A simple and rapid method for urinary acetone analysis by headspace/gas chromatography

Abstract

Urinalysis of acetone is important to monitor workers occupationally exposed to acetone and/or isopropanol, as well as in diagnosis of some diseases related to lipid metabolism impairment. This work shows a sensitive, simple and rapid static headspace-gas chromatographic procedure for quantitative determination of acetone in urine. The method was applied to measure acetone in 207 samples from general population volunteers, resulting in a mean level of 1.12 mg/L (± 0.47) and a range of 0.20 - 1.95 mg/L. The method is reproducible and reliable, making it suitable for routine analysis of acetone in urine.

acetone; isopropanol; headspace


NOTA TÉCNICA

A simple and rapid method for urinary acetone analysis by headspace/gas chromatography

Danielle Palma de OliveiraI,* * e-mail: dpalma@usp.br ; Maria Elisa Pereira Bastos de SiqueiraII

IDepartamento de Análises Clínicas, Toxicológicas e Bromatológicas, Faculdade de Ciências Farmacêuticas de Ribeirão Preto, Universidade de São Paulo, Av. do Café, s/n, 14040-903 Ribeirão Preto - SP, Brasil

IIDepartamento de Análises Clínicas, Universidade Federal de Alfenas, Rua Gabriel Monteiro da Silva, 714, 37130-000 Alfenas - MG, Brasil

ABSTRACT

Urinalysis of acetone is important to monitor workers occupationally exposed to acetone and/or isopropanol, as well as in diagnosis of some diseases related to lipid metabolism impairment. This work shows a sensitive, simple and rapid static headspace-gas chromatographic procedure for quantitative determination of acetone in urine. The method was applied to measure acetone in 207 samples from general population volunteers, resulting in a mean level of 1.12 mg/L (± 0.47) and a range of 0.20 - 1.95 mg/L. The method is reproducible and reliable, making it suitable for routine analysis of acetone in urine.

Keywords: acetone; isopropanol; headspace/gas chromatography.

INTRODUCTION

Urinary acetone (UAc) is an internal dose indicator recommended for biomonitoring workers exposed to acetone and isopropanol1 due to its good correlation with environmental solvent levels2-6. It can also be useful for the diagnosis of diseases related to the utilization of lipids such as diabetes mellitus I and for the assessment of early damage due to lipid peroxidation caused by oxidative stress4,6-8.

Acetone is a solvent widely used in various industrial processes such as the manufacture of strengthened plastic fibers and shoes, among others 5, 9, 10. The human sources of acetone emission into the environment include sewage disposal, industrial leaching, and emission from automobiles and turbines11-13.

This substance occurs as a component of normal metabolism in blood, urine and air exhaled by humans. Since the endogenous formation of acetone is closely linked to the stored fats utilization as a source of energy, endogenous levels may vary according to individual health, nutrition and activity levels5,10,14-18.

Isopropanol is a low-cost solvent with numerous industrial applications as an extracting solvent, purifying agent, and paint, varnish and oil diluent, among other19-21. In the human organism this alcohol is biotransformed to acetone through the enzyme alcohol dehydrogenase22-24.

The static headspace technique is extensively used to determine volatile substances2,3,17,25-28. This technique is simple, requires little sample handling and eliminates non-volatile interferents that might contaminate the GC injector and FID detector. The critical point is the equilibrium between the condensed phase and the vapor phase25-29 and several parameters must be optimized to guarantee maximum analyte recovery and sensitivity of the method. Since the determination of urinary acetone is of interest in various situations, the objective of the present study was to optimizing a rapid and low cost technique that could be applied in clinical and toxicological routine analysis in laboratories with limited resources, establishing the better conditions to extract AcU by static headspace and to quantify by GC/FID.

EXPERIMENTAL

Urine samples

Samples used for the determination of acetone were obtained from 207 healthy volunteers not occupationally exposed to acetone and/or isopropanol, with different gender, smoke and drinking habits, with ages varying between 18 to 70 years old. This information was obtained using a questionnaire. After collection, the samples were immediately transferred to headspace flasks containing anhydrous sodium sulfate, sealed and stored in the refrigerator until acetone determination, which was carried out on the same day.

