Pre-contoured superior locking plates offer poor bone fit for midshaft clavicle fracture fixation: cadaveric analysis of 4 commercially available systems

ABSTRACT

Objective:

The goal of this study is to verify how commercially available pre-contoured superior plates fit on clavicle midshaft fractures.

Methods:

100 cadaveric clavicles were evaluated by three distinct observers applying the clavicle congruence score and comparing four different 6 to 8-hole pre-contoured anatomic locking-plate systems.

Results:

the inter-observer agreement was considered moderate by the percentage agreement and fair by the Fleiss’ Kappa, with no significant differences between evaluations. Only 1 of the 8 plates presented an anatomic fit greater than 70%. Long plates (8 holes) presented a poor fit compared to short plates (6 or 7 holes).

Conclusions:

the overall evaluation showed that currently-available pre-contoured superior plate systems provide a poor fit on clavicles for midshaft fracture fixations. Long plates present a worse fit compared to short ones.

Keywords:
Bone Plates; Clavicle; Fracture Fixation, Internal

RESUMO

Objetivo:

verificar como as placas superiores pré-moldadas disponíveis no mercado se adaptam às fraturas do terço médio da clavícula.

Método:

100 clavículas de cadáveres foram avaliadas por três observadores distintos, aplicando-se o escore de congruência da clavícula e comparando quatro sistemas diferentes de placas bloqueadas anatômicas pré-moldadas de seis a oito furos.

Resultados:

a concordância interobservador foi considerada moderada pelo percentual de concordância e regular pelo índice Kappa de Fleiss, sem diferenças significativas entre as avaliações. Apenas uma das oito placas apresentou encaixe anatômico maior que 70%. As placas longas (oito furos) apresentaram um encaixe ruim em comparação com as placas curtas (seis ou sete furos).

Conclusões:

os sistemas de placa superior pré-moldadas disponíveis atualmente fornecem um encaixe inadequado às clavículas para fixações de fraturas do terço médio. Placas longas apresentam um encaixe pior que as curtas.

Palavras-chave:
Placas Ósseas; Clavícula; Fixação de Fratura, Interna

INTRODUCTION

Clavicle fractures correspond to about 4% of all adult fractures, with an annual incidence, in Sweden, of 50/100.000 cases11 Nowak J, Mallmin H, Larsson S. The aetiology and epidemiology of clavicular fractures. A prospective study during a two-year period in Uppsala, Sweden. Injury. 2000;31:353-8. doi: 10.1016/s0020-1383(99)00312-5.
https://doi.org/10.1016/s0020-1383(99)00...
. Approximately 80% occur in the midshaft22 Stanley D, Trowbridge EA, Norris SH. The mechanism of clavicular fracture. A clinical and biomechanical analysis. J Bone Joint Surg Br. 1988;70:461-4. doi: 10.1302/0301-620X.70B3.3372571.
https://doi.org/10.1302/0301-620X.70B3.3...
. The ideal treatment for displaced diaphyseal clavicle fractures is unknown33 Lenza M, Buchbinder R, Johnston RV, Ferrari BAS, Faloppa F. Surgical versus conservative interventions for treating fractures of the middle third of the clavicle. Cochrane Database Syst Rev. 2019;1(1):CD009363. doi: 10.1002/14651858.CD009363.pub3.
https://doi.org/10.1002/14651858.CD00936...
, but those in favor of surgical management argue that it is associated with a lower risk of nonunion and better functional outcomes44 Reito A, Launonen A, Paloneva J. Factors explaining heterogeneity in studies comparing surgical and nonsurgical treatment of midshaft clavicle fractures: a meta-regression analysis of randomized controlled trials and high-quality observational studies. J Shoulder Elbow Surg. 2020;29(8):1726-34. doi: 10.1016/j.jse.2020.02.011.
https://doi.org/10.1016/j.jse.2020.02.01...
. For this choice, the most popular surgical technique is open reduction and internal fixation using plate and screws55 Robinson CM, Goudie EB, Murray IR, Jenkins PJ, Ahktar MA, Read EO, et al. Open reduction and plate fixation versus nonoperative treatment for displaced midshaft clavicular fractures: a multicenter, randomized, controlled trial. J Bone Joint Surg Am. 2013;95(17):1576-84. doi: 10.2106/JBJS.L.00307.
https://doi.org/10.2106/JBJS.L.00307...
. However, a high implant removal rate is also described33 Lenza M, Buchbinder R, Johnston RV, Ferrari BAS, Faloppa F. Surgical versus conservative interventions for treating fractures of the middle third of the clavicle. Cochrane Database Syst Rev. 2019;1(1):CD009363. doi: 10.1002/14651858.CD009363.pub3.
https://doi.org/10.1002/14651858.CD00936...
.

