage for example, spatial resolution is of the order of 3cm. On the other hand, modulation of the beam itself limits the achievable time resolution, that is not better than 2-3ms, as it happens at LHD, CHS, TEXTOR. CCD cameras allow excellent spatial resolution in comparison to optical fiber setup (specially fiber bundle), but due to their limited frequency of data acquisition, typically 100 times higher than photomultiplier tubes, time resolution becomes compromised, excluding here their elevate cost.

This paper describes details of the design of the FNLB device installed on ETE tokamak, the calibration process and the first measurements on the edge plasma.

2 Experimental Setup

The experimental setup of the FNLB diagnostics may be divided into 3 main parts: the ion gun, the neutralization chamber and the optical detection system.

Li ions are emitted through the heating of b-eucryptite compound (Li2O.Al2O3.2SiO2) to temperature over 1100ºC in a self heated molybdenum crucible of 0.6mm diameter and 2mm depth. This crucible is housed in a Pierce gun, being submitted to a potencial of 15kV. Through the application of adequate potentials to tree electrostatic cylindrical lenses, the beam is accelerated and focused. The 3 high voltage DC power supplies being used for this purpose are connected to the cylindrical lenses through special resistors in order to limit the current delivered in the case of sparks. The heating of the crucible is made via AC voltage, where a 20kV insulation transformer is used to insulate the high voltage region from the electric distribution system. Small corrections on the trajectory of the beam may be done by polarizing the deflection plates situated straight in front of the cylindrical lenses, as it can be shown in Fig. 1.


The neutralization system, situated at a distance of 70cm from the gun, is composed of a neutralizer cell that accommodates small pieces of metallic sodium, a 12cm pass through tube and a pneumatic valve that seals the neutralizer cell. When heated to temperature above 220ºC the sodium is vaporized, filling the pass through tube when the valve is opened. Through charge exchange, the Li beam is neutralized. In order to avoid the migration of sodium to the ETE chamber and to the ion gun, two collimators with apertures of 2.5cm diameter are placed at the entrance tube and at the end of the tube of the neutralizer chamber. The neutralizer housing is also separated from the ion gun and from the tokamak window by two pneumatic gate valves, that are opened for a few seconds just when the discharge is to be initiated, as it occurs also with the pneumatic valve. Differential pumping is guaranteed by two strategically located turbomolecular pumps.

The optical detection system is composed of an adjustable camera lens focused onto the tip of a 7m fiber bundle (4 rows of 22 fibers x 3 rows of 21 fibers, f =5mm), connected to a plano-convex lens, interference filter (FW H M = 10nm) and to a photomultiplier tube. The output signal, presently acquired in a digital oscilloscope, will be captured by the data acquisition system of ETE [16], based on VME bus.

3 Beam Characterization

In order to be possible the characterization of the beam, a set of current detectors were disposed in the FNLB vacuum chamber.

The monitoring of the ion current density was performed with the aid of tree secondary electron detectors (SED1, SED2 and SED3). These SED's were positioned respectively at the end of the ion gun tube, at the neutralizer region and at the end of the flight tube of the vacuum chamber (near the entrance window of the tokamak), playing distinct roles. The maximum ion current that could be extracted from the gun was monitored by SED1. Through the variation of the potencial applied to the electrostatic lenses and the variation in the distance among them, a maximum ion current of 1mA was detected by SED1. In this case the variation of the potencial of the electrostatic lens responsible for focusing the beam did not interfere on the current detected by SED1 due to its proximity to the ion gun. Otherwise, the effects of divergence of the beam could be monitored in SED2, where the polarization of the focusing lens was indeed significant, as can be seen in the graph of Fig. 2.


In Fig. 2, it is possible to observe that a maximum current of 42 mA was detected by SED2 against an ion current detected by SED1 of 240 mA. This six fold decrease on the current detected is mainly due to the divergence of the beam, caused by space charge effects.

At the entrance window of the tokamak, the ion current density detected by SED3 is of 30mA/cm2. This is approximately the value of the ion current density of the beam that will probe the edge plasma in the tokamak.