The research protocol was approved by the Research Ethics Committee of the Federal University of Alfenas, where the samples were collected and the analyses were processed.

Standards and equipment

Solutions of acetone (Merck®) and n-propanol (Merck®) at 10 g/L were prepared in distillated water and stored at 4 ºC (stable for one month, data not shown). Standard solutions of 0.1 to 5.0 mg/mL were prepared daily, just before use. The urinary acetone concentration was analyzed using a gas chromatograph CGS® - 1000 DPC model - equipped with a Carbowax CG 745® column (30 m x 0.53 mm) and a flame ionization detector, with IQ (Integration Quality) for integration and processing of the chromatograms. After several tests, the optimized GC conditions were: temperatures of the injector and detector respectively, 150 ºC (in splitless mode) and 180 ºC. The oven temperature was kept isothermally at 90 ºC, using nitrogen as carrier gas (8 mL/min). The internal standard was n-propanol, and the relative retention time (acetone/n-propanol) was 0.47.

Optimization of the static headspace conditions

Studies were carried out to establish the best extraction conditions for the method taking into account the various factors that could affect the technique. For this step, urine samples were prepared from pooled urine from healthy individuals not exposed to chemical substances excreted as acetone and transferred to appropriate sealed flasks: a) heating time: 5, 10, 15 and 20 min were evaluated; b) heating temperature: 70, 75, 80, 85, 90, and 95 ºC were tested for samples kept in an oven for 20 min; c) effect of salt addition: samples from pooled urine were kept at 90 ºC for 20 min with no addition of salt and with anhydrous sodium sulfate or sodium chloride; d) sample volume: 2, 3 and 5 mL urine kept at 90 ºC for 20 min were tested, with the addition of anhydrous sodium sulfate.

Method

The following procedure was used for the method optimization and for the determination of acetone in real samples: 2 g anhydrous sodium sulfate and 0.2 mL internal standard were added to 2 mL urine in a headspace vial which was sealed and heated to 90 ºC for 20 min; 0.5 mL of the gas phase was injected into the GC apparatus.

The effect of the biological matrix was evaluated by comparing the curves from water and urine spiked with acetone concentrations between 0.1 and 5.0 mg/L. The analytical curve in urine was prepared for each group of ten samples analyzed, and the acetone concentration naturally present in the pooled urine used for the addition of the standards was subtracted.

The detection limit (LOD) was considered to be the minimum concentration determined with confidence, i.e., differing from zero with a signal/noise ratio = 3, and was obtained by successive dilutions of 0.1 mg/L acetone sample. The quantification limit (LOQ) was the lowest concentration that could be measured with adequate precision, i.e., with an imprecision evaluated by a coefficient of variation of less than 10% and obtained by the addition of the acetone standard to urine (0.1 mg/L) and by successive sample dilutions.

Intra-assay precision was evaluated in urine samples with no acetone addition and spiked with 1.0 and 3.0 mg/L (10 replicates/concentration). Day-by-day precision was determined in same samples for ten consecutive days.

Stability was studied in samples with no acetone addition and spiked with 1.0 and 3.0 mg/L acetone, under the following conditions: a) ambient temperature (22 ± 2 ºC) and under refrigeration (4 ºC): the samples were analyzed at time zero and 6, 12, 24 and 48 h later; b) freezer (-20 ºC): the samples were stored in aliquots and analyzed over a period of 10 consecutive days and after 15 and 30 days.

RESULTS AND DISCUSSION

The optimized GC method, allowed a satisfactory separation between the analyte and the internal standard, as can be observed in the chromatographic profile, illustrated in Figure 1.


Figure 2 shows the results for UAc headspace extraction. The area of acetone peak was the parameter used to establish the best analytical condition. A longer time of flask incubation in the oven increased the area of the acetone peak in 520% comparing 5 min to 20 min, as can be seen in Figure 2a. This time (20 min) was chosen for the analysis not only because the better recovery of the analyte, but also because it did not increased too much the time of extraction.



The heating temperature was evaluated in order to determine the ideal for acetone volatilization. The initial temperature (70 ºC) was close to the boiling acetone point (63 ºC), and the highest temperature used was 95 ºC. Figure 2b illustrates the area of the acetone peak increased in 306% with the increase of the temperature when the lowest temperature was compared to the highest. However, at 95 ºC significant water volatilization occurs, impairing the analyses and reducing performance of the chromatographic column. So, the temperature of 90 ºC was used for the analysis, showing good sensitivity and precision.

