Surface, microstructural, and adhesion strength investigations of a bioactive hydroxyapatite-titanium oxide ceramic coating applied to Ti-6Al-4V alloys by plasma thermal spraying

Abstract

Ti-6Al-4V alloy is employed in implants because of its good mechanical strength, excellent biocompatibility, and good resistance to corrosion in biological environments. Herein, a composite (HAp-TiO2, 50% by volume of both components) coating surface employing Ti-6Al-4V was characterized. Samples were analyzed post fabrication and post heat treatment to analyze the coating recrystallization, phases, crystallinity, porosity, and roughness. The coating showed rutile crystalline and amorphous Hap phases with crystallinity, porosity, and roughness of 55.6%, 13.6 ± 1.0%, and 4.2 ± 0.6 µm; whereas, after heat treatment, it showed a rutile hydroxyapatite phase and β-TCP with values 75.6%, 13.9 ± 1.9%, and 3.8 ± 0.2 µm, respectively. The composite exhibited 874 ± 26 HV100 hardness and 30 ± 2 MPa adhesion strength after heat treatment, which agree with previously reported data on other bioactive coatings. Therefore, this composite becomes much more crystalline after heating at 750 °C for 1 h.

Keywords:
Hydroxyapatite-titanium oxide coating; Ti-6Al-4V alloy; plasma thermal spraying; surface characterization


1. Introduction

Metallic materials have been used in implants to replace or repair human body parts over the last few decades. This is mainly due to two factors: higher life expectancy and a larger number of accidents involving means of transportation and extreme sports11 Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International. 2018;44(2):1250-1268..

Titanium alloys have been used as implant materials because of their desirable characteristics such as a low specific mass (as compared to stainless steel and cobalt-chromium alloys), high resistance to biocorrosion, biocompatibility, and outstanding mechanical properties. Titanium and its alloys stand out among the main metallic alloys typically used in implants, mainly the alpha-beta alloys with high mechanical strength such as Ti-6Al-4V (ASTM F67 and F136 or Ti-6Al-4V ELI). These alloys are now widely used for a number of clinical applications22 Ho WF. Effect of omega phase on mechanical properties of Ti-Mo alloys for medicals applications. Journal of Medical and Biological Engineering. 2007;28:47-51.,33 Chen Q, Thouas GA. Metallic implant biomaterials. Materials Science and Engineering. 2015;R 87:1-57..

One of the limitations of this alloy is the potential release of Al and V ions into the human body, because these ions can eventually cause long-term health issues11 Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International. 2018;44(2):1250-1268.

2 Ho WF. Effect of omega phase on mechanical properties of Ti-Mo alloys for medicals applications. Journal of Medical and Biological Engineering. 2007;28:47-51.
-33 Chen Q, Thouas GA. Metallic implant biomaterials. Materials Science and Engineering. 2015;R 87:1-57.. Recent research has focused on finding ways to prevent the release of these alloying elements and to increase the osseointegration of these metal alloys33 Chen Q, Thouas GA. Metallic implant biomaterials. Materials Science and Engineering. 2015;R 87:1-57.

4 Höhn S, Virtanen S. Effect of inflammatory conditions and H2O2on bare and coatedTi-6Al-4V surfaces: Corrosion behavior, metal ion release and Ca-Pformation under long-term immersion in DMEM. Applied Surface Science. 2015;357:101-111. doi:10.1016/j.apsusc.2015.08.261.
https://doi.org/10.1016/j.apsusc.2015.08...

5 Oliveira DP, Palmieri A, Carinci F, Bolfarini C. Gene expression of human osteoblasts cells on chemically treated surfaces of Ti-6Al-4V-ELI. Materials Science and Engineering: C. 2015;51:248-255. doi:10.1016/j.msec.2015.03.011.
https://doi.org/10.1016/j.msec.2015.03.0...
-66 Ma Z, Ren L, Liu R, Yang K, Zhang Y, Liao Z, Liu W, Qi M, Misra RDK. Effect of heat treatment on Cu distribution, antibacterial performance and cytotoxicity of Ti-6Al-4V-5Cu alloy. Journal of Materials Science and Technology. 2015;31:723-732. doi:10.1016/j.jmst.2015.04.002.
https://doi.org/10.1016/j.jmst.2015.04.0...
. Coating these alloys with bioactive ceramics is a common practice in orthopedics and dentistry, because it combines high mechanical strength, corrosion resistance, and ease of manufacturing metal implants with enhanced biocompatibility associated with bioactive ceramic films such as hydroxyapatite (HAp)11 Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International. 2018;44(2):1250-1268.

2 Ho WF. Effect of omega phase on mechanical properties of Ti-Mo alloys for medicals applications. Journal of Medical and Biological Engineering. 2007;28:47-51.
-33 Chen Q, Thouas GA. Metallic implant biomaterials. Materials Science and Engineering. 2015;R 87:1-57..

