Acessibilidade / Reportar erro

Vascular ultrasonography for follow-up of endovascular repair of abdominal aorta aneurysms

Abstracts

Background

There is little information available on follow-up of abdominal aortic aneurysm patients treated with endovascular repair using vascular ultrasonography in Brazil or on how it compares with the results of angiotomography. Since ultrasonography is an examination that is risk-free, inexpensive and widely available, it is a very attractive method. In this study we attempted to evaluate the sensitivity and specificity of vascular ultrasonography for follow-up of these patients by comparing the method with angiotomography.

Materials and methods

We conducted a prospective study from June 2012 to May 2013. We examined patients followed-up at the endovascular surgery clinic run by the University Hospital of Londrina after endovascular repair of abdominal aortic aneurysms. All patients underwent angiotomography for follow-up and were also examined using simple abdominal X-rays and vascular ultrasonography.

Results

A total of 30 patients were analyzed, with a mean age of 73 years and a mean aneurysm diameter of 6 cm. Four endoleaks were detected. Vascular ultrasonography achieved sensitivity of 75% and specificity of 96% for classification of these endoleaks, in comparison with angiotomography.

Conclusions

Vascular ultrasonography is an excellent primary method for evaluation and post-surgical follow-up of patients treated with endovascular repair of abdominal aorta aneurysms. However, when ultrasonography detects problems or is difficult to accomplish, it should be supplemented with a more specific investigation for diagnostic confirmation.

aortic aneurysm; ultrasonography; endoleaks


Contexto

As informações sobre o seguimento de pacientes submetidos à correção endovascular de aneurisma de aorta abdominal com ultrassonografia vascular no Brasil são escassas, bem como sua comparação com os resultados obtidos com a angiotomografia. Na medida em que a ultrassonografia é um exame sem riscos, de baixo custo e de grande disponibilidade, esta se torna um método bastante atraente. Na presente pesquisa, procuramos avaliar a sensibilidade e a especificidade da ultrassonografia vascular (USV) no acompanhamento desses pacientes, comparando este método com a angiotomografia.

Materiais e métodos

Realizamos estudo prospectivo durante o período de junho de 2012 a maio de 2013. Para tanto, examinamos pacientes acompanhados pelo Ambulatório de Cirurgia Endovascular do Hospital Universitário de Londrina pós-correção endovascular de aneurisma de aorta abdominal. Todos os pacientes haviam sido submetidos à angiotomografia para acompanhamento e foram também avaliados através da radiografia abdominal (raio x simples) e da ultrassonografia vascular.

Resultados

Foram analisados 30 pacientes, com média de idade de 73 anos, com diâmetro médio do aneurisma de 6 cm, detectando-se quatro vazamentos endovasculares. Na avaliação desses vazamentos endovasculares, a ultrassonografia vascular obteve uma sensibilidade de 75% e uma especificidade de 96%, em relação à angiotomografia.

Conclusão

A ultrassonografia vascular é um excelente método primário na avaliação e no acompanhamento pós-cirúrgico de pacientes submetidos à correção endovascular do aneurisma da aorta abdominal (AAA). No entanto, em caso de alteração ultrassonográfica ou dificuldade na realização do exame, uma investigação mais específica deve ser realizada para confirmação diagnóstica.