Pre-contoured locking plates were developed to fit perfectly on the shape of the clavicle, reducing surgical time, causing less soft-tissue discomfort, and increasing fixation stability due to proper bone-implant contact66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
,77 Drosdowech DS, Manwell SEE, Ferreira LM, Goel DP, Faber KJ, Johnson JA. Biomechanical analysis of fixation of middle third fractures of the clavicle. J Orthop Trauma. 2011;25(1):39-43. doi: 10.1097/BOT.0b013e3181d8893a.
https://doi.org/10.1097/BOT.0b013e3181d8...
. However, the clavicle presents a unique and complex anatomy88 Walters J, Solomons M, Roche S. A morphometric study of the clavicle. SA Orthop J. 2010;9(3):47-52. with different gender and ethnic characteristics66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
,99 Kotekar MF, Pai S, Yogesh K, Kumar MA, Shetty MS. Anatomy of clavicle in the Indian population and its applicability in pre-contoured plating. Int Orthop. 2020;44(4):699-704. doi: 10.1007/s00264-019-04390-2.
https://doi.org/10.1007/s00264-019-04390...
,1010 Yang JC-S, Lin K-J, Wei H-W, Tsai C-L, Lin K-P, Lee P-Y. Morphometric Analysis of the Clavicles in Chinese Population. BioMed Res Int. 2017;2017:8149109. doi: 10.1155/2017/8149109.
https://doi.org/10.1155/2017/8149109...
, thereby precluding the pre-molded implant from perfectly fitting the bone shape66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
, which might lead to clinical issues such as reduction failure and post-operative soft-tissue discomfort. In a Randomized Clinical Trial with clavicle fractures treated surgically using superior pre-countered plates, the hardware-removal rate was of 53%1111 Ferran NA, Hodgson P, Vannet N, Williams R, Evans RO. Locked intramedullary fixation vs plating for displaced and shortened mid-shaft clavicle fractures: A randomized clinical trial. J Shoulder Elbow Surg. 2010;19(9):783-9. doi: 10.1016/j.jse.2010.05.002.
https://doi.org/10.1016/j.jse.2010.05.00...
.

Superior clavicle-plate adaptability was first described by Huang66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
. In an attempt to clarify the adaptability of pre-countered plates on the superior surface of the clavicle, Malhas et al.1212 Malhas AM, Skarparis YG, Sripada S, Soames RW, Jariwala AC. How well do contoured superior midshaft clavicle plates fit the clavicle? A cadaveric study. J Shoulder Elbow Surg. 2016;25(6):954-9. doi: 10.1016/j.jse.2015.10.020.
https://doi.org/10.1016/j.jse.2015.10.02...
developed a plate congruence score. This study aims at comparing four pre-contoured superior locking-plate systems, with different plate sizes for midshaft clavicle fractures in terms of bone-surface accommodation, using the clavicle congruence score. The hypothesis is that there are plate fit differences according to the size and system brand

METHOD

This research project was evaluated and approved by the institution’s research and ethics committee (protocol: 80469417.0.0000.5245). This study used one hundred anatomical clavicles from skeletally mature individuals not identified by sex or race with the adjacent soft parts completely disinserted and showing no previous signs of fractures. The clavicles were numerically and sequentially cataloged from 1 to 100 and organized into ten sets of ten specimens, following the numerical sequence with no randomization of the specimens. The upper and lower surfaces and the medial and lateral extremities were identified on all clavicles. The length of each clavicle was measured using an analog caliper (FORTGPRO® - model FG8330) and its midpoint was identified. After identifying the surfaces, the clavicles were separated into left and right-side groups for sampling purposes. In all, 52 right clavicles and 48 left clavicles were separated for evaluation.

Implants from four different manufacturers - Arthrex® (Naples, FL, USA), Johnson & Johnson®/DePuySynthes® (Warsaw, IN, USA), Kanghui®/Medtronic® (Changzhou, JS, China), Hexagon® (Itapira, SP, Brazil) - were used to check how each implant fit in a hypothetical diaphyseal clavicle fracture fixation. In order to guarantee the comparison between manufacturers, the short (six or seven holes) and long (eight holes) implants were selected from each company.