Other type of current detector, made with a thin cupper rod, was used to measure the beam profile at the neutralizer region, that is strongly dependent on the divergence of the beam. The polarization of the lenses of the ion gun is adjustable in accordance with the beam profile measurement.

A typical measurement of the beam profile at the neutralizer region, in which the FWHM is equal to 2cm is shown in Fig. 3.


4 Neutralization efficiency measurement

One of the most important characteristics of FNLB diagnostics systems is related with the rate of the neutralization of the Li ions, being it strongly dependent on the energy of the beam.

The neutralization efficiency measurement for the 15KeV lithium beam in ETE was performed in two stages. As it is shown in the experimental scheme of Fig. 4, a strong electric field (200V/cm) produced by the polarization of deflection plates, totally deviates the ion beam to the secondary electron detector 3 (SED 3). When the neutralizer is on, the rate of the ion current detected by SED 3 decreases, accordingly to the increase on the temperature of the neutralizer cell. In fact, the current ID detected by SED 3 is equal to : ID = + g. Here is the beam ion current and g is the secondary electron emission coefficient. In the same way, the current detected in SED 2, after the neutralization, is equal to: ID = g, where is the beam neutral equivalent current. The rate of the currents detected by SED 3 and SED 2 is proportional to the neutralization efficiency. However, the value of g must be known in order to be possible to achieve the neutralization efficiency. The second part of the experiment is devoted to the measurement of the value of g for the beam energy of 15KeV. Now, a Faraday cup (F.C.) situated about 120cm from the neutralizer, being composed of an outer cylinder connected to the ground, a collector plate and an inner cylinder serving as a polarized grid, measures the ion current and equivalent neutral beam current passing through the 40mm diameter entrance aperture. When the neutralizer is off and the grid of the F.C. is biased at a negative voltage sufficient to suppress secondary electrons, the detected current ID is equal to . On the other hand, when the grid is biased by a positive voltage, the detector registers the current of secondary electrons released from the collector of g as well as the ion current .


The value of g = 2 was determined by comparing the beam ion current detected by the F.C. for both positive and negative polarization of the grid.

In Fig. 5 it is possible to observe the variation of the neutralization efficiency with the heating temperature of the neutralizer cell. It can be seen that a maximum of 80% of neutralization efficiency was obtained for sodium temperature of about 300ºC. The decay of the neutralization efficiency for temperature above this value is due to the excess of sodium vapor produced, that influence the ion emission.


5 The influence of stray magnetic field

In the routine operation of ETE a stray magnetic field in the poloidal and toroidal directions of about 10Gauss and 4Gauss were measured in the region of the neutralizer, respectively.

The resultant ion Larmor radius is equal to 120m for the magnetic field of 10Gauss, according to equation 1.

It implies a horizontal deviation of 6º on the trajectory of the beam until it reaches the optical detection window. Similar calculation leads to an ion Larmor radius of 300m for the spurious toroidal magnetic field case, that causes the deviation of the beam in the vertical direction of 2.4º. These deviations were observed by the change on the intensity of the light detected by the photomultiplier output signal, as can be seen in Figs. 6 and 7 below.



The increase on the intensity of the photomultiplier output signal observed in Fig. 6 is due to the upward deflection of the beam. In this case the light emitted from the beam is shifted to a position closer to the optical window, causing an increase on the signal detected. On the other hand the horizontal deflection of the beam due to the influence of the spurious poloidal field removes partially the emitted light from the observation cone of the optical detection system, causing the reduction of the intensity of the signal, as can be seen in Fig. 7.

Independent tests were made with cylinders of similar diameters to the ones used on the FNLB chamber, with permanent magnet in order to produce magnetic field varying from 10 to 100Gauss. An external shielding consisting of a 1mm thick cylinder of m-metal caused a fourfold decrease to the measured field in the shielded region.

