Hemodynamic Effects of Noninvasive Ventilation in Patients with Venocapillary Pulmonary Hypertension

Abstracts

Background:

The hemodynamic effects of noninvasive ventilation with positive pressure in patients with pulmonary hypertension without left ventricular dysfunction are not clearly established.

Objectives:

Analyze the impact of increasing airway pressure with continuous positive airway pressure on hemodynamic parameters and, in particular, on cardiac output in patients with variable degrees of pulmonary hypertension.

Methods:

The study included 38 patients with pulmonary hypertension caused by mitral stenosis without left ventricular dysfunction or other significant valvulopathy. The hemodynamic state of these patients was analyzed in three conditions: baseline, after continuous positive pressure of 7 cmH2O and, finally, after pressure of 14 cmH2O.

Results:

The population was composed of predominantly young and female individuals with significant elevation in pulmonary arterial pressure (mean systolic pressure of 57 mmHg). Of all variables analyzed, only the right atrial pressure changed across the analyzed moments (from the baseline condition to the pressure of 14 cmH2O there was a change from 8 ± 4 mmHg to 11 ± 3 mmHg, respectively, p = 0.031). Even though there was no variation in mean cardiac output, increased values in pulmonary artery pressure were associated with increased cardiac output. There was no harmful effect or other clinical instability associated with use application of airway pressure.

Conclusion:

In patients with venocapillary pulmonary hypertension without left ventricular dysfunction, cardiac output response was directly associated with the degree of pulmonary hypertension. The application of noninvasive ventilation did not cause complications directly related to the ventilation systems.

Hypertension; Pulmonary; Heart Failure; Noninvasive Ventilation; Hemodynamics


Fundamento:

Os efeitos hemodinâmicos da ventilação não invasiva com pressão positiva em pacientes com hipertensão pulmonar sem disfunção ventricular esquerda não estão bem estabelecidos.

Objetivos:

Analisar o impacto da pressurização aérea crescente através da pressão positiva contínua das vias aéreas nos parâmetros hemodinâmicos e, principalmente, no débito cardíaco de pacientes com graus variáveis de hipertensão pulmonar.

Métodos:

Foram incluídos 38 pacientes com hipertensão pulmonar causada por estenose mitral sem disfunção ventricular esquerda ou outra valvopatia significativa. O estado hemodinâmico deses pacientes foi estudado em três condições: basal, após uso de pressão positiva contínua de 7 cmH2O e , finalmente, após pressão de 14 cmH2O.

Resultados:

A população foi composta por indivíduos predominantemente jovens, do sexo feminino e com elevação significativa da pressão arterial pulmonar (pressão sistólica média de 57 mmHg). De todas as variáveis analisadas, apenas a pressão de átrio direito apresentou variação ao longo dos momentos estudados (da condição basal para a de pressão de 14 cmH2O houve aumento de 8 ± 4 mmHg para 11 ± 3 mmHg, respectivamente, p = 0,031). Apesar de não ter havido variação do valor médio do débito cardíaco, valores mais elevados de pressões na artéria pulmonar estiveram associados ao aumento do débito cardíaco. Não houve efeito deletério ou qualquer instabilidade clínica associada ao uso da pressurização das vias aéreas.

Conclusão:

Em pacientes com hipertensão venocapilar pulmonar sem disfunção ventricular esquerda, a resposta do débito cardíaco esteve diretamente associada ao grau de hipertensão pulmonar. A aplicação de ventilação não invasiva não acarretou complicações relacionadas diretamente aos sistemas de ventilação.

Hipertensão Pulmonar; Insuficiência Cardíaca; Ventilação Não Invasiva; Hemodinâmica


Introduction

Noninvasive positive pressure ventilation (CPAP) is a mode of ventilation commonly applied to a large spectrum of clinical situations. Its use has already been described both in patients in critical conditions such as acute pulmonary congestion, acute respiratory failure in patients with chronic pulmonary disease and during recovery from cardiac and non-cardiac surgeries, and in chronic patients with obstructive sleep apnea.

