Effectiveness of ozonated water in the reprocessing of blood dialyzers

Abstract

INTRODUCTION: Ozone is a potent antibacterial agent. Because ozone oxidizes organic material, it directly attacks microorganisms resulting in safe, fast and economical sterilization at low temperatures. This study evaluated the efficacy of ozonated water in the reprocessing of dialyzers obtained from a hemodialysis service in São José do Rio Preto. METHODS: A total of 23 blood dialyzers were collected at the conclusion of the morning hemodialysis shift. The equipment was collected after the maximum number of reuses, with the last disinfection performed with purified water (obtained by reverse osmosis and subsequent reverse ultrafiltration). The number and species of microorganisms in the dialyzers were evaluated before and after treatment with ozonated water. The ozonation of sterile deionized water was achieved by direct contact between the water and the ozone generating equipment. RESULTS:The mean number of microorganisms before sterilization was 1,47.10(9) colony forming units per ml (CFU/ml). After treatment with ozonated water, the number of microorganisms was 4,80.10¹ CFU/ml. CONCLUSION: Ozonated water is an effective decontaminant for most blood dialyzers.

Reprocessing; Hemodialysis; Ozone


ORIGINAL ARTICLE

Effectiveness of ozonated water in the reprocessing of blood dialyzers

Morian Lauana Miguelão CanadaI,II * * e-mail: mo_miguelao@hotmail.com , Ursulandrea Sanches AbelanI, Renato Amaro ZangaroII, Dora Inês Kozusny-AndreaniII, Roseli de Fátima Custódio YamazakiIII

ICentro Universitário de Rio Preto – UNIRP, Rua Ivete Gabriel Atique, 45, Boa Vista, CEP 15025-400, São José do Rio Preto, SP, Brasil

IIUniversidade Camilo Castelo Branco – UNICASTELO, São José dos Campos, SP, Brasil

IIIInstitute of Urology and Nephrology – IUN, São José do Rio Preto, SP, Brasil

ABSTRACT

INTRODUCTION: Ozone is a potent antibacterial agent. Because ozone oxidizes organic material, it directly attacks microorganisms resulting in safe, fast and economical sterilization at low temperatures. This study evaluated the efficacy of ozonated water in the reprocessing of dialyzers obtained from a hemodialysis service in São José do Rio Preto.

METHODS: A total of 23 blood dialyzers were collected at the conclusion of the morning hemodialysis shift. The equipment was collected after the maximum number of reuses, with the last disinfection performed with purified water (obtained by reverse osmosis and subsequent reverse ultrafiltration). The number and species of microorganisms in the dialyzers were evaluated before and after treatment with ozonated water. The ozonation of sterile deionized water was achieved by direct contact between the water and the ozone generating equipment.

RESULTS:The mean number of microorganisms before sterilization was 1,47.109 colony forming units per ml (CFU/ml). After treatment with ozonated water, the number of microorganisms was 4,80.101 CFU/ml.

CONCLUSION: Ozonated water is an effective decontaminant for most blood dialyzers.

Keywords: Reprocessing, Hemodialysis, Ozone.

INTRODUCTION

Chronic renal failure (CRF) represents a public health problem in Brazil with high rates of morbidity and mortality. According to the Brazilian Society of Nephrology, there were 47,063 patients on dialysis in 2000, and an estimated 92,091 patients on dialysis in 2010 (Sesso et al., 2011).

There are several choices for renal replacement therapy, including drug therapy, continuous ambulatory peritoneal dialysis, automated peritoneal dialysis, intermittent peritoneal dialysis, hemodialysis and renal transplantation (Barros et al., 2006; Figueiredo, 2010; Martins and Cesariano, 2005). Specialists, together with their patients, consider the degree of renal deterioration, a nutritional assessment and social and psycho economic factors to determine the appropriate treatment choice

In hemodialysis, the blood is filtered, and undesirable substances are removed. The procedure consists of the transfer of solutes between the blood and the dialysis solution through an artificial semi-permeable membrane (dialyzer or blood capillary). Solutes are transported by diffusion, ultrafiltration and convection (Barros et al., 2006; Lima and Santos, 2004).