Based on this test, the ion source chamber, the flight tubes and the neutralizer chamber were involved by a 1mm thick cylinder of m-metal. This led to a reduced vertical and horizontal deflections on the beam trajectory to 1.5º and 1º, respectively, with reduction of the variation of the output signal of the photomultiplier. These deviations are small enough not to interfere on the measurement process performed by this diagnostics.

6 Calibration process

First light signals were collected after the filling of ETE chamber with He gas and consequent interaction of the lithium atoms with neutral particles. A sweep around the lithium resonance line (6708Å) was performed in order to verify the distribution of the intensities of wavelengths detected by the optical detection system. In Fig. 8 it is possible to observe that the wavelength distribution is centered at 6708Å , in which wavelength the maximum intensity of the signal occurs, as it was expected.


The second step performed in the calibration process was done by varying the pressure inside the chamber and verifying the variation of the intensity of the output signal of the photomultiplier. The range of variation of the intensity of the light signal collected in response to the variation of the gas pressure begins at 4 × 10-6Torr, reaching pressures as high as 3 × 10-3Torr. As may be seen in Fig. 9, the maximum intensity of the signal occurred for p = 1 × 10-3Torr. The excess of collisions of lithium atoms with neutral particles existing for higher values of the pressure causes the attenuation of the beam, with consequent reduction of the intensity of the emission of light. Notice that pressure of operation in ETE is in the range of 10-5 to 10-4Torr. This fact promotes the FNLB as an excellent pressure meter for ETE discharges, since it makes it possible to measure the fast varying pressure in ETE chamber when puff valves are used for the injection of gas. In fact, several measurements with FNLB were performed by injecting gas through puff valves into ETE chamber. A typical signal showing the exact instant of the fast input of gas into the chamber and the consequent reduction of the pressure due to the vacuum pumping may be seen in Fig. 10.



7 Measurements at the plasma

The intense background radiation emitted from the plasma, mainly due to Ha emission, is just partially cut off by the optical detection system, being this noise level several times higher than the emitted line radiation (6708Å) of interest. In order, to measure the correct plasma density we must perform two sequential shots, while maintaining plasma conditions. The first one is normally made without the application of the beam, just to allow the measurement of the noise level. The difference on the intensity of both signals (with and without the application of the beam) is proportional to the electron density of the plasma under the following conditions: plasma temperature Te < 100eV and low attenuation of the Li beam. These conditions have been observed at the outer window position (R = 48cm) and for the present status of development of ETE [17] (the energy delivered by the capacitor banks represents approximately of the total energy expected for the initial stage of operation of ETE). In this radial position, the temperature measured by Thomson Scattering diagnostics remained always below 30eV [18]. The principle of density determination used for this situation was based on the comparison of the photon flux profiles emitted from the lithium atoms due to interaction with plasma electrons Nvp and due to the interaction with hydrogen gas Nvg under same discharge conditions [19]. Here, the electron density is deduced from the relation described in equation 2 [6].

where: Qg is the cross section for interaction of lithium atoms with molecular gas, Qp is the effective cross section for collision of lithium atoms with the electrons of the plasma and ng is the number density of the gas.

In the first instance the density of a glow discharge using He gas at 2 × 10-4Torr was measured. In this case the continuous intensity of the signal collected when lithium beam was injected into the discharge was 90% higher than the signal measured without the injection of the beam. From known cross section data [15] and calculation of the gas number density ng, the obtained plasma density was of 2 × 1011particles/cm3.

The time evolution of the plasma density for typical ETE discharges was also obtained for R = 48cm, using the same method described above. The graph of Fig. 11 shows the intensity of the signal collected by the photomultiplier when the beam is injected into the plasma against the background level. The resultant density profile is shown in Fig. 12.



