Influence of NiTi alloy on the root canal shaping capabilities of the ProTaper Universal and ProTaper Gold rotary instrument systems

Abstract

Objective

This study aimed to evaluate the influence of the NiTi wire in Conventional NiTi (ProTaper Universal PTU) and Controlled Memory NiTi (ProTaper Gold PTG) instrument systems on the quality of root canal preparation.

Material and Methods

Twelve mandibular molars with separate mesial canals were scanned using a high-definition microcomputed tomography system. The PTU and PTG instruments were used to shape twelve mesial canals each. The canals were scanned after preparation with F2 and F3 instruments of the PTU and PTG systems. The analyzed parameters included the remaining dentin thickness at the apical and cervical levels, root canal volume and untouched canal walls. Data was analyzed for statistical significance by the Friedman and Dunn’s tests. For the comparison of data between groups, the Mann-Whitney test was used.

Results

In the pre-operative analysis, there were no statistically significant differences between the groups in terms of the area and volume of root canals (P>.05). There was also no statistically significant difference between the systems with respect to root canal volume after use of the F2 and F3 instruments. There was no statistical difference in the dentin thickness at the first apical level between, before and after instrumentation for both systems. At the 3 cervical levels, the PTG maintained centralization of the preparation on the transition between the F2 and F3 instruments, which did not occur with the PTU. Conclusion The Conventional NiTi (PTU) and Controlled Memory NiTi (PTG) instruments displayed comparable capabilities for shaping the straight mesial root canals of mandibular molars, although the PTG was better than the PTU at maintaining the centralization of the shape in the cervical portion.

Endodontics; Micro-computed tomography; Nickel; Titanium; Root canal preparation

Introduction

Mechanical preparation of the root canal is an important step in endodontic treatment. The aim of this step is to remove vital or necrotic pulp tissue while simultaneously increasing the root canal volume to facilitate the decontamination of the root canal system by irrigants and medicaments1414 - Hülsmann M, Peters OA, Dummer PM. Mechanical preparation of root canals: shaping goals, techniques and means. Endod Topics. 2005;10:30-76..

Several types of instruments and techniques for root canal preparations have been described66 - Capar ID, Ertas H, Ok E, Arslan H, Ertas ET. Comparative study of different novel nickel-titanium rotary systems for root canal preparation in severely curved root canals. J Endod. 2014;40:852-6.,1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.,1616 - Paqué F, Peters OA. Micro-computed tomography evaluation of the preparation of long oval root canals in mandibular molars with the self-adjusting file. J Endod. 2011;37:517-21.

17 - Peters OA, Peters CI, Schönenberger K, Barbakow F. ProTaper rotary root canal preparation: effects of canal anatomy on final shape analysed by micro CT. Int Endod J. 2003;36:86-92.
-1818 - Sant'Anna Júnior A, Cavenago BC, Ordinola-Zapata R, De-Deus G, Bramante CM, Duarte MA. The effect of larger apical preparations in the danger zone of lower molars prepared using the Mtwo and Reciproc systems. J Endod. 2014;40:1855-9.. Because of the known limitations of hand-held, stainless steel instruments (such as the risk of deviations in the root canal preparation and excessive time consumption), nickel-titanium (NiTi) instruments have been widely used in endodontic practice due to their relatively greater accuracy and efficiency compared to hand-held instruments2424 - Wu H, Peng C, Bai Y, Hu X, Wang L, Li C. Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals. BMC Oral Health. 2015;15:27.. NiTi rotary files have significantly better flexibility and cutting abilities and provide higher quality in the preparation of root canals with low risk of deviation55 - Çapar ID, Arslan H. A review of instrumentation kinematics of engine-driven nickel-titanium instruments. Int Endod J. 2016;49:119-35.and higher fracture resistance to cyclic and torsional fatigue2525 - Ye J, Gao Y. Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 2012;38:105-7..