Sustainability is important to hospitals because of the high rates of waste generation in healthcare. Reprocessing blood dialyzers decreases the production of solid health waste and is a common procedure in the United States. Reprocessing is practiced by more than 75% of the Institutes of Hemodialysis in the United States and saves the country over $200 million annually (Okechukwu et al., 2000).

Dialyzer reuse is considered safe and effective for thousands of patients. This practice requires high-flux dialyzers, which are more expensive and durable. The reuse of dialyzers decreases the incidence of reactions attributed to first use, and allows dialysis treatment to be widely available (Cho et al., 2004; Miles and Friedman, 1997).

According to resolution nr 154, after June 15th 2004, dialyzers and arterial/venous lines may be used for the same client up to twelve times using manual reprocessing, or up to twenty times when using automatic reprocessing. Dialyzers and lines eligible for reuse should be chemically disinfected. Sodium hypochlorite 0.6%, hydrogen peroxide (3% or less) and mixtures of peracetic acid/peroxide hydrogen/acetic acid are often used as disinfectants (Brasil, 2004; Riella, 2010).

Ozone is hypothesized to be a powerful sterilizing agent. It has excellent oxidant action, potent bactericidal effects after a few minutes of exposure and promotes a high-level of disinfection. It is a safe and fast technology, and an economical alternative for sterilization at low temperatures (Baysan et al., 2000; Bocci, 2005; Silva et al., 2009).

The present study evaluated the efficiency of ozonated water in the reprocessing of dialyzers from a hemodialysis service in São José do Rio Preto - SP.

METHODS

Dialyzers were obtained from a hemodialysis service in São José do Rio Preto-SP. A total of 23 blood dialyzers were collected between the 2nd and 31st of January, 2013. Dialyzers were obtained at the conclusion of the morning hemodialysis shift (Monday to Friday, representing 20% of the disposal of service). The dialyzers were manufactured by the Nipro Corporation (Osaka, Japan), and their structure was composed of polycarbonate, polyurethane and polyethylene terephthalate. The inner surface did not exceed 22 mm2, and the transmembrane pressure capacity was 500 mmHg. This equipment was being evaluated at the maximum number of reuses, according to the resolution RDC nr 154, June 15th, 2004 (Brasil, 2004). The last disinfection was performed with purified water (obtained by reverse osmosis followed by reverse ultrafiltration). After removing the dialyzers from the reuse room of the hemodialysis service, the equipment was placed in individual ice boxes and sent to the laboratory for analysis. The culture media used for the internal sample of dialyzers were Sabouraud-dextrose agar (OXOID) lab-InterlabDistributor of Scientific Laboratories of São Paulo-SP and Trypticase soy agar (Imedia Laboratorius PVT limited) Mumbai, Maharashtra, Índia.). Dialyzers were evaluated for the presence and total number of mesophilic organisms and their species. The dialyzers were then subjected to internal treatment with ozonated water and the microbiological analysis was repeated. The Ozon & Life corona ozone generator was connected to the oxygen cylinder and used for the ozonation of sterile deionized water. Ozone was steadily produced by the machine and transported by a silicone tube to the diffuser at a rate of 2ppm/minL. The water was directly exposed to ozone for 3 minutes through the diffuser at a controlled temperature of 25°C, and then used for the internal sterilization of the blood dialyzers. A system consisting of a bottle with a connector adapted to the dialyzer (Figure 1) was used to wash the internal compartments of the fibers with ozonated water.


