Respiratory mechanics of neurological patients undergoing mechanical ventilation under water heated humidifier and a heat exchanger filter model

Abstracts

OBJECTIVES: In mechanically ventilated patients, humidifier devices are used to heat and moisturize the inspired gas. Heating and humidifying inspired gas may prevent complications associated with the respiratory mucosa dryness such as mucus plugging and endotracheal tube occlusion. Two devices have been commonly used to this, either heated humidifier or the heat moisture exchange filter. This study aimed to compare the effects of the heated humidifier and a model of heat moisture exchange filter on respiratory mechanics in mechanically ventilated neurological patients. METHODS: This was a randomized crossover trial, involving 31 neurological patients under mechanical ventilation randomly assigned to the humidification devices. Expired tidal volume, peak inspiratory flow, peak expiratory flow, static compliance, dynamic compliance and respiratory system resistance were evaluated. Statistical analysis used the Kolmogorov-Smirnov test and Student's t test for paired samples, in which P values < 0.05 were considered significant. RESULTS: The heat moisture exchanger filter decreased expired tidal volume, peak inspiratory flow, peak expiratory flow (p < 0.001) and dynamic compliance (p = 0.002), and increased respiratory system resistance (p < 0.001). CONCLUSION: In the studied population, the use of a heat moisture exchange filter model leaded to several changes on respiratory mechanics parameters.

Mechanical ventilation; Artificial respiration; Mechanical ventilators; Intensive care units


OBJETIVOS: Em pacientes sob ventilação mecânica, dispositivos de umidificação são utilizados para aquecer e umidificar o gás inspirado. O aquecimento e umidificação do gás inspirado podem prevenir complicações associadas ao ressecamento da mucosa respiratória, como a formação de tampão mucoso e oclusão do tubo endotraqueal. Com esse objetivo, dois dispositivos têm sido comumente utilizados: os umidificadores aquosos aquecidos e os filtros trocadores de calor e umidade. O objetivo deste estudo foi comparar o efeito da utilização do umidificador aquoso aquecido e de um modelo de filtro trocador de calor e umidade sobre a mecânica respiratória de pacientes neurocríticos sob ventilação mecânica. MÉTODOS: Trata-se de um ensaio clínico, cruzado e randomizado, onde 31 pacientes neurocríticos sob ventilação mecânica foram submetidos de forma aleatória às duas formas de umidificação. Foram avaliados o volume corrente expirado, pico de fluxo inspiratório, pico de fluxo expiratório, complacência estática, complacência dinâmica e resistência do sistema respiratório. Para análise estatística dos resultados obtidos foram utilizados os testes de Kolmogorov-Smirnov e t-Student para amostras pareadas, considerando-se a significância estatística quando observado um valor de p < 0,05. RESULTADOS: A utilização de um modelo de filtro trocador de calor e umidade promoveu a redução do volume corrente expirado, pico de fluxo inspiratório, pico de fluxo expiratório (p < 0,001) e complacência dinâmica (p = 0,002), além do aumento da resistência do sistema respiratório (p < 0,0001). CONCLUSÃO: Na população estudada, a utilização de um modelo de filtro trocador de calor e umidade promoveu a modificação de diversos parâmetros da mecânica respiratória.

Mecânica respiratória; Respiração artificial; Ventiladores mecânicos; Unidades de terapia intensiva


ORIGINAL ARTICLE

IPhysiotherapists; Cardiopulmonar Physiotherapy Specialization Students of the Faculdade Redentor - Campus Recife - Recife (PE), Brazil

IIPhysiotherapy Graduation Students of the Faculdade Integrada do Recife - FIR - Recife (PE), Brazil

IIIPhysiotherapist; PhD, Professor for the Physiotherapy Course of Universidade Federal de Pernambuco - UFPE - Recife (PE), Brazil

IVPhysiotherapist; MSc, Professor for the Physiotherapy Course of the Universidade Católica de Pernambuco - UNICAP - Recife (PE), Brazil

VPhysiotherapist; MSc, Professor for the Faculdade Integrada do Recife - FIR and Faculdade Pernambucana - FAPE - Recife (PE), Brazil

VIPhysiotherapist; MSc, Professor for the Physiotherapy Course of the Universidade Católica de Pernambuco - UNICAP - Recife (PE), Brazil and Faculdade Integrada do Recife - FIR - Recife (PE), Brazil

Author for correspondence

ABSTRACT

OBJECTIVES: In mechanically ventilated patients, humidifier devices are used to heat and moisturize the inspired gas. Heating and humidifying inspired gas may prevent complications associated with the respiratory mucosa dryness such as mucus plugging and endotracheal tube occlusion. Two devices have been commonly used to this, either heated humidifier or the heat moisture exchange filter. This study aimed to compare the effects of the heated humidifier and a model of heat moisture exchange filter on respiratory mechanics in mechanically ventilated neurological patients.