NiTi wires have undergone various modifications in fabrication, including different types of thermic treatments that have been employed during the manufacturing process, which are designed to optimize physical characteristics that favor greater elasticity as well as increased flexibility and resistance2020 - Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod. 2013;39:163-72.. Among the new NiTi wires now available on the market with the above-mentioned modifications, the CM wire is highlighted in the present study.

According to the manufacturers, CM wire instruments have a lower percentage of Nickel (by weight) than other NiTi instruments1010 - Elnaghy AM. Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J. 2014;47:1034-9.,2222 - Uygun AD, Kol E, Topcu MK, Seckin F, Ersoy I, Tanriver M. Variations in cyclic fatigue resistance among ProTaper Gold, ProTaper Next and ProTaper Universal instruments at different levels. Int Endod J. 201;49:494-9.. Additionally, because of the thermic treatments used during the manufacturing process, CM wire instruments do not bounce (rebound) to their original shape after being flexed. This characteristic, which is associated with greater flexibility, can improve the quality of the root canal preparation by reducing the risk of deviation, instrument fracture, and perforations1515 - Ninan E, Berzins DW. Torsion and bending properties of shape memory and superelastic nickel-titanium rotary instruments. J Endod. 2013;39:101-4..

Among the NiTi systems, the PTU (Dentsply Maillefer, Ballaigues, Switzerland) is a rotary system of conventional NiTi wire that has been widely used and studied11 - Arias A, Singh R, Peters OA. Torque and force induced by ProTaper universal and ProTaper next during shaping of large and small root canals in extracted teeth. J Endod. 2014;40:973-6.,66 - Capar ID, Ertas H, Ok E, Arslan H, Ertas ET. Comparative study of different novel nickel-titanium rotary systems for root canal preparation in severely curved root canals. J Endod. 2014;40:852-6.,1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.,2424 - Wu H, Peng C, Bai Y, Hu X, Wang L, Li C. Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals. BMC Oral Health. 2015;15:27.. It has a variable taper along the length of the instrument, a convex triangular cross-section, and a sharp tip77 - Ceyhanli KT, Erdilek N, Tatar I, Cetintav B. Comparative micro-computed tomography evaluation of apical root canal transportation with the use of ProTaper, RaCe and Safesider systems in human teeth. Aust Endod J. 2014;40:12-6.,1313 - Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:1134-8.. Another system, the PTG (Dentsply Maillefer, Ballaigues, Switzerland), was recently introduced to the market and has characteristics that are generally similar to those of the PTU, except that the PTG uses CM wire, while the PTU uses conventional NiTi wire. Because of this distinction, the PTG exhibits higher flexibility and greater resistance to cyclic fatigue than the PTU system1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.,1313 - Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:1134-8..

The aim of the present study was to evaluate the influence of the NiTi wire in PTU and PTG instruments on the quality of root canal preparation, as assessed by microcomputed tomography (micro CT). The null hypothesis evaluated was that the type of NiTi wire has no influence on the root canal preparation.

Material and methods

Selection of the teeth

One hundred extracted human mandibular molars with complete apexes were selected. The teeth were scanned using a high-definition micro CT system (Skyscan 1174v2, Bruker-microCT, Kontich, Belgium). Twelve teeth were selected based on the following attributes: root canal anatomy classified as Vertucci type IV2323 - Vertucci FJ. Root canal anatomy of the human permanent teeth. Oral Surg Oral Med Oral Pathol. 1984;58:589-99., with distinct main foramens, and maximum curvature of 5°1919 - Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32:271-5..

Micro CT scanning procedures

The teeth were scanned as follows: 19 µm voxel size, 50 kV, 800 µA, and 0.8° step-size rotation using a 1024x1304 resolution. The obtained images were reconstructed with dedicated software (NRecon v.1.6.3; Bruker-microCT) and saved in axial sections in BMP format. The images were used to observe the root canal configuration and its initial volume.