Coatings are applied to metal substrates through various methods such as sol-gel77 Milella E, Cosentino F, Licciulli A, Massaro C. Preparation and characterisation of titania/hydroxyapatite composite coatings obtained by sol-gel process. Biomaterials. 2001;22:1425-1431. doi:10.1016/S0142-9612(00)00300-8.
https://doi.org/10.1016/S0142-9612(00)00...
, biomimetic11 Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International. 2018;44(2):1250-1268., electrolytic11 Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International. 2018;44(2):1250-1268., sputtering ion coating11 Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International. 2018;44(2):1250-1268., physical vapor deposition88 Rahmati B, Sarhan AAD, Zalnezhad E, Kamiab Z, Dabbagh A, Choudhury D, Abas WABW. Development of tantalum oxide (Ta-O) thin film coating on biomedical Ti-6Al-4V alloy to enhance mechanical properties and biocompatibility. Ceramics International. 2015;42:466-480. doi:10.1016/j.ceramint.2015.08.133.
https://doi.org/10.1016/j.ceramint.2015....
, and plasma thermal spray methods99 Tsui YC, Doyle C, Clyne TW. Plasma sprayed hydroxyapatite coatings on titanium substrates. Part 1: Mechanical properties and residual stress levels. Biomaterials. 1998;19:2015-2029. doi:10.1016/S0142-9612(98)00103-3.
https://doi.org/10.1016/S0142-9612(98)00...

10 Lu YP, Li MS, Li ST, Wang ZG, Zhu RF. Plasma-sprayed hydroxyapatite+titania composite bond coat for hydroxyapatite coating on titanium substrate. Biomaterials. 2004;25:4393-4403. doi:10.1016/j.biomaterials.2003.10.092.
https://doi.org/10.1016/j.biomaterials.2...
-1111 Chou B, Chang E. Interface investigation of plasma-sprayed hydroxyapatite coating on titanium alloy with ZrO2 intermediate layer as bond coat. Scripta Materialia. 2001;45:487-493.. The plasma spray process has the best chemical control, resistance to biocorrosion, and process efficiency among the aforementioned methods99 Tsui YC, Doyle C, Clyne TW. Plasma sprayed hydroxyapatite coatings on titanium substrates. Part 1: Mechanical properties and residual stress levels. Biomaterials. 1998;19:2015-2029. doi:10.1016/S0142-9612(98)00103-3.
https://doi.org/10.1016/S0142-9612(98)00...

10 Lu YP, Li MS, Li ST, Wang ZG, Zhu RF. Plasma-sprayed hydroxyapatite+titania composite bond coat for hydroxyapatite coating on titanium substrate. Biomaterials. 2004;25:4393-4403. doi:10.1016/j.biomaterials.2003.10.092.
https://doi.org/10.1016/j.biomaterials.2...
-1111 Chou B, Chang E. Interface investigation of plasma-sprayed hydroxyapatite coating on titanium alloy with ZrO2 intermediate layer as bond coat. Scripta Materialia. 2001;45:487-493..

The main concerns regarding the use of hydroxyapatite as a coating are related to substrate/coating interface instability (adhesion) and coating longevity in the physiological environment; studies have only assessed the coating efficiency over a short term (maximum of three years)1010 Lu YP, Li MS, Li ST, Wang ZG, Zhu RF. Plasma-sprayed hydroxyapatite+titania composite bond coat for hydroxyapatite coating on titanium substrate. Biomaterials. 2004;25:4393-4403. doi:10.1016/j.biomaterials.2003.10.092.
https://doi.org/10.1016/j.biomaterials.2...
,1212 Morris WF. Hydroxyapatite-coated implants: a case for their use. Journal of Oral and Maxillofacial Surgery. 1998;56:1303-1311..

Therefore, recent studies have sought to develop bioactive ceramic coatings with improved adhesion between the coating and metal implants to thus increase the service life of implants66 Ma Z, Ren L, Liu R, Yang K, Zhang Y, Liao Z, Liu W, Qi M, Misra RDK. Effect of heat treatment on Cu distribution, antibacterial performance and cytotoxicity of Ti-6Al-4V-5Cu alloy. Journal of Materials Science and Technology. 2015;31:723-732. doi:10.1016/j.jmst.2015.04.002.
https://doi.org/10.1016/j.jmst.2015.04.0...
,99 Tsui YC, Doyle C, Clyne TW. Plasma sprayed hydroxyapatite coatings on titanium substrates. Part 1: Mechanical properties and residual stress levels. Biomaterials. 1998;19:2015-2029. doi:10.1016/S0142-9612(98)00103-3.
https://doi.org/10.1016/S0142-9612(98)00...
,1010 Lu YP, Li MS, Li ST, Wang ZG, Zhu RF. Plasma-sprayed hydroxyapatite+titania composite bond coat for hydroxyapatite coating on titanium substrate. Biomaterials. 2004;25:4393-4403. doi:10.1016/j.biomaterials.2003.10.092.
https://doi.org/10.1016/j.biomaterials.2...
,1111 Chou B, Chang E. Interface investigation of plasma-sprayed hydroxyapatite coating on titanium alloy with ZrO2 intermediate layer as bond coat. Scripta Materialia. 2001;45:487-493.

12 Morris WF. Hydroxyapatite-coated implants: a case for their use. Journal of Oral and Maxillofacial Surgery. 1998;56:1303-1311.

13 Chalisgaonkar V, Das M, Balla VK. Laser processing of Ti composite coatings reinforced with hydroxyapatite and bioglass. Additive Manufacturing. 2018;20:134-143.
-1414 Li H, Khor KA, Cheang P. Thermal sprayed hydroxyapatite splats: nanostructures, pore formation mechanisms and TEM characterization. Biomaterials. 2004;25:3463-3471. Doi:10.1016/j.biomaterials.2003.10.051
https://doi.org/10.1016/j.biomaterials.2...
.

