Biosensor Based on Cysteine Monolayer and Monoclonal Antibody for Specific Detection of Aflatoxin B1 in Rice

Abstract

An electrochemical immunosensor for detection of aflatoxin B1 (AFB1) in food samples was developed. The sensor is composed by cysteine monolayer immobilized on gold electrode surface with subsequent bind to monoclonal antibody (MAb-Cys-modified electrode). The AFB1 detection was evaluated by cyclic voltammetry and impedance spectroscopy in the frequency range of 100 mHz-100 kHz. Samples of rice were spiked with AFB1 to evaluate the sensitivity of the sensor. Impedance spectra could be fitted to a Randles equivalent circuit containing a constant phase element. Atomic force microscope (AFM) images showed MAb-AFB1 complex immobilized across the electrode surface. The electron transfer resistance (Rct) increase was attributed to a decrease in the charge permeability of the MAb-AFB1-Au surface to a redox probe K4[Fe(CN)6]4-/K3[Fe(CN)6]3-. The surface coverage exhibited a linear relationship as function of toxin concentration and is found to be 0.75 at 30 µg mL-1. The obtained sensor is a promising candidate for detection of AFB1 in rice with sensitivity and specificity.

Keywords:
biosensor; impedance; cyclic voltammetry; aflatoxin


Introduction

Toxins are substances produced by plants, animals and microorganisms that cause adverse effects in human being. Also, toxins are a serious problem for food industry and affect the economy of many countries by interfering with production and exportation of foods.11 Leung, M. C. K.; Diaz-Llano, G.; Smith, T. K.; J. Agric. Food Chem. 2006, 54, 9623.,22 Bhat, R.; Rai, R. V.; Karim, A. A.; Compr. Rev. Food Sci. Food Saf. 2010, 9, 57. Mycotoxins are toxic compounds synthesized by fungi and easily found throughout the food chain and causing harm to human health.33 Rai, M. K.; Bonde, S. R.; Ingle, A. P.; Gade, A. K.; J. Pharm. Educ. Res. 2012, 3, 22.

Aflatoxin B1 (AFB1) is a secondary metabolite produced mainly by fungus Aspergillus flavus and Aspergillus parasiticus with capability to colonize food and feed during harvesting, storage and processing.44 Ding, X. X.; Li, P. W.; Bai, Y. Z.; Zhou, H. Y.; Food Control 2012, 23, 143. AFB1 is considered the most toxic of all aflatoxins due to the carcinogenic, teratogenic, mutagenic and immunosuppressive properties.55 Awad, W. A.; Ghareeb, K.; Böhm, J.; J. Anim. Physiol. Anim. Nutr. 2010, 94, 486. European legislation establishes 2 mg kg-1 as the maximum amount of AFB1 in food, which above of this level toxic manifestations may arise, leading to liver cancer.66 Anklam, E.; Stroka, J.; Boenke, A.; Food Control 2002, 13, 173. Therefore, a previous identification of the contaminated samples is important to ensure quality and safety in the food chain.

Diverse methods, such as liquid chromatography-mass spectrometry (LC-MS), liquid chromatography-tandem mass spectrometry (LC-MS/MS), enzyme-linked immunosorbent assay (ELISA), thin-layer chromatography (TLC) and high-performance liquid chromatography (HPLC), have been used in order to identify AFB1 in diverse samples.77 Nonaka, Y.; Saito, K.; Hanioka, N.; Narimatsu, S.; Kataoka, H.; J. Chromatogr. A 2009, 1216, 4416.

