Analysis of the high dose rate brachytherapy protocols of quality assurance programs of some local services, based on TG40, TG56 and ARCAL XXX.

High dose rate brachytherapy has been increasingly recognized in most countries, and radiotherapy services using this equipment are encouraged to have a very efficient quality assurance program to ensure protection for patients, workers and other personnel involved. The objective of this paper was to determine the types of tests for high dose rate equipment required by official protocols (TG40, TG56 and ARCAL XXX) and to compare them with the types of tests utilized by some radiotherapy services. We concluded that: a) the protocol TG56 is more extensive and complete than the other official protocols (TG40 and ARCAL XXX); b) the protocols used by the services evaluated on this study were based on the protocol TG56, and were concordant with the other official protocols. In these protocols annual tests were frequently replaced by tests performed quarterly or twice a year. This study established the types of test used and their frequency of utilization, and permitted the design of an optimized protocol that may help in the implementation of basic and indispensable tests in order to ensure patient adequate treatment and safety to personnel involved, and consequently improve high dose rate brachytherapy quality assurance.

Dosimetry; Brachytherapy; High dose rate brachytherapy (HDR); Quality control