Salt addition was studied to determine the salting out effect on analyte extraction, as indicated by some investigators who use headspace technique for several volatile chemicals extraction26,27,29,30. According to Seto29 the solubility of many non-electrolyte compounds is reduced by salt addition due to an increase in the activity coefficient (y). Both anhydrous sodium sulfate and sodium chloride elicited this effect, but sodium sulfate showed higher sensitivity, with an increase of 26% in the peak acetone area compared to no salt addition, as opposed to 7% for sodium chloride (Figure 2c).

The sample volume of 2 mL was chosen because it presented a more effective volatilization of the analytes compared to the other volumes tested (Figure 2d), probably due to the greater free space in the vial available for the vapors. The variation observed between 3 and 5 mL was small, only 5%, but sensitivity increased by 28% when 2 mL was used compared to 5 mL.

The complex composition of the biological matrix can strongly affect the headspace technique since the various components can alter volatilization of the analytes26,27. The effect of matrix on UAc determination can be observed in Figure 3. Biological matrix interference (Figure 3) was found: linear coefficients (intercepts) and the angular coefficients (slope) from curves prepared in water and urine were not similar. Thus, the calibration curves for acetone quantification in the general population were obtained using standards added to urine.


The method was linear from 0.1 to 5.0 mg/L acetone in urine (Figure 3). The concentration range was chosen according to the backgrounds levels variation5,17,26,28,31-34.

The LOQ obtained was considered satisfactory since all samples showed results above this value. The LOD, equal to the LOQ, was of 0.1 mg/L, with a coefficient of variation of 7% for ten replicates. The mean value of the CV's were: 4.8% for the intra-assay precision and 6.8% for day-by-day precision.

Acetone in urine showed a slight fall (mean 17% in 48 h) at room temperature. At this temperature peaks of other components were observed, probably formed in the sample, with retention times close to acetone, leading to overlap, with loss of resolution and difficulty in area integration. Acetone remains stable for at least 12 hours when storing urine at 4 ºC and for at least ten days when kept at – 20 ºC (mean CV = 6%), temperature recommended by some authors3,5,15,26,31,33,35.

Urinary acetone in general population (Table 1), expressed in mg/L1,15, was similar to the background values published by some investigators in another countries, ranging from 0.84 to 3.85 mg/L25,28,31,32,34.

In conclusion, the proposed method is rapid, simple, low cost and permits its application with confidence to urine acetone determination in several situations where this analysis is requested.

