Feasibility of Use of the Microwave Induced Plasma Atomic Emission Detector as a Compound Independent Detector for Quantitative Chromatographic Analysis

The independence of the response of a Microwave Induced Plasma Atomic Emission Detector (GC-MIP) for Gas Chromatography with the chemical structure of the analyte has been exploited using as teste samples nine organochlorine compounds. With signal monitoring at the chlorine 479.5 nm emission line and after proper adjustment of the operational conditions of the GC-MIP the agreement of the detector response for the nine compounds was of 8.5% or less, which is adequate for the use of the compound-independent calibration method. It is proposed that the reliability of the quantitation after applying the compound-independent calibration methodology may be enhanced if a set of calibration compounds is used instead of a single calibration compound.

microwave induced plasma; chlorine specific detection; compound-independent calibration method