The degree of accommodation of the implant was evaluated according to the clavicle congruence score (CCS), described by Malhas et al.1212 Malhas AM, Skarparis YG, Sripada S, Soames RW, Jariwala AC. How well do contoured superior midshaft clavicle plates fit the clavicle? A cadaveric study. J Shoulder Elbow Surg. 2016;25(6):954-9. doi: 10.1016/j.jse.2015.10.020.
https://doi.org/10.1016/j.jse.2015.10.02...
Graded from one to three, where a rating of three (3) represents an anatomic fit, with the plate perfectly adapted to the bone, two (2) represents a good fit, with anterior or posterior plate protrusion, but with each plate hole centralized on the bone, and one (1) represents a poor fit, with a complete discrepancy between the plate and the clavicle, with one or more plate holes without screws on the bone.

Figures 1A and 1B demonstrate the CCS for the same implant, in anatomic and poor fits, respectively.

Figures 1A and 1B
Anatomic and poor degree of CCS.

Each examiner evaluated all clavicles on the same day for a single plate sequentially by sets of ten specimens. The fit of the eight plates was verified for each clavicle. The examiner was allowed to start the evaluation of the next plate in the subsequent day. The adequacy of the implants was checked freehand, without the use of reduction clamps, to avoid damaging the anatomical specimens. Each plate was placed individually in the upper position, with the correct correspondence of the medial and lateral edges of the implant, according to the manufacturer’s description. Starting with the first plate, the examiner assessed the CCS one time for the first clavicle; then, he repeated this two more times for this same clavicle, totaling three assessments with this implant for that clavicle. Next, the same plate was maintained, used again and the second clavicle was evaluated. Therefore, for each plate, the examiner performed 300 evaluations. Since there were eight implants, a total of 2400 evaluations were performed per examiner.

The midpoint longitudinal mark simulated a transverse fracture, and the short implants were positioned to ensure three free holes on each side of the plate. The multifragmented fracture pattern was estimated when selecting the long implant to be evaluated. The length of the multifragmented fracture line was not predetermined, so it was presumed that the use of a longer implant would be necessary for this fracture pattern. In these cases, the examiner was instructed not to consider the two center plate holes and position the three holes on either side in the most distant portions of the midpoint.

The evaluations were carried out by three different examiners with varying degrees of experience: an orthopedic surgeon (evaluator 1) with more than eight years of surgical experience in orthopedic trauma; two medical residents in an orthopedic-surgery post-graduation program: one in the third year (evaluator 2), and another in the first (examiner 3).

The descriptive analysis of the data aimed at recording the distribution of the clavicle length measurements and analyzing the CCS frequency. The inferential analysis aimed at evaluating the statistical significance of the differences observed between the left and right clavicle measurements and the significance of the differences found in the CCS frequency distributions for the clavicles of different subgroups. The hypothesis of normality in the clavicle length distribution was verified by the Kolmogorov-Smirnov (KS) tests and the Shapiro-Wilk test (SW). The difference-significance analysis found in the CCS frequency distributions for different subgroups was analyzed by the Chi-square test. The intra and inter-observer agreement was assessed by the percentage agreement and the Fleiss’ Kappa, both interpreted as follows [13]: values ≤0 indicated no agreement, 0.01-0.20 no to slight agreement, 0.21-0.40 fair agreement, 0.41-0.60 moderate agreement, 0.61-0.80 substantial agreement, and 0.81-1.00 an almost perfect agreement. All statistical tests were 2-sided, with a p-value set to 0.05. The data were analyzed using statistical analysis software R, version 3.1.0, and with the IBM SPSS program (Statistical Package for the Social Sciences), version 22.0.

RESULTS

The clavicles measured 11.50 to 17.5cm, with an average length of 14.27cm (SD=1.16) (Table 1). The variability of the sample was very low. The percentiles showed that only 5% of the clavicles were less than 12.50cm long, and only 5% were over 16.00cm: 90% measured between 12.50 and 16.00cm. Clavicle length measurements followed a normal distribution, given the normality test p-values, both overall and in the right- and left-side clavicle subgroups. Levene’s test results did not show any significant difference between clavicle measurements in both groups (p-value=0.776). There was no significant difference between right and left mean sizes (p-value=0.329), according to the Student’s t-test.

Table 1
The main statistics on the lengths of the clavicles evaluated, by side and overall.