The aim of the present study was to characterize the surface of 50% - 50% by volume HAp-TiO2 coatings applied to the Ti-6Al-4V alloy by plasma thermal spraying.

2. Materials and Methods

2.1 Hydroxyapatite Powder (HAp)

The hydroxyapatite (HAp) used in this work was kindly provided by Inside Materiais Avançados Ltda., Belo Horizonte County, Minas Gerais State, Brazil. X-ray fluorescence assays (XRF) were conducted to identify the chemical composition in the aforementioned powder. The XRF assay was performed with the Rigaku RIX 3100 X-ray fluorescence spectrometer in the Materials Engineering Department (DEMA) at the Unicamp Mechanical Engineering School (FEM) - Campinas County, São Paulo State, Brazil. The XRF assay results are presented in Table 1.

Table 1
XRF data of the as-received hydroxyapatite powder.

The hydroxyapatite used in this work had a mean particle diameter of 8.67 ± 0.14 µm and D90 = 17.34 ± 0.58 µm.

2.2 Titanium oxide (TiO2)

The titanium oxide used was Metco 102 (Sulzer Metco), which was kindly provided by the Mechanical Engineering School of Unicamp. This titanium oxide is intended for plasma spraying applications, and its particle size ranges from 7.8 to 88 µm (nominal value). The TiO2 was characterized by XRF, particle size analysis, morphology, and scanning electron microscopy (SEM) prior to plasma spraying. The observed characteristics of TiO2 are presented in Table 2.

Table 2
Characteristics of the titanium oxide powder (Metco 102).

Powder granulometry is one of many variables that influences the quality of the plasma-sprayed coating due to the high heat extraction that is inherent to plasma thermal spraying. Very fine particles, of the order of 1066 Ma Z, Ren L, Liu R, Yang K, Zhang Y, Liao Z, Liu W, Qi M, Misra RDK. Effect of heat treatment on Cu distribution, antibacterial performance and cytotoxicity of Ti-6Al-4V-5Cu alloy. Journal of Materials Science and Technology. 2015;31:723-732. doi:10.1016/j.jmst.2015.04.002.
https://doi.org/10.1016/j.jmst.2015.04.0...
ºC/s, can solidify before they touch the substrate. These particles are then incorporated and encapsulated by the lamellar structure of the coating after they solidify, leading to a possible drop in the adhesion strength of the coating1111 Chou B, Chang E. Interface investigation of plasma-sprayed hydroxyapatite coating on titanium alloy with ZrO2 intermediate layer as bond coat. Scripta Materialia. 2001;45:487-493.. Therefore, it is recommended to use powders with a particle size ranging from 20 to 100 µm1616 McPherson R. The relationship between the mechanism of formation, microstructure and properties of plasma-sprayed coatings. Thin Solid Films. 1981;83:297-310. doi:10.1016/0040-6090(81)90633-7.
https://doi.org/10.1016/0040-6090(81)906...
. However, issues concerning the incorporation of solid particles by the coating can be addressed by adjusting operational variables such as the distance from the torch to the substrate, primary/secondary gas flow, and electric current in the plasma.

2.3 Ti-6Al-4V alloy

The Ti-6Al-4V alloy substrate was manufactured by the National Institute of Science and Technology in Biofabrication (BIOFABRIS - Brazil) by direct metal laser sintering (DMLS). The substrate had a diameter of 25.4 mm and height of 4.0 mm.

The chemical composition data obtained by XRF were provided by the alloy powder supplier as well as the ASTM F136 standard (which specifies the Ti-6Al-4V composition for application in surgical implants), and are presented in Table 3.

Table 3
Chemical composition of Ti-6Al-4V.

2.4 Determination of the composition of the HAp-TiO2 composite

Initially, the composition HAp-TiO2 was chosen because it was expected to adhere well to the metallic substrate. The ratio between Hap and TiO2 took into account the variation between the coefficient of thermal expansion of the coating and that of the substrate. Equations (6.8) and (6.9) by Ashby and Jones were applied for the elasticity modulus, whereas Equations [1] and [2] by Garmong and Shepard were applied for the elasticity modulus1717 Garmong G, Shepard LA. Matrix Strengthening Mechanisms of an Iron Fiber-Copper Matrix Composite as a Function of Fiber Size and Spacing. Metallurgical Transactions. 1971;2(1):175-180.:

(1) α cu = f · α HAp + f · α TiO 2

(2) α cl = 1 f α HAp + f α TiO 2

where α is the coefficient of thermal expansion, indices cu and cl represent the upper and lower limits of the composite, HAp represents hydroxyapatite, TiO2 represents titanium oxide, and f represents the volume fraction of the assessed material. The coefficient of thermal expansion and density of hydroxyapatite and titanium oxide are presented in Table 4.

Table 4
Thermal expansion coefficient and density of hydroxyapatite and titanium oxide.

2.5 Thermal plasma spraying process

The HAp-TiO2 composite was mixed by tumbling. Prior to coating, the substrate specimens were prepared by grinding with coarse alumina (Al2O3) particles (grain size of 60 mesh at 75 psi pressure, 90° angle, and grinding time 30 s) in order to improve substrate roughness, eliminate possible oxides that can influence coating adhesion, and facilitate mechanical anchoring of the HAp-TiO2 composite. The specimens were cleaned with compressed air after blasting to remove loose alumina from the blasted surface and to prepare them for thermal spraying.