8 Bacaloni, A.; Cavaliere, C.; Cucci, F.; Foglia, P.; Samperi, R.; Lagana, A.; J. Chromatogr. A 2008, 1179, 182.

9 Li, P. W.; Zhang, Q.; Zhang, W.; TrAC, Trends Anal. Chem. 2009, 28, 1115.

10 Otta, K. H.; Papp, E.; Bagocsi, B.; J. Chromatogr. A 2000, 882, 11.

11 Fernandez, A.; Belio, R.; Ramos, J. J.; Sanz, M. C.; Saez, T.; J. Sci. Food Agric. 1997, 74, 161.
-1212 Jaimez, J.; Fente, C. A.; Vazquez, B. I.; Franco, C. M.; Cepeda, A.; Mahuzier, G.; Prognon, P.; J. Chromatogr. A 2000, 882, 1. In spite of these methods have high sensitivity and specificity, they are time-consuming, require experienced technicians, extensive sample pre-treatment and, in some cases, are expensive equipments.1313 Stroka, J.; Anklam, E.; TrAC, Trends Anal. Chem. 2002, 21, 90.,1414 Sharma, A.; Matharu, Z.; Sumana, G.; Solanki, P. R.; Kim, C. G.; Malhotra, B. D.; Thin Solid Films 2010, 519, 1213. Thus, there is an urgent need for the development of new sensor strategies applied for AFB1 screening in food industry and inspection agencies. Immunosensors represent an alternative as a screening tool due to the specificity, sensitivity and in-field detection of AFB1 in foodstuffs.1515 Ricci, F.; Volpe, G.; Micheli, L.; Palleschi, G.; Anal. Chim. Acta 2007, 605, 111.

In recent years, electrochemical immunosensors have been developed for specific detection of toxic small molecules using monoclonal antibodies (MAbs).1616 Tan, Y.; Chu, X.; Shen, G. L.; Yu, R. Q.; Anal. Biochem. 2009, 387, 82.

17 Bacher, G.; Pal, S.; Kanungo, L.; Bhand, S.; Sens. Actuators, B 2012, 168, 223.
-1818 Perrotta, P. R.; Arevalo, F. J.; Vettorazzi, N. R.; Non, M. A.; Fernandez, H.; Sens. Actuators, B 2012, 162, 327. In addition, cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) are useful techniques for the development of immunosensors due to their high sensitivity, low cost and possibility of instrument miniaturization.1919 Ronkainen, N. J.; Halsall, H. B.; Heineman, W. R.; Chem. Soc. Rev. 2010, 39, 1747. EIS is based on the application of a small alternating current (AC) potential as a function of time to measure the current generated. CV technique is used to acquire qualitative information about electrochemical processes relating the magnitude of the current generated by electron transfer during the redox process with the amount of analyte present in the electrode-solution interface.2020 Brett, A. M. O.; Brett, C. M. A.; Eletrochemistry: Principles, Methods, and Applications; Oxford University Press: Oxford, 1993.,2121 Bard, A. J.; Faulkner, L. R.; Electrochemical Methods: Fundamentals and Applications, 2nd ed.; John Wiley & Sons: Oxford, 2000.

Self-assembled monolayers (SAM) have been used to obtain a well-ordered ultrathin layer to the development of bio-devices with high sensitivity and reproducibility.2222 Subramanian, A.; Irudayaraj, J.; Ryan, T.; Biosens. Bioelectron. 2006, 21, 998.,2323 Arya, S. K.; Solanki, P. R.; Datta, M.; Malhotra, B. D.; Biosens. Bioelectron. 2009, 24, 2810. In this study, we modified the bare gold electrode (BGE) surface with a SAM of cysteine (Cys), an amino acid formed by three functional groups (-SH, -COOH and -NH2) in its structure.2424 Jiang, D.; Ji, J.; An, L.; Sun, X. L.; Zhang, Y.; Zhang, G.; Tang, L.; Biosens. Bioelectron. 2013, 50, 150. Subsequently, SAM of Cys were chemically activated using 1-[3-(dimethylamino)propyl]-3-ethylcarbodiimide hydrochloride (EDC) and N-hydroxysuccinimide (NHS) for antibody immobilization. The fabrication process of the biosensor is shown in Figure 1. Atomic force microscopy (AFM), CV and EIS techniques were used to study the process of the electrode modification and detection of AFB1 in samples of rice contaminated with AFB1.