Aneurisma aórtico; ultrassonografia; vazamento endovascular


INTRODUCTION

Endovascular aneurysm repair, introduced by Parodi 1. Greenhalgh RM, Powell JT. Endovascular repair of abdominal aortic aneurysm. N Engl J Med. 2008;388:494-501. PMid:18234753. http://dx.doi.org/10.1056/NEJMct0707524
http://dx.doi.org/10.1056/NEJMct0707524...
,2. Stavropoulos SW, Charagundla SR. Imaging techniques for detection and management of endoleaks after endovascular aortic aneurysm repair. Radiology. 2007;243:641-55. PMid:17517926. http://dx.doi.org/10.1148/radiol.2433051649
http://dx.doi.org/10.1148/radiol.2433051...
in 1991, is an option that is widely used for open surgical repair in patients with aneurysms of the abdominal aorta (AAA). The technique was developed in order o avoid expansion and rupture of the aneurysm, by placement of an endoprosthesis within the aneurysm sac, forming a channel for blood to flow through.1. Greenhalgh RM, Powell JT. Endovascular repair of abdominal aortic aneurysm. N Engl J Med. 2008;388:494-501. PMid:18234753. http://dx.doi.org/10.1056/NEJMct0707524
http://dx.doi.org/10.1056/NEJMct0707524...
,2. Stavropoulos SW, Charagundla SR. Imaging techniques for detection and management of endoleaks after endovascular aortic aneurysm repair. Radiology. 2007;243:641-55. PMid:17517926. http://dx.doi.org/10.1148/radiol.2433051649
http://dx.doi.org/10.1148/radiol.2433051...
Compared with conventional surgery, endovascular repair offers low rates of perioperative mortality and morbidity, but requires rigorous surveillance for the rest of the patient's life because of a relevant incidence of long-term complications, such as endovascular leakage (endoleaks), aneurysm growth, structural failures, graft migration, thrombosis, occlusion of branches and infections.1. Greenhalgh RM, Powell JT. Endovascular repair of abdominal aortic aneurysm. N Engl J Med. 2008;388:494-501. PMid:18234753. http://dx.doi.org/10.1056/NEJMct0707524
http://dx.doi.org/10.1056/NEJMct0707524...
,3. Chaer RA, Gushchin P, Rhee R, et al. Duplex ultrasound as the sole long-term surveillance method postendovascular aneurysm repair: a safe alternative for stable aneurysm. Vasc Surg. 2009;49:845-50. PMid:19341877. http://dx.doi.org/10.1016/j.jvs.2008.10.073
http://dx.doi.org/10.1016/j.jvs.2008.10....

. Kranokpirakasa P, Kaufman JA. Follow-up of endovascular aneurysm repair: plain radiography, ultrasound, CT/CT angiography, MR imaging/MR angiography, or what? J Vasc Interv Radiol. 2008;19:S27-36. PMid:18502384. http://dx.doi.org/10.1016/j.jvir.2008.03.009
http://dx.doi.org/10.1016/j.jvir.2008.03...
-5. EVAR trial participants. Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial. Lancet. 2005;365:2179-86. http://dx.doi.org/10.1016/S0140-6736(05)66627-5
http://dx.doi.org/10.1016/S0140-6736(05)...

According to Geller SC,6. Geller SC. Imaging guidelines for abdominal aortic aneurysm repair with endovascular stent grafts. J Vasc Interv Radiol. 2003;14:S263-4. PMid:14514830. follow-up with imaging studies after endovascular repair should assess three major parameters: (1) aneurysmal diameter; (2) detection and classification of endoleaks, and (3) detection of morphological changes to endoprostheses. Graft occlusion, infection and other potential complications should be assessed.4. Kranokpirakasa P, Kaufman JA. Follow-up of endovascular aneurysm repair: plain radiography, ultrasound, CT/CT angiography, MR imaging/MR angiography, or what? J Vasc Interv Radiol. 2008;19:S27-36. PMid:18502384. http://dx.doi.org/10.1016/j.jvir.2008.03.009
http://dx.doi.org/10.1016/j.jvir.2008.03...
Angiotomography has become considered the gold-standard imaging technique for investigating all of these complications,1. Greenhalgh RM, Powell JT. Endovascular repair of abdominal aortic aneurysm. N Engl J Med. 2008;388:494-501. PMid:18234753. http://dx.doi.org/10.1056/NEJMct0707524
http://dx.doi.org/10.1056/NEJMct0707524...
but some studies over recent years have demonstrated an increased risk of radiation-induced cancer and nephropathy caused by the contrast after repeated exposures to tomography.7. Brenner DJ, Hall EJ. Computed tomography an increasing source of radiation exposure. N Engl J Med. 2007;357:2277-84. PMid:18046031. http://dx.doi.org/10.1056/NEJMra072149
http://dx.doi.org/10.1056/NEJMra072149...
,8. Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298:317-23. PMid:17635892. http://dx.doi.org/10.1001/jama.298.3.317
http://dx.doi.org/10.1001/jama.298.3.317...