Root canal preparation

A single operator performed all procedures. The access cavities were conventionally performed, and the root canal lengths were determined by the insertion of #10 K-files (Dentsply Maillefer, Ballaigues, Switzerland) until visualization in the apical foramen with a stereomicroscope (Carl Zeiss Micro-imaging, Göttingen, Germany). The working length (WL) was established as being 1 mm shorter than the real root canal length (RL). The #10 and #15 K-Files (Dentsply Maillefer, Ballaigues, Switzerland) were used in the WL before instrumentation with the rotary systems. Next, 6 mesio-buccal and 6 mesio-lingual canals in different teeth (n=12) were instrumented using the PTU system as follows: S1 (18/.02) and SX (19/.035) instruments in the cervical preparation and the S1 (18/.02), S2 (20/.04), F1 (20/.07), and F2 (25/.08) instruments were used in the WL, according to the manufacturer’s instructions.

The remaining root canals (n=12) were instrumented using the PTG system in the same sequence as the one used for the PTU group. Subsequently, the roots were scanned again as in the initial evaluation. After scanning, the root canals were prepared with the F3 (30/.09) instrument of each system. The X-Smart Plus motor (Dentsply Maillefer, Ballaigues, Switzerland) was used with a specific torque and velocity for each instrument, in accordance with the manufacturer’s instructions, and the instrumentation was performed with in-and-out motion strokes in an apical direction. For the irrigation procedure, a 2.5% sodium hypochlorite solution was used; the root canals were irrigated before, during, and after instrumentation. Two milliliters of solution were used between each instrument change. The final irrigation was made with 3 mL of 17% EDTA for 3 minutes, and then with 3 mL of saline solution. Next, a new scan of the roots was performed using previously set parameters.

Micro-CT measurements and evaluations

The CTAn v.1.12 software (Bruker-microCT) was used to measure the canal volumes as well as the dentin thickness in the mesial and distal walls in the cervical and apical portions. For pre- and post-operative measurements, images of 3 different slices of the cervical third were evaluated; the first slice was 1 mm below the level of bifurcation, and the other 2 slices followed apically at intervals of 1 mm. In the apical third, the first slice was obtained at 1 mm from the apical foramen and the other 2 slices followed cervically at 1 mm intervals. The minimal dentin thickness of the mesial and distal walls of the root canals was measured in millimeters. The CTAn v.1.12 software was used to measure the canal volume between the 1 mm apical level and at 1mm below the root furcation.

For analysis of the un-instrumented area, 3D models were created with color codes, and the pre- and post-operative images were recorded using automatic image registration. Prepared areas and stripes of the canals were compared using the CTVol v. 2.2.1 software (Bruker-microCT). The intact surface area of the canal was determined based on the number of static objects; i.e., markings present in the same position on the surface of the root canal before and after instrumentation1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.. The data was then converted and presented as: (a) percentage (%) of dentinal removal, (b) un-instrumented area, and (c) relative volume increase between groups.

The Shapiro-Wilks test was used to assess normality of the non-parametric data. Results of the volume increase, dentin thickness, and un-instrumented areas were compared using the Friedmann and Dunn’s tests for the intra-group analyses. For analysis of the percentage of dentinal wear, un-instrumented areas, and volume increase between the groups, the Mann-Whitney test was used. The level of significance was established at 5% (P<0.05).

