Survey on quality control of radiopharmaceutical dose calibrators in nuclear medicine units in the city of São Paulo, SP, Brazil

OBJECTIVE: To perform a survey on routine quality control tests of dose calibrators at nuclear medicine units in the city of São Paulo, SP, Brazil. To evaluate the accuracy of measurements of seven dose calibrators activities, utilizing sources of clinically significant radionuclides at the calibration laboratory of Instituto de Pesquisas Energéticas e Nucleares. MATERIALS AND METHODS: The survey on quality control of dose calibrators has been based on questionnaires answered by the nuclear medicine units. Seven dose calibrators, Ga-67, Tc-99m and Tl-201 sources and a secondary standard instrument have been utilized in the accuracy study. RESULTS: The survey results on the quality control tests of the dose calibrators showed some inappropriateness, for example, the absence of daily reproducibility tests in all of the units. The accuracy tests for the seven dose calibrators showed results within the acceptable limit in compliance with the national regulations (±10%). CONCLUSION: According to the few nuclear medicine units participating in the survey, the dose calibrators quality control is unsatisfactory. The accuracy study of seven dose calibrators has not demonstrated any performance faults, and has established the calibration of these instruments for the utilized sources.

Nuclear medicine; Dose calibrators; Calibration; Quality control


Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem Av. Paulista, 37 - 7º andar - conjunto 71, 01311-902 - São Paulo - SP, Tel.: +55 11 3372-4541, Fax: 3285-1690, Fax: +55 11 3285-1690 - São Paulo - SP - Brazil
E-mail: radiologiabrasileira@cbr.org.br