Recebido em 31/5/06; aceito em 9/11/06; publicado na web em 6/8/07

  • 1
    ACGIH - American Conference of Governmental Industrial Hygienists; Threshold limit values for chemical substances and physical agents. Biological exposure indices, ACGIH: Cincinnati, 2001.
  • 2. DiVincenzo, G. D.; Yanno, F. J.; Astill, B. D.; Am. Ind. Hyg. Assoc. J. 1973, 34, 329.
  • 3. Ghittori, S.; Imbriani, M.; Pezzagno, G.; Capodaglio, E.; Am. Ind. Hyg. Assoc. J. 1987, 48, 786.
  • 4. Dong, I.; Shen, X.; Deng, C.; Anal. Chim. Acta 2006, 569, 91.
  • 5. Kawai, T.; Yasugi, T.; Horiguchi, S.; Uchida, Y.; Iwami, O.; Iguchi, H.; Int. Arch. Occup. Environ. Health. 1990, 62, 409.
  • 6. Bagchi, D.; Bagchi, M.; Hassoun, E.; Stohs, S. J.; Toxicology 1992, 75, 81.
  • 7. Deng, C.; Zhang, J.; Yu, X.; Zhang, W.; Zhang, X.; J.Chromatogr., B: Anal. Technol. Biomed. Life Sci. 2004, 810, 269.
  • 8. Dezwart, L. L.; Venhorst, J.; Groot, M.; Commandeur, J. N. M.; Hermanns, R. C. A.; Meerman, J. H. M.; J. Chromatogr., B: Anal. Technol. Biomed. Life Sci. 1997, 694, 277.
  • 9. Encyclopedia of chemical technology, John Willey & Sons: New York, 1991, vol. 1, p. 176.
  • 10. Krasavage, W. J.; O'donoghue, J. L.; Divincenzo, G. D.; Em Patty's Industrial Hygiene and Toxicology; Clayton, G. D.; Clayton, F. E., eds.; Wiley-Interscience: New York, 1982, p. 4721.
  • 11. Jonsson, A.; Persson, K. A.; Grigoriadis, V.; Environ. Int. 1985, 11, 383.
  • 12. Lipari, F.; Dasch, J. M.; Scruggs, W. F.; Environ. Sci. Technol. 1984, 18, 326.
  • 13. Zervas, E.; Montagne, X.; Lahaye, J.; Atmos. Environ. 2001, 35, 1301.
  • 14. Ernstgard, L.; Sjogren, B.; Warholm, M.; Johanson, G.; Toxicol. Appl. Pharmacol. 2003, 193, 158.
  • 15. Fujino, A.; Satoh, T.; Takebayashi, T.; Nakashima, H.; Sakurai, H.; Higashi, T.; Br. J. Ind. Med 1992, 49, 654.
  • 16. Jones, A. W.; J. Anal. Toxicol. 2000, 24, 8.
  • 17. Wang, G.; Maranelli, G.; Perbellini, L.; Raineri, E.; Brugnone, F.; Int. Arch. Occup. Environ. Health. 1994, 65, 285.
  • 18. Wigaeus, E.; Holm, S.; Astrand, I.; Scand. J. Work Environ. Health 1981, 7, 84.
  • 19. Brugnone, F.; Perbellini, L.; Apostoli, P.; Bellomi, M.; Caretta, D.; Br. J. Ind. Med. 1983, 40, 160.
  • 20. Treon, J. F. Em Industrial Hygiene and toxicology; Patty, F. A., ed.; Wiley-Interscience: New York:, 1960, p. 4561.
  • 21
    World Health Organization; Isopropanol [Environmental health criteria nº 103], WHO: Geneva, 1990.
  • 22. Bailey, D. N.; Clin. Toxicol 1990, 28, 4459.
  • 23. Pappas, A. A.; Ackerman, B. H.; Olsen, K. M.; Clin. Toxicol. 1991, 29, 11.
  • 24. Zakhari, S.; Leibowitz, M.; Levy, P.; Aviado, D. M.; Isopropanol and ketones in the environment, CRC Press: Cleveland, 1982, p. 3.
  • 25. Ioffe, B. V.; J. Chromatogr 1984, 290, 363.
  • 26. Jones, A. W.; Sagarduy, E.; Arnqvist, H. J.; J. Anal. Toxicol. 1993, 17,182.
  • 27. Nunez, A. J.; González, L. F.; Janák, J.; J. Chromatogr. 1984, 300,127.
  • 28. Pezzagno, G.; Imbriani, M.; Ghittori, S.; Capodaglio, E.; Am. Ind. Hyg. Assoc. J. 1988, 49, 546.
  • 29. Seto, Y.; J. Chromatogr., A 1994, 674, 25.
  • 30. Vitenberg, A. G.; J. Chromatogr. 1991,556, 1.
  • 31. Brega, A.; Villa, P.; Quadrini, G.; Quadri, A.; Lucarelle, C.; J. Chromatogr. 1991, 553, 249.
  • 32. Ghittori, S.; Maestri, L.; Maraccini, P.; Imbriani, M.; Ind. Health 1996, 34, 409.
  • 33. Kawai, T.; Yasugi, T.; Mizunuma, K.; Horiguchi, S.; Iguchi, H.; Ikeda, M.; Toxicol. Lett. 1992a, 62, 85.
  • 34. Kawai, T.; Yasugi, T.; Uchida, Y.; Iwami, O.; Ikeda, M.; Int. Arch. Occup. Environ. Health. 1992b, 62, 165.
  • 35. Satoh, T.; Omae, K.; Nakashima, H.; Takebayashi, T.; Matsumura, H.; Kawai, T.; Int. Arch. Occup. Environ. Health. 1996, 68, 147.
  • *