High dose-rate endobronchial brachytherapy: a dosimetric study

OBJECTIVE: To evaluate the dose distribution in different situations of high dose-rate endobronchial brachytherapy, focusing especially on high-dose volumes, and try to identify better or worse situations in terms of dose distribution to aid as guidance in the clinical practice. MATERIALS AND METHODS: Theoretical study simulating high dose-rate endobronchial brachytherapy utilizing two catheters, varying the loading extent, angle between the catheters, prescription depth, and source step. With a prescription dose of 7.5 Gy, the volumes involved by the 100%, 150% and 200% isodoses (V100, V150 and V200, respectively) and V150/V100 and V200/V100 ratios were calculated. RESULTS: There was a volume enhancement with larger loaded lengths, increase in prescription depth and angles, with a tendency towards a proportionally smaller increase with larger angulations. In general, the V150/V100 and V200/V100 ratios were homogeneous, respectively around 0.50 and 0.30. CONCLUSION: Overall, the dose distribution in the standard situation was appropriate. No specific parameter that could be related to a higher toxicity was identified. The authors recommend a swift evaluation of the treatment quality through the analysis of the V150/V100 and V200/V100 ratios.

Radiotherapy; Endobronchial brachytherapy; Dosimetry