METHODS: This was a randomized crossover trial, involving 31 neurological patients under mechanical ventilation randomly assigned to the humidification devices. Expired tidal volume, peak inspiratory flow, peak expiratory flow, static compliance, dynamic compliance and respiratory system resistance were evaluated. Statistical analysis used the Kolmogorov-Smirnov test and Student's t test for paired samples, in which P values < 0.05 were considered significant.

RESULTS: The heat moisture exchanger filter decreased expired tidal volume, peak inspiratory flow, peak expiratory flow (p < 0.001) and dynamic compliance (p = 0.002), and increased respiratory system resistance (p < 0.001).

CONCLUSION: In the studied population, the use of a heat moisture exchange filter model leaded to several changes on respiratory mechanics parameters.

Keywords: Mechanical ventilation; Artificial respiration; Mechanical ventilators; Intensive care units

INTRODUCTION

Artificial airway (AAW) and mechanical ventilation (MV) are indicated in the presence of imbalance between demand and ventilatory capacity. Upper airways have a very important role for the inspired air heating, humidification and filtration, and are responsible for about 65% of alveolar gas humidity.(1,2)

Prolonged inhalation of inappropriately conditioned gases may lead to hypothermia, bronchial secretions thickening, airways epithelium destruction, atelectasis, tracheal wall inflammation, dysphagia, laryngeal narrowing, in addition to vocal cords, oral cavity and tracheal injuries.(1,3-5)

The use of an AAW contributes to inhalation of a large volume of gas at low temperature and poor humidity, being the use of humidifier system coupled to the inspiratory line of the mechanical ventilator to provide the inspired air conditioning.(1,4,6-9)

Two types of devices have been commonly used during MV aiming to humidify and heat the inspired air, namely heated humidifiers (HH) and heat moisture exchange filter (HMEF). HH promote inspired air heating and humidification by passing the air through a partially filled with heated distilled water chamber, while HMEF are devices adapted between the AAW and the ventilator circuit, and able to store part of the heat and water vapor from the exhaled air, rendering them available for a new inspiration.(1,8,10,11)

Some disadvantages are associated to the use of both humidifiers. During HH use, large amounts of condensate are frequent, increasing the incidence of ventilator-associated pneumonia, patient-ventilator asynchrony, requires electricity use, and constant water replacement. Thus, prolonged HMEF use (> 24 hours) may increase the respiratory workload due to the increased resistance to the inspired air, in addition to alveolar ventilation changes.(3,8,9,12-14)

Although some studies(3,8,10,12-16) analyzed the humidification systems effectiveness regarding viral and bacterial colonization and ability to provide appropriate humidification and heating, few studies(10,14,15) evaluated the effects of their use on respiratory mechanics and respiratory muscles workload.

Considering the above discussed, this study aimed to evaluate the effects of HH and a HMEF model used on critically ill neurological patients under MV.

METHODS

This was a controlled, crossover, randomized study (randomized using ruffle card) involving critically ill neurological patients admitted in the intensive care unit from the Hospital da Restauração between October 2008 and October 2009.

This study was approved by the Institutional Ethic Committee and the legally accepted representatives of the subjects signed an informed consent form, according to the Brazilian Health Authorities requirements.

Inclusion criteria were patients older than 18 years; without previous pulmonary disease or neurological diagnosis; less than 72 hours under mechanical ventilation and sedation and absence of respiratory effort.

As exclusion criteria we considered: contraindications for HMEF use (i.e. abundant or thick tracheal secretion); increased minute volume (above 10 L/min); less than 300 mL or more than 1100 mL tidal volume; bronchopleural fistulae and hypothermia (temperature below 32ºC).(17,18)

The HH Misty-3® (Intermed®, São Paulo - Brazil), filled with 280 mL distilled water was used, adjusted on the device's level 5. The HMEF used was the Hygrobac "S"® (Mallinckrodt DAR® - Tyco Healthcare, Mirandola - Italy) electrostatic model, just unpacked, with dead space of 51 mL and weight of 28 g (according to information provided by the manufacturer).