The composite was sprayed onto the Ti-6Al-4V substrate using a plasma thermal spray gun (9MBII METCO), and the coatings were manually applied in passes. This pistol model used herein has a radial dust feed, works at low to medium power levels, which can reach a maximum of 40 kW (500 A and 80 V). The operational deposition conditions that were used are presented in Table 5. After the aspersion thermal process, some samples underwent heat treatment at 750 °C for one hour, in order to recrystallize and release stress.

Table 5
Parameters in the thermal plasma spray process.

2.6 Coating characterization

Optical microscopy (OM), SEM, and X-ray diffraction (XRD) were used to characterize the coating. Samples were prepared by sanding, polishing, and etching, as well as sanding with silicon carbide (SiC) meshes with granulates ranging from 220 to 1200. Next, mechanical polishing was performed using a nylon cloth soaked in 1 µm granulation diamond paste, followed by washing with acetone in an ultrasonic bath for 20 min. Finally, the Keller's reaction was performed. The diffractograms presented herein were collected with a voltage of 40 kV and a current of 10 mA in steps of 0.05° over an observation range of 20° ≤ 2θ ≤ 90°.

The crystallinity of the deposited coatings was assessed with Difract Suite Eva® 2010 software, version 1.3, which automatically calculated the crystallinity by simply selecting the boundary between the typical broad peak of an amorphous pattern and the discrete diffraction peaks. The crystallinity index that was calculated by the software is based on Equation (3), wherein Acris is the area of the crystalline peaks of the coating, because the XRD background represents the amorphous pattern area.

(3) I c = A cris A cris + A amor · 100 %

The coating roughness and porosity were evaluated to assess the physical properties of the coating. Porosity images were recorded with OM and captured with a digital camera coupled to the microscope. Next, these images were edited in the Adobe Photoshop® CS2 software to increase the contrast between the coating and pores. The porosity was assessed with a GSA (Semi-Automatic Granulometer) Image Analyzer. This software operates by counting pores at the intersection of lines that are pre-established by the analyzer.

2.7 Hardness tests

Coating hardness was assessed with the micro-hardness method according to the Vickers scale. A Shimadzu hardness tester, model HMV-2000, was used for the analysis at a 100-g load for 15 s. Five measurements were performed per sample, and the mean coating hardness of these measurements was considered.

2.8 Adhesion tests

The adhesion test was performed according to the ASTM C633-08 standard. Ten cylindrical specimens with a diameter of 25.4 mm and a length of 25.4 mm, as well as a self-aligning device, were produced. The self-aligning device had two degrees of freedom in order to avoid shear stress during the test. The device shown in Figure 1 was manufactured according to ASTM C633-08 recommendations. High-adhesion Scotch-Weld 2214 3M epoxy resin was used to bond the two metallic cylindrical pieces.

Figure 1
Fixing device used in the adhesion test. Adapted from ASTM C633-08.

The set was drawn into an EMIC conventional tensile testing machine with a 100 kN capacity. A 100 kN load cell was used in the study and the deformation rate was found to be 0.02 mm/s.

3. Results and Discussion

3.1 Determination of volumetric composition of the composite

The plot in Figure 2 illustrates the influence of titanium oxide addition on the thermal expansion coefficient of the formed composite by applying the values shown in Table 3 to Equations (1) and (2).

Figure 2
Influence of TiO2 on the thermal expansion coefficient of the HAp-TiO2 composite.

Figure 2 shows that the addition of titanium oxide to hydroxyapatite reduced the thermal expansion coefficient of the formed composite. The range of TiO2 volume varied from 47% to 57% near the thermal expansion coefficient of the substrate (Ti-6Al-4V), which was 9.2.10-6 °C-1. A composition with a higher volumetric fraction of hydroxyapatite was chosen for better composite biocompatibility and lesser difference in thermal expansion between the substrate and coating. This is because hydroxyapatite is a bio-active material whereas titanium oxide is bioinert1414 Li H, Khor KA, Cheang P. Thermal sprayed hydroxyapatite splats: nanostructures, pore formation mechanisms and TEM characterization. Biomaterials. 2004;25:3463-3471. Doi:10.1016/j.biomaterials.2003.10.051
https://doi.org/10.1016/j.biomaterials.2...
,2222 Willmann G, Richter H, Wimmer M. Rotational bending test of hydroxyapatite plasma-coated models of Ti-6Al-4V. Biomedical Engineering/Biomedizinische Technik. 1993;38:6-14..

3.2 Characterization of ceramic powder before deposition

Figure 3 shows the morphology of the ceramic powders used in the present study, which exhibits granulometric dispersion as well as natural agglomeration of the particles. Pawlowski2323 Warashina H, Sakano S, Kitamura S, Yamauchi KI, Yamaguchi J, Ishiguro N, Hasegawa Y. Biological reaction to alumina, zirconia, titanium and polyethylene particles implanted onto murine calvaria. Biomaterials. 2003;24:3655- 3661. doi: 10.1016/S0142-9612(03)00120-0.
https://doi.org/10.1016/S0142-9612(03)00...
emphasized the importance of assessing the powders through SEM, since information about particle agglomeration is not recorded in the laser light scattering (LLS) process; accordingly, the two trials were performed. Titanium oxide, on the other hand, had an angular morphology with a well-defined shape and size.

