Acessibilidade / Reportar erro

Development of a low cost device for chemiluminescence measurements

A simple and low cost device (ca. US$ 150) that comprises two photodiodes fixed in lab-made Perspex flow cell is proposed for chemiluminescence measurements. The characteristics of the device (large observation window and reduced thickness) allow maximizing the amount of the emitted radiation detected. A sensitivity improvement of ca. 50 % was observed by employing two photodiodes for signal measurements. The performance of the device was assessed by the oxidation of luminol by hydrogen peroxide, yielding a linear response within the range of 2.50 to 500 µmol L-1 H2O2. The detection limit was estimated as 0.8 µmol L-1 hydrogen peroxide which is comparable with those obtained by using equipments based on photomultipliers.

flow analysis; chemiluminescence; hydrogen peroxide; photodiode


Sociedade Brasileira de Química Secretaria Executiva, Av. Prof. Lineu Prestes, 748 - bloco 3 - Superior, 05508-000 São Paulo SP - Brazil, C.P. 26.037 - 05599-970, Tel.: +55 11 3032.2299, Fax: +55 11 3814.3602 - São Paulo - SP - Brazil