Figure 1
Schematic representation of the immunosensor fabrication. Steps of electrode modification consisting in cysteine (Cys) adsorption, COOH activation, antibody immobilization, remaining biding sites blockage by bovine serum albumin (BSA) and sample evaluation.

Experimental

Materials

Bovine serum albumin (BSA), L-cysteine, AFB1 and anti-AFB1 rabbit monoclonal antibody (MAb-AFB1), 1-[3-(dimethylamino)propyl]-3-ethylcarbodiimide hydrochloride (EDC) and N-hydroxysuccinimide (NHS) were purchased from Sigma-Aldrich (St. Louis, MO, USA). Potassium ferri- and ferrocyanide were obtained from Vetec (Rio de Janeiro, RJ, Brazil). All chemicals and solvents were of analytical-grade and used as received, without further purification. High-purity water was obtained after a Milli-Q plus treatment (Millipore, Billerica, MA, USA).

Preparation of the biosensor system

Initially, BGE was cleaned using piranha solution rinsed thoroughly with ultrapure water, and dried at room temperature (r.t.). SAM formation was preceded by adding a small volume of Cys (1 µL, 25 mM) diluted in phosphate buffer (PB) on BGE surface and allowed to dry for 10 min. Then, the electrode was washed with PB to remove any unbound Cys molecules. Subsequently, 1 µL of 0.4 mol L-1 EDC and 0.1 mol L-1 NHS at a ratio of 1:1 (v/v) was dropwise on the Cys-modified BGE surface and waiting for 10 min. After that, 1 µL of MAb-AFB1 solution (25 mg mL-1) was dropped on the Cys-EDC-NHS-modifiedBGE surface and incubated for 10 min at r.t.; in order to block the remaining active sites, it was added 1 μL BSA (10%, m/v) to the BGE surface for 10 min.

Sample preparation

Uncontaminated rice used in this study was obtained from a local market. Initially, the rice was ground in a household blender. Subsequently, 1 mg of ground rice sample was spiked with AFB1 (150 mg mL-1) and mixed in a vortex mixer. After that, 5 mL of methanol (80%, v/v) was added to the sample, followed by shaking during 45 min and centrifuged at 5000 rpm for 10 min. The supernatant was carefully removed and diluted in PB to obtain different concentrations of AFB1 (0.5 and 1 ng mL-1; 1, 5, 10, 20 and 30 mg mL-1).1616 Tan, Y.; Chu, X.; Shen, G. L.; Yu, R. Q.; Anal. Biochem. 2009, 387, 82.

Biorecognition experiments

The experiments of biorecognition were carried out by the addition of l µL AFB1 at different concentrations (0.5 and 1 ng mL-1; 1, 5, 10, 20 and 30 mg mL-1) on the sensor surface for 10 min. Finally, experiments using samples of rice containing different concentrations of AFB1 were performed.

Electrochemical measurements

Electrochemical measurements were carried out on a PGSTAT 128N potentiostat/galvanostat (Metrohm Autolab B.V., Utrecht, The Netherlands). An electrochemical cell containing three electrodes, platinum wire as auxiliary electrode, Ag/AgCl (saturated with KCl) electrode as reference electrode, and gold disk electrode as working electrode was used. The analysis were performed in the presence of a 10 mmol L-1 K4[Fe(CN)6]4-/K3[Fe(CN)6]3- (1:1) redox pair solution in 10 mmol L-1 PB (pH 7.4). Nyquist plots were obtained in a frequency range varying between 100 mHz to 100 kHz with amplitude of the applied sine wave potential of 10 mV and CV plots were obtained at a potential range between -0.2 and +0.7 V in a scan rate of 50 mV s-1.