Figure 3
Ceramic morphology assessment. (a) Titanium oxide powder; and (b) Hydroxyapatite powder.

3.3 Microstructural assessment and coating crystallinity

Microscopical analysis of the coatings formed through thermal spraying allowed for verification of several important structural features such as thickness and coating adhesion to the substrate. The thickness of the deposited coating was uniform and observed to be 66.7 ± 1.8 µm.

Complete overlap between the parts was observed in the substrate-coating interaction, without voids or cracks on the interface. There were fine lines parallel to the substrate in the structure, which indicates that the coating was formed by lamellar deposition, and also signifies the possible presence of diffusion processes that resulted in better coating adhesion (Figure 4 (a)).

Figure 4
Micrographs of Ti-6Al-4V alloy (a) substrate/coating interaction; amorphous and crystalline HAp-TiO2 phases after plasma spraying (500X), without etching (optical microscopy); (b) SEM image of the morphological aspect of the coating surface with splash (white arrows), and the resulting lamellae as a result of molten projection or semi-cast droplets (1000X), without etching.

Figure 4 (b) shows the microstructure formed by the flattened, solidified HAp-TiO2 droplets which were arranged in successive layers forming the coating. It also shows the addition of well-scattered lamellae, which suggests a higher composite plasticity state at the moment they reach the surface. The mechanism responsible for film formation is triggered by the accumulation of successive layers of bioceramic powder droplets that existed in the plasma torch environment. Finally, these droplets were launched against the substrate (at a speed in excess of 400 m/s) which resulted in the liquid coating. The coating was flattened and experienced partial elastic recoil at the moment of impact, subsequently solidifying to form the lamellar microstructure1515 Höhn S, Virtanen S. Biocorrosion of TiO2 nanoparticle coating of Ti-6Al-4V in DMEM under specific in vitro conditions. Applied Surface Science. 2015;329:356-362. doi:10.1016/j.apsusc.2014.12.114.
https://doi.org/10.1016/j.apsusc.2014.12...
.

The coating phases are shown in Figure 5 (a). Only peaks characteristic of the high-crystallinity rutile phase were observed before heat treatment. There were no expected Hap peaks or peaks in phases derived from hydroxyapatite decomposition. Amorphous poplars at angles 29°, 32°, 33°, 34°, and 39° were observed, and these angles correspond with the angles where Hap peaks are expected; this indicates a low crystallinity index in the Hap portion of the composite.

Figure 5
Diffractogram of HAp-TiO2 composite showing (a) the phases before the heat treatment; (b) the phases after the heat treatment.

The distinct behavior that was observed in the two ceramics after the thermal process in relation to crystallinity seems to be associated with the difference in the lattice parameter of the HA phase (a = 9.4 Å), which was significantly higher than that of TiO2 (4.6 Å). Since TiO2 has lower network parameter, it also has a lower tendency than HAp to become amorphous, due to the higher network parameter of HAp as compared with that of titanium oxide.

However, in addition to β-TCP formation and the presence of rutile, hydroxyapatite recrystallization was observed after the thermal treatment (Figure 4 (b)). No phase was formed from the reaction between HAp and TiO2.

Thus, there was a greater amount of the amorphous phase after deposition without heat treatment. In quantitative terms, the crystallinity under the deposited condition was 55.6%, whereas that after heat treatment was 75.6%.

3.4 Physical properties of coating

The rough surface on Ti-6Al-4V was necessary to impart good mechanical anchoring of HAp-TiO2 to the substrate. Roughness was generated by an abrasive alumina blasting process. Table 6 presents the surface roughness of the substrate and ceramic coating.

Table 6
Surface roughness of substrate and ceramic coating

The substrate and coating roughness were similar, because the ceramic-composite deposition resulted in successive lamellar layers. Apart from promoting good mechanical coating anchoring, the substrate roughness promoted roughness in the coating. Surface roughness is known to have a significant positive influence on bone growth2424 Hayashi K, Inadome T, Tsurnura H, Nakashima Y, Sugioka Y. Effect of surface roughness of hydroxyapatite-coated titanium on the bone-implant interface shear strength. Biomaterials. 1994;15(14):1187-1191..

Our results corroborate preliminary studies2525 Braceras I, Onate JI, Goikoetxea L, Viviente JL, Alava JI, de Maeztu MA. Bone cell adhesion on ion implanted titanium alloys. Surface and Coatings Technology. 2005;196:321-326. doi:10.1016/j.surfcoat.2004.08.201.
https://doi.org/10.1016/j.surfcoat.2004....
; 1) An HAp coating roughness of 3.8 ± 0.4 µm was recorded for a blasted surface and 4.23 ± 0.46 µm for the coatings, 2) a TiO2 film roughness of 2.8 ± 0.6 µm was recorded for a blasted surface, and 3) 4.6 ± 0.4 µm for the coatings. All the cited authors used plasma thermal spraying for coating deposition2626 Levingstone TJ, Ardhaoui M, Benyounis K, Looney L, Stokes JT. Plasma sprayed hydroxyapatite coatings: Understanding process relationships using design of experiment analysis. Surface and Coatings Technology. 2015;283:29-36..

