Diagnostic Performance of Digital Radiograph and Low-Dose Computed Tomography for the Diagnosis of Fishbone Retention in the Oropharynx

Abstract

Introduction

Fishbone foreign body retention is one of the most common problem with various clinical manifestations from asymptomatic, abscess formation, and perforation to mediastinitis with subsequence morbidity and mortality. Accurately identifying the location of the fishbone leads to precise removal, which, in turn, prevents serious consequences. Digital radiographs have been widely used for diagnosis, but many studies show poor sensitivity.

Object

The present study was designed to compare the diagnostic performances of digital radiograph and low-dose computed tomography (CT) for fishbone retention and to demonstrate the radiation dose of the two modalities.

Methods

We collected 2 pieces of fishbone from each of the 15 species commonly eaten in Southeast Asia. We embedded each fishbone in a fresh pig’s neck, then subjected the pig’s neck to lateral soft tissue neck digital radiograph. The locations to embed included tonsil, base of tongue, and upper esophagus. Then, we subjected the same specimen to a CT scan. Two experienced radiologists interpreted each image.

Results

Visibility in the digital radiograph group was 13%, and in CT images group, it was 87% regardless of the locations. The average radiation dose from digital radiographs was 0.4 mGy (radiation dose field), while from CT images it was 8.6 mGy (CT dose index).

Conclusion

Most of the common fishbones in Southeast Asia could not be visualized by digital radiograph when embedded in the neck. Computed tomography scans demonstrated better diagnostic performance of fishbone retention compared to digital radiographs, regardless of the embedded location.

Keywords
fishbone; oropharyngeal; esophageal; foreign body; diagnostic performance; CT; digital radiograph

Introduction

There are variations in the causes of food-borne foreign body disease among geographic regions and cultures. Oropharyngeal fishbone retention is one of the most common cause, particularly in the Asian culture and in coastal areas (42–76%).11 Kim HU. Oroesophageal fishbone foreign body. Clin Endosc 2016; 49(04):318–326,22 Klein A, Ovnat-Tamir S, Marom T, Gluck O, Rabinovics N, Shemesh S. Fishbone foreign body: the role of imaging. Int Arch Otorhinolaryngol 2019;23(01):110–115

The condition has spectrums of clinical manifestation from mild to fatal. The symptoms consist of foreign body sensation, irritation, sore throat, dysphagia, odynophagia, abscess, mediastinitis, and perforation.33 Dray X, Cattan P. Foreign bodies and caustic lesions. Best Pract Res Clin Gastroenterol 2013;27(05):679–689, 44 Venkatesh SH, Venkatanarasimha Karaddi NK. CT findings of accidental fishbone ingestion and its complications. Diagn Interv Radiol 2016;22(02):156–160, 55 Kim JP, Kwon OJ, Shim HS, Kim RB, Kim JH, Woo SH. Analysis ofclinical feature and management of fishbone ingestion of upper gastrointestinal tract. Clin Exp Otorhinolaryngol 2015;8(03):261–267 Approximately 80 to 90% of ingested foreign bodies are passed spontaneously.33 Dray X, Cattan P. Foreign bodies and caustic lesions. Best Pract Res Clin Gastroenterol 2013;27(05):679–689 About 10 to 20% require intervention for removal.55 Kim JP, Kwon OJ, Shim HS, Kim RB, Kim JH, Woo SH. Analysis ofclinical feature and management of fishbone ingestion of upper gastrointestinal tract. Clin Exp Otorhinolaryngol 2015;8(03):261–267 Frequent lodging sites in the oropharynx are the tonsils, base of the tongue, valleculae, and pyriform recesses. The tonsils are the most common site, followed by the base of the tongue. The most common location in the esophagus is the upper esophageal sphincter.11 Kim HU. Oroesophageal fishbone foreign body. Clin Endosc 2016; 49(04):318–326,22 Klein A, Ovnat-Tamir S, Marom T, Gluck O, Rabinovics N, Shemesh S. Fishbone foreign body: the role of imaging. Int Arch Otorhinolaryngol 2019;23(01):110–115