In response to these findings, many researchers have proposed substituting tomography for follow-up of these patients with other imaging methods, including vascular ultrasonography (USV), with or without contrast. This is relatively inexpensive and is an attractive option for postoperative surveillance, since it is also a noninvasive diagnostic method that is generally available.9. Conrad MF, Adams AB, Guest JM, et al. Secondary intervention after endovascular abdominal aortic aneurysm repair. Ann Surg. 2009;250:383-9. PMid:19652592.

10 . Bendick PJ, Bove PG, Long GW, Zelenock GB, Brown OW, Shanley CJ. Efficacy of ultrasound scan contrast agents in the noninvasive follow-up of aortic stent grafts. J Vasc Surg. 2003;37:381-5. PMid:12563210. http://dx.doi.org/10.1067/mva.2003.17
http://dx.doi.org/10.1067/mva.2003.17...
-1111 . Collins JT, Boros MJ, Combs K. Ultrasound surveillance of endovascular aneurysm repair: a safe modality versus computed tomography. Ann Vasc Surg. 2007;21:671-5. PMid:17980791. http://dx.doi.org/10.1016/j.avsg.2007.07.009
http://dx.doi.org/10.1016/j.avsg.2007.07...
Several different studies have therefore been conducted to compare the precision of vascular ultrasonography with that of angiotomography for classification of endoleaks.1212 . Cantisani V, Ricci P, Grazhdani H, et al. Prospective comparative analysis of colour-Doppler ultrasound, contrast-enhanced ultrasound, computed tomography and magnetic resonance in detecting endoleak after endovascular abdominal aortic aneurysm repair. Eur J Endovasc Surg. 2011;41:186-92. PMid:21095141. http://dx.doi.org/10.1016/j.ejvs.2010.10.003
http://dx.doi.org/10.1016/j.ejvs.2010.10...

13 . Carrafiello G, Laganà D, Recaldini C, et al. Comparison of contrast-enhanced ultrasound and computed tomography in classifying endoleaks after endovascular treatment of abdominal aorta aneurysms: preliminary experience. Cardiovasc Intervent Radiol. 2006;29:969-74. PMid:16897267. http://dx.doi.org/10.1007/s00270-005-0267-x
http://dx.doi.org/10.1007/s00270-005-026...
-1414 . Ten Bosch JA, Rouwet EV, Peters CTH, et al. Contrast-enhanced ultrasound versus computed tomographic angiography for surveillance of endovascular abdominal aortic aneurysm repair. J Vasc Interv Radiol. 2010;21:638-43. PMid:20363153. http://dx.doi.org/10.1016/j.jvir.2010.01.032
http://dx.doi.org/10.1016/j.jvir.2010.01...
The objective of this study was to determine the correlation, sensitivity and specificity of the ultrasonography method, with relation to angiotomography, for measurement of aneurysm diameter, identification of tortuosity and for detection and classification of endoleaks.

MATERIALS AND METHODS

Study design

The study was conducted from June 2012 to May 2013, while under consideration by the Ethics Committee, which gave its approval. All patients Domingos Moraes Filho, Fernando Barbosa Trevisan et al. treated with endovascular repair of abdominal aortic aneurysms (AAA) and seen at the Endovascular Surgery Clinic of the University Hospital in Londrina were enrolled. The postoperative follow-up protocol included angiotomography, ultrasonography and abdominal X-ray after 1, 6 and 12 months, during the first year, and annually thereafter. Patients who did not comply with the follow-up protocol or whose USV and angiotomography were not conducted within 30 days of each other were excluded.

X-ray protocol

Simple abdominal X-rays, anteroposterior projection, were taken of all patients in follow-up after endovascular repair of abdominal aortic aneurysms. The X-rays were taken to check for morphological changes to the endoprostheses, such as fractures, tortuosity or displacement of segments.