Results

In the pre-operative analysis, there were no statistically significant differences in the area or volume of root canals between the PTU and PTG groups (P>0.05), indicating adequate pairing of the root canals. Table 1 shows the median, minimum, and maximum values of the root canal volume (mm33 - Brunson M, Heilborn C, Johnson DJ, Cohenca N. Effect of apical preparation size and preparation taper on irrigant volume delivered by using negative pressure irrigation system. J Endod. 2010;36:721-4.) and surface area (mm22 - Bramante CM, Berbert A, Borges RP. A methodology for evaluation of root canal instrumentation. J Endod. 1987;13:243-5.) before and after use of the F2 and F3 instruments. Also presented in Table 1 are the percentage of volume increase and the percentage of static voxel after use of the F2 and F3 instruments. No statistically significant differences were observed between the PTU and PTG systems with respect to the root canal volumes obtained after use of the F2 and F3 instruments (P>0.05). However, upon intra-group analysis of the PTU group, a statistically significant difference was observed between the pre-operative and post-operative volumes, but only after use of the F3 instrument (P<0.05). For the PTG group, a statistically significant difference was found between the volumes before and after use of the F2 and F3 instruments (P<0.05). Figure 1 depicts the three-dimensional reconstructions of mesial root canals before instrumentation (Figure 1A), and after instrumentation with F2 (Figure 1B) and F3 (Figure 1C) instruments.

Table 1
Median, minimum and maximum of the volume values (mm3), percentage of volume increase, surface area (mm2) and percentage of static voxel

Figure 1
Three-dimensional reconstructions of mesial root canals of a mandibular molar: (A) before instrumentation (green), (B) after instrumentation with the ProTaper Gold F2 (light red), and the ProTaper Universal F2 file (light blue). (C) Shows the mesial canals after instrumentation with ProTaper Gold F3 (dark red) and the ProTaper Universal F3 file (dark blue)

Table 2 shows the median, minimum and maximum values (presented as percentages) of dentin wear in the mesial and distal walls (mm) at the 3 apical and 3 cervical levels, before and after use of the F2 and F3 instruments for the PTU and PTG groups, respectively. There was no statistical difference between the percentage of dentin wear at the first apical level before and after instrumentation for both systems (P>0.05). At the 3 cervical levels, the PTG system maintained centralization of the preparation on transition between the F2 and F3 instruments, in contrast to the PTU system where this did not occur.

Table 2
Median, minimum and maximum percentage of mesial and distal wall wear at 3 apical and 3 cervical levels proportioned by the ProTaper Universal and ProTaper Gold instruments

For all segments that were evaluated, a dentin thickness of less than 0.4 mm was found regardless of the instrument and system used (Figure 2).

Figure 2
Graphic representation of the median, minimum and maximum values of the initial dentin thickness at the apical and cervical portions after use of the F2 and F3 instruments

Discussion

With respect to the specific root canal anatomy that was targeted for analysis in the present study, the data supports the null hypothesis because there were no statistically significant differences between the PTU and PTG instruments with respect to root canal enlargement or un-instrumented areas.

After a micro-CT evaluation, the mandibular molars to be examined were selected based on whether they: (a) had two canals in the mesial roots, (b) presented distinct foramens and (c) had no isthmuses related to type IV of Vertucci’s classification2323 - Vertucci FJ. Root canal anatomy of the human permanent teeth. Oral Surg Oral Med Oral Pathol. 1984;58:589-99.. This experimental model allowed us to compare the capabilities of two instrument systems based on the same anatomical condition1818 - Sant'Anna Júnior A, Cavenago BC, Ordinola-Zapata R, De-Deus G, Bramante CM, Duarte MA. The effect of larger apical preparations in the danger zone of lower molars prepared using the Mtwo and Reciproc systems. J Endod. 2014;40:1855-9..

Several methodologies such as histologic sections, serial axial sections, and scanning electron microscopy have been used to evaluate root canal preparations22 - Bramante CM, Berbert A, Borges RP. A methodology for evaluation of root canal instrumentation. J Endod. 1987;13:243-5.,44 - Burroughs JR, Bergeron BE, Roberts MD, Hagan JL, Himel VT. Shaping ability of three nickel-titanium endodontic file systems in simulated S-shaped root canals. J Endod. 2012;38:1618-21.,88 - De-Deus G, Barino B, Zamolyi RQ, Souza E, Fonseca A Jr, Fidel S, et al. Suboptimal debridement quality produced by the single-file F2 ProTaper technique in oval-shaped canals. J Endod. 2010;36:1897-900.; however, these methods compromise the integrity of the samples because they require the cutting of the specimens, which limits specimen use to a single analysis. By contrast, micro-CT is a nondestructive method that allows for the progressive evaluation and observation of the preparation instruments with different kinetics1818 - Sant'Anna Júnior A, Cavenago BC, Ordinola-Zapata R, De-Deus G, Bramante CM, Duarte MA. The effect of larger apical preparations in the danger zone of lower molars prepared using the Mtwo and Reciproc systems. J Endod. 2014;40:1855-9.. The evaluations conducted in the present study were made after the use of the F2 and F3 instruments.