The baseline patients' evaluation included personal data collection (name, age, gender), and clinical data (cause for intubation, comorbidities, and total MV time) obtained from the medical chart. The previously used ventilatory parameters and ventilation mode were directly checked on the mechanical ventilator (Inter 5-Plus® - Intermed®, São Paulo, Brazil).

All the patients were positioned in 30º elevated dorsal decubitus, which was measured with a goniometer (Carci®, São Paulo, Brazil), and underwent tracheal aspiration. After aspiration, were monitored the heart rate (HR), mean blood pressure (MBP) and peripheral oxygen saturation (SpO2) using the DX 2010® (Dixtal®, São Paulo - Brazil) monitor.

All patients were initially ventilated in pressure controlled ventilation (PCV) mode for 15 minutes, pressure range of 20 cmH2O (ΔP), respiratory rate (RR) of 12 ipm, inspiratory time (Tins) of 1 second, inspiratory sensitivity (Sens) set on -2cmH2O, positive end-expiratory pressure (PEEP) and inspired oxygen fraction (FiO2) kept on the previous level. After these parameters were obtained, the expired tidal volume (TVexp), peak expiratory flow (PEF), and peak inspiratory flow (PIF) were measured using the Inter-GMX Slim® monitor (Intermed®, São Paulo - Brazil).

Then, the ventilatory mode was changed to volume controlled ventilation (VCV) with square flow wave, tidal volume (TV) of 8 mL/kg from the predicted body weight (from the formulas 50 + 0.91 x [height (cm) - 152.4] for men and 45.5 + 0.91 x [height (cm) - 152.4] for women),(19) PIF 60 Lpm, inspiratory pause of 2 seconds, RR of 12 ipm, PEEP and FiO2 kept on previous levels, obtaining a peak pressure (Ppeak) and plateau pressure (Pplat), and calculated the dynamic compliance - Cdyn (TV/Ppeak - PEEP), static compliance - Cstat [TV/(Pplat - PEEP)] and respiratory system resistance - Rrs [(Ppeak - Pplat)/ PIF].(20)

Following Cdyn, Cstat and Rrs evaluations, the patients were ventilated in the PCV mode (parameters above mentioned), and the protocol repeated 15 minutes later, using the other humidification device.

For the statistical analysis, Kolmogorov-Smirnov test was used to measured variables normality, and for the inter-groups analysis the Student's t test for paired samples was applied. The GraphPad Prism 4 and Microsoft Excel 2007 softwares were used, and values with p<0.05 were considered significant.

RESULTS

Thirty one patients with different neurological conditions were evaluated. The subjects' characteristics, including gender, age, anthropometric data and intubation days are shown on table 1.

Table 2 shows the means ± standard deviations for the respiratory mechanic parameters evaluated in this study. It can be seen that patients under HMEF significantly reduced TVexp (p<0.001), PEF (p<0.001), PIF (p<0.001), and Cdyn (0.002). Regarding Rrs, a significant increase was seen in the HMEF group (p<0.001). Cest was not significantly different between groups.

DISCUSSION

The use of HMEF has increased in the last years due to their reduced operational costs, easy handling and possible clinical benefits such as reduced circuit condensate volume and the incidence of ventilator-associated pneumonia.(21,22) However, its use may be associated with increased resistance and dead space, and consequently increased respiratory muscle workload, reduced minute volume and carbon dioxide (CO2) retention.(10, 23-25)

These HMEF adverse effects have been described with its partial obstruction by tracheobronchial secretions and prolonged use, and are associated to its increased hygroscopicicity and weight, not expecting major changes, while using the dry filter.(23,24)

In this study we chose to evaluate a HMEF effect over the respiratory mechanics just after starting its use, and found a mean Rrs increase according to the manufacturer recommendation (±2.5 cmH2O/L/sec, with PIF = 60 Lpm), reaching a mean value above (14.84 cmH2O/L/sec) to the considered acceptable for mechanically ventilated patients (12 cmH2O/L/sec), which could be even greater with longer duration of use.

In our sample composed of critically ill neurological patients, a significant TVexp was identified with HMEF used during assisted-controlled ventilation mode in neurological patients.

