Performance evaluation of the Sysmex® XE-2100D automated hematology analyzer

Avaliação do desempenho analítico do analisador hematológico Sysmex® XE-2100D

Introduction:

The Sysmex® XE-2100D is a multiparameter hematology analyzer designed for hematology testing in samples with ethylenediamine tetraacetic acid (EDTA).

Objectives:

Considering the importance of this hematology analyzer for clinical and laboratory practice, the objective of this study was to evaluate its analytical performance, comparing the obtained results with quality specifications described in literature.

Material and method:

In the evaluation of analytical performance, according to recommendations of the document H26-A2 of the Clinical and Laboratory Standards Institute (CLSI), intra-run imprecision, inter-run imprecision, linearity, carryover, autosampler evaluation, clinical sensitivity of the atypical lymphocytes flag (n = 400 samples) were included, as well as the comparison between automated and manual leukocyte differential count (n = 400 samples), based on an adaptation of the document H20-A2 of CLSI.

Results:

Repeatability, reproducibility, linearity and carryover were satisfactory according to the manufacturer's specifications. The clinical sensitivity of the atypical lymphocytes flag showed efficiency, sensitivity and specificity of 92.5%, 65.2% and 94.1% respectively. The correlation coefficients between the automated and manual differential counts of neutrophils, lymphocytes, monocytes, eosinophils and basophils were 0.991, 0.99, 0.872, 0.974 and 0.557, respectively.

Conclusions:

The results were in accordance with quality specifications described in literature, indicating reliability in Sysmex® XE-2100D. This fact ensures certainty to both laboratory professionals and medical staff. We conclude that the Sysmex® XE-2100D showed excellent analytical performance, and is useful to provide reliable hematology data.

automation; hematology analyzer; Sysmex® XE-2100d; analytical performance; quality control