The results for coating porosity are presented in Table 7. These values are similar to those recorded for the porosity of HAp coatings2727 Cannillo V, Lusvarghi L, Sola A. Production and characterization of plasma-sprayed TiO2-hydroxyapatite functionally graded coatings. Journal of the European Ceramic Society. 2008;28:2161-2169.. Lower porosity levels are a desirable property for coatings, because low porosity is known to minimize toxic metal ion dissolution and release from the substrate into the human body2727 Cannillo V, Lusvarghi L, Sola A. Production and characterization of plasma-sprayed TiO2-hydroxyapatite functionally graded coatings. Journal of the European Ceramic Society. 2008;28:2161-2169..

Table 7
Coating porosity data.

Porosity levels close to those recorded in this study have been reported in the literature. Khor et al.2929 Khor KA, Gu YW, Quek CH, Cheang P. Plasma spraying of functionally graded hydroxyapatite/Ti-6Al-4V coatings. Surface and Coatings Technology. 2003;168:195-201. doi:10.1016/S0257-8972(03)00238-X.
https://doi.org/10.1016/S0257-8972(03)00...
studied HAp coatings on Ti-6Al-4V alloys, and reported a coating porosity of 19%. Sun et al.3030 Sun L, Berndt CC, Grey CP. Phase, structural and microstructural investigations of plasma sprayed hydroxyapatite coatings. Materials Science and Engineering: A. 2003;360:70-84. doi:10.1016/S0921-5093(03)00439-8.
https://doi.org/10.1016/S0921-5093(03)00...
investigated the microstructure, structure, and phases of different HA coatings, and obtained coating at a porosity of up to 12%. The aforementioned authors attributed this result to the low power of the plasma equipment used in the study (27.5 kW). The power of the equipment was increased to 42.0 kW, which reduced the porosity level to 7%. The explanation for the porosity reduction to 7% is related to the incorporation of unfused particles, which decreases the coating porosity level. A plasma power of 28.8 kW was used in the present study and may have contributed to the observed increase in porosity.

3.5 Hardness

The developed composite exhibited a hardness of 523 ± 10 HV100 before and 874 ± 26 HV100 after the heat treatment. The higher hardness recorded after heat treatment is due to the enhanced crystallinity of the coating. A comparison with previous studies showed that the addition of titanium oxide to hydroxyapatite provided a significant enhancement in coating hardness, as shown in Table 8. The authors therein used deposition variables similar to those used in the present study, including the same metallic substrate.

Table 8
Hardness of HAp-TiO2 compared to values reported in the literature.

3.6 Adhesion strength

Tensile tests were performed to analyze the adhesion strength of the coating. The composite showed adhesion strength of 16.6 ± 2.0 MPa prior to heat treatment and 30.0 ± 2.0 MPa after thermal treatment. The coating herein presented better adherence than previously reported coatings which contained only hydroxyapatite. Table 9 presents the results from preliminary studies conducted over the last ten years.

Table 9
Adhesion strength of HAp-TiO2 compared to those reported in the literature

A mixed fracture, indicative of adhesive failure of the coating, was observed in the non-heat-treated coatings. However, cohesive failure was also observed in the adhesive. Figure 6 (a) shows the fracture; the region representing the failed adhesive is labeled with the letter A whereas that representing cohesive failure in the coating is labeled with the letter B. A fracture showing predominantly cohesive failure in the adhesive, as well as a small region showing adhesive failure between the coating and substrate, was observed in the specimens subjected to thermal treatment. Figure 6 (b) shows the adhesive failure region marked by the letter A.

Figure 6
(a) Non-heat-treated coating after failure; (b) Heat-treated coating after failure.

4. Conclusions

The best range for the HAp-TiO2 composition is between 47% and 57% for the addition of TiO2 to hydroxyapatite. This is so that the coefficient of linear thermal expansion of the composite is close to that of the Ti-6Al-4V alloy substrate. Thus, the HAp-TiO2 composite with 50% by volume of both components was chosen because it displayed the crystalline phase of amorphous TiO2 and HAp after spraying. Moreover, the rutile, recrystallized hydroxyapatite and β-TCP phases after thermal treatment showed no phases resulting from the reaction between hydroxyapatite and titanium oxide. Without heat treatment, 55.6% crystallinity was observed in the as-deposited specimen, whereas 75.6% crystallinity was observed after the heat treatment. Thus, heat treatment enabled an approximately 36% increase in crystallinity in the composite. There was no significant variation in roughness and porosity, which indicates that these properties were not influenced by the heat treatment. The composite had a HV100 hardness of 874 ± 26 and adhesion strength of 30.0 ± 2.0 MPa after the heat treatment. Both these values are higher than those described in the literature.

5. Acknowledgements

The authors would like to thank INCT-Biofabris for the manufacture of the test specimens through rapid prototyping, to Inside Ltda. for providing the hydroxyapatite, Labiomec/DEMA/FEM/UNICAMP for providing the titania, and CEFET-MG for the scanning electron microscope that was used in the microstructural analysis and Federal Institute of Espirito Santo (IFES) by the financial contribution intended for the revision of the article.