Atomic force microscopy analysis

Atomic force microscopy measurements were performed using a PicoSPM II microscope (Molecular Imaging, Tempe, AZ, USA). Cantilevers with a silicon AFM probe (Multi 75 Al, resonant frequency = 75 KHz, force constant = 3 N m-1) were used for the noncontact mode AFM in air at r.t. (ca. 25 ºC). Lateral resolution was set at 512 × 512 pixels in a scan area of 5 × 5 µm. Images were obtained from at least three macroscopically separated areas to ensure a good distribution and analyzed using AFM Gwyddion software.2525 Nečas, D.; Klapetek, P.; Cent. Eur. J. Phys. 2012, 10, 181.

Results and Discussion

Morphological characterization

The topographical and morphological analyses were performed by means of AFM. Atomic force microscopy is an excellent complementary technique to evaluate biosensors, since the electrochemical response depends on the morphology, such as surface roughness, porosity of films and defects.2626 Galal, A.; Atta, N. F.; El-Ads, E. H.; Talanta 2012, 93, 264. Typical topographical (2D) and 3D AFM images of the substrates surfaces containing Cys, Cys-MAb and Cys-MAb-AFB1 obtained by contact mode are shown in Figure 2. Figure 2a shows the topography of the electrode surface after Cys modification. Our results for Cys height are in accordance with a previous work.2727 Huayhuas-Chipana, B. C.; Gomero, J. C. M.; Sotomayor, M. D. P. T.; J. Braz. Chem. Soc. 2014, 25, 1737. Of note, it can be seen a complete and homogeneous layer of the surface without the presence of aggregates or defects. After obtaining the Cys-MAb sensor system, it is visualized an increase in the surface roughness (Figure 2b). Cys-MAb sensor system shows a surface roughness of 24 ± 2 nm. Immobilized MAb has a more globular looking (Figure 2c), demonstrating its protein nature.2828 Yang, L. J.; Li, Y. B.; Biosens. Bioelectron. 2005, 20, 1407. Some authors found a typical height for antibody layer ranging from 6 to 12 nm.2828 Yang, L. J.; Li, Y. B.; Biosens. Bioelectron. 2005, 20, 1407.,2929 Hao, C.; Yan, F.; Ding, L.; Xue, Y. D.; Ju, H. X.; Electrochem. Commun. 2007, 9, 1359. When comparing the AFM images of the sensor system before and after AFB1 recognition, an obvious difference could be observed, indicating that the AFB1 was successfully recognized.

Figure 2
(a) Atomic force microscope (AFM) images of the substrate surfaces containing Cys; (b) Cys-MAb; and (c) Cys-MAb-AFB1 using contact mode. Scan area of 5 × 5 µm.

Electrochemical characterization of the immunosensor

A previous study varying the concentration of Cys (10, 15, 20, 25 and 30 mmol L-1) was performed aiming to determine its optimal concentration for electrode modification (data not shown). At a fixed adsorption time (10 min), it was observed that after 25 mmol L-1 of Cys, the electrode surface saturation occurred. In addition, we performed a study varying the MAb-AFB1 concentration (18, 25, 50 and 75 mg mL-1) adsorbed on the electrode surface using an incubation time of 10 min. We observed a saturation behavior of the electrode surface at 25 mg mL-1 of MAb-AFB1.