Special attention should be available to the control of alveolar ventilation in this population, and in hypercapnic patients while using HMEF, particularly COPD patients and those with a history of bronchospasm, in order to avoid unwanted PaCO2 fluctuations and cerebral hemodynamics changes.(26-28)

Patients undergoing weaning from MV may have respiratory muscles overload while using HMEF due to increased resistance. Several studies compared the effects of HMEF in patients weaning from MV, observing reduced TV, increased PaCO2 and respiratory rate, and minute volume.(4,10,15,23)

In our study the PIF, PEF and Cdyn values were significantly reduced, while Rrs increased significantly using the HHEF. These results showed an increased airflow resistance, which may increase the ventilatory demand and consumption oxygen, difficult the mechanical ventilation weaning process.(10,23,29,30)

Chiaranda et al. reported significant Rsr increase in 83% of the HMEF users after 24 hours.(31) However, in a randomized controlled trial, Ricard et al. evaluated the effect of usage time on the HMEF performance in 45 mechanically ventilated patients, and observed increased Rrs after the initial moment of its use, not observing differences in the Rrs after 7 days to the first day of use.

In our study we used one single HMEF hygroscopic model. Lucato et al. evaluated the effects of different types of HMEF (hygroscopic x hydrophobic x mixed), and found a significant Rsr increase in the hygroscopic HMEF, which was correlated to the excessive humidity accumulation and increased filter weight.(14) Our study did not allow a comparison of several hygroscopic filters, preventing the extrapolation of our results to other models commercially available.

Iotti et al. noted that the increased Rrs from the HMEF may promote increased RR, decreased expiratory time, auto-PEEP development, and consequently increased respiratory workload.(29) Auto-PEEP is associated with increased intrathoracic pressure and reduced cerebral venous return,(32) predisposing to intracranial pressure increase in extensive brain injury patients.(27,28)

Conti et al., evaluating healthy volunteers that spontaneous breathing, noted that HMEF lead to increased Rsr, and that the respiratory muscles workload may be reduced by expiratory time prolongation.(33)

In this study we found no clinical signs of increased respiratory work, as all patients were under sedation and/or anesthesia, and controlled ventilation.

In previous studies, the effect of using HMEF on Cdyn is controversial. Iotti et al. did not observed significant change in Cdyn in patients using HMEF,(29) while Morán et al. found significant reduction in Cdyn during employment of HMEF due to an increased restrictive pressure,(34) similarly to our results.

Reduction in PEF found in this study during HMEF use may difficult bronchial secretions displacement to upper airways in mechanically ventilated patients. According to Gosselink et al and Volpe et al., increase PEF is associated to center of mass displacement and better removal of bronchial mucus excess, thus contributing to lower risks of pulmonary complications such as atelectasis and respiratory infections.(35,36)

In this study, HMEF did not cause significant change in the Cdyn. Macintyre et al., studied 26 patients under mechanical ventilation using HMEF more than 24, also found no significant changes in Cest, which could be explained due to no change in alveolar pressure during during application of HMEF.(37)

Although changes in respiratory mechanics have been observed in our sample, it is not possible to assess the clinical impact, because we did not analized its effect on variables as such mechanical ventilation time and ICU length of stay and hospital mortality. Further more studies should be addressed using others HMEF models (hygroscopic, hydrophobic and mixed) to assess alveolar ventilation and cerebral hemodynamics in critically ill neurological patients, and the respiratory muscles workload in patients weaning from mechanical ventilation.

CONCLUSION

Our results suggest that the use of a HMEF model may lead to significant changes in the respiratory mechanics (TVexp, PIF, PEF, Cdyn and Rsr) in mechanically ventilated neurological patients.

ACKNOWLEDGEMENTS

The authors would like to thank the Hospital da Restauração - PE's Intensive Care Unit and Physiotherapy Team coordinators, and Professor Andrezza Lemos for their kind cooperation.