We classified the sample by length, and according to the 25th and 75th distribution percentile values, they were categorized into small, regular, and large. A clavicle was considered small (24 specimens) if its length was less than 14cm (25th percentile); regular (58 specimens) if its length was greater than or equal to the 25th percentile and less than or equal to the 75th percentile; and large (18 specimens) if its size was greater than 15cm (75th percentile).

In the percentage agreement evaluation, the intra-observer analysis showed that evaluator 1 agreed in 0.68 (SD=0.08) of the cases (substantial agreement); evaluator 2 agreed in 0.71 (SD=0.10) (substantial agreement), and evaluator 3 agreed in 0.81 (SD=0.05) (almost perfect agreement). In the homogeneity analysis, all showed low variability. The inter-observer results for the three evaluators were interpreted as moderate agreement [evaluator 1=0.45 (SD=0.135), evaluator 2=0.41 (SD=0.13), and evaluator 3=0.41 (SD=0.11)], with no significant difference between the evaluations.

In the analysis according to the Fleiss’ Kappa, the intra-observer analysis showed that evaluator 1 had “substantial” agreement (Kappa=0.63; 95% CI 0.60-0.66), evaluator 2 had “moderate” agreement (Kappa=0.60; 95% CI=0.57-0.63), and evaluator 3 had “substantial” agreement, (Kappa=0.76; 95% CI=0.71-0.81). In the inter-observer analysis, the agreement between the three evaluators was ‘fair’, with the following values per evaluator: evaluator 1: Kappa=0.27; 95% CI=0.24-0.30; evaluator 2: Kappa=0.24; 95% CI=0.21-0.27; evaluator 3: Kappa=0.25; 95% CI=0.22-0.28. There were no significant differences between evaluations.

Table 2 shows the distribution per plate, per evaluator, and overall. The results were different between the evaluators. For evaluators 1 and 2, the plate with the lowest “Poor Fit” percentage was the Kanghui®/Medtronic® (KG) 6-hole plate, and for evaluator 3 it was the Johnson&Johnson®/DePuy Synthes® (J&J) 6-hole plate. For evaluator 1, the highest “Anatomic Fit” percentage rating was found for the KG 6-hole plate. For evaluators 2 and 3, the highest “Anatomic Fit” ratings were found for the J&J 6-hole plate. In the overall analysis, the Arthrex 7-hole plate had the lowest “Poor Fit” percentage and the J&J large 8-hole plate had the highest “Poor Fit” percentage classification. The J&J short 6-hole plate had the highest “Anatomic Fit” percentage classification. The Arthrex large 8-hole and Hexagon 8-hole plates had the lowest “Anatomic Fit” percentages.

Table 2
Fit data, according to CCS, by plate, evaluator and overall rating, highlighted the lowest and the highest CCS.

Figure 2 shows the fit classification frequencies without discriminating the evaluator, obtained for the different plate brands and models. When comparing the fit classification of plates of the same brand on a clavicle using the chi-square test, the p-value=0.000 was obtained for all classification comparisons (Arthrex 7 vs. Arthrex 8 holes; J&J 6 vs. J&J 8 holes; KG 6 vs. KG 8 holes; Hexagon 6 vs. Hexagon 8 holes), that is, the fit rating of a plate from the same brand was significantly associated with the size of the plate.

Figure 2
Distribution of frequencies of overall CCS of the eight different plates.

When comparing the frequencies of plates of the same brand, it was found that there was a significant difference between the poor, good, and anatomic CCS of small and large Arthrex (p-value=0.000), J&J (p-value=0.000), and KG (p-value=0.000) plates. There was no significant difference between the Hexagon plates (p-value=0.544). The fit classification of several brands without discriminating the size of the plate can be seen in Figure 3. Altogether, without discriminating the evaluator, Arthrex had the lowest (13.3%) and J&J had the highest “Anatomic fit” rating percentages (59.2%).

Figure 3
Distribution of overall rating of plate’s fit by different systems.

Regarding the aggregated distribution data for short and long plates (Figure 4), it was observed that the CCS scores were significantly different and that the small plates had better anatomical fit performance than the large ones (p-value=0.000).

Figure 4
Differences in adaptation according to plate length.

