Assessment of superficial rugosity of cast titanium in fluoride solutions and artificial salive

Abstract

The purpose of this study was to investigate the superficial rugosity of commercially pure titanium cast, under the action of different fluoride dental solutions and synthetic saliva. All specimens received the standard metallography procedure for titanium and were divided in groups: Group 1) synthetic saliva pH 7.0; Group 2) 1.23% APF gel pH 3.5;, Group 3) 2% NaF gel pH 6; Group 4) 0.05% NaF solution pH 4.0; and Group 5) 0.05% NaF solution pH 7.5. The specimens were exposed to these solutions for 16 minutes and immersed in synthetic saliva for 96 hours. The surfaces were analyzed in a roughness device and the results were statistically treated - Snededor 'F' and Bonferroni test. In addition, scanning electron microscopy was employed to observe the surface morphology after the test. The Group 2 was the only one which had a significant increase in the superficial roughness. Authors concluded that the fluoride solutions of dental use that have a high concentration of fluoride and low pH are harmful to the surface of cast titanium.

titanium; corrosion; fluoride solutions


LETTER TO THE EDITOR

Assessment of superficial rugosity of cast titanium in fluoride solutions and artificial salive

Viviane Fernandes BarrosI,* * e-mail: vivianefbarros@gmail.com ; Mauro Sayão de MirandaI; José Brant de CamposII

IDentistry Faculty, Rio de Janeiro State University, Av. 28 de setembro, 157, Vila Izabel, 20551-030 Rio de Janeiro - RJ, Brazil

IINational Institute of Technology, Av. Venezuela, 82, 20081-312 Rio de Janeiro - RJ, Brazil

ABSTRACT

The purpose of this study was to investigate the superficial rugosity of commercially pure titanium cast, under the action of different fluoride dental solutions and synthetic saliva. All specimens received the standard metallography procedure for titanium and were divided in groups: Group 1) synthetic saliva pH 7.0; Group 2) 1.23% APF gel pH 3.5;, Group 3) 2% NaF gel pH 6; Group 4) 0.05% NaF solution pH 4.0; and Group 5) 0.05% NaF solution pH 7.5. The specimens were exposed to these solutions for 16 minutes and immersed in synthetic saliva for 96 hours. The surfaces were analyzed in a roughness device and the results were statistically treated - Snededor 'F' and Bonferroni test. In addition, scanning electron microscopy was employed to observe the surface morphology after the test. The Group 2 was the only one which had a significant increase in the superficial roughness. Authors concluded that the fluoride solutions of dental use that have a high concentration of fluoride and low pH are harmful to the surface of cast titanium.

Keywords: titanium, corrosion, fluoride solutions

1. Introduction

Titanium has been developed with distinction in Dentistry, for being a light metal, biocompatible, with low thermal conductivity and excellent resistance to corrosion5. The resistance of metals to corrosion is directly related mainly to characteristics of their surface and to the medium in which the material is inserted. In relation of titanium, the presence of stable and dense oxide (TiO2) formed on this surface is responsible for its corrosion resistance4,5. Although the oxide layer is not destroyed under physiological conditions, solutions containing fluoride ions may decrease the resistance to corrosion of titanium3.

In dental treatment, titanium casts are commonly used to manufacture crowns, bridge frameworks, implant-supported structures, and full and partial denture frameworks with low thermal conductivity4. The success of dental treatment is dependent on the maintenance of good oral hygiene and cavity control. Fluoride has been used to help the professional in controlling the incidence of new caries and to decrease the dentin sensibility. It has been added to toothpaste, prophylactic agents, and dental rinse. However, a high fluoride concentration used in dental treatment might be harmful to titanium devices such as implants or restorations6,10.