The coefficients of variation of the stepwise assembly immunosensor are shown in Figure 3a. The redox reaction of K4[Fe(CN)6]4-/K3[Fe(CN)6]3- exhibits a reversible CVs for BGE. Cys is used to modify the electrode surface through a sulfhydryl group. The adsorption of the Cys on the electrode surface promoted a reduction in peak currents. Cys molecule can increase the resistance due to electrostatic repulsion between the carboxylic group of the Cys and negative charges of the probe redox Fe(CN)64-/3-.3030 Luna, D. M. N.; Avelino, K. Y. P. S.; Cordeiro, M. T.; Andrade, C. A. S.; Oliveira, M. D. L.; Sens. Actuators, B 2015, 220, 565. After the co-addition of NHS and EDC coupling agents, the activated negatively charged terminal carboxylic group of Cys was replaced by NHS ester on the electrode surface. Therefore, an increase of amperometric response was observed due to an electrostatic attraction. The neutrally/positively charged NHS ester promoted the transfer of the negative redox probe to the electrode surface.3131 Geng, P.; Zhang, X.; Meng, W.; Wang, Q.; Zhang, W.; Jin, L.; Feng, Z.; Wu, Z.; Electrochim. Acta 2008, 53, 4663. EDC-NHS coupling agents are important to facilitate binding of the antibody to the sensor system. Then, MAb-AFB1 was added to the sensor system to form a sensing layer. The MAb-AFB1 immobilization on the sensor system promotes a decrease in the anodic and cathodic peak current of K4[Fe(CN)6]4-/K3[Fe(CN)6]3-. Subsequently, 10% BSA was added to block any nonspecific response, increasing the specificity of the sensor.1414 Sharma, A.; Matharu, Z.; Sumana, G.; Solanki, P. R.; Kim, C. G.; Malhotra, B. D.; Thin Solid Films 2010, 519, 1213.

Figure 3
(a) Cyclic voltammograms and (b) Nyquist plots of the stepwise immobilization of Cys, Cys-EDC-NHS, Cys-MAb-AFB1 and Cys-MAb-AFB1-BSA. The impedance spectra were taken for 10 mM K4[Fe(CN)6]4-/K3[Fe(CN)6]3- (1:1) in 10 mM phosphate buffer (pH 7.4) in the frequency range from 100 mHz to 100 kHz. All electrochemical measurements were performed in triplicate using three different samples.

Figure 3b shows the Nyquist diagram corresponding to the modification of the BGE for obtaining the sensor electrode layer. The curve obtained for BGE reveals a small semicircle associated to a very low electron transfer resistance (Rct). The assemblage of Cys on the electrode surface induced an increase in the Rct as an indication of the electron transfer blockage. The modification of the Cys monolayer using EDC-NHS coupling agents resulted in a decrease of the Rct. Coupling agents causes a decrease in resistance due to the attraction between the positively charged groups NHS and negative charge of the redox probe.3232 Ding, L.; Bond, A. M.; Zhai, J.; Zhang, J.; Anal. Chim. Acta 2013, 797, 1. After immobilization of the MAb-AFB1 on Cys-modified electrode, it was obtained an increase in the Rct. As expected, with the addition of BSA on MAb-AFB1-Cys-modified electrode, a new Rct increase was observed. The impedimetric response is increased after complete assembly of the system proportionally to the quantity of material immobilized on the electrode surface, demonstrating that the sensing layer was successfully obtained.

Aflatoxin B1 detection

Figure 4a shows the biorecognition process of the immunosensor for five different concentrations of AFB1 (0.5 and 1 ng mL-1; 1, 5, 10, 20 and 30 mg mL-1) and negative control. The CVs of different concentrations of AFB1 reveal a gradual decrease of the peak current and an increase in the peak-to-peak separation with increasing AFB1 concentration. The calibration curve is shown in Figure 4b. The curve exhibits a toxin detection ranging from 0.5 ng mL-1 to 30 mg mL-1. Based on this, we found that the calibration result is aligned with clinical reference range and can be applicable for identification of toxins in food. In addition, the coefficient of variation from this assay was found approximately within 2%, supporting the applicability of this method for real field analysis.

Figure 4
(a) Cyclic voltammograms; (b) calibration curve obtained from cyclic voltammetry (CV); and (c) Nyquist plots of the Cys-MAb-AFB1-BSA, system-control negative and system-AFB1 in the concentrations of 0.5 and 1 ng mL-1, 1, 2, 5, 10, 20 and 30 mg mL-1. Insert in (c) represents the equivalent circuit adopted to fit the impedance data, where RΩ is the ohmic resistance of the electrolyte solution, Q the phase constant element, W the Warburg impedance and Rct the electron-transfer resistance. Supporting electrolyte: 10 mM K4[Fe(CN)6]4-/K3[Fe(CN)6]3- (1:1) in phosphate buffer 10 mM, pH 7.4 solution; scan rate of 50 mV s-1.