There is still no clear algorithm for the diagnosis and management of fishbone retention.66 Chiu HS, Chung CH. Management of foreign bodies in throat: an emergency department’s perspective. Hong Kong J Emerg Med 2002;9:126–130

InThailand, digital radiographs are still the imaging modality of choice for initial evaluation in the treatment of fishbone retention. However, many studies have revealed poor sensitivity and specificity of this imaging modality.11 Kim HU. Oroesophageal fishbone foreign body. Clin Endosc 2016; 49(04):318–326,22 Klein A, Ovnat-Tamir S, Marom T, Gluck O, Rabinovics N, Shemesh S. Fishbone foreign body: the role of imaging. Int Arch Otorhinolaryngol 2019;23(01):110–115,77 Sanei-Moghaddam A, Sanei-Moghaddam A, Kahrobaei S. Lateral soft tissue x-ray for patients with suspected fishbone in oropharynx, a thing in the past. Iran J Otorhinolaryngol 2015;27(83):459–462, 88 Kasemsiri P, Mahawerawat K, Ratanaanekchai T, Puttarak W, Munkong W. The accuracy of digital radiography for diagnosis of fishbone foreign bodies in the throat. Int Arch Otorhinolaryngol 2017;21(03):255–258, 99 Qureshi TA, Awan MS, Hussain M, Wasif M. Effectiveness of plain X-ray in detection of fish and chicken bone foreign body in upper aerodigestive tract. J Pak Med Assoc 2017;67(04):544–547, 1010 Woo SH, Kim KH. Proposal for methods of diagnosis of fishbone foreign body in the Esophagus. Laryngoscope 2015;125(11): 2472–2475 And the fact that the bones ofdifferent species of fish vary in degree of opacification under plain film leads to doubt about the usefulness of this modality. Computed tomography (CT) has emerged as an optional imaging modality to evaluate foreign body retention. As technology progressed, there is now better CT imaging quality, and it can produce lower radiation dose1111 Schmid AI, Uder M, Lell MM. Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation. Dentomaxillofac Radiol 2017;46(01): 20160131. Doi: 10.1259/dmfr.20160131. The present study intended to evaluate the diagnostic performance of plain films compared to CT scans, and the added value of CT scans to plain films.

Methods

We collected two pieces of fishbone from each of the 15 species commonly eaten in Southeast Asia, particularly in Thailand (►Table 1). Each piece measured approximately 2cm. A radiograph of all 30 pieces of fishbone was taken (►Fig. 1).

Table 1
Visibility of each fishbone species by imaging modality and embedded site

Fig. 1
Radiograph of 30 fishbones 1 = Common Silver Barb; 2 = Nile Tilapia; 3 = Red Tilapia; 4 = Tongue Sole; 5 = Short-bodied Mackerel; 6 = Catfish; 7 = Black Kingfish; 8 = Asian Seabass; 9 = Salmon; 10 = Areolated Grouper; 11 = Longtail Tuna; 12 = Silver Pomfret; 13 = Snakeskin Gourami; 14 = Spanish Mackerel; 15 = Sablefish

Next, we embedded each fishbone in a fresh pig’s neck, whose neck circumference and laryngeal framework was similar to that of an adult human.1212 Gao N, Cui X, Sun G, Zhang G, Zhou G, Zhao X. Comparative anatomy of pig arytenoid cartilage and human arytenoid cartilage. J Voice 2019;33(05):620–626. Doi: 10.1016/j.jvoice.2018.02.015
https://doi.org/10.1016/j.jvoice.2018.02...
For tonsils and the base of the tongue, we used an alligator forceps to hold the fishbone and embedded it under direct laryngoscope. For the esophageal location, we used the longer alligator forceps, under direct laryngoscope and with fluoroscopic guidance. (►Fig. 2)

Fig. 2
Demonstrate method of embedding fishbones in the pig neck.