Although rotary NiTi instruments are flexible, they can suffer flexural and/or torsional fatigue1313 - Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:1134-8.,1414 - Hülsmann M, Peters OA, Dummer PM. Mechanical preparation of root canals: shaping goals, techniques and means. Endod Topics. 2005;10:30-76.. With technological advancement, however, new generations of instruments have been developed with different cross sections, diameters, tapers, and blades that have various cutting angles. Interestingly, the most recent advances have been directed towards improvements of the NiTi wire, whereby subtle changes in the proportion of metals, and in thermomechanical treatments have enabled the development of instruments with greater flexibility and fracture resistance when compared to their previous generations44 - Burroughs JR, Bergeron BE, Roberts MD, Hagan JL, Himel VT. Shaping ability of three nickel-titanium endodontic file systems in simulated S-shaped root canals. J Endod. 2012;38:1618-21.,1010 - Elnaghy AM. Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J. 2014;47:1034-9.,1313 - Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:1134-8.. For instance, the CM wire is widely used in rotary instruments because it has greater flexibility than conventional NiTi wires1010 - Elnaghy AM. Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J. 2014;47:1034-9.,1515 - Ninan E, Berzins DW. Torsion and bending properties of shape memory and superelastic nickel-titanium rotary instruments. J Endod. 2013;39:101-4.,2121 - Testarelli L, Plotino G, Al-Sudani D, Vincenzi V, Giansiracusa A, Grande NM, et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod. 2011;37:1293-5.,2222 - Uygun AD, Kol E, Topcu MK, Seckin F, Ersoy I, Tanriver M. Variations in cyclic fatigue resistance among ProTaper Gold, ProTaper Next and ProTaper Universal instruments at different levels. Int Endod J. 201;49:494-9.. The CM wire is manufactured using a complex heat treatment, with variations in wire composition; i.e., it has a lower percentage (52% by weight) of nickel than the most commonly used wire, which is composed of 54.5 to 57% of nickel (by weight)2121 - Testarelli L, Plotino G, Al-Sudani D, Vincenzi V, Giansiracusa A, Grande NM, et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod. 2011;37:1293-5.. The PTU and PTG systems share a similar design, but they vary in the composition of the NiTi wire. One study assessed the shaping ability of these two systems, and found that the PTG produced less transportation and maintained more dentin than the PTU1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24..

Apical preparations with larger diameters have been suggested based on studies that showed that this approach facilitates more efficient irrigation in the apical region33 - Brunson M, Heilborn C, Johnson DJ, Cohenca N. Effect of apical preparation size and preparation taper on irrigant volume delivered by using negative pressure irrigation system. J Endod. 2010;36:721-4., better infection control and improved quality of the root canal filling99 - ElAyouti A, Kiefner P, Hecker H, Chu A, Löst C, Weiger R. Homogeneity and adaptation of endodontic fillings in root canals with enlarged apical preparation. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;108:e141-6.. A previous report showed that, in the mesial roots of mandibular molars, the increase of apical preparation before use of the 0.40 mm instrument significantly increased the root canal volume without significantly reducing the dentin thickness in the danger zone1818 - Sant'Anna Júnior A, Cavenago BC, Ordinola-Zapata R, De-Deus G, Bramante CM, Duarte MA. The effect of larger apical preparations in the danger zone of lower molars prepared using the Mtwo and Reciproc systems. J Endod. 2014;40:1855-9.. In the present study, the apical preparation was performed until use of the F3 (30.09) instrument, which promoted a significant increase in the root canal volume. In both groups the preparation was centralized at the last apical millimeter; however, in the 3 cervical levels evaluated, the PTG system maintained centralized shaping on transition between the F2 and F3 instruments, which did not occur with the PTU. Nonetheless, both systems maintained significant degrees of mesial and distal dentin wall thicknesses.

