Study of behavior alloy Ti and 316L in to simulated body fluid by electrochemical techniques

ABSTRACT

In this work, Ti-316L stainless steel was produced, and its electrochemical behavior was characterized. Ti-steel alloys were produced using an induction furnace, by mixing commercial Ti and 316L stainless steel. X-ray diffraction analyses showed the presence of both the materials, Ti and 316L, in the samples produced. The elemental composition of the materials was determined by optical emission spectroscopy and energy-dispersive spectroscopy, which showed similar quantities of both the elements. Both commercial 316L stainless steel and Ti–steel were studied in simulated biological fluid for imitate a like composition of body blood plasma. Electrochemical experiments conducted at 37°C indicated the stable passive polarization behavior of the Ti-steel alloy. Furthermore, the electrochemical behavior of the 316L stainless steel was also analyzed for comparison purposes. Corrosion velocities were determined using the Tafel method and corrosion resistances were obtained using the electrochemical impedance spectroscopy. The Ti-steel exhibited better protection capacity, compared with the commercial 316L steel. The corrosion velocity and the passive current density of the Ti–steel alloy were lower than those exhibited by the 316L stainless steel.

Keywords
XRD; elemental composition; Tafel curves; corrosion velocity

1. INTRODUCTION

Currently, biomaterials play an important role in the improvement of human life quality [11 OKAZAKI, Y., RAO, S., TATEISHI, T., et al., “Cytocompatibility of various metal and development of new titanium alloys for medical implants”, Materials Science and Engineering: A., v. 243, n.1-2, pp. 250–256, Mar. 1998. http://www.sciencedirect.com/science/article/pii/S0921509397008095
http://www.sciencedirect.com/science/art...
, 22 MAHYUDIN, F., WIDHIYANTO, L., HERMAWAN, H., “Biomaterials in orthopedics”, In: Advanced Structured Materials. , v. 58, Biomaterials and Medical Devices., Springer, Cham, pp. 161–181, 2016. https://link.springer.com/chapter/10.1007/978-3-319-14845-8_7
https://link.springer.com/chapter/10.100...
]. For biomedical implants, metallic biomaterials have been widely used in surgical procedures such as replacement of bone joints and dental implants. Although the materials technology has increased the performance and longevity of the implants, materials employed in these implants still exhibit lack of efficiencies, regarding the biological compatibility and the mechanical performance. Because of the advances in the study of Ti compounds, it is expected that this material can play an important role in the development of a new generation of medical implants [33 MISHNAEVSKY, L., et al., “Nanostructured titanium-based materials for medical implants : Modeling and development”, Materials Science and Engineering: R., v. 81, pp. 1–19, Jul. 2014. http://www.sciencedirect.com/science/article/pii/S0927796X14000515
http://www.sciencedirect.com/science/art...
, 44 STATHI, K., TARANTILI, P., POLYZOIS, G., “The effect of accelerated ageing on performance properties of addition type silicone biomaterials”, Materials Science: Materials in Medicine, v. 21, n. 5, pp. 1403–1411, Jan. 2010. https://link.springer.com/article/10.1007/s10856-010-3991-y
https://link.springer.com/article/10.100...
].

