Analysis of gap formation at tooth-composite resin interface: effect of C-factor and light-curing protocol

Abstract

OBJECTIVE: The aim of this study was to evaluate the effect of C-factor and light-curing protocol on gap formation in composite resin restorations. Material and METHODS: Cylindrical cavities with 5.0 mm diameter and three different depths (A=1.0, B=2.0 and C=3.0 mm) were prepared on the occlusal surface of 30 human molars and restored in a single increment with P 60. The composite resin was light-cured according to two protocols: standard - 850 mW/cm² / 20 s and gradual - 100 up to 1000 mW/cm² / 10 s + 1000 mW/cm² / 10 s. After storage in distilled water (37°C/7 days), the restorations were cut into three slices in a buccolingual direction and the gap widths were analyzed using a 3D-scanning system. The data were submitted to ANOVA and Student-Newman-Keuls test (alpha=0.05). RESULTS: ANOVA detected a significant influence for the C-factor and light-curing protocol as independent factors, and for the double interaction C-factor vs. light-curing protocol. Cavities with higher C-factor presented the highest gap formation. The gradual light-curing protocol led to smaller gap formation at cavity interfaces. CONCLUSIONS: The findings of this study suggest that the C-factor played an essential role in gap formation. The gradual light-curing protocol may allow relaxation of composite resin restoration during polymerization reaction.

Composite resins; C-Factor; Light-curing protocol; Gap formation; Polymerization shrinkage


ORIGINAL ARTICLES

Analysis of gap formation at tooth-composite resin interface: effect of C-factor and light-curing protocol

Gustavo Oliveira dos SantosI; Antônio Henrique Monteiro da Fonseca Thomé da SilvaII; José Guilherme Antunes GuimarãesIII; Alexandre de Araújo Lima BarcellosIII; Eduardo Martins SampaioIV; Eduardo Moreira da SilvaIII

IMSc, Department of Restorative Dentistry, School of Dentistry, Federal Fluminense University, Niterói, RJ, Brazil

IIMSc, Department of Mechanical Engineering, School of Engineering, State University of Rio de Janeiro, Nova Friburgo, RJ, Brazil

IIIPhD, Associate Professor, Department of Restorative Dentistry, School of Dentistry, Federal Fluminense University, Niterói, RJ, Brazil

IVPhD, Associate Professor, Department of Mechanical Engineering, School of Engineering, State University of Rio de Janeiro, Nova Friburgo, RJ, Brazil

Corresponding address Corresponding address: Dr. Eduardo Moreira da Silva Universidade Federal Fluminense Faculdade de Odontologia Rua Presidente Pedreira, nº 97 / 904 24210-470, - Ingá, Niterói, RJ, Brasil Phone: 55 21 2629-9832 - Fax: 55 21 2622-5739 e-mail: emsilva@vm.uff.br

ABSTRACT

OBJECTIVE: The aim of this study was to evaluate the effect of C-factor and light-curing protocol on gap formation in composite resin restorations. Material and

METHODS: Cylindrical cavities with 5.0 mm diameter and three different depths (A=1.0, B=2.0 and C=3.0 mm) were prepared on the occlusal surface of 30 human molars and restored in a single increment with P 60. The composite resin was light-cured according to two protocols: standard - 850 mW/cm2 / 20 s and gradual - 100 up to 1000 mW/cm2 / 10 s + 1000 mW/cm2 / 10 s. After storage in distilled water (37°C/7 days), the restorations were cut into three slices in a buccolingual direction and the gap widths were analyzed using a 3D-scanning system. The data were submitted to ANOVA and Student-Newman-Keuls test (a=0.05).

RESULTS: ANOVA detected a significant influence for the C-factor and light-curing protocol as independent factors, and for the double interaction C-factor vs. light-curing protocol. Cavities with higher C-factor presented the highest gap formation. The gradual light-curing protocol led to smaller gap formation at cavity interfaces.