We subjected the pig’s neck to lateral soft tissue neck digital radiograph, with 318 mA, 70 kVp, using a GE Discovery XR656, model AL01C II (GE Healthcare, Chicago, IL, USA). The locations used include the tonsil, base of the tongue, and the upper esophagus. Scout films of the pig’s neck were also obtained as control images.

Then, we subjected the same specimen to a CTscan, using Optima CT 660 (GE Healthcare), with standard plain CT protocol of the neck. The images were obtained and recorded by codes. Two experienced radiologists interpreted each image as “visualized” or “not visualized”. Whenever there was difference in the interpretation, the two radiologists got together to find a consensus.

We defined the diagnostic performance of each imaging modality per different location using nonparametric receiver operating characteristics (ROC) curveanalysis with Bamber and Hanley confidence intervals for the area under the ROC curve. To evaluate the additional benefit of CT images over digital radiographs, we also analyzed the added value of the ROC curve following each location where the fishbone was embedded.

Result

The visibility of the digital radiograph and CT images of the bone of each fish species is shown in table 1. Two pieces of fishbone from each species provided the same visibility results in both digital radiograph and CT images.

Visibility in the digital radiograph group was 13%, and, in the CT images group, it was 87%, regardless of the locations.

The average radiation dose from digital radiographs was 0.4 mGy (radiation dose field), while from CT images it was 8.6 mGy (CT dose index).

The accuracy of digital radiographs for the diagnosis of fishbone foreign bodies ranged from 53.3 to 58.3%, with lodging site in the upper esophagus showing the highest accuracy and lodging site in the tonsils showing the poorest diagnostic performance (►Table 2).

Table 2
Sensitivity, specificity, and accuracy with confidence intervals from digital radiographs in diagnosis of each site of fishbone retention

Computed tomography images demonstrated accuracy for the diagnosis of fishbone retention of 96.7% in all locations. The added diagnostic value of CT scan over digital radiographs was statistically significant in all lodging sites (►Table 3). Diagnostic performance of digital radiographs and CT images according to the lodging site is shown in ROC curves (►Fig. 3)

Table 3
Sensitivity, specificity, and accuracy with confidence intervals from computed tomography in diagnosis of each site of fishbone retention

Fig. 3
A receiver operator characteristics curve for diagnostic performance of digital radiograph (blue line) compared with computed tomography images (redline) at tonsillar region. 3b Receiver operator characteristics curve for diagnostic performance of digital radiograph (blue line) compared with computed tomography images (red line) at base of tongue. 3c Receiver operator characteristics curve for diagnostic performance of digital radiograph (blue line) compared with computed tomography images (red line) at upper esophagus.

Discussion

Digital radiograph has been widely used due to its availability and reasonable cost. In the past, the use of CT scans for bony foreign body identification was not common, because it is well known that bony structures can be seen with radiographs. However, unlike most bony foreign bodies, fishbones are small and show variation in opacity. Most of the common fishbones in Southeast Asia, particularly in Thailand, cannot be not visualized by digital radiograph when embedded in the neck. The tonsillar region shows the lowest sensitivity in visualization of embedded fishbones, probably due to surrounding air and overlying bony structures (►Fig. 4). However, some particular fishbones with greater opacification, such as catfish bones (number 6 in ►Fig. 1), can be visualized even in the tonsillar region (►Fig. 5).

Fig. 4
Tongue sole bone embedded at tonsillar region which is not visualized by digital radiograph.

Fig. 5
Catfishbone embedded at tonsillar region, visible by digital radiograph (arrow).

Overlying bony structures or surrounding air do not affect the visibility of fishbones in CT scan, differently from digital radiograph. Therefore, the locations of the embedded fishbone were accurately identified (►Fig. 6). However, some fishbone species which do not or faintly opacified cannot be visualized by CTscan (►Fig. 7), as demonstrated in ►Figure 1 numbers 11 (Longtail tuna) and 12 (Silver pomfret). The bones of these species are thin and show subtle opacification even when not embedded in soft tissue.