To evaluate the specificity of the immunosensor, it was used a mycotoxin ochratoxin A as negative control. Viewing curve b (negative control) it is observed non-significant variations in the anodic and cathodic peaks of the voltammogram, demonstrating the specificity of the biosensor.

Figure 4c shows a successive increase in the Rct values with increasing of AFB1 concentration. EIS results are in accordance with coefficient of variation analysis. The reduction of the amperometric response and gradual increase in the Rct values (Figures 4a, 4b and 4c, respectively) results from a progressive blockage of the electron passage in the electrode/solution interface. This blockage is related to the amount of the antibody-antigen complex (MAb-AFB1) formed on the electrode surface. In Figure 4c, we observed an insignificant Rct value due to the interaction of the immunosensor with negative control, demonstrating the specificity of the immunosensor.

Additionally, the Nyquist plots were submitted to analysis of data through EQUIVCRT program.3333 Boukamp, B. A.; Sol. St. Ionics 1996, 18, 136. The experimental data were fitted using a modified Randles equivalent circuit (insert of Figure 4c), which allowed to obtain theoretical curves (Figure 4c). Warburg impedance (W) represents the impedance of semi-infinite diffusion of the redox probe to the electrode. The ohmic resistance of the solution (RΩ) is associated with the bulk properties of the electrolyte solution. Constant-phase element (Q) represents the behavior of a double layer for a non-homogeneous system. The phase angle of the Q element is related to n parameter, where n is related to the dispersion of relaxation time through the surface due to a non-homogeneity in the capacitance.3434 Silva, D. J. R.; Diniz, F. B.; Electrochim. Acta 2014, 119, 99. Electron transfer resistance controls the electron transfer kinetics of the redox-probe at the electrode/electrolyte interface.

Table 1 shows the equivalent circuit parameters of the fitting curves for the immunosensor preparation followed by antibody interaction with different concentrations of AFB1 in samples.

Table 1
Values of the equivalent circuit elements from fitted impedance results

The performance of the immunosensor for aflatoxin detection was evaluated through the relative variation of the Rct (DRct), according to the equation 1:

where Rct(toxin) is the value of the electron transfer resistance after recognition of the toxin and Rct(MAbA-FB1) corresponds to the Rct value of the sensor layer. ΔRct increases as function of toxin concentration (Figure 5a), indicating an interaction between antibody and AFB1. Thus, our results demonstrated that the modified electrode is effective on AFB1 detection. The filling of the recognition sites by toxin surface coverage (θ) can be calculated by (equation 2):

Figure 5
(a) Variation of electron transfer resistance (DRct); and (b) toxin surface coverage (q) as a function of aflatoxin B1 (AFB1) concentration. The experimental data used to calculate q is shown in Table 1, including the standard deviation. All experiments were performed in triplicate using three different samples.

where RB is the charge transfer resistance for Cys-Mab-AFB1 biosystem and RC is the charge transfer resistance obtained for different concentrations of the AFB1 after interaction with the immunosensor. Figure 5b shows a plot of θ as a function of concentration of toxin linkage to the immunosensor. The values of θ exhibit a linear relationship with toxin concentration and is found to be 75% for 30 mg mL-1 of toxin.