REFERENCES

  • 1. Ribeiro DC, Lopes CR, Lima RZ, Teixeira YCN. Comparação da resistência imposta pelos filtros artificiais durante a ventilação mecânica. Arq Med ABC. 2007;32(Supl 2): S42-6.
  • 2. Kirton OC, DeHaven CB, Morgan JP, Windsor J, Civetta JM. Elevated imposed work of breathing masquerading as ventilator weaning intolerance. Chest. 1995;108(4):1021-5.
  • 3. Thiéry G, Boyer A, Pigné E, Salah A, De Lassence A, Dreyfuss D, Ricard JD. Heat and moisture exchanger in mechanically ventilated intensive care unit patients: a plea for an independent assessment of their performance. Crit Care Med. 2003;31(3):699-704.
  • 4. Galvão AM, Galindo Filho VC, Marinho PEM, Gomes R, França EET, Brandão DC, et al. Estudo comparativo entre os sistemas de umidificação aquoso aquecido e trocador de calor e de umidade na via aérea artificial de pacientes em ventilação mecânica invasiva. Rev Bras Fisioter. 2006;10(3):303-8.
  • 5. Scanlan CL, Wilkins RL, Stoller JK. Fundamentos da terapia respiratória de Egan. 7a ed. São Paulo: Manole; 2000. Cap. 31. p.683-703.
  • 6. Branson RD. Humidification for patients with artificial airways. Respir Care.1999;44(6):630-42.
  • 7. Chiumello D, Pelosi P, Park G, Candiani A, Bottino N, Storelli E, et al. In vitro and in vivo evaluation of a new active heat moisture exchanger. Crit Care. 2004;8(5):R281-8.
  • 8. Kollef MH, Shapiro SD, Boyd V, Silver P, Von Harz B, Trovillion E, Prentice D. A randomizer trial comparing an extended-use hygroscopic condenser humidifier with heated-water humidification in mechanically ventilated patients. Chest. 1998;133(3):759-67.
  • 9. Lucato JJ, Adams AB, Souza R, Torquato JA, Carvalho CR, Marini JJ. Evaluating humidity recovery efficiency of currently available heat and moisture exchangers: a respiratory system model study. Clinics (Sao Paulo). 2009;64(6):585-90.
  • 10. Girault C, Breton L, Richard JC, Tamion F, Vandelet P, Aboab J, et al. Mechanical effects of airway humidification devices in difficult to wean patients. Crit Care Med. 2003;31(5):1306-11.
  • 11. Branson RD, Campbell R, Johannigmam J, Ottaway M, Davis K Jr, Luchette FA, Frame S. Comparison of conventional heated humidification with a new active hygroscopic heat and moisture exchanger in mechanically ventilated patients. Respir Care. 1999;44(8):912-7.
  • 12. Koening SM, Truwit JD. Ventilator-associated pneumonia: diagnosis, treatment, and prevention. Clin Microbiol Rev. 2006;19(4):637-57.
  • 13. Diaz RB, Barbosa DA, Bettencourt AR, Vianna LAC, Gir E, Guimarães T. Avaliação do uso de filtros umidificadores higroscópicos para prevenção de pneumonia hospitalar. Acta Paul Enferm. 2002;15(4):32-44.
  • 14. Lucato JJ, Tucci MR, Schettino GP, Adams AB, Fu C, Forti G Jr, et al. Evaluation of resistance in 8 different heat-and-moisture exchangers: effects of saturation and flow rate/profile. Respir Care. 2005;50(5):636-43.
  • 15. Le Bourdellès G, Mier L, Fiquet B, Djedaïni K, Saumon G, Coste F, Dreyfuss D. Comparison of the effects of heat and moisture exchangers and heated humidifiers on ventilation and gas exchange during weaning trials from mechanical ventilation. Chest.1996;110(5):1294-8.
  • 16. Turnbull D, Fisher PC, Mills GH, Morgan-Hughes NJ. Performance of breathing filters under wet conditions: a laboratory evaluation. Br J Anaesth. 2005;94(5):675-82.
  • 17. Branson RD, Davis K Jr, Campbell RS, Johnson DJ, Porembka DT. Humidification in the intensive care unit. Prospective study of new protocol utilizing heated humidification and a hygroscopic condenser humidifier. Chest.1993;104(6):1800-5.
  • 18. Ricard JD, Le Mière E, Markowicz P, Lasry S, Saumon G, Djedaïni K, et al. Efficiency and safety of mechanical ventilation with a heat and moisture exchanger changed only once a week. Am J Respir Crit Care Med. 2000;161(1):104-9.
  • 19. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med. 2000;342(18):1301-8.
  • 20. Emmerich JC. Monitorização respiratória: fundamentos. 2a. ed. Rio de Janeiro: Revinter; 2001. p.159-81.
  • 21. Masterton RG, Galloway A, French G, Street M, Armstrong J, Brown E, et al. Guidelines for the management of hospital-acquired pneumonia in the UK: report of the working party on hospital-acquired pneumonia of the British Society for Antimicrobial Chemotherapy. J Antimicrob Chemother. 2008;62(1):5-34. Review.