Ultrasonography protocol

Vascular ultrasonography scans were conducted in mode B, with color, spectral analysis and power mode, of the whole abdominal aorta and the iliac arteries, using an ESAOTE 3-5MHz convex abdominal transducer, for all patients in follow-up after endovascular repair of abdominal aortic aneurysms. Ultrasound scans were conducted after 8 hours' fasting, with appropriate intestinal preparation and 40 drops of dimethicone every 8 hours, starting on the day before the examination. All examinations were conducted by the same Vascular Surgeon, qualified in Vascular Echography. These examinations were conducted in order to evaluate the diameter of aneurysms and patency of endoprostheses, and to identify and classify any endoleaks. These were assessed by aneurysm measurements in anteroposterior and lateral projections, and by detection of blood flow outside of the endoprosthesis using Doppler.

Tomography protocol

Angiotomography was conducted on patients in follow-up according to our routine protocol, as follows: full abdominal angiotomography in three steps, first without contrast; then with 100-120 mL of contrast, and once more (late) 180 seconds after injection of the contrast, using a 16-channel tomography machine and 1.25 mm abdominal and pelvic slices. These scans were also used to detect and classify any endoleaks present, identified by contrast seen outside of endoprostheses, to evaluate anteroposterior and lateral aneurysm diameter measurements, and to check for patency Vascular ultrasonography for aorta aneurysms and structural changes to endoprostheses visible on three-dimensional reconstructions. In order to avoid interference from the ultrasonographic examination, a different examiner, also qualified, was responsible for this procedure and only conducted angiotomography for this study. Patients with a history of allergy and/or renal failure were duly prepared with antiallergic drugs and corticosteroids and given adequate hydration before and after examinations.

Statistics

Comparisons of mean size of aneurysm on USV and angiotomography were accomplished by calculating coefficients of variation. The coefficient of variation was determined by dividing the standard deviation of the difference between the two measurements by the mean of means of the study population. Pearson's correlation coefficient was used to determine the degree of agreement between USV and angiotomography.

RESULTS

Between June 2012 and May 2013, a total of 62 patients were examined using the follow-up protocol for endovascular AAA repair. Three of these patients had been treated for endovascular AAA repair in 2010, 11 in 2011, 13 in 2012 and three in 2013. The remaining 32 patients were excluded because they had not complied with the follow-up protocol and did not have their ultrasonography and angiotomography scans close enough in time.

The sample comprised 25 men (84%) and five women (16%), with a mean age of 73 years (56-84). Hypertension, smoking and diabetes were the most common comorbidities. On angiotomography, the diameter of abdominal aneurysms found after endovascular repair ranged from 3.0 to 10.5 cm, with a mean of 5.8 cm. Aneurysms of iliac arteries were also identified by angiotomography in nine patients, 30% of cases, five of which (17% of the total) were bilateral.

Analysis of the abdominal X-rays found tortuosity of the endoprosthesis structure in just one patient, without any significant hemodynamic alterations when viewed on vascular ultrasonography (Figure 1). No X-rays showed structural fractures or disconnections.

Figure 1
Standard abdominal X-ray showing tortuosity of an endoprosthesis branch.

With relation to USV findings, four endoleaks were identified in the 30 patients examined, two type IA and two type II. When compared with angiotomography, which also found four endoleaks, three USV-detected leaks were confirmed. In just one angiotomography a leak was identified that had not been diagnosed using ultrasonography. Therefore, the statistical analysis resulted in sensitivity and specificity for endoleaks of 75% and 96% respectively, and positive predictive and negative predictive values of 75% and 96%. However, we did not observe the same relationship for classification of endoleaks, since the classification only coincided in one case; the other three cases were two misclassifications and one false positive.

Only one of the USV examinations detected abnormalities on the spectral analysis. In this case, angiotomography confirmed the ultrasonographic diagnosis of occlusion of the external branch of a sandwich graft in the iliac arteries (Figure 2). This represents sensitivity and specificity of 100%, despite the small sample.

Figure 2
(a) Vascular ultrasonography with monophasic flow in right external iliac. (b) Angiotomography with occlusion right external iliac branch.

We observed a variation of three millimeters between aneurysmal diameters measured on ultrasonography and the equivalent angiotomography measurements, with a strong positive linear correlation (Table 1).

Table 1
Correlation between analyses of aneurysm diameter: USV vs. angiotomography.