DISCUSSION

In the present study, we evaluated the accommodation of 8 different clavicle plates according to the score from Malhas et al.1212 Malhas AM, Skarparis YG, Sripada S, Soames RW, Jariwala AC. How well do contoured superior midshaft clavicle plates fit the clavicle? A cadaveric study. J Shoulder Elbow Surg. 2016;25(6):954-9. doi: 10.1016/j.jse.2015.10.020.
https://doi.org/10.1016/j.jse.2015.10.02...
, and the results showed that the majority of implants did not present high anatomic fit levels, and that the long ones where deemed to have poor performance compared to short ones. This is relevant information since the surgeon must be aware that some plate adjustments will be needed to achieve proper fracture fixation. Careful preoperative planning with adequate imaging evaluation is recommended aiming at correctly understanding the patient anatomy, thereby saving surgical time. Bauer et al.1414 Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, et al. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755-8. doi: 10.1016/j.otsr.2018.01.013.
https://doi.org/10.1016/j.otsr.2018.01.0...
, analyzing the fit of a long (8-hole) clavicle plate, described that it was poor compared to a 3.5mm reconstruction plate. In spite of the higher fatigue resistance of anatomical implants in comparison to recon plates in a superior position66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
, even in the simulation of comminuted fractures1515 Uzer G, Yildiz F, Batar S, Bozdag E, Kuduz H, Bilsel K. Biomechanical comparison of three different plate configurations for comminuted clavicle midshaft fracture fixation. J Shoulder Elbow Surg. 2017;26(12):2200-5. doi: 10.1016/j.jse.2017.06.034.
https://doi.org/10.1016/j.jse.2017.06.03...
, a careful evaluation by the surgeon is advised when choosing an implant with a better fit and less resistance1414 Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, et al. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755-8. doi: 10.1016/j.otsr.2018.01.013.
https://doi.org/10.1016/j.otsr.2018.01.0...
,1616 Robertson C, Celestre P, Mahar A, Schwartz A. Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Shoulder Elbow Surg. 2009;18(2):204-9. doi: 10.1016/j.jse.2008.10.002.
https://doi.org/10.1016/j.jse.2008.10.00...
or a more resistant anatomical implant with lower fit probability1414 Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, et al. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755-8. doi: 10.1016/j.otsr.2018.01.013.
https://doi.org/10.1016/j.otsr.2018.01.0...
,1616 Robertson C, Celestre P, Mahar A, Schwartz A. Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Shoulder Elbow Surg. 2009;18(2):204-9. doi: 10.1016/j.jse.2008.10.002.
https://doi.org/10.1016/j.jse.2008.10.00...
.

The low profile pre-contoured locking plate is the most popular implant for displaced midshaft clavicle fracture fixation66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
,1212 Malhas AM, Skarparis YG, Sripada S, Soames RW, Jariwala AC. How well do contoured superior midshaft clavicle plates fit the clavicle? A cadaveric study. J Shoulder Elbow Surg. 2016;25(6):954-9. doi: 10.1016/j.jse.2015.10.020.
https://doi.org/10.1016/j.jse.2015.10.02...
, but instead of an anatomic accommodation to the shape of the clavicle, the data describe the poor performance of the studied implants1414 Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, et al. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755-8. doi: 10.1016/j.otsr.2018.01.013.
https://doi.org/10.1016/j.otsr.2018.01.0...
,1717 Vancleef S, Herteleer M, Carette Y, Herijgers P, Duflou JR, Nijs S, et al. Why off-the-shelf clavicle plates rarely fit: anatomic analysis of the clavicle through statistical shape modeling. J Shoulder Elbow Surg. 2019;28(4):631-8. doi: 10.1016/j.jse.2018.09.018.
https://doi.org/10.1016/j.jse.2018.09.01...
,1818 Naimark M, Dufka FL, Han R. Plate fixation of midshaft clavicular fractures: patient-reported outcomes and hardware-related complications. J Shoulder Elbow Surg. 2016;25(5):739-46. doi: 10.1016/j.jse.2015.09.029.
https://doi.org/10.1016/j.jse.2015.09.02...
. Our results demonstrated that for all brands evaluated, the poor fit rate varied up to 19.2%. It is noteworthy that, in terms of anatomic characteristics and anthropometric measurements of the sample, our findings are quite similar to previous studies with different populations, demonstrating that even with many anatomic similarities, a low appropriate plate-accommodation level still persists99 Kotekar MF, Pai S, Yogesh K, Kumar MA, Shetty MS. Anatomy of clavicle in the Indian population and its applicability in pre-contoured plating. Int Orthop. 2020;44(4):699-704. doi: 10.1007/s00264-019-04390-2.
https://doi.org/10.1007/s00264-019-04390...
,1414 Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, et al. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755-8. doi: 10.1016/j.otsr.2018.01.013.
https://doi.org/10.1016/j.otsr.2018.01.0...
,1717 Vancleef S, Herteleer M, Carette Y, Herijgers P, Duflou JR, Nijs S, et al. Why off-the-shelf clavicle plates rarely fit: anatomic analysis of the clavicle through statistical shape modeling. J Shoulder Elbow Surg. 2019;28(4):631-8. doi: 10.1016/j.jse.2018.09.018.
https://doi.org/10.1016/j.jse.2018.09.01...
.