The concern of many researchers has been with the effect of fluoride on titanium. Several authors have reported that titanium is easily degraded while in contact with fluoride acid solutions1-3,7,8,10-12. Such a phenomenon is interpreted as being a result from incorporation of fluoride ions in the oxide layer, in which the protective properties are the considerably reduced2. The formation of titanium fluoride oxide leads to many structure defects in such a layer. Furthermore the degradation of the titanium surface may increase the degree of rugosity of said surface, favouring the accumulation of bacterial plaque, thus impairing the treatment of restoration as well as the patient's oral health1. The aim of the present study, in vitro, was to assess the superficial behavior of cast titanium under the effect of different fluoride dental solution, by means of an analysis of superficial rugosity. In addition, scanning electron microscopy (SEM) was employed to observe the surface morphology after the test.

2. Materials and Methods

2.1. Specimen preparation and test solutions

Cylindrical specimen (20 mm x 5 mm, n = 5) consisting of commercially pure titanium ASTM®/grade 2 (Ti2 - Dentaurum J.P. Winkelstroeter KG- Pforzheim, Germany) were casted in the Rematitan system (Rematitan, Dentaurum J.P. Winkerstroeter KG, Pforzheim, Germany) by Vaiazzi Dental Laboratory (São Paulo - SP, Brazil). All specimens were cold mounted with epoxy resin and then received the standard metallography procedure for titanium. The specimens were mechanically polished on wet SiC paper grade 100-1000 (Strues, Denmark), and on successively finer grades of diamond paste 6, 3 and 1 μm (Strues, Denmark). Exposed surface areas of the specimens were 90 to 100 mm2. They were then degreased in ethanol and rinsed with distilled water. The polished surfaces were examined by optical microscopy (BX600, Olympus®) before fluoride attack.

All test solutions were prepared by Mil Fórmulas pharmacy (Rio de Janeiro - RJ, Brazil). The specimens (n = 5) were divided into 5 groups (one for each group) of different tested solutions: Group 1) synthetic saliva with pH 7.0 (control); Group 2) 1.23% acidulated phosphate fluoride (APF) gel with pH 3.5; Group 3) 2% NaF gel with pH 6.5; Group 4) 0.05% NaF solution with pH 4.0; and Group 5) 0.05% NaF solution with pH 7.5. The specimens were exposed to these solutions for 16 minutes. To simulate an intra oral condition, after attack the specimens were immersed in synthetic saliva waved for 96 hours in magnetic stirrer. The composition of test solutions used is in "Table 1". Test solutions were maintained at room temperature.

2.2. Analysis of rugosity and statistic treatment

The assessment of the rugosity of surfaces was performed with the help of a roughness device (SJ-201 P, Mitutoyo®). Such an apparatus has a kneading tip with 0.005 mm range size which moves at speed of 0.5 mm/min, connected with reading of the surface. The roughness device was duly calibrated according to the instruction guidebook which standardizes the initial reading over a standard surface provided by the manufacturer, in order to thereafter proceed with the subsequent reading of the specimens. A grid with 10 x 10 points was created, where the all specimens (n = 5) were scanned in a 1 cm course. The parameter used was Ra, which registers the average between the peaks and valleys of rugosity in μm.

To compare the rugosity of the specimens as received and of said specimen after being polished, the Student "t" was used. In the comparison of the polished specimen rugosity with the five attack groups, the Snedecor "F" test was used and when its value was meaningful the Bonferroni test was utilized.

2.3. SEM surface morfology

After the specimens were exposed to solutions, the surfaces were observed under the scanning electron microscopy (Zeiss® DSM 940, Germany).

3. Results

3.1. Analysis of roughness

  • Student "t" test for the specimens comparation as received vs. polished. t = 39.26 (p = 0.000) - meaningful at 1% level.

  • Snedecor "F" test for the specimens comparation as polished vs. Group 1 vs. Group 2 vs. Group 3 vs. Group 4 vs. Group 5 F = 14.19 (p = 0.000) - meaningful at 1% level. There was a meaningful difference between: Group 2 and polished; Group 2 and Group 1; Group 2 and Group 3; Group 2 and Group 4; Group 2 and Group 5.

