Acessibilidade / Reportar erro

Development of a titrator based on drop counting

TitDrop II, a practice low-cost titrator based on drop counting is proposed. A microcomputer counts drops in an opto-switch and it receives pH values from a commercial pHmeter with RS-232 serial port. The volume of drops must kept constant and the amount of base in each drop is obtained by standardization in the titrator. A linear behavior between height of reservoir of titrant and drop frequency was observed, but there is no influence of low height on drop volume. Acetic acidity in samples of vinager was compared with volumetric titration, and deviation between -2.6 to 3.9% was observed.

titrator; potentiometric titration; low-cost titrator


Sociedade Brasileira de Química Secretaria Executiva, Av. Prof. Lineu Prestes, 748 - bloco 3 - Superior, 05508-000 São Paulo SP - Brazil, C.P. 26.037 - 05599-970, Tel.: +55 11 3032.2299, Fax: +55 11 3814.3602 - São Paulo - SP - Brazil
E-mail: quimicanova@sbq.org.br