CONCLUSIONS: The findings of this study suggest that the C-factor played an essential role in gap formation. The gradual light-curing protocol may allow relaxation of composite resin restoration during polymerization reaction.

Uniterms: Composite resins; C-Factor; Light-curing protocol; Gap formation; Polymerization shrinkage.

INTRODUCTION

The volumetric contraction that accompanies resin-based composite (RBC) polymerization, typically to the order of 1.5-5%13, generates stress at the tooth-restoration interface and may lead to marginal gap formation, marginal discoloration, postoperative sensitivity and secondary caries15. These clinical consequences of polymerization shrinkage constitute the main reasons for replacing RBC restorations, and explain why polymerization shrinkage is regarded as the main limitation of current RBCs22.

Factors such as polymerization shrinkage, flexural modulus, flow capacity and adhesion to the cavity walls may influence the stress produced in RBC restorations22. Strategies to reduce these factors have been studied13. Davidson, et al.7 (1984), reported that cavity shape is of great importance in conserving the RBC-dentin bond. The shape is described by the configuration factor 'C-factor', which is defined as the ratio of the bonded to unbonded (free) restoration surfaces10. It was observed that in most clinically relevant cavity configurations, the stress relieving flow is not sufficient to preserve adhesion to dentin by dentin-bonding agents. Flow capacity may depend on the physical properties of RBCs3, C-factor and polymerization kinetics4.

Clinically, the light-curing protocol and the irradiance provided by the light source can influence the polymerization shrinkage stress developed in the material3. Activation with a high irradiance creates a rapid light-curing process, leading to higher shrinkage stress in the composite16,32. On the other hand, although activation with low irradiance may reduce the development of leakage at tooth-restorative material interface12,27, the degree of conversion and mechanical properties of composites can be affected14,21,26,28,31. Some recent studies have shown that the use of gradual polymerization, in which the composite is first submitted to a low light irradiance, followed by an increase of light intensity, can promote a decrease in shrinkage stress without interfering in the degree of conversion and mechanical properties of the material1,17,20,21,25.

The aim of this study was to investigate the influence of the C-factor and light-curing protocol on gap formation in RBC restorations, and to test a new adhesive interface assessment methodology. The research hypotheses were: (1) the higher the C-factor, the higher the gap formation and (2) the gradual light-curing protocol would reduce gap formation.

MATERIALS AND METHODS

A commercially available minifilled hybrid RBC was tested in this study (P60, 3M/ESPE, St. Paul, MN, USA). The composition of the materials used in the restorative procedures is described in Table 1. All specimens were light cured with a quartz-tungsten-halogen unit (Optilux 501; Kerr, Danbury, CT, USA). Two light-curing protocols were used: standard (S) – 850 mW/cm2 for 20 s (17 J/cm2) and gradual (G) – 100 to 1000 mW/cm2 for 10 s + 1000 mW/cm2 for 10 s (@17 J/cm2). The radiant exposure was calculated as the product of the curing unit irradiance measured with a radiometer (model 100; Demetron Inc., Danbury, CT, USA) by the irradiation time. For G protocol, the radiant exposure was obtained by the sum of mean irradiance within the first 10 s multiplied by 10 s with 10 J/cm2, corresponding to the radiant exposure in the last 10 s of light exposure.

Tooth Selection and Preparation

The steps to gap analysis are illustrated in Figure 1. Thirty human molars free of cracks and structural defects, stored in a 0.5% chloramine solution for 15 days and frozen in distilled water for less than 3 months were used in this study. All occlusal surfaces were wet ground in a polishing machine (DPU-10, Struers, Copenhagen, Denmark), with 150 and 600-grit SiC papers until flat dentin surfaces were obtained. The roots were embedded in polyester resin in PVC cylinders (0.5 inch in diameter) with the flat dentin surfaces parallel to the cylinder borders, and the teeth were randomly assigned to three groups (n = 10), according to cavity depth. Cylindrical cavities measuring 5.0 mm in diameter and with three different depths (A = 1.0, B = 2.0 and C = 3.0 mm) were prepared on the flat dentin surfaces with a diamond bur (# 4054, KG Sorensen, São Paulo, SP, Brazil) in a high-speed handpiece fixed in a special sample aligning device. Cavity depths were controlled by using a digital caliper (MPI/E-101, Mitutoyo, Tokyo, Japan). The C-factor was obtained using the following equation:


where r is the cavity radius and h is the cavity depth. Therefore, C-factor for each cavity depth was: A=1.8; B=2.6 and C=3.4.