It is known that, in general, application of airway pressure may benefit the hemodynamic condition of patients with pulmonary congestion and left ventricular systolic dysfunction primarily by increasing cardiac output1Baratz DM, Westbrook PR, Shah PK, Mohsenifar Z. Effect of nasal continuous positive pressure on cardiac output and oxygen delivery in patients with congestive heart failure. Chest. 1992;102(5):1397-401.

Naughton MT, Rahman MA, Hara K, Floras JS, Bradley TD. Effect of continuous positive airway pressure on intrathoracic and left ventricular transmural pressures in patients with congestive heart failure. Circulation. 1995;91(6):1725-31.
-3Bradley TD, Holloway RM, McLaughlin PR, Ross BL, Walters J, Liu PP. Cardiac output response to continuous positive airway pressure in congestive heart failure. Am Rev Respir Dis. 1992;145(2 Pt 1):377-82.. This effect has been attributed mainly to an elevation in intrathoracic pressure and consequent reduction in left ventricular afterload, improving the systolic function of this chamber2Naughton MT, Rahman MA, Hara K, Floras JS, Bradley TD. Effect of continuous positive airway pressure on intrathoracic and left ventricular transmural pressures in patients with congestive heart failure. Circulation. 1995;91(6):1725-31.,4Buda AJ, Pinsky MR, Ingeis NB, Daughters GJ, Stinson EB, Alderman EL. Effect of intrathoracic pressure on left ventricular performance. N Engl J Med. 1979;301(9):453-9.. The ultimate impact from use of CPAP in the cardiac performance of patients with previous systolic dysfunction is of a complex nature and reflects the interaction between the compromised heart and the intrathoracic vasculature. In fact, this intricate relationship can be even more complex in the presence of venocongestive pulmonary hypertension.

In contrast, in the presence of pulmonary hypertension, invasive mechanical ventilation has been associated with unfavorable hemodynamic effects5Zamanian RT, Haddad F, Doyle RL, Weinacker AB. Management strategies for patients with pulmonary hypertension in the intensive care unit. Crit Care Med. 2007;35(9):2037-50.. Therefore, patients with pulmonary hypertension due to pulmonary venocapillary congestion may be subject to unpredictable hemodynamic effects during application of airway pressure.

Still today, the hemodynamic effects of CPAP in patients without left ventricular dysfunction have not been entirely described. Particularly in patients with pulmonary hypertension, a subclass of patients often encountered in daily clinical practice, the impact of CPAP has been poorly characterized. Even though its use is widely accepted as benefical in patients with sleep apnea and some elevation in pulmonary pressure, its effects have not yet been completely described in a population with marked pulmonary hypertension6Sajkov D, Cowie RJ, Thornton AT, Espinoza HA, McEvoy RD. Pulmonary hypertension and hypoxemia in obstructive sleep apnea syndrome. Am J Respir Crit Care Med. 1994;149(2 Pt 1):416-22.,7Arias MA, Garcia-Rio F, Alonso-Fernandez A, Martinez I, Villamor J. Pulmonary hypertension in obstructive sleep apnoea: effects of continuous positive airway pressure: a randomized, controlled cross-over study. Eur Heart J. 2006;27(9):1106-13..

The aim of this study, therefore, is to analyse the impact of increasing airway pressure with CPAP on hemodynamic parameters in patients with variable degrees of pulmonary hypertension. In particular, the primary aim is to evaluate the effects of CPAP on the cardiac output of patients with chronic pulmonary hypertension.

Methods

Studied cohort

This study included a total of 38 patients with pulmonary hypertension (resting systolic pulmonary arterial pressure equal to or higher than 30 mmHg) confirmed by cardiac catheterization. In order to homogenize the cohort, the analysis included only patients with pulmonary hypertension caused by mitral stenosis who were selected to undergo balloon catheter mitral valvuloplasty (BCMV). The selection was conducted according to medical history, physical examination and echocardiogram with two-dimensional color Doppler. This last was conducted in all patients up to 3 months before the invasive evaluation. Patients with left ventricular dysfunction, any greater than mild valvular heart disease, and left atrial thrombus were excluded. The consent term was obtained from all patients and the study was approved by the Ethics Committee of our institution.