DISCUSSION

Serial X-rays are useful in follow-up of endovascular AAA treatment for assessment of structural changes, especially when interest lies in detection of fractures and kinking of abdominal endoprostheses.1515 . Fearn S, Lawrence-Brown MM, Semmens JB, Hartley D. Follow-up after endovascular aortic aneurysm repair: the plain radiograph has an essential role in surveillance. J Endovasc Ther. 2003;10:894-901. http://dx.doi.org/10.1583/1545-1550(2003)010<0894:FAEAAR>2.0.CO;2
http://dx.doi.org/10.1583/1545-1550(2003...
These X-rays are simple and inexpensive to obtain and are widely available. Additionally, their sensitivity for detection of graft migration is similar to that offered by computed tomography,1515 . Fearn S, Lawrence-Brown MM, Semmens JB, Hartley D. Follow-up after endovascular aortic aneurysm repair: the plain radiograph has an essential role in surveillance. J Endovasc Ther. 2003;10:894-901. http://dx.doi.org/10.1583/1545-1550(2003)010<0894:FAEAAR>2.0.CO;2
http://dx.doi.org/10.1583/1545-1550(2003...
although an understanding of endoprosthesis construction and the placement of radiopaque markers is indispensable for correct assessment and interpretation of these X-rays.1616 . Hodgson R, McWilliams RG, Simpson A, et al. Migration versus apparent migration: Importance of error due to positioning variation in plain radiographic follow-up of aortic stentgrafts. J Endovasc Ther. 2003;10:902-10. http://dx.doi.org/10.1583/1545-1550(2003)010<0902:MVAMIO>2.0.CO;2
http://dx.doi.org/10.1583/1545-1550(2003...
,1717 . Murphy M, Hodgson R, Harris PL, McWilliams RG, Hartley DE, Lawrence-Brown MM. Plain radiographic surveillance of abdominal aortic stent-grafts: the Liverpool/Perth protocol. J Endovasc Ther. 2003;10:911-2. http://dx.doi.org/10.1583/1545-1550(2003)010<0911:PRSOAA>2.0.CO;2
http://dx.doi.org/10.1583/1545-1550(2003...
However, X-rays are subject to two important limitations: they cannot be used to measure aneurysm diameter nor to detect endoleaks.

On the other hand, USV can be used both for monitoring of aneurysm diameters for screening for endoleaks after endovascular AAA repair.3. Chaer RA, Gushchin P, Rhee R, et al. Duplex ultrasound as the sole long-term surveillance method postendovascular aneurysm repair: a safe alternative for stable aneurysm. Vasc Surg. 2009;49:845-50. PMid:19341877. http://dx.doi.org/10.1016/j.jvs.2008.10.073
http://dx.doi.org/10.1016/j.jvs.2008.10....
,4. Kranokpirakasa P, Kaufman JA. Follow-up of endovascular aneurysm repair: plain radiography, ultrasound, CT/CT angiography, MR imaging/MR angiography, or what? J Vasc Interv Radiol. 2008;19:S27-36. PMid:18502384. http://dx.doi.org/10.1016/j.jvir.2008.03.009
http://dx.doi.org/10.1016/j.jvir.2008.03...
Additionally, USV is convenient, noninvasive, inexpensive and portable, does not involve radiation exposure and does not need contrast. However, there is a relatively higher rate of interobserver and intraobserver variability with ultrasonography than with angiotomography,1818 . Ederle FA, Wilson SE, Johnson GR, et al. Variability in measurement of abdominal aortic aneurysms. J Vasc Surg. 1995;21:945-52. http://dx.doi.org/10.1016/S0741-5214(95)70222-9
http://dx.doi.org/10.1016/S0741-5214(95)...
but this variability is no larger than 4 mm, on average.1919 . Singh K, Bønaa KH, Solberg S, Sørlie DG, Bjørk L. Intra- and interobserver variability in ultrasound measurements of abdominal aortic diameter. The Tromsø Study. Eur J Vasc Endovasc Surg. 1998;15:497-504. http://dx.doi.org/10.1016/S1078-5884(98)80109-3
http://dx.doi.org/10.1016/S1078-5884(98)...
Additionally, ultrasonography is also able to provide a better approximation to the true perpendicular diameter of an abdominal aortic aneurysm, since it is not affected by the angle of the aorta, which impacts on tomography results at angles greater than 25°.2020 . Sprouse LR, Meier GH, Parent FN, DeMasi RJ, Glickman MH, Barber GA. Is ultrasound more accurate than axial computed tomography for determination of maximal abdominal aortic aneurysm diameter? Eur J Vasc Endovasc Surg. 2004;28:28-35. PMid:15177228. http://dx.doi.org/10.1016/j.ejvs.2004.03.022
http://dx.doi.org/10.1016/j.ejvs.2004.03...