Sample analysis

After the initial analysis, the proposed immunosensor was applied to the determination of AFB1 in rice to test its performance in food samples. Figure 6shows the electrochemical behavior of the immunosensor using samples. In Figure 6a, we showed the CVs of the biosystem at different concentrations of AFB1 in rice. After the assemblage of the system on the electrode surface and the process of linkage between the antibody and AFB1 present in rice, it is possible to observe a decrease in the peaks current of the redox probe followed by an increase in the peak-to-peak separation. A more detailed study of the sensitivity of the immunosensor was conducted by EIS. Distinct concentrations of AFB1 in rice were analyzed (0.5 and 1 ng mL-1; 1, 5, 10, 20 and 30 mg mL-1) and the impedance spectra reveals an increase in the Rct proportional to AFB1 concentration (Figure 6b). At higher concentration, it can be observed an increased Rct value after recognition process of AFB1. In Figure 6c we observed a positive linear relationship between the DRct and AFB1 concentration available in contaminated rice. Therefore, our data indicates that the proposed sensor could detect AFB1 in spiked rice samples in small concentrations (1 mg mL-1), demonstrating that the established biosensing system could be applied for AFB1 determination in real agriculture products.

Figure 6
(a) Cyclic voltammograms; (b) Nyquist plots; and (c) variation of electron transfer resistance (DRct) of the immunosensor before and after interaction with aflatoxin B1 (AFB1) in rice at different concentrations. Cys-Mab-AFB1-BSA; system-control negative; and system-AFB1 at concentrations of 0.5 and 1 ng mL-1; 1, 2, 5, 10, 20 and 30 mg mL-1. Supporting electrolyte: 10 mM K4[Fe(CN)6]4-/K3[Fe(CN)6]3- (1:1) in 10 mM phosphate buffer (pH 7.4) solution; scan rate of 50 mV s-1.

Conclusions

A highly selective label-free impedimetric immunosensor based on cysteine monolayer and monoclonal antibody for AFB1 detection in rice was obtained. Cyclic voltammetry and electrochemical impedance techniques were used to investigate the immobilization of MAb on cysteine layer. EDC-NHS system was usefull to immobilize MAb without lost in the protein function. The biorecognition process was revealed by the blockage of the charge transfer. Obtained results demonstrated a good filling of the recognition sites by toxin surface coverage. The resulting device exhibited good sensitivity and reproducibility and could be used to detect toxins in food.

Acknowledgments

The authors are grateful for support from the Rede de Nanobiotecnologia/CAPES, Instituto Nacional de Ciência e Tecnologia para Inovação Farmacêutica (INCT-IF), CNPq (grant No. 310305/2012-8 and 310361/2012-5) and FACEPE.