Protocol

Patients were evaluated in the catheterization laboratory immediately before BCMV. Analgesia was only performed with local lidocaine infiltration on the puncture site and sedative medications were avoided. Hemodynamic parameters were obtained in the supine position on three moments: 1) at baseline, in ambient air; 2) 5 minutes after CPAP set at 7 cmH2O; and 3) 5 minutes after CPAP set at 14 cmH2O. Noninvasive ventilation was delivered with a CPAP equipment (Solo LX Deluxe, Respironics Inc, USA) and oronasal mask (model Comfortfull, Respironics Inc, USA).

In each of the three conditions, heart rate and rhythm, as well as digital oximetry were measured continuously with a multiparameter monitor. Right atrial pressure, pulmonary arterial pressure and pulmonary capillary pressure were measured with a 7-Fr Swan-Ganz catheter (Edwards Laboratories, Santa Ana, USA), inserted via the right femoral vein. Systemic arterial and left ventricular pressures were measured with a 5-Fr pigtail catheter inserted via the left femoral artery. All measurements were performed at the end of expiration and the pressure transducers were zeroed at the level of the mid-axillary line.

Cardiac output was calculated by thermodilution as a mean of three measurements obtained after injection of 10 mL of 5% dextrose solution randomly performed during the respiratory cycle. If one of the measurements differed in more than 10% of the mean, two other cardiac output measurements were determined and the highest and lowest values were discarded for calculation of the mean. Samples of arterial and venous blood gases were collected from the aorta and the pulmonary artery, respectively. The derived hemodynamic and oxymetric variables were calculated from standard formulas.

Statistical analysis

Categorical variables were shown in tables with absolute (n) and relative (%) frequencies, whereas quantitative variables were presented as mean and standard deviation.

Quantitative variables, measured on each of the three conditions (baseline, CPAP 7 cmH2O and CPAP 14 cmH2O), were presented in tables as means and standard deviations and were evaluated with analysis of variance with repeated measurements. When significant, contrasts of maximum likelihood were conducted to discriminate the differences.

Correlation coefficients were calculated to evaluate the variation in cardiac output with echocardiographic and hemodynamic parameters.

P values < 0.05 were considered statistically significant.

Sample size was calculated using the software G Power 3.1 with the formula for calculation of samples of repeated measures. A statistical power of 80%, significance level of 5%, and an effect (f) size of 0.25 were estipulated. This way, the minimum number of individuals was defined as 33.

Results

Thirty-eight patients with mitral stenosis and pulmonary hypertension followed as outpatients were recruited for this research. The population was composed of individuals predominantly young, females and with significant elevation in pulmonary arterial pressure, with a mean pulmonary arterial pressure of 57 mmHg. Table 1 presents the clinical and echocardiographic characteristics of the cohort, whereas Table 2 presents hemodynamic and blood gases data (Tables 1 and 2).

Table 1
Demographic and echocardiographic features
Table 2
Hemodynamic and blood gases data before and after CPAP

Variation in cardiac output

Table 2 also shows the stability of the mean value of the variables that represent left ventricular systolic capacity, cardiac output, cardiac index and systolic volume. Even though the mean cardiac output values of the cohort showed no variation across all three conditions, there was a significant variation among the patients. It was not possible, therefore, to identify the factors that modulated its variation during different moments of the study. The interactions between baseline characteristics and the effect of CPAP on the cardiac output are presented in Table 3. In the analysis of the use of CPAP set at 7 cmH2O, it was not possible to identify at baseline variables associated with the variation in cardiac output. However, in the analysis of the use of CPAP set at 14 cmH2O, the pressures in the pulmonary system and the oxygen saturation were associated with changes in cardiac output. The pressures in the pulmonary system were the main determinants of the changes in cardiac output, with increase in cardiac output in those with increased levels of pulmonary hypertension (Graphs 1 to 3). The arterial oxygen saturation of hemoglobin was also associated with changes in cardiac output across the studied conditions (Graph 4).

Table 3
Correlation between baseline characteristics and effect of CPAP on cardiac output variation
Graph 1
Percentage change in cardiac output from baseline to CPAP set at 14 cmH2O according to baseline systolic pulmonary pressure.
Graph 2
Percentage change in cardiac output from baseline to CPAP set at 14 cmH2O according to baseline diastolic pulmonary pressure.
Graph 3
Percentage change in cardiac output from baseline to CPAP set at 14 cmH2O according to baseline pulmonary capillary pressure.
Graph 4
Percentage change in cardiac output from baseline to CPAP set at 14 cmH2O according to baseline arterial oxygen saturation.