One of the complications after surgical fixation of displaced clavicle fractures is the high implant removal rate. The advantage of the clavicle plate would be its ability to cause less soft tissue discomfort66 Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
https://doi.org/10.2106/JBJS.G.00111...
, diminishing the need for plate removal. The results of our study show that, in many evaluations, a considerable portion of hardware lay beyond the bone. This may be associated with the persistence of soft tissue irritation and the need for implant removal, which is supported by the literature with the removal rate for pre-countered plates ranging from 12.71818 Naimark M, Dufka FL, Han R. Plate fixation of midshaft clavicular fractures: patient-reported outcomes and hardware-related complications. J Shoulder Elbow Surg. 2016;25(5):739-46. doi: 10.1016/j.jse.2015.09.029.
https://doi.org/10.1016/j.jse.2015.09.02...
to 53%1111 Ferran NA, Hodgson P, Vannet N, Williams R, Evans RO. Locked intramedullary fixation vs plating for displaced and shortened mid-shaft clavicle fractures: A randomized clinical trial. J Shoulder Elbow Surg. 2010;19(9):783-9. doi: 10.1016/j.jse.2010.05.002.
https://doi.org/10.1016/j.jse.2010.05.00...
.

The CCS might help surgeons in choosing the appropriate implant. Despite its reproducibility being questionable, especially due to the fair inter-observer agreement, our results are similar to other studies that replicated the method99 Kotekar MF, Pai S, Yogesh K, Kumar MA, Shetty MS. Anatomy of clavicle in the Indian population and its applicability in pre-contoured plating. Int Orthop. 2020;44(4):699-704. doi: 10.1007/s00264-019-04390-2.
https://doi.org/10.1007/s00264-019-04390...
. In a clinical perspective, the score can be used in preoperative planning with 3D-printed replicas of the fractured clavicle, increasing the ability to better assess the fit of the implants and reduce plate-related discomfort1919 van Doremalen RFM, van der Linde RA, Kootstra JJ, van Helden SH, Hekman EEG. Can 3D printing avoid discomfort related implant removal in midshaft clavicle fractures? A four year follow up. Arch Ortop Trauma Surg. 2021;141(11):1899-1907. doi: 10.1007/s00402-020-03654-6.
https://doi.org/10.1007/s00402-020-03654...
.

The choice of two statistics for the evaluation of CCS among the three examiners with different levels of experience in orthopedic surgery is supported by the literature. According to McHugh, in situations where varying levels of training or guesswork are expected among observers, it is recommended to evaluate both the percentage agreement and the Kappa1313 McHugh ML. Interrater reliability: the kappa statistic. Biochem Med (Zagreb). 2012;22(3):276-82..

Another matter deserves to be pointed out in the present study: to overcome a potential bias when comparing implants that were developed for a specific population, we compared four short and four long plates manufactured in North America, Asia, and Latin America. Nevertheless, only one implant (J&J) presented an anatomic classification in over 70% of the cases. Moreover, even observers with different knowledge levels in orthopedic surgery presented similar behaviors, in line with studies published about the inadequate fit of pre-contoured clavicle plates99 Kotekar MF, Pai S, Yogesh K, Kumar MA, Shetty MS. Anatomy of clavicle in the Indian population and its applicability in pre-contoured plating. Int Orthop. 2020;44(4):699-704. doi: 10.1007/s00264-019-04390-2.
https://doi.org/10.1007/s00264-019-04390...
,1212 Malhas AM, Skarparis YG, Sripada S, Soames RW, Jariwala AC. How well do contoured superior midshaft clavicle plates fit the clavicle? A cadaveric study. J Shoulder Elbow Surg. 2016;25(6):954-9. doi: 10.1016/j.jse.2015.10.020.
https://doi.org/10.1016/j.jse.2015.10.02...
,1717 Vancleef S, Herteleer M, Carette Y, Herijgers P, Duflou JR, Nijs S, et al. Why off-the-shelf clavicle plates rarely fit: anatomic analysis of the clavicle through statistical shape modeling. J Shoulder Elbow Surg. 2019;28(4):631-8. doi: 10.1016/j.jse.2018.09.018.
https://doi.org/10.1016/j.jse.2018.09.01...
.