Restorative Procedure

The cavities were bonded with Single Bond 2 adhesive system (3M/SPE, St. Paul, MN, USA) following manufacturers' instructions, and the composite resin was inserted in a single increment using a flat-sided instrument (Suprafill #1, SSWhite, Rio de Janeiro, RJ, Brazil). The composite resin was covered with a polyester strip and light cured, in accordance with the two light-curing protocols under study. Six experimental groups were formed according to C-factor and light curing protocol (n=5). After storage in distilled water at 37°C for 7 days, finishing and polishing procedures were performed with sequential Sof Lex discs (3M, St Paul., MN, USA).

The teeth were longitudinally sectioned in a buccolingual direction through the restorations with a diamond disk (KG Sorensen, São Paulo, SP, Brazil) and 3 slices were obtained for each restoration. The sectioned surfaces were polished with 600- and 1200-grit SiC abrasive paper (DPU-10, Struers, Copenhagen, Denmark) and ultrasonicated in distilled water for 5 min (Model 750 USC; Unique, Rio de Janeiro, RJ, Brazil).

Gap Measurement

Both sides of the polished surfaces were analyzed using a 3D-scanning system (Talyscan 150, Taylor Hobson, Leicester, England) with a space of 1µm in the scanning direction (x) and 60 µm in the direction (y) at a scanning speed of 1000 µm/s. The images obtained were leveled and roughness profiles of the tooth-restoration surfaces were obtained. Gap width was analyzed at 10 different positions for each slice floor.

Statistical Analysis

Statistical analysis was performed using Statgraphics 5.1 Software (Manugistics, Rockville, MD, USA). Data were analyzed by two-way ANOVA and one-way ANOVA with Student-Newman-Keuls test for multiple comparisons. All statistical analyses were performed at a significance level of a=0.05.

RESULTS

Two-way ANOVA detected a significant influence for the C-factor and light-curing protocol as independent factors (p<0.0001), as well as for the double interaction C-factor vs. light-curing protocol (p<0.0001). The G light-curing protocol yielded the smallest gap formation (Figure 2). Student-Newman-Keuls test showed that the cavities with C-factor = 3.4 had a higher gap formation than those with C-factors of 2.6 and 1.8 (p<0.0001), which did not differ significantly to each other (p>0.05) (Figure 3). Representative photomicrographs of samples are presented in Figure 4.




DISCUSSION

RBC polymerization reaction involves the conversion of C=C bonds in individual monomer molecules and the formation of C-C bonds to form polymer chains, causing volume reduction, as covalent bonds are created and molecular distances and free volume are reduced13. The shrinkage generated during this process may cause debonding at tooth cavity-RBC interface, as well as fractures of the enamel prisms along the cavosurface margin, thus leading to clinical failure of the restoration15. The polymerization reaction involves three phases: pre-gel, gel and post-gel. In the pre-gel phase, the composite presents a viscous behavior and shrinkage stresses can be released by the material flow relaxation2,6,7,9. Polymer chains are distributed in a linear mode and have mobility that permits tensions induced by polymerization shrinkage to be dissipated by flowing.13 As the reaction progresses, the post-gel phase starts the first cross links between chains, making flow difficult and simultaneously promoting an increase in mechanical properties and flexural modulus, which involve inducing stresses in the restoration13.