There was no association between the variation in cardiac output and the nonparametric variables functional class (p = 0.07 for CPAP 7 cmH2O and p = 0.14 for CPAP 14 cmH2O) and presence of atrial fibrillation (p = 0.77 for CPAP 7cmH2O and p = 0.52 for CPAP 14 cmH2O).

Variation in hemodynamic parameters across the conditions

As for hemodynamic modifications during the different moments of the study, only right atrial pressure showed variation. From baseline to CPAP 14 cmH2O, there was an increase from 8 ± 4 mmHg to 11 ± 3 mmHg, respectively (p = 0.031). In contrast, compared with baseline, the increase with CPAP 7 cmH2O was not significant (p = 0.993). Despite a trend towards an increase, there was no expressive variation in values of systemic arterial and pulmonary arterial pressures. Other clinically relevant parameters, such as heart rate and oxygen saturation, also did not show modifications along the study.

It is important to highlight that no harmful effect was observed. There was no occurrence of clinical instability associated with use of CPAP. In addition, all patients, after a brief explanation about the use of CPAP, tolerated its application satisfactorily during the study.

Discussion

The main result of this study was the association of the pressures in the pulmonary system as the main determinants of the variation in cardiac output with CPAP use in a population of patients with pulmonary hypertension. As expected, the hemodynamic changes were more evident with application of higher levels of airway pressure. In the analysis during CPAP set at 7 cmH2O, none of the variables correlated with changes in cardiac output. Following that, the evaluation during CPAP set at 14 cmH2O showed the importance of the pulmonary hypertension in cardiac output response. Higher values of systolic, diastolic and occlusion pulmonary arterial pressures were associated with increased cardiac output.

The association between higher values of pulmonary artery occlusion pressure (PAOP) and positive response to use of CPAP has been previously documented by other authors in a population with a very distinct hemodynamic profile. Baratz et al1Baratz DM, Westbrook PR, Shah PK, Mohsenifar Z. Effect of nasal continuous positive pressure on cardiac output and oxygen delivery in patients with congestive heart failure. Chest. 1992;102(5):1397-401., during a study with patients with idiopathic myocardiopathy after use of CPAP set at 5 cmH2O, identified two groups of patients. The first group, with PAOP equal to or higher than 12 mmHg, responded with an increase of 17% in cardiac output. The second group of patients with PAOP below 12 mmHg, presented a reduction of 8% in cardiac output. Bradley et al3Bradley TD, Holloway RM, McLaughlin PR, Ross BL, Walters J, Liu PP. Cardiac output response to continuous positive airway pressure in congestive heart failure. Am Rev Respir Dis. 1992;145(2 Pt 1):377-82. confirmed this concept, observing that the use of CPAP set at 5 to 10 cmH2O caused an elevation in cardiac output proportional to the level of pressure offered in patients with cardiac failure and elevated PAOP. These studies have in common with the present study the association between higher PAOP values and positive response to use of CPAP. However, only patients with left ventricular dysfunction were included, which was a condition excluded in the current study. It is known that the reduction in left ventricular afterload imposed by CPAP may be particularly advantageous in the presence of systolic dysfunction. In contrast, in patients with pulmonary hypertension and respiratory failure, it is known that mechanical ventilation with positive pressure can impose negative hemodynamic effects. The increase in pulmonary volume and decrease in functional residual capacity can increase the pulmonary vascular resistance and, consequently, the right ventricular output impedance, mainly in patients with pre-existent right ventricular failure5Zamanian RT, Haddad F, Doyle RL, Weinacker AB. Management strategies for patients with pulmonary hypertension in the intensive care unit. Crit Care Med. 2007;35(9):2037-50.. The acute effects of noninvasive ventilation in patients with extreme pulmonary hypertension are not well described. Even though the use of CPAP has been extensively studied in patients with sleep apnea syndrome, this condition is usually associated with a more mild pulmonary hypertension than that observed in the current study6Sajkov D, Cowie RJ, Thornton AT, Espinoza HA, McEvoy RD. Pulmonary hypertension and hypoxemia in obstructive sleep apnea syndrome. Am J Respir Crit Care Med. 1994;149(2 Pt 1):416-22.,7Arias MA, Garcia-Rio F, Alonso-Fernandez A, Martinez I, Villamor J. Pulmonary hypertension in obstructive sleep apnoea: effects of continuous positive airway pressure: a randomized, controlled cross-over study. Eur Heart J. 2006;27(9):1106-13.. In a study that included 20 patients without cardiac or pulmonary pathologies, only five of these patients fulfilled the criteria of pulmonary hypertension. In this study, the mean pulmonary artery pressure of the group was 16.8 mmHg6Sajkov D, Cowie RJ, Thornton AT, Espinoza HA, McEvoy RD. Pulmonary hypertension and hypoxemia in obstructive sleep apnea syndrome. Am J Respir Crit Care Med. 1994;149(2 Pt 1):416-22.. In 40 patients with obstructive sleep apnea and pulmonary hypertension studied by Laks et al8Laks L, Lehrhaft B, Grunstein RR, Sullivan CE. Pulmonary hypertension in obstructive sleep apnea. Eur Respir J. 1995;8(4):537-41., the pulmonary artery systolic pressure estimated by ecochardiogram ranged from 20 to 52 mmHg and the mean was only 29 mmHg8Laks L, Lehrhaft B, Grunstein RR, Sullivan CE. Pulmonary hypertension in obstructive sleep apnea. Eur Respir J. 1995;8(4):537-41.. A study with only nine patients with respiratory failure due to decompensation of chronic obstructive pulmonary disease were studied with hemodynamic monitoring during application of noninvasive ventilation with pressure support and positive end-expiratory pressure (PEEP)9Ambrosino N, Nava S, Torbick A, Riccardi G, Fracchia C, Opasich C, et al. Haemodynamic effects of pressure support and PEEP ventilation by nasal route in patients with stable chronic obstructive pulmonary disease. Thorax. 1993;48(5):523-8.. There was a significant decrease in cardiac output associated with use of PEEP set at 5 cmH2O. However, even though all patients presented at least moderate pulmonary hypertension, the mean pulmonary artery pressure was 27 mmHg, which was much below the value found in our study.