References

  • 1
    Leung, M. C. K.; Diaz-Llano, G.; Smith, T. K.; J. Agric. Food Chem 2006, 54, 9623.
  • 2
    Bhat, R.; Rai, R. V.; Karim, A. A.; Compr. Rev. Food Sci. Food Saf 2010, 9, 57.
  • 3
    Rai, M. K.; Bonde, S. R.; Ingle, A. P.; Gade, A. K.; J. Pharm. Educ. Res. 2012, 3, 22.
  • 4
    Ding, X. X.; Li, P. W.; Bai, Y. Z.; Zhou, H. Y.; Food Control 2012, 23, 143.
  • 5
    Awad, W. A.; Ghareeb, K.; Böhm, J.; J. Anim. Physiol. Anim. Nutr. 2010, 94, 486.
  • 6
    Anklam, E.; Stroka, J.; Boenke, A.; Food Control 2002, 13, 173.
  • 7
    Nonaka, Y.; Saito, K.; Hanioka, N.; Narimatsu, S.; Kataoka, H.; J. Chromatogr. A 2009, 1216, 4416.
  • 8
    Bacaloni, A.; Cavaliere, C.; Cucci, F.; Foglia, P.; Samperi, R.; Lagana, A.; J. Chromatogr. A 2008, 1179, 182.
  • 9
    Li, P. W.; Zhang, Q.; Zhang, W.; TrAC, Trends Anal. Chem 2009, 28, 1115.
  • 10
    Otta, K. H.; Papp, E.; Bagocsi, B.; J. Chromatogr. A 2000, 882, 11.
  • 11
    Fernandez, A.; Belio, R.; Ramos, J. J.; Sanz, M. C.; Saez, T.; J. Sci. Food Agric 1997, 74, 161.
  • 12
    Jaimez, J.; Fente, C. A.; Vazquez, B. I.; Franco, C. M.; Cepeda, A.; Mahuzier, G.; Prognon, P.; J. Chromatogr. A 2000, 882, 1.
  • 13
    Stroka, J.; Anklam, E.; TrAC, Trends Anal. Chem 2002, 21, 90.
  • 14
    Sharma, A.; Matharu, Z.; Sumana, G.; Solanki, P. R.; Kim, C. G.; Malhotra, B. D.; Thin Solid Films 2010, 519, 1213.
  • 15
    Ricci, F.; Volpe, G.; Micheli, L.; Palleschi, G.; Anal. Chim. Acta 2007, 605, 111.
  • 16
    Tan, Y.; Chu, X.; Shen, G. L.; Yu, R. Q.; Anal. Biochem 2009, 387, 82.
  • 17
    Bacher, G.; Pal, S.; Kanungo, L.; Bhand, S.; Sens. Actuators, B 2012, 168, 223.
  • 18
    Perrotta, P. R.; Arevalo, F. J.; Vettorazzi, N. R.; Non, M. A.; Fernandez, H.; Sens. Actuators, B 2012, 162, 327.
  • 19
    Ronkainen, N. J.; Halsall, H. B.; Heineman, W. R.; Chem. Soc. Rev 2010, 39, 1747.
  • 20
    Brett, A. M. O.; Brett, C. M. A.; Eletrochemistry: Principles, Methods, and Applications; Oxford University Press: Oxford, 1993.
  • 21
    Bard, A. J.; Faulkner, L. R.; Electrochemical Methods: Fundamentals and Applications, 2nd ed.; John Wiley & Sons: Oxford, 2000.
  • 22
    Subramanian, A.; Irudayaraj, J.; Ryan, T.; Biosens. Bioelectron 2006, 21, 998.
  • 23
    Arya, S. K.; Solanki, P. R.; Datta, M.; Malhotra, B. D.; Biosens. Bioelectron 2009, 24, 2810.
  • 24
    Jiang, D.; Ji, J.; An, L.; Sun, X. L.; Zhang, Y.; Zhang, G.; Tang, L.; Biosens. Bioelectron 2013, 50, 150.
  • 25
    Nečas, D.; Klapetek, P.; Cent. Eur. J. Phys. 2012, 10, 181.
  • 26
    Galal, A.; Atta, N. F.; El-Ads, E. H.; Talanta 2012, 93, 264.
  • 27
    Huayhuas-Chipana, B. C.; Gomero, J. C. M.; Sotomayor, M. D. P. T.; J. Braz. Chem. Soc 2014, 25, 1737.
  • 28
    Yang, L. J.; Li, Y. B.; Biosens. Bioelectron 2005, 20, 1407.
  • 29
    Hao, C.; Yan, F.; Ding, L.; Xue, Y. D.; Ju, H. X.; Electrochem. Commun 2007, 9, 1359.
  • 30
    Luna, D. M. N.; Avelino, K. Y. P. S.; Cordeiro, M. T.; Andrade, C. A. S.; Oliveira, M. D. L.; Sens. Actuators, B 2015, 220, 565.
  • 31
    Geng, P.; Zhang, X.; Meng, W.; Wang, Q.; Zhang, W.; Jin, L.; Feng, Z.; Wu, Z.; Electrochim. Acta 2008, 53, 4663.
  • 32
    Ding, L.; Bond, A. M.; Zhai, J.; Zhang, J.; Anal. Chim. Acta 2013, 797, 1.
  • 33
    Boukamp, B. A.; Sol. St. Ionics 1996, 18, 136.
  • 34
    Silva, D. J. R.; Diniz, F. B.; Electrochim. Acta 2014, 119, 99.