Polymerization shrinkage stresses may increase as light intensity generated by the light source increases30. Versluis, et al.29 (1994), reported that the gel point of RBCs that were light cured with a standard light-curing protocol was reached 1.5-2 s after the polymerization reaction started. Moreover, a previous study demonstrated that the highest stress development associated with polymerization shrinkage occurs during the first 30 to 40 s of light irradiation30. Based on this, several studies8,11,12 have proposed the use of protocols that generated low initial irradiance for light-curing RBCs. A slower curing process can extend the pre-gel phase, allowing relief of polymerization shrinkage stress11,12. In the present study, the G light-curing protocol presented lower gap formation at the tooth-P60 interfaces than S protocol (Figure 2). This is certainly related to the low initial irradiance (100 up to 1000 mW/cm2 for10 s) provided by G protocol. In agreement with these findings, Feilzer, et al.12 (1995), demonstrated that light curing with low irradiance was related to better marginal sealing of cavities restored with light-cured RBCs. Some previous studies have shown that light curing with low irradiance may decrease the degree of conversion and influence negatively the mechanical properties of RBCs24,27,31. In spite of the differences in light-curing protocols, the final radiant exposure used in this study for both techniques was approximately the same (17 J/cm). According to Rueggeberg, et al.24 (1993), this radiant exposure would be adequate to light cure RBCs. Based on this, it may be assumed that gradual protocols that generate at least 17 J/cm2 could be indicated for polymerizing RBCs.

The influence of the confinement conditions imposed on the composite resin (usually expressed as the bonded to unbonded ratio, known as C-factor) plays an essential role in gap formation at cavity-composite interfaces. According to Feilzer, et al.10 (1987), most clinical restorations have C-factor values of approximately 1 to 2. Improvements in adhesive systems and composite resins have, however, encouraged dental practitioners to build deeper restorations that have a high C-factor. As a result, a cavity with C-factor of 3.4 was introduced in the present study, to simulate a more realistic clinical situation. Braga, et al.4 (2006), analyzing the influence of cavity dimensions on shrinkage stress and microleakage in composite restorations, showed that cavity depth had a stronger influence on both responses than diameter. Based on this, the cavity diameters in the present study were kept constant (5.0 mm) and the C-factor was varied as a function of cavity depths (1.0, 2.0 and 3.0 mm).

Gap formation was higher in cavities with C-factor = 3.4 than in those with 2.6 and 1.8 C-factors. Moreover, all cavities with C-factor = 3.4 presented a gap at the tooth-composite resin interface (Figure 4). On the other hand, in the groups with C-factor 1.8 and 2.6, only slices from one cavity per group showed gap formation at the tooth-composite resin interface. From the clinical point of view, this finding is extremely important because the absence of gap would increase the longevity of composite resin restorations. Interpretation of these results may be based on the fact that in cavities with C-factor 1.8 and 2.6, the composite relaxation provided by the unbonded surface, which was the same for both cavities, was more efficient for relieving shrinkage stress generated during the polymerization reaction10,18. On the other hand, the greater depth of the cavities with C-factor 3.4 increased the bonded interface. Thus, the wall-to-wall shrinkage was increased and so was the gap formation. These results agree with previous studies. Yoshikawa, et al.32 (2001), found that cavity-wall gap formation significantly increased when the C-factor increased from 2.3 to 3, and concluded that a C-factor of 2.3 can be considered low. Furthermore, Loguercio, et al.19 (2004), showed that the linear polymerization shrinkage and the gap width were higher when the C-factor increased from 0.3 to 3.0. An important aspect was that these authors measured the linear polymerization shrinkage of the composite resin inside the restored cavities. In addition, some previous studies have shown that the increase in the C-factor also has a harmful effect on the bond strength of adhesive systems to dentin19,23.

Figure 4 shows representative photomicrographs of specimens. In (c), great gap formation at the internal angles in a 3-mm-deep cavity can be seen. This feature was observed in all specimens that presented gap formation at the tooth-P60 interface. This finding could be explained by the high shrinkage stress concentration generated by the wall-to-wall bonding competition in these areas, i.e. floor vs. axial wall cavity4. On the other hand, it can be seen that in the cavosurface region the sealing was more satisfactory. Probably, the relaxation of P60 free surface could have allowed more stress relief in these areas, allowing the adhesive system to support the debonding phenomenon6. These findings explain why in the present study the gap measurement was analyzed only on the cavity floor.