A potential limitation of this study is that in the sample of 100 clavicles it was supposed that the ethnic characteristics were of a single population, even if they were not identified according to gender or race. Another limitation is that the previous sample size calculation according to the measurement of agreement was not performed because the consulting statistician considered that the number of evaluations per examiners was large enough for the pragmatic purpose of the study.

As a strength of our study, we believe that the method of evaluating the sample with different plates, different brands and examiners with different levels of experience simulated the surgical practice of choosing the implant and demonstrated that there is a difference in the fit of all implants.

Since the clavicle is a unique bone with its own singularities and a complex anatomy, designing an ideal implant remains a challenging obstacle. With the continuous implant-engineering development, future directions towards customized implants for clavicle fractures2020 Cronskär M, Rännar LE, Bäckström M, Nilsson KG, Samuelsson B. Patient-Specific Clavicle Reconstruction Using Digital Design and Additive Manufacturing. J. Mech. Des. 2015;137(11):1114-18. doi: 10.1115/1.4030992.
https://doi.org/10.1115/1.4030992...
individualized according to the patient anatomy can be a promising alternative.

CONCLUSION

The vast majority of implants tested presented less than 70% anatomic plate fit, according to the clavicle congruence score. Long implants showed a poor fit, compared to the short plates. These results point out that an ideal implant aiming at the anatomical fit on different clavicle shapes is unlikely. Future tendencies towards customized implants may be an effective alternative to decrease surgical time and minimize implant removal rates.