The need to seek new materials for medical implants has developed to avoid the drawbacks of the currently used materials. Although metallic biomaterials have been widely used for biomedical implants because of their mechanical resistance and capacity of deformation, they exhibit the disadvantage of low corrosion resistance [55 CHENG, Z., LIAN, J., HUI Y., et al., “Biocompatible DCPD coating formed on az91d magnesium alloy by chemical deposition and its corrosion behaviors in SBF”, Journal Bionic Engineering, v. 11, n. 4, pp. 610–619,Oct. 2014. http://www.sciencedirect.com/science/article/pii/S167265291460072X
http://www.sciencedirect.com/science/art...
]. For this reason, it was found that a great number of metals and alloys are not useful for biomedical implants because they are highly reactive when they are placed into the human body. Therefore, in general, metals used for biomedical implants are made of austenitic stainless steels, cobaltchrome-based alloys, and titanium and its alloys [66 KOHN, D. H., “Metals in medical applications”, Current Opinion in Solid State and Materials Science , v. 3, n. 3, pp. 309–316, Jun. 1998. http://www.sciencedirect.com/science/article/pii/S1359028698801071
http://www.sciencedirect.com/science/art...
, 77 GOTMAN, I., “Characteristics of Metals Used in Implants”, Journal Endourology, v. 11, n. 6, pp. 383–389, Mar.1997. http://online.liebertpub.com/doi/pdf/10.1089/end.1997.11.383
http://online.liebertpub.com/doi/pdf/10....
]. Ti is considered one of the most promising materials for the new generation of biomedical implants because of its low elasticity modulus (two times lower than that of stainless steel and Co–Cr alloys), decreased stress shielding, and decreased osseous reabsorption [88 DAVIS, J., et al., “Handbook of Materials for Medical Devices,” Ohio, ASM Internacional, 2003. http://www.asminternational.org/documents/10192/1849770/06974G_Frontmatter.pdf
http://www.asminternational.org/document...
]– [1010 NIINOMI, M., NAKAI, M., HIEDA, J., “Development of new metallic alloys for biomedical applications”, Acta Biomaterialia, v. 8, n. 11. pp. 3888–3903, Nov. 2012. http://www.sciencedirect.com/science/article/pii/S1742706112002942
http://www.sciencedirect.com/science/art...
]. Furthermore, because of the high biocompatibility of Ti, it has become striking for implantation applications [1111 LI, Y., YANG, C., ZHAO, H., et al., “New developments of ti-based alloys for biomedical applications”, Materials, v. 7, n. 3. pp. 1709–1800, Mar. 2014. http://www.mdpi.com/1996-1944/7/3/1709/htm
http://www.mdpi.com/1996-1944/7/3/1709/h...
]–[1313 YETIM, T., “Corrosion Behavior of Ag-doped TiO2 Coatings on Commercially Pure Titanium in Simulated Body Fluid Solution”, Journal Bionic Engineering, v. 13, n.3, pp. 397–405, Jul. 2016. http://www.sciencedirect.com/science/article/pii/S1672652916603116
http://www.sciencedirect.com/science/art...
]. The chemical biocompatibility of Ti is largely because of its high corrosion resistance, as it avoids the release of cations to the physiological environment. This protection is given by the TiO2 layer formed naturally on the surface of Ti [1414 PAN, J., THIERRY, D., LEYGRAF, C., “Electrochemical impedance spectroscopy study of the passive oxide film on titanium for implant application”, Electrochimica Acta, v. 41, n. 7–8, pp. 1143–1153, May. 1996. http://www.sciencedirect.com/science/article/pii/0013468695004653
http://www.sciencedirect.com/science/art...
, 1515 CZARNOWSKA, E., WIERZCHOŃ, T., MARANDA-NIEDBAŃA, A., “Properties of the surface layers on titanium alloy and their biocompatibility in in vitro tests,” Journal of Materials Processing Technology, v. 92–93, pp. 190–194, Aug. 1999. http://www.sciencedirect.com/science/article/pii/S0924013699002289
http://www.sciencedirect.com/science/art...
]. Several studies have been conducted regarding Ti implants. Echavarría et al. [1616 ECHAVARRIA, A. I., ARROYAVE, C. E., “Electrochemical assesment of some titanium and stainless steel implant dental alloys,” Revista Metalurgia, v.39, pp. 174–181, Sep.2003.http://revistademetalurgia.revistas.csic.es/index.php/revistademetalurgia/article/view/1116/1128
http://revistademetalurgia.revistas.csic...
] used electrochemical impedance spectroscopy for evaluating the behavior of commercially pure Ti that was used in dental implants. They concluded that Ti exhibit a similar biocompatible behavior as that of stainless steel. Yong Li et. al [1717 LI, Y., et al., “Enhanced osseointegration and antibacterial action of zinc-loaded titania-nanotube-coated titanium substrates: In vitro and in vivo studies,” Journal of Biomedical Materials Research Part A, v. 102, n. 11, pp. 3939–3950, Dec. 2014. http://onlinelibrary.wiley.com/doi/10.1002/jbm.a.35060/full
http://onlinelibrary.wiley.com/doi/10.10...
] produced coatings of zinc-reinforced Ti nanotubes on Ti substrates, using the anodization and hydrothermal treatment methods. The produced materials exhibited a better osseous formation that improves the osteointegration between the bone and the implant; moreover, they inhibited bacterial growth.