The ozonated water flowed for 10 minutes inside the dialyzers. The collection of the internal material of the dialyzers was performed in laminar flow (Pachane brand) with a micropipettor. A quantity of 0.1 mL was removed from the internal contents of the dialyzers and used for inoculation of Petri plates containing the previously described culture media distributed through sterile glass beads. The Petri dishes were incubated in a BOD incubator at 37 °C for 24-48 hours to cultivate bacteria and yeast, and for two weeks to cultivate filamentous fungi. Microorganisms were quantified and expressed as colony forming units per mL (CFU/mL). The morphology of the colonies was using Gram staining and observed by light microscopy. After staining, colonies were transferred to different agarized media to maintain pure cultures. Gram-positive bacteria were grown on Baird-Parker Agar (Oxoid) with TSA medium, and Gram-negative bacteria were grown on Eosin Methylene Blue (EMB, Oxoid). Samples were incubated at 37°C for 24 hours. The yeasts were transferred to Sabouraud dextrose agar. This procedure was repeated three times before performing different biochemical tests.

Gram-negative bacteria were identified by the API20E system (Analytical Profile Index, BioMérieux). Coagulase and catalase, NaCl 5%, oxidase, novobiocin and DNAse tests were used for Gram-positive bacteria. Yeasts were grown on CHROMagar (Difco) and subsequently identified by germ tube formation, urease tests and fermentation of carbohydrates (maltose, sucrose, lactose, galactose, xylose and dextrose).

Data were analyzed by the Mann-Whitney nonparametric test and the chi-square test with 0.05 (5%) probability. Minitab 15

software was used for statistical analysis.

RESULTS

In total, 41 occurrences of microorganisms were verified in the blood dialyzers (34 before treatment with ozonated water, and seven after the treatment). Table 1 shows the number and species of microorganisms found in the dialyzers before and after exposure to ozonated water.

There was a significant presence of C. albicans in the evaluated dialyzers. Few bacteria were found, with the isolated species including Escherichia coli, Streptococcus sp, Micrococcus sp and Proteus mirabilis. Filamentous fungi were not isolated. Table 2 demonstrates significant differences in the total count of mesophilic microorganisms when comparing the dialyzers in relation to treatment with ozonated water (P < 0.001).

DISCUSSION

Improper maintenance of hemodialysis water systems and reprocessing of dialyzers may result in reactions in hemodialysis patients (Borges et al., 2007). The quality of dialysis fluids depends on complex systems with devices and procedures that allow for effective treatment. If the water is not treated properly, potentially harmful chemical or biological contaminants may be transferred to patients. This exposure may lead to the appearance of pyrogenic reactions and sepsis (Calderaro and Heller, 2001; Varo et al., 2007). Even after undergoing some treatment, the water may still be contaminated with microorganisms depending on the efficiency of the disinfection method used. Ineffective disinfection leads to biofilm formation in hemodialysis systems (Cappelli et al., 2006).

In the present study, microorganisms including Candida albicans, Escherichia Coli, Streptococcus sp, Micrococcus sp and Proteus mirabilis were identified before the reprocessing of blood dialyzers with ozonated water. These colonies corresponded to 34 patients with Candida albicans and seven occurrences of Micrococus after dialysis treatment (Table 1). Studies show that approximately 90% of the bacteria isolated in hemodialysis water are Gram-negative. This bacteria multiplies rapidly, reaching concentrations greater than 105 CFU/ml in less than 24 hours (Borges et al., 2007). Research has identified that the main microorganisms isolated in hospitals are Staplylococcus aureus, Streptococcus pneumoniae, Enterococcus, Pseudomonas sp, Escherichia coli and Candida albicans. The most common (95%) etiological agent associated with urinary tract infections (UTI) in elderly women is Escherichia coli. In men, Proteus mirabilis is the most commonly identified microorganism in hospitalized patients and Escherichia coli in outpatients (Abelan et al., 2013; Corrêa and Montalvão, 2010; Silva et al., 2008).