REFERENCES

  • 1
    Nowak J, Mallmin H, Larsson S. The aetiology and epidemiology of clavicular fractures. A prospective study during a two-year period in Uppsala, Sweden. Injury. 2000;31:353-8. doi: 10.1016/s0020-1383(99)00312-5.
    » https://doi.org/10.1016/s0020-1383(99)00312-5
  • 2
    Stanley D, Trowbridge EA, Norris SH. The mechanism of clavicular fracture. A clinical and biomechanical analysis. J Bone Joint Surg Br. 1988;70:461-4. doi: 10.1302/0301-620X.70B3.3372571.
    » https://doi.org/10.1302/0301-620X.70B3.3372571
  • 3
    Lenza M, Buchbinder R, Johnston RV, Ferrari BAS, Faloppa F. Surgical versus conservative interventions for treating fractures of the middle third of the clavicle. Cochrane Database Syst Rev. 2019;1(1):CD009363. doi: 10.1002/14651858.CD009363.pub3.
    » https://doi.org/10.1002/14651858.CD009363.pub3
  • 4
    Reito A, Launonen A, Paloneva J. Factors explaining heterogeneity in studies comparing surgical and nonsurgical treatment of midshaft clavicle fractures: a meta-regression analysis of randomized controlled trials and high-quality observational studies. J Shoulder Elbow Surg. 2020;29(8):1726-34. doi: 10.1016/j.jse.2020.02.011.
    » https://doi.org/10.1016/j.jse.2020.02.011
  • 5
    Robinson CM, Goudie EB, Murray IR, Jenkins PJ, Ahktar MA, Read EO, et al. Open reduction and plate fixation versus nonoperative treatment for displaced midshaft clavicular fractures: a multicenter, randomized, controlled trial. J Bone Joint Surg Am. 2013;95(17):1576-84. doi: 10.2106/JBJS.L.00307.
    » https://doi.org/10.2106/JBJS.L.00307
  • 6
    Huang JI, Toogood P, Chen MR, Wilber JH, Cooperman DR. Clavicular anatomy and the applicability of precontoured plates. J Bone Joint Surg Am. 2007;89(10):2260-5. doi: 10.2106/JBJS.G.00111.
    » https://doi.org/10.2106/JBJS.G.00111
  • 7
    Drosdowech DS, Manwell SEE, Ferreira LM, Goel DP, Faber KJ, Johnson JA. Biomechanical analysis of fixation of middle third fractures of the clavicle. J Orthop Trauma. 2011;25(1):39-43. doi: 10.1097/BOT.0b013e3181d8893a.
    » https://doi.org/10.1097/BOT.0b013e3181d8893a
  • 8
    Walters J, Solomons M, Roche S. A morphometric study of the clavicle. SA Orthop J. 2010;9(3):47-52.
  • 9
    Kotekar MF, Pai S, Yogesh K, Kumar MA, Shetty MS. Anatomy of clavicle in the Indian population and its applicability in pre-contoured plating. Int Orthop. 2020;44(4):699-704. doi: 10.1007/s00264-019-04390-2.
    » https://doi.org/10.1007/s00264-019-04390-2
  • 10
    Yang JC-S, Lin K-J, Wei H-W, Tsai C-L, Lin K-P, Lee P-Y. Morphometric Analysis of the Clavicles in Chinese Population. BioMed Res Int. 2017;2017:8149109. doi: 10.1155/2017/8149109.
    » https://doi.org/10.1155/2017/8149109
  • 11
    Ferran NA, Hodgson P, Vannet N, Williams R, Evans RO. Locked intramedullary fixation vs plating for displaced and shortened mid-shaft clavicle fractures: A randomized clinical trial. J Shoulder Elbow Surg. 2010;19(9):783-9. doi: 10.1016/j.jse.2010.05.002.
    » https://doi.org/10.1016/j.jse.2010.05.002
  • 12
    Malhas AM, Skarparis YG, Sripada S, Soames RW, Jariwala AC. How well do contoured superior midshaft clavicle plates fit the clavicle? A cadaveric study. J Shoulder Elbow Surg. 2016;25(6):954-9. doi: 10.1016/j.jse.2015.10.020.
    » https://doi.org/10.1016/j.jse.2015.10.020
  • 13
    McHugh ML. Interrater reliability: the kappa statistic. Biochem Med (Zagreb). 2012;22(3):276-82.
  • 14
    Bauer DE, Hingsammer A, Schenk P, Vlachopoulos L, Imam MA, Fürnstahl P, et al. Are commercially-available precontoured anatomical clavicle plating systems offering the purported superior optimum fitting to the clavicle? A cadaveric analysis and review of literature. Orthop Traumatol Surg Res. 2018;104(6):755-8. doi: 10.1016/j.otsr.2018.01.013.
    » https://doi.org/10.1016/j.otsr.2018.01.013
  • 15
    Uzer G, Yildiz F, Batar S, Bozdag E, Kuduz H, Bilsel K. Biomechanical comparison of three different plate configurations for comminuted clavicle midshaft fracture fixation. J Shoulder Elbow Surg. 2017;26(12):2200-5. doi: 10.1016/j.jse.2017.06.034.
    » https://doi.org/10.1016/j.jse.2017.06.034
  • 16
    Robertson C, Celestre P, Mahar A, Schwartz A. Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Shoulder Elbow Surg. 2009;18(2):204-9. doi: 10.1016/j.jse.2008.10.002.
    » https://doi.org/10.1016/j.jse.2008.10.002
  • 17
    Vancleef S, Herteleer M, Carette Y, Herijgers P, Duflou JR, Nijs S, et al. Why off-the-shelf clavicle plates rarely fit: anatomic analysis of the clavicle through statistical shape modeling. J Shoulder Elbow Surg. 2019;28(4):631-8. doi: 10.1016/j.jse.2018.09.018.
    » https://doi.org/10.1016/j.jse.2018.09.018
  • 18
    Naimark M, Dufka FL, Han R. Plate fixation of midshaft clavicular fractures: patient-reported outcomes and hardware-related complications. J Shoulder Elbow Surg. 2016;25(5):739-46. doi: 10.1016/j.jse.2015.09.029.
    » https://doi.org/10.1016/j.jse.2015.09.029
  • 19
    van Doremalen RFM, van der Linde RA, Kootstra JJ, van Helden SH, Hekman EEG. Can 3D printing avoid discomfort related implant removal in midshaft clavicle fractures? A four year follow up. Arch Ortop Trauma Surg. 2021;141(11):1899-1907. doi: 10.1007/s00402-020-03654-6.
    » https://doi.org/10.1007/s00402-020-03654-6
  • 20
    Cronskär M, Rännar LE, Bäckström M, Nilsson KG, Samuelsson B. Patient-Specific Clavicle Reconstruction Using Digital Design and Additive Manufacturing. J. Mech. Des. 2015;137(11):1114-18. doi: 10.1115/1.4030992.
    » https://doi.org/10.1115/1.4030992
  • Funding source:

    none.