Up to where an extensive review of the literature allows us to consider, the evaluation of the acute hemodynamic effects of noninvasive ventilation in patients with such marked pulmonary hypertension had not yet been conducted. Although our sample, composed of patients with severe mitral stenosis, may represent a profile of patients with preserved left ventricular systolic performance associated with venocapillary pulmonary hypertension, it is not possible to determine whether other etiologies of pulmonary hypertension without left ventricular dysfunction would respond similarly to use of CPAP.

Another important result of this study was that use of CPAP was safe in the conditions applied in the current study, similarly to what has been described by other authors. Only right atrial pressure presented a significant variation during application of airway pressure. In the conditions evaluated in this study, there were no significant variations in relevant clinical parameters, such as heart rate, systemic arterial pressure, cardiac output, pulmonary arterial pressure and arterial oxygen saturation. After the researchers explained the procedures to the patients, including details of aspects related to noninvasive ventilation, use of CPAP was well tolerated, even at a pressure of 14 cmH2O. There were no cases in which noninvasive ventilation had to be removed by request from the patient. It should also be highlighted that there was no register of complication related directly to the use of pressurized ventilation systems.

The use of noninvasive ventilation has been shown capable to reduce the need for intubation, length of stay and ICU mortality in patients with hypoxemic acute respiratory failure. This is due to its effects on the hemodynamic status and pulmonary mechanics. With respect to the respiratory system, its application has as main physiological benefits the decrease in respiratory work, in addition to an improve in oxygenation. As for the clinical applicability of the current study, considering that the respiratory benefits of CPAP would be maintained, the small variation in relevant hemodynamic parameters favors use of CPAP in patients with hypoxemic respiratory failure and hemodynamic profile similar to that in this study.

