Recommendations for the Post-Mortem Management of Cardiac Implantable Electronic Devices

Abstract

The management of cardiac implantable electronic devices after death has become a source of controversy. There are no uniform recommendations for such management in Brazil; practices rely exclusively on institutional protocols and regional custom. When the cadaver is sent for cremation, it is recommended to remove the device due to the risk of explosion and damage to crematorium equipment, in addition to other precautions. Especially in the context of the SARS-CoV-2 pandemic, proper guidance and organization of hospital mortuary facilities and funeral services is essential to minimize the flow of people in contact with bodily fluids from individuals who have died with COVID-19. In this context, the Brazilian Society of Cardiac Arrhythmias has prepared this document with practical guidelines, based on international publications and a recommendation issued by the Brazilian Federal Medical Council.

Cardiac Resynchronization Therapy Device/complications; Principle Based Ethics; Autopsy/methods

Resumo

O manejo de dispositivos cardíacos eletrônicos implantáveis de pacientes que evoluem a óbito tem sido motivo de controvérsia. Em nosso meio, não há recomendações uniformes, estando baseadas exclusivamente em protocolos institucionais e em costumes regionais. Quando o cadáver é submetido para cremação, além de outros cuidados, recomenda-se a retirada do dispositivo devido ao risco de explosão e dano do equipamento crematório. Principalmente no contexto da pandemia causada pelo SARS-Cov-2, a orientação e organização de unidades hospitalares e serviços funerários é imprescindível para minimizar o fluxo de pessoas em contato com fluidos corporais de indivíduos falecidos por COVID-19. Nesse sentido, a Sociedade Brasileira de Arritmias Cardíacas elaborou este documento com orientações práticas, tendo como base publicações internacionais e recomendação emitida pelo Conselho Federal de Medicina do Brasil.

Dispositivos Cardíacos Eletrônicos Implantáveis/complicações; Ética Baseada em Princípios; Autópsia/métodos

The number of cardiac implantable electronic devices (CIEDs) placed, such as pacemakers, multisite devices, implantable cardioverter-defibrillators, and implantable loop recorder, has increased substantially in recent decades due to major diagnostic and therapeutic advances, as well as to the evident increases achieved in patient longevity (older adults may experience a longer life span as a result of CIED placement). The growing range of clinical situations in which these devices are used has been followed by an increase in complications and other related issues.

One major concern in relation to CIEDs—especially in the elderly population, due to the high mortality rates related to SARS-Cov-2 infection (COVID-19)—is their handling after death:

  • Should the CIED be turned off?

  • Should the CIED be removed before the patient’s burial?

  • If the body will be cremated, which precautions should be taken?

  • If the decision is made to remove the CIED, who should perform the procedure?

In response to these questions, the Brazilian Society of Cardiac Arrhythmias (SOBRAC) has developed a series of recommendations based on international evidence11. Gale CP, Mulley GP. Pacemaker explosions in crematoria: problems and possible solutions. J R Soc Med. 2002;95(7):353-5.,22. Pitcher D, Soar J, Hogg K, Linker N, Chapman S, Batle J, et al. Cardiovascular implanted electronic devices in people towards the end of life, during cardiopulmonary resuscitation and after death: guidance from the Resuscitation Council (UK), British Cardiovascular Society and National Council for Palliative Care. Heart. 2016;102 (Suppl 7):A1-A17., which we believe to be essential guidance for physicians, hospitals, and funeral services, which are faced with this situation on a daily basis:

  1. In cases of sudden death, the CIED should be electronically interrogated whenever the attending physician believes further elucidation of the cause of death is warranted, as the capability of these devices to record and store heart rhythm can be useful for this purpose.

  2. In cadaver with CIEDs who will be buried rather than cremated, neither reprogramming nor device removal is necessary.

  3. Conversely, cadaver with CIEDs whose remains will be cremated must always undergo device removal due to the risk of explosion and damage to crematorium facilities secondary to overheating (the device is exposed to a temperature of approximately 1300°C for 90 minutes).

  4. Implantable cardioverter-defibrillators (ICDs) in cadaver who will be cremated or will undergo autopsy must be disabled by telemetry prior to the start of either of these procedures, so as to prevent inadvertent activation and delivery of a shock to forensic or mortuary professionals during manipulation. Device reprogramming via telemetry should be performed by a physician with expertise in cardiac implantable electronic devices or by a qualified technician. For autopsy, one alternative that can be considered under exceptional circumstances is magnet placement over the ICD generator instead of reprogramming via telemetry. In such cases, the magnet must be kept over the generator even after device removal, since the risk of shock persists until the ICD is disabled by reprogramming.

  5. According to a guidance letter issued by the Federal Medical Council (see Appendix) on May 5, 2020, in response to a SOBRAC request, post-mortem removal of CIEDs should be carried out “preferably by a physician and duly noted in the patient’s medical record, as a safeguard during the time of the COVID-19 pandemic; hence, the importance of involving upper technical management so as to create the optimal protocol for each institution” (letter #2,628/2020 – DEPCO). SOBRAC warns that non-medical personnel (whether funeral directors or hospital staff) who may be assigned to this task must be properly trained to perform it, even though such a procedure is technically simple and does not endanger the life of the professional who performs it.

Understanding the aspects related to handling of CIEDs after death during the COVID-19 pandemic is essential to reducing the risk of disease spread.

The recommended CIED removal technique is as follows:

  • don appropriate personal protective equipment in order to avoid contamination by bodily fluids;

  • using a scalpel blade, deeply incise the skin overlying the pulse generator;

  • divide the deep tissues bluntly (with fingers) or sharply (with scissors);

  • pull out the generator and divide the electrodes with scissors;

  • close the incision with gauze and tape—there is no need for sutures;

  • dispose of the device safely, following the relevant technical recommendations and institutional protocols.

Referências

  • 1
    Gale CP, Mulley GP. Pacemaker explosions in crematoria: problems and possible solutions. J R Soc Med. 2002;95(7):353-5.
  • 2
    Pitcher D, Soar J, Hogg K, Linker N, Chapman S, Batle J, et al. Cardiovascular implanted electronic devices in people towards the end of life, during cardiopulmonary resuscitation and after death: guidance from the Resuscitation Council (UK), British Cardiovascular Society and National Council for Palliative Care. Heart. 2016;102 (Suppl 7):A1-A17.
  • Study Association
    This study is not associated with any thesis or dissertation work.
  • Sources of Funding .There were no external funding sources for this study.