This study verified that the investigated samples contained high microbial load prior to treatment with ozonated water (Table 2). The pretreatment median values of CFU/ml were 2x109, and after disinfection with ozonated water, they were 0 (p < 0.001). Therefore, the ozonated water was effective in decontaminating blood dialyzers. Santos et al. (2007) compared ozone to peracetic acid in hydraulic circuits and hemodialysis machines. The average number of bacteria in the dialysis solution after peracetic acid treatment was 17 CFU/ml, and it was 10 CFU/ml (p < 0.05) when using ozone.

In this study, after the dialyzers were exposed to ozonated water, there was total elimination of the following bacterial species: E. coli, Streptococcus and Proteus mirabilis. A small number of the species Microccus sp, remained in the dialyzer (Table 1). Prabakaran et al. (2012) also reported that E.coli was highly sensitive to ozone treatment. According to Bocci et al. (2009), the antimicrobial action of ozone results from the inhibition of the metabolic activity of microorganisms. Ozone is therefore capable of broad spectrum disinfection by destroying the viral DNA and acting in the formation of free radicals (Stübinger et al., 2006).

The incidence of fungal infections has increased considerably in hospital settings, and the species Candida albicans has become a predominant pathogen for critically ill patients. Candida albicans results in high morbidity and mortality accounting for approximately 80% of documented hospital infections. Changes in host defense mechanisms are often the result of immunosuppression induced by medical procedures (Dignani et al., 2003; Hamill and Hollander, 1997; Penna, 1999). Several factors contribute to the virulence and pathogenicity of C. albicans including its ability to adhere to the epithelium and mucosa, the dimorphism that aids in tissue invasion, significant thermotolerance and its capacity to form biofilms (Ross et al., 2006).

In this study, the presence of Candida albicans in the 23 dialyzers evaluated prior to the disinfection was significant. After treatment with ozonated water, only six dialyzers were contaminated with C. albicans. However, a high reduction of CFU/mL (initially 2x109 CFU/ml, Table 2) was found. Abelan et al. (2013) determined that a 15 minute application of ozonated water was capable of sterilizing nail pliers contaminated with C. albicans.

According to the resolution RDC nr 154, of June 15th 2004 (Brazil, 2004), the water used in preparing the solution for dialysis services must have its quality guaranteed at all stages of treatment, storage and distribution through the monitoring of microbiological and physico-chemical parameters. A quantity of up to 200 CFU/ml of heterotrophic bacteria is acceptable with the absence in 100 mL of fecal coliforms and 2 EU/mL of endotoxins. These standards are associated with the prevention of clinical complications arising from bacterial contamination of the dialysis fluid. In this study, there was one dialyzer contaminated by Escherichia coli (Table 1), and a high incidence of mesophilic microorganisms above the parameters accepted by legislation (Lonneman, 2010).

The correct use of ozone dissolved in water may be a safe, practical, inexpensive and effective method for removing bacteria and biofilms. Ozone has a high oxidative capacity, affecting enzymatic systems and cell membranes. The oxidation causes cell lysis and prevents bacterial growth (Santos et al., 2007; Silva et al., 2009).

Thanomsub et al. (2002) found that the ozone treatment was able to modify and destroy the structure of Gram-positive and Gram-negative bacteria. After 30 minutes of ozone exposure, the number of bacteria in cultures of 103, 104, and 105 CFU/ml decreased. However, at higher bacterial loads (concentrations of 106 and 107 CFU/ml), there was a gradual decline in the survival of the cells, but the ozone was not effective against all bacterial cells even after a longer treatment period (150 minutes). Adequate concentrations of ozone can effectively result in antimicrobial activity by destroying the bacterial cell membrane and resulting in intracellular leakage and cell lysis (Thanomsub et al., 2002).

Currently, dialysis treatment seeks to reverse uremic problems by using modern machines for cardiopulmonary bypass and a filter that matches the blood dialyzer. This treatment leads to a decreased risk of morbidity and improves the quality of life and social reintegration of these patients. The use of ozonated water as the sterilizing agent for reprocessing dialyzers requires additional research. It is an innovative practice that is straightforward, inexpensive and benefits patients, the health team and the environment.