3.2. Descriptive analysis of images under SEM

In the analysis performed under SEM different morphological aspects of the titanium surface were observed according to the solution exposed in different groups.

Group 1: A homogeneous surface was observed with no morphological aspect characteristic of corrosion (Figure 1a);


Group 2: A totally corroded surface was observed in alls its length, characteristic of a severe and uniform corrosion (Figure 1b);

Group 3: The specimen showed a formation of corrosion micro areas on the surface (Figure 1c); and

Group 4 and 5: The specimens showed a quite similar behavior among them, with generalized formation of pits all over their surface (Figure 1d).

4. Discussion

To Dentistry, the use of the superficial rugosity analysis of restoration materials, both of direct use and of indirect use, is of paramount importance, since an increase in the restoration will be more susceptible to the adherence of bacterial plaque. This could impair the patient's oral hygiene and health, which may cause the recurrence of cavity lesions and periodontal diseases.

In the present study, the acidulated phosphate fluoride gel entailed a meaningful increase in the superficial rugosity of cast titanium. Nevertheless, in others solutions studied, there was no meaningful difference in the superficial rugosity. Such findings are in accordance with others studies6,11. However Pröbster et al.6 found higher averages of rugosity both before, in the polished specimen (3.57 μm) as after the attack with the solution (6.19 μm). On the other hand, the arithmetic averages found in the present study for the same type of gel, with attack time of 16 minutes, were of before (0.26 μm, Table 2) and after the attack (0.52 μm, Table 2).

When analyzing the surface of the specimens under naked eye, soon after using the acidulated phosphate fluoride gel, a change in its color was observed, which was initially a mirrored metallic brilliant color. After the immersion, the specimen showed a grayish color with total loss of brilliance and there was a drastic increase in the gray shade with new uses of the gel. Such a discoloring was observed by the other authors3,12. Lausmaa et al.3 suggest that the acceleration of the corrosion products, caused by the attack of fluoride on the acceleration of the corrosion products, caused by the attack of fluoride on the titanium surface, entails the discoloring of the surface. However, Horasawa and Marek2 verified through a color sensor (MCR-A, MINOLVA) a discoloring of titanium even in the presence of low concentration of fluoride ions resulting from glass ionomer cement. Such results were not found in the preset study and no change was verified in the coloring of the surface in contact with the solutions of low fluoride concentration and of neutral gels.

In the present study, the artificial saliva solution chosen was regarded as control as several authors point out that titanium presents high resistance to corrosion in such a medium7,8.

According to the results of this study, the morphological aspect of titanium surface, when in contact with fluoride ions may vary from the presence of few corrosion pits in localized areas (Figure 1d) to a severe uniform corrosion (Figure 1b), comprising all the surface area of specimens with the destruction of all the oxided titanium film. Such morphological variations in corrosion generated by different fluoride solutions used in this study may be ascribed mainly to the difference of fluoride concentrations of the solutions, as well as to the pH solution which may even have influenced.

The fluoride solutions chosen for test medium in such a study regarding the concentration of fluoride and pH of said solutions are most commonly used in Dentistry. Yet, instead of using ready solutions available on the market, which present varied manufacturing dates as well as some unknown time on shelves, which may entail a reduction in the concentration shown on labels, we have opted for solutions compounded in pharmacies, especially for the research, preservative and sent free solutions and with fluoride concentrations and pH under control, which provides the study with a standardization, thus reducing the risk of undesirable variables.

The sixteen minutes time used in this study is equivalent to sixteen applications or rinses with fluoride solutions. As regard rinses, this would correspond to an interval in time of sixteen days, considering that patient would only have a rinse a day. As regard the gel of professional use, it would correspond to a much longer period of time, considering that, at least, such applications would be repeated weekly, being either monthly or with as even longer interval in time. In the literature studied, the time of contact of titanium with the fluoride solutions varies from minutes6 up to weeks2,9 of specimens in immersion in the attack medium.

