Determination of propylthiouracil in pharmaceuticals by differential pulse voltammetry using a cathodically pretreated boron-doped diamond electrode

A simple procedure is described for the determination of propylthiouracil (PTU) by differential pulse voltammetry (DPV) using a cathodically pretreated boron-doped diamond (BDD) electrode. Cyclic voltammetry studies indicate that the oxidation of PTU is irreversible at a peak potential of 1.42 V (vs. Ag/AgCl (3.0 mol L-1 KCl)) in a Britton-Robinson (BR) buffer solution (pH 2.0). Under optimized conditions, the obtained analytical curve was linear (r = 0.9985) for the PTU concentration range of 1.0 to 29.1 mmol L-1 in a BR buffer solution (pH 2.0), with a detection limit of 0.90 mmol L-1. The proposed method was successfully applied in the determination of PTU in pharmaceutical samples, with results in agreement at a 95% confidence level with those obtained using an official titration method.

propylthiouracil determination; 6-propyl-2-thiouracil; BDD electrode; cathodic pretreatment; differential pulse voltammetry