Based on the obtained results, it may be assumed that gap formation is a multifactorial phenomenon, dependent on several factors22 related to the restorative material, C-factor4,6,10,32 and light-curing protocol12,20,25-28. Furthermore, it is also important to study other factors, such as composite resin incremental technique and the use of lining materials13 in order to improve restoration sealing. Further studies should be conducted in order to yield better sealing of cavities restored with light-cured RBCs.

CONCLUSIONS

The results of this in vitro study supported the research hypotheses. It may be concluded that: 1. High C-factor values produced the highest gap formation; 2. The gradual light-curing protocol was effective in reducing gap formation.

ACKNOWLEDGEMENTS

This study was supported by the Adhesion and Adherence Laboratory (LAA), from the Department of Mechanical Engineering of the State University of Rio de Janeiro, Brazil. The authors would like to thank 3M/ESPE for supplying the materials used in this study.

Received: July 21, 2006

Modification: March 09, 2007

Accepted: May 25, 2007

  • 1- Barros GK, Aguiar FH, Santos AJ, Lovadino JR. Effect of different intensity light curing modes on microleakage of two resin composite restorations. Oper Dent. 2003;28:642-6.
  • 2- Bausch JR, de Lange K, Davidson CL, Peters A, de Gee AJ. Clinical significance of polymerization shrinkage of composite resins. J Prosthet Dent. 1982;48:59-67.
  • 3- Bouschlicher MR, Rueggeberg FA, Boyer DB. Effect of stepped light intensity on polymerization force and conversion in a photoactivated composite. J Esthet Dent. 2000;12:23-32.
  • 4- Braga RR, Boaro LC, Kuroe T, Azevedo CL, Singer JM. Influence of cavity dimensions and their derivatives (volume and 'C' factor) on shrinkage stress development and microleakage of composite restorations. Dent Mater. 2006;22:818-23.
  • 5- Carvalho RM, Pereira JC, Yoshiyama M, Pashley DH. A review of polymerization contraction: the influence of stress development versus stress relief. Oper Dent. 1996;21:17-24.
  • 6- Davidson CL, de Gee AJ. Relaxation of polymerization contraction stresses by flow in dental composites. J Dent Res. 1984;63:146-8.
  • 7- Davidson CL, de Gee AJ, Feilzer A. The competition between the composite-dentin bond strength and the polymerization contraction stress. J Dent Res. 1984;63:1396-9.
  • 8- Davidson CL, Feilzer AJ. Polymerization shrinkage and polymerization shrinkage stress in polymer-based restoratives. J Dent. 1997;25:435-40.
  • 9- Emami N, Soderholm KJ, Berglund LA. Effect of light power density variations on bulk curing properties of dental composites. J Dent. 2003;31:189-96.
  • 10- Feilzer AJ, de Gee AJ, Davidson CL. Setting stress in composite resin in relation to configuration of the restoration. J Dent Res. 1987;66:1636-9.
  • 11- Feilzer AJ, de Gee AJ, Davidson CL. Setting stresses in composites for two different curing modes. Dent Mater. 1993;9:2-5.
  • 12- Feilzer AJ, Dooren LH, de Gee AJ, Davidson CL. Influence of light intensity on polymerization shrinkage and integrity of the restoration-cavity interface. Eur J Oral Sci. 1995;103:322-6.
  • 13- Ferracane JL. Developing a more complete understanding of stresses produced in dental composites during polymerization. Dent Mater. 2005;21:36-42.
  • 14- Halvorson RH, Erickson RL, Davidson CL. Energy dependent polymerization of resin-based composite. Dent Mater. 2002;18:463-9.
  • 15- Hickel R, Manhart J. Longevity of restorations in posterior teeth and reasons for failure. J Adhes Dent. 2001;3:45-64.
