Acessibilidade / Reportar erro

Development and mechanical properties of a locking T-plate

Desenvolvimento e as propriedades mecânicas de uma placa bloqueada em formato de T.

ABSTRACT:

This study aimed to develop a locking T-plate and to evaluate its mechanical properties in synthetic models. A titanium 2.7mm T-plate was designed with a shaft containing three locked screw holes and one dynamic compression hole, and a head with two locked screw holes. Forty T-shaped polyurethane blocks, and 20 T-plates were used for mechanical testing. Six bone-plate constructs were tested to failure, three in axial compression and three in cantilever bending. Fourteen bone-plate constructs were tested for failure in fatigue, seven in axial compression and seven in cantilever bending. In static testing higher values of axial compression test than cantilever bending test were observed for all variables. In axial compression fatigue testing all bone-plate constructs withstood 1,000,000 cycles. Four bone-plate constructs failure occurred before 1,000,000 cycles in cantilever bending fatigue testing. In conclusion, the locking T-plate tested has mechanical properties that offer greatest resistance to fracture under axial loading than bending forces.

INDEX TERMS:
Bone-plate; implant; fracture; mechanical study

Colégio Brasileiro de Patologia Animal - CBPA Pesquisa Veterinária Brasileira, Caixa Postal 74.591, 23890-000 Rio de Janeiro, RJ, Brasil, Tel./Fax: (55 21) 2682-1081 - Rio de Janeiro - RJ - Brazil