Influence of sodium hypochlorite on cyclic fatigue resistance in niti controlled memory endodontics rotary instruments: an experimental evaluation

Abstract

Introduction

It is important to know whether sodium hypochlorite (NaOCl) influences the cyclic fatigue resistance of nickel-titanium (NiTi) files.

Objective

To evaluate the influence of NaOCl 2.5% on the cyclic fatigue resistance of two NiTi systems.

Material and method

Forty rotary instruments – 20 TruNatomy® (TRU, Dentsply Sirona, Maillefer, Ballaigues, Switzerland) and 20 Prodesign Logic2® (PDL2, Bassi, Belo Horizonte, Minas Gerais, Brazil) – were distributed randomly across four experimental groups (n=10) and submerged in distilled water (H2O) or 2.5% NaOCl at 37°C according to allocation. Cyclic fatigue testing was then performed, measuring the number of cycles to fracture (NCF), and post-test surfaces were analysed by scanning electron microscopy (SEM). Statistical Analysis: Analysis of variance (ANOVA) was applied for between-group analysis, followed by Tukey’s post-hoc test.

Result

A significant difference was observed in all groups (P<0.05). PDL2 instruments showed higher fracture resistance under H2O and 2.5% NaOCl conditions compared to TRU. Analysis of all instrument groups showed that exposure to 2.5% NaOCl decreased the NCF compared to H2O.

Conclusion

Cyclic fatigue resistance of the TRU and PDL2 instruments was decreased by exposure to 2.5% NaOCl. PDL2 instruments were more resistant to fracture than TRU instruments.