  • 16- Jain P, Pershing A. Depth of cure and microleakage with high-intensity and ramped resin-based composite curing lights. J Am Dent Assoc. 2003;134:1215-23.
  • 17- Koran P, Kurschner R. Effect of sequential versus continuous irradiation of a light-cured resin composite on shrinkage, viscosity, adhesion and degree of polymerization. Am J Dent. 1998;11:17-22.
  • 18- Li Y, Swartz ML, Phillips RW, Moore BK, Roberts TA. Effect of filler content and size on properties of composites. J Dent Res. 1985;64:1396-401.
  • 19- Loguercio AD, Reis A, Ballester RY. Polymerization shrinkage: effects of constraint and filling technique in composite restorations. Dent Mater. 2004;20:236-43.
  • 20- Mehl A, Hickel R, Kunzelmann KH. Physical properties and gap formation of light-cured composites with and without 'softstart-polymerization'. J Dent. 1997;25:321-30.
  • 21- Muangmingsuk A, Senawongse P, Yudhasaraprasithi S. Influence of different softstart polymerization techniques on marginal adaptation of Class V restorations. Am J Dent. 2003;16:117-9.
  • 22- Peutzfeldt A, Asmussen E. Determinants of in vitro gap formation of resin composites. J Dent. 2004;32:109-15.
  • 23- Price RB, Dérand T, Andreou P, Murphy D. The effect of two configuration factors, time, and thermal cycling on resin to dentin bond strengths Biomaterials. 2003;24:1013-21.
  • 24- Rueggeberg FA, Caughman WF, Curtis JW Jr, Davis HC. Factors affecting cure at depths within light-activated resin composites. Am J Dent. 1993;6:91-5.
  • 25- Sakaguchi RL, Berge HX. Reduced light energy density decreases post-gel contraction while maintaining degree of conversion in composites. J Dent. 1998;26:695-700.
  • 26- Uno S, Asmussen E. Marginal adaptation of a restorative resin polymerized at a reduced rate. Scand J Dent Res. 1991;99:440-4.
  • 27- Unterbrink GL, Muessner R. Influence of light intensity on two restorative systems. J Dent. 1995;23:183-9.
  • 28- Vandewalle KS, Ferracane JL, Hilton TJ, Erickson RL, Sakaguchi RL. Effect of energy density on properties and marginal integrity of posterior resin composite restorations. Dent Mater. 2004;20:96-106.
  • 29- Verluis A, Sakagushi RL, Douglas WH. Stress development in composite resins during polymerization [abstract 1000]. J Dent Res. 1994;73:226.
  • 30- Watts DC, Cash AJ. Determination of polymerization shrinkage kinetics in visible-light-cured materials: methods development. Dent Mater. 1991;7:281-7.
  • 31- Yap AU, Seneviratne C. Influence of light energy density on effectiveness of composite cure. Oper Dent. 2001;26:460-6.
  • 32- Yoshikawa T, Burrow M.F, Tagami J. The effects of bonding system and light curing method on reducing stress of different C-factor cavities. J Adhes Dent. 2001;17:177-83.
  • Corresponding address:
    Dr. Eduardo Moreira da Silva
    Universidade Federal Fluminense
    Faculdade de Odontologia
    Rua Presidente Pedreira, nº 97 / 904
    24210-470, - Ingá, Niterói, RJ, Brasil
    Phone: 55 21 2629-9832 - Fax: 55 21 2622-5739
    e-mail:
  • Publication Dates

    • Publication in this collection
      06 Nov 2007
    • Date of issue
      Aug 2007

    History

    • Accepted
      25 May 2007
    • Reviewed
      09 Mar 2007
    • Received
      21 July 2006
    Faculdade De Odontologia De Bauru - USP Serviço de Biblioteca e Documentação FOB-USP, Al. Dr. Octávio Pinheiro Brisolla 9-75, 17012-901 Bauru SP Brasil, Tel.: +55 14 32358373 - Bauru - SP - Brazil
    E-mail: jaos@usp.br