6. References

  • 1
    Harun WSW, Asri RIM, Alias J, Zulkifli FH, Kadirgama K, Ghani SAC, Shariffuddin JHM. A comprehensive review of hydroxyapatite-based coatings adhesion on metallic biomaterials. Ceramics International 2018;44(2):1250-1268.
  • 2
    Ho WF. Effect of omega phase on mechanical properties of Ti-Mo alloys for medicals applications. Journal of Medical and Biological Engineering 2007;28:47-51.
  • 3
    Chen Q, Thouas GA. Metallic implant biomaterials. Materials Science and Engineering 2015;R 87:1-57.
  • 4
    Höhn S, Virtanen S. Effect of inflammatory conditions and H2O2on bare and coatedTi-6Al-4V surfaces: Corrosion behavior, metal ion release and Ca-Pformation under long-term immersion in DMEM. Applied Surface Science 2015;357:101-111. doi:10.1016/j.apsusc.2015.08.261.
    » https://doi.org/10.1016/j.apsusc.2015.08.261
  • 5
    Oliveira DP, Palmieri A, Carinci F, Bolfarini C. Gene expression of human osteoblasts cells on chemically treated surfaces of Ti-6Al-4V-ELI. Materials Science and Engineering: C 2015;51:248-255. doi:10.1016/j.msec.2015.03.011.
    » https://doi.org/10.1016/j.msec.2015.03.011
  • 6
    Ma Z, Ren L, Liu R, Yang K, Zhang Y, Liao Z, Liu W, Qi M, Misra RDK. Effect of heat treatment on Cu distribution, antibacterial performance and cytotoxicity of Ti-6Al-4V-5Cu alloy. Journal of Materials Science and Technology 2015;31:723-732. doi:10.1016/j.jmst.2015.04.002.
    » https://doi.org/10.1016/j.jmst.2015.04.002
  • 7
    Milella E, Cosentino F, Licciulli A, Massaro C. Preparation and characterisation of titania/hydroxyapatite composite coatings obtained by sol-gel process. Biomaterials 2001;22:1425-1431. doi:10.1016/S0142-9612(00)00300-8.
    » https://doi.org/10.1016/S0142-9612(00)00300-8
  • 8
    Rahmati B, Sarhan AAD, Zalnezhad E, Kamiab Z, Dabbagh A, Choudhury D, Abas WABW. Development of tantalum oxide (Ta-O) thin film coating on biomedical Ti-6Al-4V alloy to enhance mechanical properties and biocompatibility. Ceramics International 2015;42:466-480. doi:10.1016/j.ceramint.2015.08.133.
    » https://doi.org/10.1016/j.ceramint.2015.08.133
  • 9
    Tsui YC, Doyle C, Clyne TW. Plasma sprayed hydroxyapatite coatings on titanium substrates. Part 1: Mechanical properties and residual stress levels. Biomaterials 1998;19:2015-2029. doi:10.1016/S0142-9612(98)00103-3.
    » https://doi.org/10.1016/S0142-9612(98)00103-3
  • 10
    Lu YP, Li MS, Li ST, Wang ZG, Zhu RF. Plasma-sprayed hydroxyapatite+titania composite bond coat for hydroxyapatite coating on titanium substrate. Biomaterials 2004;25:4393-4403. doi:10.1016/j.biomaterials.2003.10.092.
    » https://doi.org/10.1016/j.biomaterials.2003.10.092
  • 11
    Chou B, Chang E. Interface investigation of plasma-sprayed hydroxyapatite coating on titanium alloy with ZrO2 intermediate layer as bond coat. Scripta Materialia 2001;45:487-493.
  • 12
    Morris WF. Hydroxyapatite-coated implants: a case for their use. Journal of Oral and Maxillofacial Surgery 1998;56:1303-1311.
  • 13
    Chalisgaonkar V, Das M, Balla VK. Laser processing of Ti composite coatings reinforced with hydroxyapatite and bioglass. Additive Manufacturing 2018;20:134-143.
  • 14
    Li H, Khor KA, Cheang P. Thermal sprayed hydroxyapatite splats: nanostructures, pore formation mechanisms and TEM characterization. Biomaterials 2004;25:3463-3471. Doi:10.1016/j.biomaterials.2003.10.051
    » https://doi.org/10.1016/j.biomaterials.2003.10.051
  • 15
    Höhn S, Virtanen S. Biocorrosion of TiO2 nanoparticle coating of Ti-6Al-4V in DMEM under specific in vitro conditions. Applied Surface Science 2015;329:356-362. doi:10.1016/j.apsusc.2014.12.114.
    » https://doi.org/10.1016/j.apsusc.2014.12.114
  • 16
    McPherson R. The relationship between the mechanism of formation, microstructure and properties of plasma-sprayed coatings. Thin Solid Films 1981;83:297-310. doi:10.1016/0040-6090(81)90633-7.
    » https://doi.org/10.1016/0040-6090(81)90633-7
  • 17
    Garmong G, Shepard LA. Matrix Strengthening Mechanisms of an Iron Fiber-Copper Matrix Composite as a Function of Fiber Size and Spacing. Metallurgical Transactions 1971;2(1):175-180.
  • 18
    Lu X, Lin X, Chiumenti M, Cervera M, Li J, Ma L, Wei L, Hu Y, Huang W. Finite element analysis and experimental validation of the thermomechanical behavior in laser solid forming of Ti-6Al-4V. Additive Manufacturing 2018;21:30-40.
  • 19