  • 22. American Thoracic Society; Infectious Diseases Society of America. Guidelines for the management of adults with hospital-acquired, ventilator-associated, and healthcare-associated pneumonia. Am J Respir Crit Care Med. 2005;171(4):388-416.
  • 23. Campbell RS, Davis K Jr, Johannigman JA, Branson RD. The effects of passive humidifier dead space on respiratory variables in paralyzed and spontaneously breathing patients. Respir Care. 2000;45(3):306-12.
  • 24. Manthous C, Schmidt GA. Resistive pressure of a condenser humidifier in mechanically ventilated patients. Crit Care Med. 1994;22(11):1792-5.
  • 25. Kola A, Eckmanns T, Gastmeier P. Efficacy of heat and moisture exchangers in preventing ventilator-associated pneumonia: meta-analysis of randomized controlled trials. Intensive Care Med. 2005;31(1):5-11.
  • 26. Stocchetti N, Maas AI, Chieregato A, van der Plas AA. Hyperventilation in head injury: a review. Chest. 2005;127(5):1812-27.
  • 27. Mascia L, Majorano M. Mechanical ventilation for patients with acute brain injury. Curr Opin Crit Care. 2000;6(1):52-6.
  • 28. Oliveira-Abreu M, Almeida ML. Manuseio da ventilação no trauma cranioencefálico: hiperventilação e pressão positiva expiratória final. Rev Bras Ter Intensiva. 2009;21(1):72-9.
  • 29. Iotti GA, Olivei MC, Braschi A. Mechanical effects of heat-moisture exchangers in ventilated patients. Critical Care. 1999;3(5):R77-82.
  • 30. Pelosi P, Solca M, Ravagnan I, Tubiolo D, Ferrario L, Gattinoni L. Effects of heat and moisture exchangers on minute ventilation, ventilatory drive, and work of breathing during pressure-support ventilation in acute respiratory failure. Crit Care Med.1996;24(7):1184-8.
  • 31. Chiaranda M, Verona L, Pinamonti O, Dominioni L, Minoja G, Conti G. Use of heat and moisture exchanging (HME) filters in mechanically ventilated ICU patients: influence on airway low- resistance. Intensive Care Med. 1993;19(8):462-6.
  • 32. Blanch L, Bernabé F, Lucangelo U. Measurement of air trapping, intrinsic positive end-expiratory pressure, and dynamic hyperinflation in mechanically ventilated patients. Respir Care. 2005;50(1):110-23; discussion 123-4.
  • 33. Conti G, De Blasi RA, Rocco M, Pelaia P, Antonelli M, Bufi M, et al. Effects of the heat-moisture exchangers on dynamic hyperinflation of mechanically ventilated COPD patients. Intensive Care Med. 1990;16(7):441-3.
  • 34. Morán I, Bellapart J, Vari A, Mancebo J. Heat and moisture exchangers and heated humidifiers in acute lung injury/acute respiratory distress syndrome patients. Effects on respiratory mechanics and gas exchange. Intensive Care Med. 2006;32(4):524-31.
  • 35. Gosselink R, Bott J, Johnson M, Dean E, Nava S, Norrenberg M, et al. Physiotherapy for adult patients with critical illness: recommendations of the European Respiratory Society and European Society of Intensive Care Medicine Task Force on Physiotherapy for Critically Ill Patients. Intensive Care Med. 2008;34(7):1188-99.
  • 36. Volpe MS, Adams AB, Amato MB, Marini JJ. Ventilation patterns influence airway secretion movement. Respir Care. 2008;53(10):1287-94.
  • 37. Macintyre NR, Anderson HR, Silver RM, Schuler FR, Coleman RE. Pulmonary function on mechanically-ventilated patients during 24-hour use of a hygroscopic condenser humidifier. Chest. 1983;84(5):560-4.
  • Respiratory mechanics of neurological patients undergoing mechanical ventilation under heated humidifier and a heat exchanger filter model

    Tiago Branco SiqueiraI; Juliana Cristina Gomes de Freitas CostaI; Isnar Campos TavaresII; Priscilla Muniz TorresII; Maria do Amparo AndradeIII; Eduardo Ériko Tenório de FrançaIV; Valdecir Castor Galindo FilhoV; Flávio Maciel Dias AndradeVI
  • Publication Dates

    • Publication in this collection
      20 Oct 2010
    • Date of issue
      Sept 2010

    History

    • Received
      01 Mar 2010
    • Accepted
      02 Aug 2010
    Associação de Medicina Intensiva Brasileira - AMIB Rua Arminda, 93 - Vila Olímpia, CEP 04545-100 - São Paulo - SP - Brasil, Tel.: (11) 5089-2642 - São Paulo - SP - Brazil
    E-mail: rbti.artigos@amib.com.br