Another study by Gagliardi, et al.1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.(2015) showed that the PTG was associated with less deviation than the PTU, which could be explained by the curvature of the specimens, as that study used moderately curved (25°–35°) mesial canals of mandibular molars, in contrast to the present study that used straight root canals. Therefore, the degree of curvature of the root canals appears to be a key determinant of the performance of the NiTi CM wire.

For both groups, un-instrumented areas of the root canals were observed in all specimens, demonstrating that the PTU and the PTG instruments are not as capable of performing a complete mechanical cleaning of the dentin walls as other NiTi rotary instruments1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.,1212 - Gergi R, Osta N, Bourbouze G, Zgheib C, Arbab-Chirani R, Naaman A. Effects of three nickel titanium instrument systems on root canal geometry assessed by micro-computed tomography. Int Endod J. 2015;48:162-70.,2626 - Zhao D, Shen Y, Peng B, Haapasalo M. Root canal preparation of mandibular molars with 3 nickel-titanium rotary instruments: a micro-computed tomographic study. J Endod. 2014;40:1860-4.. The post-preparation percentage of un-instrumented areas prior to use of the F2 instrument was comparable to that obtained in a previous study that employed similar instruments and methodologies1111 - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.; however, in the present study, the root canals were enlarged before use of the F3 instrument, which promoted a significant reduction in un-instrumented areas. This increase in diameter after preparation may be associated with better mechanical cleaning. The dentin thickness at the cervical levels was significantly reduced with the use of the F3 instrument in the PTU group, and the distal wall thickness appeared to be the most at risk. However, after use of the F3 instrument, the dentin thickness at the distal wall, in the cervical portion, was, on average, larger than 0.4 mm in both groups.

Conclusions

The ProTaper Universal and ProTaper Gold systems exhibited similar capabilities for shaping the straight mesial root canals of mandibular molars. In the specific anatomical condition examined here, the final preparation with the F3 instrument, regardless of the system, enhanced the root canal volume and reduced the un-instrumented area. In conclusion, theControlled Memory NiTi ProTaper Gold PTG system was better at maintaining the centralization of shape in the cervical portion of the straight mesial root canals of mandibular molars than the conventional NiTi ProTaper Universal PTU system.

Acknowledgements

The authors wish to thank the State of São Paulo Research Foundation for funding support (FAPESP No. 2015/03829-1).