    Sheppard LM. Surge in electronic materials continues. American Ceramic Society Bulletin 1992;70(9):1467.
  • 20
    Guo HL, Zhao XP. Preparation of a kind of red encapsulated electrophoretic ink. Optical Materials 2004;26:297-300.
  • 21
    Castañeda L, Alonso JC, Ortiz A, Andrade E, Saniger JM, Bañuelos JG. Spray pyrolysis deposition and characterization of titanium oxide thin films. Materials Chemistry and Physics 2002;77:938-944. doi:10.1002/crat.200710918.
    » https://doi.org/10.1002/crat.200710918
  • 22
    Willmann G, Richter H, Wimmer M. Rotational bending test of hydroxyapatite plasma-coated models of Ti-6Al-4V. Biomedical Engineering/Biomedizinische Technik 1993;38:6-14.
  • 23
    Warashina H, Sakano S, Kitamura S, Yamauchi KI, Yamaguchi J, Ishiguro N, Hasegawa Y. Biological reaction to alumina, zirconia, titanium and polyethylene particles implanted onto murine calvaria. Biomaterials 2003;24:3655- 3661. doi: 10.1016/S0142-9612(03)00120-0.
    » https://doi.org/10.1016/S0142-9612(03)00120-0
  • 24
    Hayashi K, Inadome T, Tsurnura H, Nakashima Y, Sugioka Y. Effect of surface roughness of hydroxyapatite-coated titanium on the bone-implant interface shear strength. Biomaterials 1994;15(14):1187-1191.
  • 25
    Braceras I, Onate JI, Goikoetxea L, Viviente JL, Alava JI, de Maeztu MA. Bone cell adhesion on ion implanted titanium alloys. Surface and Coatings Technology 2005;196:321-326. doi:10.1016/j.surfcoat.2004.08.201.
    » https://doi.org/10.1016/j.surfcoat.2004.08.201
  • 26
    Levingstone TJ, Ardhaoui M, Benyounis K, Looney L, Stokes JT. Plasma sprayed hydroxyapatite coatings: Understanding process relationships using design of experiment analysis. Surface and Coatings Technology 2015;283:29-36.
  • 27
    Cannillo V, Lusvarghi L, Sola A. Production and characterization of plasma-sprayed TiO2-hydroxyapatite functionally graded coatings. Journal of the European Ceramic Society 2008;28:2161-2169.
  • 28
    Singh G, Singh S, Prakash S. Surface characterization of plasma sprayed pure and reinforced hydroxyapatite coating on Ti-6Al-4V alloy. Surface and Coatings Technology 2011;205:4814-4820. doi:10.1016/j.surfcoat.2011.04.064.
    » https://doi.org/10.1016/j.surfcoat.2011.04.064
  • 29
    Khor KA, Gu YW, Quek CH, Cheang P. Plasma spraying of functionally graded hydroxyapatite/Ti-6Al-4V coatings. Surface and Coatings Technology 2003;168:195-201. doi:10.1016/S0257-8972(03)00238-X.
    » https://doi.org/10.1016/S0257-8972(03)00238-X
  • 30
    Sun L, Berndt CC, Grey CP. Phase, structural and microstructural investigations of plasma sprayed hydroxyapatite coatings. Materials Science and Engineering: A 2003;360:70-84. doi:10.1016/S0921-5093(03)00439-8.
    » https://doi.org/10.1016/S0921-5093(03)00439-8
  • 31
    Morks MF, Kobayashi A. Influence of spray parameters on the microstructure and mechanical properties of gas-tunnel plasma sprayed hydroxyapatite coatings. Materials Science and Engineering: B 2007;139:209-215. doi:10.1016/j.mseb.2007.02.008.
    » https://doi.org/10.1016/j.mseb.2007.02.008
  • 32
    Roy M, Balla VK, Bandyopadhyay A, Bose A. Compositionally graded hydroxiapatite/tricalcium phosphate coating on Ti by laser and induction plasma. Acta Biomaterialia 2011;7:866-873. doi: 10.1016/j.actbio.2010.09.016
    » https://doi.org/10.1016/j.actbio.2010.09.016
  • 33
    Gadow R, Killinger A, Stiegler N. Hydroxyapatite coatings for biomedical applications deposited by different thermal spray techniques. Surface and Coatings Technology 2010;205:1157-1164. doi:10.1016/j.surfcoat.2010.03.059
    » https://doi.org/10.1016/j.surfcoat.2010.03.059
  • 34
    Kweh SWK, Khor KA, Cheang P. An in vitro investigation of plasma sprayed hydroxyapatite (HA) coatings produced with flame-spheroidized feedstock. Biomaterials 2002;23:775-785. doi: 10.1016/S0142-9612(01)00183-1.
    » https://doi.org/10.1016/S0142-9612(01)00183-1
  • 35
    Yang XYC, Chang E. Influence of residual stress on bonding strength and fracture of plasma-sprayed hydroxyapatite coatings on Ti-6Al-4V substrate. Biomaterials 2001;22:1827-1836. doi:10.1016/S0142-9612(00)00364-1.
    » https://doi.org/10.1016/S0142-9612(00)00364-1
  • 36
    Zheng X, Huang M, Ding C. Bond strength of plasma-sprayed hydroxyapatite/Ti composite coatings. Biomaterials 2000;21:841-849. doi:10.1016/S0142-9612(99)00255-0.
    » https://doi.org/10.1016/S0142-9612(99)00255-0