This study determined that ozonated water can effectively decontaminate most blood dialyzers resulting in a total elimination of microorganisms. The average microbial load before sterilization was 1,47.109 colony forming units per ml (CFU/ml). After treatment with ozonated water, the number of microorganisms was 4,80.101 CFU/ml. These results suggest that this biocide can be used in the control of pathogenic micro-organisms in health care, but should be considered microbial species involved.

ACKNOWLEDGEMENTS

We thank the Centro Universitário de São José do Rio Preto-SP (Unirp) and the Universidade Camilo Castelo Branco (UNICASTELO) for their support.

Received: 03 September 2013

Accepted: 05 May 2014

  • Abelan US, Zangaro RA, Kozusny-Andreani DI. Avaliação da atividade antimicrobiana da água ozonizada em alicates utilizados por manicures. In: Anais do I Encontro de Pós Graduação e Iniciação Científica UNICASTELO; 2013; Fernandópolis.
  • Barros EM, Manfro RC, Thomé FS, Goncalves LFS. Nefrologia: rotinas, diagnóstico e tratamento. 3. ed. Porto Alegre: Artmed; 2006.
  • Baysan A, Whiley RA, Lynch E. Antimicrobial effect of a novel ozone-generation device on micro-organisms associated with primary root carious lesions in vitro. Caries Research. 2000; 34(6):498-501.
  • Bocci VA. Ozone: a new medical drug. 2nd ed. Springer: The Netherlands; 2005.
  • Bocci VA, Borelli E, Travagli V, Zanardi I. The ozone paradox: ozone was a Strong oxidant as well as a medical drug. Medicinal Research Reviews. 2009; 29(4):646-82.
  • Borges CR, Lascowski KMS, Filho NR, Pelayo JS. Microbiological quality of water and dialysate in a hemodialysis unit in Ponta Grossa-PR, Brazil. Journal of Applied Microbiology. 2007; 103(5):1791-7.
  • Brasil. Agência Nacional de Vigilância Sanitária - Anvisa. Resolução RDC nº 154 de 15 de junho de 2004. Estabelece o regulamento técnico para o funcionamento dos serviços de diálise. Diário Oficial da União, Brasília, 17 jun. 2004.
  • Calderaro RVV, Heller L. Surto de reações hemolíticas associado a residuais de cloro e cloraminas na água de hemodiálise. Revista de Saúde Pública. 2001; 35(5):481-6.
  • Cappelli G, Riccardi M, Perrone S, Bondi M, Ligabue G, Albertazzi, A. Water treatment and monitor disinfection. Hemodialysis International. 2006; 10(1):13-8.
  • Corrêa EF, Montalvão ER. Infecções do trato urinário em geriatria. Estudos. 2010; 37(7):625-35.
  • Cho HK, Shin GT, Kim H. Status of dialyser reuse in Korea. Nephrology. 2004; 9(4):212-6.
  • Dignani MC, Solomkin, JS, Anaissie E. Candida. In: Anaissi E, Mc Ginnis MR, Pfaller MA, editors. Medical mycology. Filadélfia: Churchill Livingstone; 2003. p. 195-239.
  • Figueiredo NM. Práticas de enfermagem: ensinando a cuidar de clientes em situações clínicas e cirúrgicas. 1. ed. São Paulo: Yendis; 2010.
  • Lima EX, Santos I. Atualização em enfermagem em nefrologia. Rio de Janeiro: SOBEN; 2004.
  • Lonneman G. The quality of dialysate: an integrated approach. Kidney International. 2010; 58:112-9.
  • Hamill RJ, Hollander H. Infectious diseases: mycotic. In: Tierney Jr LM, McPhee SJ, Papadakis MA, editors. Connecticut: Appleton & Lange; 1997. p 1356-7.
