Energy and angular dependence of thermoluminescent materials to beta monitoring

Energy and angular dependences of different thermoluminescent materials were studied with the objective to verify which type of detector would be the most appropriate for beta monitoring of workers. Three types of CaSO4:Dy + teflon pellets were studied. The energy dependence was evaluated using standard beta radiation sources (147Pm, 204Tl and 90Sr+90Y). For the angular dependence study, the pellets were exposed to beta radiation of the 90Sr+90Y source, varying the incidence angle between 0° and 90°. In relation to the studied characteristics, the CaSO4:Dy + 10% C dosimeters were the most adequate for use in beta monitoring of workers.

Thermoluminescent dosimetry; Beta radiation; Angular dependence; Energetic dependence