8 Conclusion

Several phases of the development of the FNLB diagnostics were overcame. First light signals were collected after the filling of ETE chamber with He gas at 10-4Torr, being possible the optimization of the diagnostics and the adjustment of the optical detection system setup. The measurement of the fast varying pressure during injection of gas in ETE chamber by puff valves was obtained. A 15keV lithium beam of 2cm diameter with equivalent current of 30mA/cm2, 80% neutralized, is being used to probe the edge plasma of the ETE tokamak. Even though we realize that some important modifications must be done to improve the measurements, the first results showed to be very satisfactory. The main modification to be implemented is the increase of the channels for optical detection, which will allow the measurement of the density for other radial positions. The interference filter being used of 10nm pass-band will also be substituted by a 3nm pass-band one. This will allow a higher signal to noise ratio during measurements.

In summary, a neutral lithium beam probe system for measurement of the time evolution of the plasma density in ETE has been developed. The measurement of the temporal evolution of the density of the plasma edge with FNLB, of the order of 1012cm-3, must be confronted with data from Langmuir probe for the region near the limiter and with data from Thomson Scattering for the inner region. The preparations for this experiment is underway.

Received on 2 February, 2004; revised version received on 30 April, 2004

  • [1] K. Nakamura, H. Iguchi, M. Ueda, T. Morisaki, M. Isobe, K. Ida, C. Takahashi, S. Okamura, and K. Matsuoka, Journal of Nuclear Materials, 725 (2003).
  • [2] M. Ueda et al., Journal of Nuclear Materials, 923 (1992).
  • [3] P. Varela and M. E. Manso, Rev. Sci. Instrum. 74, 1493 (2003).
  • [4] W. M. Stacey, Fusion Technology 36, 38 (1999).
  • [5] E. Hintz, Transactions of Fusion Technology 25, 333 (1994).
  • [6] H. Iguchi, K. Kadota, K. Takasugi, T. Shoji, M. Hosokawa, M. Fujiwara, and H. Ikegami, Rev. Sci. Instrum. 56, 1050 (1985).
  • [7] A. Komori, O. Mitarai, K. Yamagiwa, C. Honda, K. Kadota, and J. Fujita, Nucl. Fusion 28, 1460 (1988).
  • [8] D. M. Thomas, Rev. Sci. Instrum. 66, 806 (1995).
  • [9] T. Morisaki, Plasma Phys. Controlled Fusion 37, 787 (1995).
  • [10] T. Morisak, A. Komori, O. Motojima and LHD Experimental Group, Rev. Sci. Instrum. 74, 1865 (2003).
  • [11] K. McCormick, M. Kick, ''The ASDEX 100 keV Neutral Lithium Beam Diagnostic Gun'', IPP Report IP III/85, Max-Planck Institute, Garching (1983).
  • [12] G. O. Ludwig, E. Del Bosco, J.G. Ferreira, L.A. Berni, R.M. Oliveira, M.C.R. Andrade, C.S. Shibata, M. Ueda, L.F.W. Barbosa, J.J. Barroso, P.J. Castro, and H. Patire Jr, Braz. J. Phys. 33, 848 (2003).
  • [13] M. Ueda, R. R. Silva, R. M. Oliveira, H. Iguchi, J. Fujita, and K. Kadota, J. Phys. D: Appl. Phys. 30, 2711 (1997).
  • [14] R. M. Oliveira, J. A. N. Gonçalves, M. Ueda, and A. F. Beloto, Japanese Journal of Applied Physics 43, 1154 (2004).
  • [15] K. Kadota, K. Tsuchida, Y. Kawasumi, and J. Fujita, Plasma Physics 20, 1011 (1978).
  • [16] R. M. Oliveira, L. F. W. Barbosa, J. G. Ferreira, and C. S. Shibata, Fusion Engineering and Design 60, 403 (2002).
  • [17] L. A. Berni, E. Del Bosco, R. M. Oliveira, and M. P. Alonso, Braz. J. Phys. 34, 1572 (2004).
  • [18] L. A. Berni, M. Ueda, E. Del Bosco, J. G. Ferreira, R. M. Oliveira, and W. A. Vilela, Rev. Sci. Instrum. 74, 1200 (2003).
  • [19] R. M. Oliveira, M. Ueda, and W. A. Vilela, Braz. J. Phys. 32, 76 (2002).