In this work, a study of the corrosion resistance of Ti-reinforced 316L stainless steel in simulated biological fluid (SBF) was performed. An induction furnace was used to produce the alloy. Chemical, morphological, and electrochemical analyses were performed using the Electrochemical Impedance Spectroscopy and Tafel curves. Although several studies of the biocompatible behavior of the stainless steel can be found in the literature, it is necessary to deepen and widen the knowledge about the biocompatibility of steels when they are combined with other elements as titanium, to improve their anticorrosive properties and osteointegration in living beings.

2. EXPERIMENTAL DETAILS

Ti-steel alloys were produced using an induction furnace Fornax® T 200-240 VAC, 50/60 Hz, where the 316L stainless steel mixed with Ti was melted in a crucible by applying an alternating current. Figure 1 shows the schematic representation of the production process of Ti-steel. The synthetization was performed by introducing steel and Ti into the crucible; then, an external magnetic field was applied. This magnetic field induced an a.c. current that melts and mixes the materials. Table 1 presents the parameters used for materials processing.

Figure 1
Schematic representation of Ti-steel alloy production using an induction furnace
Table 1
Parameters for the Ti-steel alloy production using an induction furnace.

The chemical composition was analyzed using an optical emission spectrometer, ARL 3460 Metal Analyzer. These analyses were performed under an argon environment, with 99.99% purity and a flux of 3.5 L/min. SEM analyses were performed using an FEI-CUANTA 250 equipment; EDX analyses were performed to determine the elemental composition of steel and the Ti-steel alloy; a D8 Bruker AXS Diffractometer was used for X-ray diffraction characterization. It has an X-ray source of Cu Kα, with λ=1.5406 Å and a secondary monochromator of graphite.

Electrochemical measurements were made using a potentiostat Gamry 1000 employing a cell of three electrodes, with graphite as the counter electrode and an Ag/AgCl electrode as the reference. These measurements were performed at 37°C, with an exposition area of 1 cm2. The tests were conducted using SBF (pH 7.4) as electrolyte. The table 2 shows the compositions and concentrations of SBF, this electrolyte was made according to the methodology established of Kokubo [1818 KOKUBO, T., TAKADAMA, H., “How useful is SBF in predicting in vivo bone bioactivity,” Biomaterials, v. 27, n. 15, pp. 2907–2915, May. 2006. http://www.sciencedirect.com/science/article/pii/S0142961206000457
http://www.sciencedirect.com/science/art...
], which is used by many researchers. The critical factors for performance of the SBF are functioning in a proper temperature range, pH adjustment, and avoidance of precipitation. Electrochemical impedance spectroscopy (EIS) measurements were conducted with a stabilization time of 500 s and a frequency scan between 105 and 0.01 Hz. In addition, the Tafel polarization technique was performed by applying a potential of ±250 mV, according to the ASTM G59-97 standard [1919 ASTM G59-97, “Standard Test Method for Conducting Potentiodynamic Polarization Resistance Measurements”, ASTM International, West Conshohocken, PA, 1991 https://www.astm.org/DATABASE.CART/HISTORICAL/G59-97.htm
https://www.astm.org/DATABASE.CART/HISTO...
].

Table 2
List of reagents for the simulated biological fluid preparation and their ion concentration in the SBF and the blood plasma

3. RESULTS AND DISCUSSION

Figures 2(a) and (b) show the diffractograms of the 316L stainless steel and the commercial pure Ti, respectively. Figure 2(c) presents the diffractogram of the Ti-316L stainless steel. Table 2 presents the peaks and plains of the Ti-steel sample and intensity ratio between peaks (111)/ (110). It can be observed that the diffractogram exhibited peaks only 316L austenitic steel phase. However, the change of intensities between 316L steel and Ti-steel can be observed, the calculation of intensities ratio of peaks (111)/ (110) show begin from 3.03 to 0.49. This change is attributed of formation of titanium as solid solution in to lattice of 316L steel, occupying substitutional places in cell.