The result of the current study show that the corrosion of titanium by fluoride is a important factor to be considered by the professional before the use of highly concentrated solutions, in patients who have titanium casting restorations or any other piece of titanium in the oral cavity. The fluoride agents must be carefully selected when used in such a type of patient. This shall become even more relevant with the increase of the number of patients treated with a titanium restoration. The superficial defects caused by the corrosion may entail an increase in the accumulation of bacterial plaque and discoloring in the titanium restorations, which may harm its aesthetic as well as its function. Such patients shall be instructed so as to avoid fluoride solutions, mainly those of high concentration, as well as dentist surgeons shall be more criterious as regard the use of topic applications of gels with high fluoride concentration.

5. Conclusions

The acidulated phosphate fluoride gel entailed a meaningful increase in the rugosity on the surface of the cast titanium, yet the other solutions tested did not show a meaningful difference in the superficial rugosity.

Received: July 15, 2008;

Revised: April 26, 2009

  • 1. Harzer W, Schröter A, Gedrange T, Muschter F. Sensitivity of titanium brackets to the corrosive influence of fluoride-containing toothpaste and tea. Angle Orthodontist 2001; 71(4):318-323.
  • 2. Horasawa N, Marek M. Effect of fluoride from glass ionomer on discoloration and corrosion of titanium. Journal of Dental Research 2001; 80:546.
  • 3. Lausmaa J, Kasemo B, Hansson S. Accelerated oxide growth on titanium implants during autoclaving caused by fluorine contamination. Biomaterials 1985; 6(1):23-28.
  • 4. Meyer JM, Lüthy H. Titan in der Zahnheilkunde- Welche Entwicklung glbt es bis heute? Quintessenz Zahntechv 1995; 21:681-712.
  • 5. Parr GR, Gardner LK, Toth RW. Titanium: the mystery metal of implant dentistry. Dental materials aspects. The Journal of Prosthetic Dentistry 1995; 54(3):410-414.
  • 6. Pröbster L, Lin W, Hüttemann H. Effect of fluoride prophylactic agents on titanium surfaces. The International Journal of Oral Maxillofacial Implants 1992; 7(3):390-394.
  • 7. Reclaru L, Meyer JM. Effects of fluorides on titanium and other dental alloys in dentistry. Biomaterials 1998; 19(1-3):85-92.
  • 8. Schiff N, Grosgogeat B, Lissac M, Dalard F. Influence of fluoride content and pH on the corrosion resistance of titanium and its alloys. Biomaterials. 2002; 23(9):1995-2002.
  • 9. Strietzel R, Hösch A, Kalbfleisch H, Buch D. In vitro corrosion of titanium. Biomaterials. 1998; 19(16):1495-1499.
  • 10. Toumelin-Chemia F, Rouelle F, Burdairon G. Corrosive properties of fluoride-containing odontologic gels against titanium. Journal of Dentistry. 1996; 24(1-2):109-115.
  • 11. Watanabe I, Watanabe E. Surface changes induced by fluoride prophylactic agents on titanium-based orthodontic wires. American Journal of Orthodontics and Dentofacial Orthopedics. 2003; 123(6):653-656.
  • 12. Wiig P, Ellingsen JE, Videm K. Corrosion of titanium by fluoride. Journal of Dental Research. 1993; 72:195.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      24 July 2009
    • Date of issue
      June 2009

    History

    • Accepted
      26 Apr 2009
    • Received
      15 July 2008
    ABM, ABC, ABPol UFSCar - Dep. de Engenharia de Materiais, Rod. Washington Luiz, km 235, 13565-905 - São Carlos - SP- Brasil. Tel (55 16) 3351-9487 - São Carlos - SP - Brazil
    E-mail: pessan@ufscar.br