References

  • 1
    - Arias A, Singh R, Peters OA. Torque and force induced by ProTaper universal and ProTaper next during shaping of large and small root canals in extracted teeth. J Endod. 2014;40:973-6.
  • 2
    - Bramante CM, Berbert A, Borges RP. A methodology for evaluation of root canal instrumentation. J Endod. 1987;13:243-5.
  • 3
    - Brunson M, Heilborn C, Johnson DJ, Cohenca N. Effect of apical preparation size and preparation taper on irrigant volume delivered by using negative pressure irrigation system. J Endod. 2010;36:721-4.
  • 4
    - Burroughs JR, Bergeron BE, Roberts MD, Hagan JL, Himel VT. Shaping ability of three nickel-titanium endodontic file systems in simulated S-shaped root canals. J Endod. 2012;38:1618-21.
  • 5
    - Çapar ID, Arslan H. A review of instrumentation kinematics of engine-driven nickel-titanium instruments. Int Endod J. 2016;49:119-35.
  • 6
    - Capar ID, Ertas H, Ok E, Arslan H, Ertas ET. Comparative study of different novel nickel-titanium rotary systems for root canal preparation in severely curved root canals. J Endod. 2014;40:852-6.
  • 7
    - Ceyhanli KT, Erdilek N, Tatar I, Cetintav B. Comparative micro-computed tomography evaluation of apical root canal transportation with the use of ProTaper, RaCe and Safesider systems in human teeth. Aust Endod J. 2014;40:12-6.
  • 8
    - De-Deus G, Barino B, Zamolyi RQ, Souza E, Fonseca A Jr, Fidel S, et al. Suboptimal debridement quality produced by the single-file F2 ProTaper technique in oval-shaped canals. J Endod. 2010;36:1897-900.
  • 9
    - ElAyouti A, Kiefner P, Hecker H, Chu A, Löst C, Weiger R. Homogeneity and adaptation of endodontic fillings in root canals with enlarged apical preparation. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;108:e141-6.
  • 10
    - Elnaghy AM. Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J. 2014;47:1034-9.
  • 11
    - Gagliardi J, Versiani MA, Sousa-Neto MD, Plazas-Garzon A, Basrani B. Evaluation of the shaping characteristics of ProTaper Gold, ProTaper NEXT, and ProTaper Universal in curved canals. J Endod. 2015;41:1718-24.
  • 12
    - Gergi R, Osta N, Bourbouze G, Zgheib C, Arbab-Chirani R, Naaman A. Effects of three nickel titanium instrument systems on root canal geometry assessed by micro-computed tomography. Int Endod J. 2015;48:162-70.
  • 13
    - Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:1134-8.
  • 14
    - Hülsmann M, Peters OA, Dummer PM. Mechanical preparation of root canals: shaping goals, techniques and means. Endod Topics. 2005;10:30-76.
  • 15
    - Ninan E, Berzins DW. Torsion and bending properties of shape memory and superelastic nickel-titanium rotary instruments. J Endod. 2013;39:101-4.
  • 16
    - Paqué F, Peters OA. Micro-computed tomography evaluation of the preparation of long oval root canals in mandibular molars with the self-adjusting file. J Endod. 2011;37:517-21.
  • 17
    - Peters OA, Peters CI, Schönenberger K, Barbakow F. ProTaper rotary root canal preparation: effects of canal anatomy on final shape analysed by micro CT. Int Endod J. 2003;36:86-92.
  • 18
    - Sant'Anna Júnior A, Cavenago BC, Ordinola-Zapata R, De-Deus G, Bramante CM, Duarte MA. The effect of larger apical preparations in the danger zone of lower molars prepared using the Mtwo and Reciproc systems. J Endod. 2014;40:1855-9.
  • 19
    - Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32:271-5.
  • 20
    - Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod. 2013;39:163-72.
  • 21
    - Testarelli L, Plotino G, Al-Sudani D, Vincenzi V, Giansiracusa A, Grande NM, et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod. 2011;37:1293-5.
  • 22
    - Uygun AD, Kol E, Topcu MK, Seckin F, Ersoy I, Tanriver M. Variations in cyclic fatigue resistance among ProTaper Gold, ProTaper Next and ProTaper Universal instruments at different levels. Int Endod J. 201;49:494-9.
  • 23
    - Vertucci FJ. Root canal anatomy of the human permanent teeth. Oral Surg Oral Med Oral Pathol. 1984;58:589-99.
  • 24
    - Wu H, Peng C, Bai Y, Hu X, Wang L, Li C. Shaping ability of ProTaper Universal, WaveOne and ProTaper Next in simulated L-shaped and S-shaped root canals. BMC Oral Health. 2015;15:27.
  • 25
    - Ye J, Gao Y. Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 2012;38:105-7.
  • 26
    - Zhao D, Shen Y, Peng B, Haapasalo M. Root canal preparation of mandibular molars with 3 nickel-titanium rotary instruments: a micro-computed tomographic study. J Endod. 2014;40:1860-4.