Figure 2
Diffractograms of (a) commercial 316L stainless steel, (b) commercial Ti, and (c) Ti-steel alloy
Table 3
Peaks and planes of Ti-steel alloys determined by X-ray diffraction.

To analyze the chemical composition of the Ti-steel and corroborating the inclusion of Ti in the steel, quantitative elemental composition of the steel and Ti-steel was analyzed using optical emission spectroscopy and energy-dispersive spectroscopy (EDS). Table 4 presents the values of atomic percentages for both steel and Ti-steel, using both the techniques. The main results of this analysis were as follows: (i) the presence of Ti in the Ti-steel sample, indicating the incorporation of Ti on the steel; (ii) the difference in the elemental composition observed by the two techniques. This difference is mainly because EDS cannot determine the presence of S, Nb, and Sn because of their low concentrations, considering that the instrument used in this technique has a limit of detection of 0.1%. Furthermore, for the case of low-atomic number elements such as carbon and phosphor, the accuracy of the EDS technique is lower for lighter elements than that for heavier elements [2020 CICILEO, G., GASSA, L., et al., “Corrosión de aleaciones de aluminio de aplicación aeroespacial”, Congreso Internacional de Ciencia y Tecnología de Metalurgia y Materiales 2013, Puerto Iguazú, Argentina, Aug. 2013. https://www.researchgate.net/profile/Gabriela_Cicileo/publication/273121214_CORROSION_DE_ALEACIONES_DE_ALUMINIO_DE_APLICACION_AEROESPACIAL/links/54f73ce20cf2ccffe9daf365.pdf
https://www.researchgate.net/profile/Gab...
]. This could be the reason for the large difference in values of the elemental composition of these two elements obtained for the two methods.

Table 4
Elemental composition of the commercial 316L stainless steel and the Ti-steel alloy determined by optical emission spectroscopy and EDS.

Figure 3(a) shows the Nyquist diagram of the 316L steel. In this figure, two semicircles can be observed; the first one, at higher frequencies, can be attributed to the interactions between the electrolyte and the passive layer of the 316L steel. This layer is responsible for protecting the steel against corrosion. Subsequently, charge transfer is inhibited due to its high resistance; however, the semicircle at lower frequencies can be due to the interactions between the electrolyte and the metal (steel).

Figure 3
Nyquist diagrams of (a) commercial 316L steel and (b) Ti-steel alloy.

Figure 3(b) shows the Nyquist diagram for the Ti-steel. From this figure, we can observe the semicircular tendency of the Ti-steel and an inflexion point. This semicircular tendency is attributed to the interactions between the electrolyte and the Ti-steel. The inflection point can be attributed to the diffusion that was a consequence of the interactions between the electrolyte and the electrodes, thereby causing changes in concentrations.

A comparison between Figures 3(a) and (b) showed that the Ti-steel exhibits greater resistance than the 316L steel. Because the diameters semicircles represent the materials resistance, these figures indicate that the resistance of Ti-steel is in the order of kΩ while that of the commercial 316L steel is in the Ω.

For a better analysis of the corrosion resistance by using SBF as the electrolyte, the curves (semicircles) of both measures were fitted using equivalent circuits that represent the obtained experimental data; for performing this procedure, the Gamry Echem Analyst software as used.

