Acessibilidade / Reportar erro

Neural Recovery Function of the Auditory Nerve in Cochlear Implant Surgery: Comparison between Prelingual and Postlingual Patients

Abstract

Introduction

Cochlear implants (CIs) enable objective measures of the neural function in implanted patients through the measurements of the neural response telemetry (NRT) and of the Auditory nerve Recovery Function (REC). These measurements help in programming the speech processor and understanding the auditory system.

Objective

To compare the NRT and the REC in prelingual and postlingual implanted patients.

Methods

An observational, descriptive and prospective study was carried out. The NRT and the REC (through the T0, A, and tau parameters) were evaluated in individuals submitted to CI surgery, who were divided into two groups: prelingual and postlingual patients.

Results

In total, 46 patients were evaluated. Data analysis showed no statistically significant difference between the NRT measurements and the T0, A, and Tau of the REC in the comparison between the two groups, except for the NRT in the basal cochlear region.

Conclusion

There was no statistically significant difference in the REC in pre- and postlingual patients.

Keywords:
cochlear implants; prelingual hearing loss; postlingual hearing loss; neural response telemetry

Fundação Otorrinolaringologia R. Teodoro Sampaio, 483, 05405-000 São Paulo/SP Brasil, Tel.: (55 11) 3068-9855, Fax: (55 11) 3079-6769 - São Paulo - SP - Brazil