Study limitations

The main limitation of the study was the fact that the patients were clinically stable during the study procedures. Usually, the application of noninvasive ventilation is indicated in patients with respiratory failure. Therefore, with regard to the best scientific methodology, the ideal would be to evaluate the hemodynamic conditions of patients with pulmonary congestion before any treatment and during noninvasive ventilation. However, this approach was not possible due to a series of factors. First, it would be risky to keep a patients with pulmonary congestion without adequate treatment while invasive monitoring is provided. Second, the great majority of the patients with pulmonary congestion and/or acute pulmonary edema do not require invasive monitoring as part of the therapy. Monitoring is related to a number of complications that would increase the risk to the patient and would, therefore, imply in a questionable ethical judgment. Finally, this strategy would limit substantially the number of selectable patients, jeopardizing the viability of the study. It is also worth mentioning that even though the study was not performed in patients with clinical instability, nothing prevents its value to be extended to this situation.

Conclusion

In patients with mitral stenosis, venocapillary pulmonary hypertension and without left ventricular dysfunction, the cardiac output response to use of CPAP was directly associated with the degree of pulmonary hypertension. The application of CPAP did not incur in complications directly related to the ventilation systems.

References

  • 1
    Baratz DM, Westbrook PR, Shah PK, Mohsenifar Z. Effect of nasal continuous positive pressure on cardiac output and oxygen delivery in patients with congestive heart failure. Chest. 1992;102(5):1397-401.
  • 2
    Naughton MT, Rahman MA, Hara K, Floras JS, Bradley TD. Effect of continuous positive airway pressure on intrathoracic and left ventricular transmural pressures in patients with congestive heart failure. Circulation. 1995;91(6):1725-31.
  • 3
    Bradley TD, Holloway RM, McLaughlin PR, Ross BL, Walters J, Liu PP. Cardiac output response to continuous positive airway pressure in congestive heart failure. Am Rev Respir Dis. 1992;145(2 Pt 1):377-82.
  • 4
    Buda AJ, Pinsky MR, Ingeis NB, Daughters GJ, Stinson EB, Alderman EL. Effect of intrathoracic pressure on left ventricular performance. N Engl J Med. 1979;301(9):453-9.
  • 5
    Zamanian RT, Haddad F, Doyle RL, Weinacker AB. Management strategies for patients with pulmonary hypertension in the intensive care unit. Crit Care Med. 2007;35(9):2037-50.
  • 6
    Sajkov D, Cowie RJ, Thornton AT, Espinoza HA, McEvoy RD. Pulmonary hypertension and hypoxemia in obstructive sleep apnea syndrome. Am J Respir Crit Care Med. 1994;149(2 Pt 1):416-22.
  • 7
    Arias MA, Garcia-Rio F, Alonso-Fernandez A, Martinez I, Villamor J. Pulmonary hypertension in obstructive sleep apnoea: effects of continuous positive airway pressure: a randomized, controlled cross-over study. Eur Heart J. 2006;27(9):1106-13.
  • 8
    Laks L, Lehrhaft B, Grunstein RR, Sullivan CE. Pulmonary hypertension in obstructive sleep apnea. Eur Respir J. 1995;8(4):537-41.
  • 9
    Ambrosino N, Nava S, Torbick A, Riccardi G, Fracchia C, Opasich C, et al. Haemodynamic effects of pressure support and PEEP ventilation by nasal route in patients with stable chronic obstructive pulmonary disease. Thorax. 1993;48(5):523-8.
  • Author contributions
    Conception and design of the research: Bento AM, Cardoso LF, Tarasoutchi F, Sampaio RO, Kajita LJ; Acquisition of data: Bento AM, Cardoso LF, Sampaio RO, Kajita LJ; Analysis and interpretation of the data and Statistical analysis: Bento AM, Cardoso LF, Lemos Neto PA; Writing of the manuscript: Bento AM, Cardoso LF; Critical revision of the manuscript for intellectual content: Bento AM, Tarasoutchi F, Sampaio RO, Lemos Neto PA.
  • Sources of Funding
    There were no external funding sources for this study.
  • Study Association
    This article is part of the thesis of Doctoral submitted by André Moreira Bento, from Instituto do Coração do Hospital das Clínicas da Universidade de São Paulo.

Publication Dates

  • Publication in this collection
    10 Oct 2014
  • Date of issue
    Nov 2014

History

  • Received
    14 Apr 2014
  • Reviewed
    14 July 2014
  • Accepted
    22 July 2014
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br