  • Martins MRI, Cesariano CB. Qualidade de vida de pessoas com doença renal crônica em tratamento hemodialítico. Revista Latino-Americano Enfermagem. 2005; 13(5):670‑6. http://dx.doi.org/10.1590/S0104-11692005000500010
  • Miles AMV, Friedman EA. A review of hemodialyzer reuse. Seminars in Dialysis. 1997; 10(1):32-7.
  • Okechukwu CN, Orzol SM, Held PJ, Pereira BJ, Agodoa LY, Wolfe RA, Port FK. Characteristics and Treatment of Patients Not Reusing Dialyzers in Reuse Units. American Journal of Kidney Diseases. 2000; 36(5):991-9.
  • Prabakaran M, Tamil S, Merinal S, Panneerselvam A. Effect of ozonation on pathogenic bactéria. Advances in Applied Science Research. 2012; 3(1):299-302.
  • Penna GO. Doenças infecciosas e parasitárias: aspectos clínicos, de vigilância epidemiológica e de controle; guia de bolso. Brasília: Ministério da Saúde; 1999.
  • Riella MC. Princípios de nefrologia e distúrbios hidroeletroliticos. 5. ed. Rio de Janeiro: Guanabara; 2010.
  • Ross C, Quesada RMB, Girardello R, Rogeri LMS, Calixto LA, Pelayo JS. Análise microbiológica de pontas de cateteres venosos centrais provenientes de pacientes internados no Hospital Universitário da Universidade Estadual de Londrina. Semina: Ciências Biológicas e da Saúde. 2006; 27(2):117‑23. http://dx.doi.org/10.5433/1679-0367.2006v27n2p117
  • Santos F, Biernat JC, Santos AMG, Souza MELS, Raubach AS, Demin MSS. Desinfecção de máquinas de hemodiálise com ozônio. Jornal Brasileiro de Nefrologia. 2007; 24(1):14‑8.
  • Sesso R, Lopes AA, Thomé FS, Lugon JR, Santos DN. Relatório do censo brasileiro de diálise de 2010. Jornal Brasileiro de Nefrologia. 2011; 33(4):442-7.
  • Silva J, Ferreira S, Costa E, Resende A, Ramos MH. Agentes etiológicos e contaminantes em hemoculturas. Revista Portuguesa de Ciências Biomédicas. 2008; 111(3):18-21.
  • Silva RA, Garotti JER, Silva RSB, Navarini A, Pacheco AM Jr. Analysis of the bactericidal effect of ozone pneumoperitoneum. Acta Cirurgica Brasileira. 2009; 24(2):124-7.
  • Stübinger S, Saber R, Fillipi. A. The use of ozone in dentistry and maxillofacial surgery: a review. Quintessence International. 2006; 37(50):353-9.
  • Thanomsub B, Anupunpisit V, Chanphetch T. Effects of ozone treatment on cell growth and ultrastructural changes in bacteria. Journal of General and Applied Microbiology. 2002; 48(4):193-9.
  • Varo SD, Martins CHG, Cardoso MJO, Sartori FG, Montanari LB, Goncalves RHP. Isolamento de fungos filamentosos em água utilizada em uma unidade de hemodiálise. Revista da Sociedade Brasileira de Medicina Tropical. 2007; 40(3):326-31.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      24 Sept 2014
    • Date of issue
      Sept 2014

    History

    • Received
      03 Sept 2013
    • Accepted
      05 May 2014
    SBEB - Sociedade Brasileira de Engenharia Biomédica Sociedade Brasileira de Engenharia Biomédica, Centro de Tecnologia, , bloco H, sala 327 - Cidade Universitária, 21941-914 Rio de Janeiro - RJ , Tel./Fax.:: (55 21) 2562-8591 - Rio de Janeiro - RJ - Brazil
    E-mail: rbeb@rbeb.org.br