Determination of low levels of copper in natural waters by flame atomic absorption spectrometry using a SIA system with a poly(ethylenimine) column

This paper describes a sequential injection analysis (SIA) set-up coupled to a flame atomic absorption spectrometer (FAAS) to accomplish the determination of low concentrations of copper in drinking waters. Copper is first retained under neutral media in an on-line 29x1.6 mm column filled with poly(ethylenimine) immobilised on silica gel. The retained analyte is then eluted by flowing through the column 250 <FONT FACE=Symbol>m</font>L of a nitric acid solution. The selection of 3.85 ml of sample enabled to obtain a detection limit of 0.27 mug/L and a sampling rate of about 24 samples/h. There was a good agrement between the results of 12 samples furnished by the proposed procedure and by electrothermal atomic absorption spectrometry. Repeatability assessment gave a relative standard deviation of 1.3 % after ten replicate analysis of a sample containing about 70 mug/L in copper..

SIA/FAAS; copper; poly(ethylenimine)