Fig. 6
6a Asian seabass bone embedded at tonsillar region, visualized by computed tomography (arrow). 6b Salmon bone at base of tongue, visualized by computed tomography (arrow).

Fig. 7
Long tail tuna bone, not visualized by computed tomography.

With developing technology, we can reduce the radiation dosage while maintaining diagnostic performance. Nonetheless, the CT radiation dosage is still about 20 times higher than that of digital radiograph. Radiation dosage is one of the factors that influences the choice of imaging modalities.

Conclusion

Computed tomography scans demonstrate better diagnostic performance of fishbone retention compared to digital radiographs, regardless of the embedded location. However, despite the decreased radiation dosage from CT scans, the total dose per study isstill 20 times higher than thatofdigital radiographs.

  • Funding
    Suranaree University of Technology.

References

  • 1
    Kim HU. Oroesophageal fishbone foreign body. Clin Endosc 2016; 49(04):318–326
  • 2
    Klein A, Ovnat-Tamir S, Marom T, Gluck O, Rabinovics N, Shemesh S. Fishbone foreign body: the role of imaging. Int Arch Otorhinolaryngol 2019;23(01):110–115
  • 3
    Dray X, Cattan P. Foreign bodies and caustic lesions. Best Pract Res Clin Gastroenterol 2013;27(05):679–689
  • 4
    Venkatesh SH, Venkatanarasimha Karaddi NK. CT findings of accidental fishbone ingestion and its complications. Diagn Interv Radiol 2016;22(02):156–160
  • 5
    Kim JP, Kwon OJ, Shim HS, Kim RB, Kim JH, Woo SH. Analysis ofclinical feature and management of fishbone ingestion of upper gastrointestinal tract. Clin Exp Otorhinolaryngol 2015;8(03):261–267
  • 6
    Chiu HS, Chung CH. Management of foreign bodies in throat: an emergency department’s perspective. Hong Kong J Emerg Med 2002;9:126–130
  • 7
    Sanei-Moghaddam A, Sanei-Moghaddam A, Kahrobaei S. Lateral soft tissue x-ray for patients with suspected fishbone in oropharynx, a thing in the past. Iran J Otorhinolaryngol 2015;27(83):459–462
  • 8
    Kasemsiri P, Mahawerawat K, Ratanaanekchai T, Puttarak W, Munkong W. The accuracy of digital radiography for diagnosis of fishbone foreign bodies in the throat. Int Arch Otorhinolaryngol 2017;21(03):255–258
  • 9
    Qureshi TA, Awan MS, Hussain M, Wasif M. Effectiveness of plain X-ray in detection of fish and chicken bone foreign body in upper aerodigestive tract. J Pak Med Assoc 2017;67(04):544–547
  • 10
    Woo SH, Kim KH. Proposal for methods of diagnosis of fishbone foreign body in the Esophagus. Laryngoscope 2015;125(11): 2472–2475
  • 11
    Schmid AI, Uder M, Lell MM. Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation. Dentomaxillofac Radiol 2017;46(01): 20160131. Doi: 10.1259/dmfr.20160131
  • 12
    Gao N, Cui X, Sun G, Zhang G, Zhou G, Zhao X. Comparative anatomy of pig arytenoid cartilage and human arytenoid cartilage. J Voice 2019;33(05):620–626. Doi: 10.1016/j.jvoice.2018.02.015
    » https://doi.org/10.1016/j.jvoice.2018.02.015

Publication Dates

  • Publication in this collection
    04 Nov 2022
  • Date of issue
    Jul-Sep 2022

History

  • Received
    05 Mar 2021
  • Accepted
    04 Aug 2021
  • Published
    08 Nov 2021
Fundação Otorrinolaringologia R. Teodoro Sampaio, 483, 05405-000 São Paulo/SP Brasil, Tel.: (55 11) 3068-9855, Fax: (55 11) 3079-6769 - São Paulo - SP - Brazil
E-mail: iaorl@iaorl.org