Figures 4(a) and (b) show the equivalent circuits that fitted the experimental data obtained, for both the 316L steel and Ti-steel. The nomenclature and description of the components are listed in Table 5. The polarization resistance (Rp) is an important parameter for determining the corrosion behavior of the materials when they are immersed in SBF. Values obtained for Rp were 57.22 ± 0.71 kΩ and 103.5 ± 3.436 Ω for the Ti-steel and the 316L steel, respectively. The greater resistance of Ti-steel is due to the addition of Ti to the 316L steel. This protection can be attributed to the formation of a passive layer of TiO2. The protection effect of oxide films produced in the titanium alloys is generally caused by (i) changes in the surface potential, (ii) the suppression of the activity of aggressive ion adsorption, and (iii) the decrease charge in mobility at the metal–electrolyte interface[2121 WANG, Y., YU, H., CHEN, C., et al., “Review of the biocompatibility of micro-arc oxidation coated titanium alloys,” Materials and Design, v. 85, pp. 640–652, Nov. 2015. http://www.sciencedirect.com/science/article/pii/S0264127515301611
http://www.sciencedirect.com/science/art...
], [2222 YEROKHIN. A. L., NIE, X., LEYLAND, A., et al., “Characterization of oxide films produced by plasma electrolytic oxidation of a Ti–6Al–4V alloy,” Surface and Coatings Technology, v. 130, n. 2-3, pp. 195–206, Aug. 2000. http://www.sciencedirect.com/science/article/pii/S0257897200007192
http://www.sciencedirect.com/science/art...
]

Figure 4
Equivalent circuits that fit the Nyquist diagrams for (a) the 316L steel and (b) Ti-steel
Table 5
Description of the components used in the equivalent circuit for fitting the Nyquist diagram

The polarization Tafel curves for the 316L steel and the Ti-steel are presented in figure 5. These curves include a cathodic branch, which is represented by the most negative potential, and an anodic part, which is attributed to the most positive potential. Values of kinetic corrosion parameters such as the corrosion potential (Ecorr), corrosion current density (jcorr), and corrosion velocity were obtained from the Tafel curves.

Figure 5
Polarization Tafel curves for (a) the 316L steel and (b) the Ti-steel

The corrosion parameter values obtained from the Tafel curves are summarized in Table 6. The potentiodynamic curve of the Ti-steel shifted toward lower currents, compared with that of the 316L steel.

Table 6
Values of corrosion potential, corrosion current, and corrosion rate, obtained from the Tafel curves, for the commercial 316L steel and the Ti-steel

When the samples were immersed in SBF, the jcorr for the Ti-steel was lesser than the 316L steel. This behavior can be explained for the greater corrosion resistance of Ti-steel.

As mentioned earlier, this increase in the corrosion resistance of Ti-steel is attributed to the TiO2 passive layer that inhibits the absorption of corrosive ions. This passive layer also changes the surface potential and reduces the number of charge carriers in the interlayer between the metal and the electrolyte, which in turn protects the substrate from corrosion damages [2222 YEROKHIN. A. L., NIE, X., LEYLAND, A., et al., “Characterization of oxide films produced by plasma electrolytic oxidation of a Ti–6Al–4V alloy,” Surface and Coatings Technology, v. 130, n. 2-3, pp. 195–206, Aug. 2000. http://www.sciencedirect.com/science/article/pii/S0257897200007192
http://www.sciencedirect.com/science/art...
], [2323 SHOKOUHFAR, M., DEHGHANIAN C., MONTAZERI, M., et al., “Preparation of ceramic coating on Ti substrate by plasma electrolytic oxidation in different electrolytes and evaluation of its corrosion resistance: Part II,” Applied Surface Science, v. 258, n. 7, pp. 2416–2423, Jan. 2012. http://www.sciencedirect.com/science/article/pii/S0169433211016333
http://www.sciencedirect.com/science/art...
].

In summary, this study clearly shows the contribution of Ti to the 316L steel in producing a functional biomaterial. In addition, from the electrochemical impedance spectroscopy and polarization curves, it can be concluded that Ti-steel exhibits a better protection capacity against corrosion than the pure 316L steel.

This enhancement in the corrosion resistance behavior of Ti-steel is because the additional passive layer of Ti alloys overcomes the protective conditions of the passive layer formed in stainless steels [99 GIL, F., PLANELL, J., “Aplicaciones biomédicas del titanio v sus aleaciones,” Biomecánica, v.1, n.1, pp. 34–42, Dec.1993. https://upcommons.upc.edu/bitstream/handle/2099/6814/Article04.pdf
https://upcommons.upc.edu/bitstream/hand...
], [2424 MANERO, J., et al., “Wear behavior of the pair Ti-6Al-4V-UHMWPE of acrylic bone cements containing different radiopaque agents”, Journal of Biomaterials Applications, v. 18, n. 4, pp. 305–319, Apr. 2004. http://journals.sagepub.com/doi/abs/10.1177/0885328204041531
http://journals.sagepub.com/doi/abs/10.1...
], despite both are formed because of the metal interaction with oxygen. The passive layer of Ti alloys is composed of TiO2 [2121 WANG, Y., YU, H., CHEN, C., et al., “Review of the biocompatibility of micro-arc oxidation coated titanium alloys,” Materials and Design, v. 85, pp. 640–652, Nov. 2015. http://www.sciencedirect.com/science/article/pii/S0264127515301611
http://www.sciencedirect.com/science/art...
], while the passive layer of stainless steels is composed of Cr2O3 [2323 SHOKOUHFAR, M., DEHGHANIAN C., MONTAZERI, M., et al., “Preparation of ceramic coating on Ti substrate by plasma electrolytic oxidation in different electrolytes and evaluation of its corrosion resistance: Part II,” Applied Surface Science, v. 258, n. 7, pp. 2416–2423, Jan. 2012. http://www.sciencedirect.com/science/article/pii/S0169433211016333
http://www.sciencedirect.com/science/art...
], as shown in Figure 6.

Figure 6
Passive layer formation in (a) Ti-steel and (b) 316L steel

Nevertheless, the passive layer of Ti-steel (TiO2) is more protective than that of stainless steel (Cr2O3). This behavior can be attributed to several factors. One of these factors is the valence used by each element at the time of bond formations with oxygen. Ti and Cr exhibit the oxidation states of +4 and +3, respectively; these materials can be easily oxidized in the presence of water, air, or other reagents. Second, the energy required for removing four electrons (in the case of Ti4+) is higher than that required for removing three electrons (in the case of Cr3+). Third, TiO2 (Figure 7(a)) has a covalent nature, while Cr2O3 (Figure 7(b)) has an ionic nature, which results in stronger and less reactive bonds in TiO2 [2626 COTTON, A., WILKINSON, G., Advanced Inorganic Chemistry, 4 Ed., New York, John Wiley & Sons, 1980.http://web.uni-plovdiv.bg/plamenpenchev/mag/books/inorgchem/Cotton-Wilkinson%20-%20Advanced%20Inorganic%20Chemistry_file1.pdf
http://web.uni-plovdiv.bg/plamenpenchev/...
].

Figure 7
Schematic representation of a (a) TiO2 molecule and (b) Cr2O3 molecule

4. CONCLUSIONS

  • Ti-steel alloy was produced using an induction furnace. The structural, morphological, and electrochemical behaviors of the alloy were characterized by comparing their properties with those of the commercial 316L stainless steel.

  • Using X-ray diffraction, peaks of the 316L stainless steel and Ti were found in Ti-steel, with possible overlapping of some peaks in both the materials.

  • Optical emission spectrometry and EDS were used for determining the elemental composition of both the 316L steel and the Ti-steel, corroborating the presence of Ti in the samples produced.

  • The electrochemical characterization of both the 316L steel and the Ti-steel were performed using EIS and Tafel polarization curves; these results showed that the Ti-steel exhibited better protection capacity, compared with the commercial 316L steel.

  • The corrosion resistance and corrosion velocity in SBF of both the samples was obtained from EIS and Tafel, and it was observed that the Ti-steel alloys exhibit higher corrosion resistance and lower corrosion velocity.

ACKNOWLEDGMENTS

The authors gratefully acknowledge financial support from the Dirección Nacional de Investigaciones of the National University.

BIBLIOGRAPHY

Publication Dates

  • Publication in this collection
    16 Sept 2019
  • Date of issue
    2019

History

  • Received
    20 Sept 2018
  • Accepted
    16 May 2019
Laboratório de Hidrogênio, Coppe - Universidade Federal do Rio de Janeiro, em cooperação com a Associação Brasileira do Hidrogênio, ABH2 Av. Moniz Aragão, 207, 21941-594, Rio de Janeiro, RJ, Brasil, Tel: +55 (21) 3938-8791 - Rio de Janeiro - RJ - Brazil
E-mail: revmateria@gmail.com