Endoleaks can be detected by ultrasound using color spectral analysis to show blood flow. While published specificity rates for detection of endoleaks with Doppler ultrasonography are high (89%-97%),2121 . Sun Z. Diagnostic value of color duplex ultrasonography in the follow-up of endovascular repair of abdominal aortic aneurysm. J Vasc Interv Radiol. 2006;17:759-64. PMid:16687740. http://dx.doi.org/10.1097/01.RVI.0000217944.36738.02
http://dx.doi.org/10.1097/01.RVI.0000217...
the sensitivity and diagnostic power of Doppler US for detecting endoleaks is still debatable, when compared with angiotomography.1111 . Collins JT, Boros MJ, Combs K. Ultrasound surveillance of endovascular aneurysm repair: a safe modality versus computed tomography. Ann Vasc Surg. 2007;21:671-5. PMid:17980791. http://dx.doi.org/10.1016/j.avsg.2007.07.009
http://dx.doi.org/10.1016/j.avsg.2007.07...
,2222 . Fletcher J, Saker K, Batiste P, Dyer S. Color Doppler diagnosis of perigraft flow following endovascular repair of abdominal aortic aneurysm. Int Angiol. 2000;19:326-30. PMid:11305731.

23 . D'Audiffret A, Desgranges P, Kobeiter H, Becquemin JP. Follow-up evaluation of endoluminally treated abdominal aortic aneurysms with duplex ultrasonography: validation with computed tomography. J Vasc Surg. 2001;33:42-50. PMid:11137922. http://dx.doi.org/10.1067/mva.2001.112215
http://dx.doi.org/10.1067/mva.2001.11221...

24 . McLafferty RB, McCrary BS, Mattos MA, et al. The use of color-flow duplex scan for the detection of endoleaks. J Vasc Surg. 2002;36:100-4. PMid:12096265. http://dx.doi.org/10.1067/mva.2002.123089
http://dx.doi.org/10.1067/mva.2002.12308...

25 . Bendick PJ, Bove PG, Long GW, Zelenock GB, Brown OW, Shanley CJ. Efficacy of ultrasound scan contrast agents in the non-invasive follow-up of aortic stent grafts. J Vasc Surg. 2003;37:381-5. PMid:12563210. http://dx.doi.org/10.1067/mva.2003.17
http://dx.doi.org/10.1067/mva.2003.17...

26 . Raman KG, Missig-Carroll N, Richardson T, Muluk SC, Makaroun MS. Color-flow duplex ultrasound scan versus computed tomographic scan in the surveillance of endovascular aneurysm repair. J Vasc Surg. 2003;38:645-51. http://dx.doi.org/10.1016/S0741-5214(03)00909-1
http://dx.doi.org/10.1016/S0741-5214(03)...

27 . Elkouri S, Panneton JM, Andrews JC, et al. Computed tomography and ultrasound in follow-up of patients after endovascular repair of abdominal aortic aneurysm. Ann Vasc Surg. 2004;18:271-9. PMid:15354627. http://dx.doi.org/10.1007/s10016-004-0034-5
http://dx.doi.org/10.1007/s10016-004-003...

28 . AbuRahma AF, Welch CA, Mullins BB, Dyer B. Computed tomography versus color duplex ultrasound for surveillance of abdominal aortic stent-grafts. J Endovasc Ther. 2005;12:568-73. PMid:16212456. http://dx.doi.org/10.1583/05-1575MR.1
http://dx.doi.org/10.1583/05-1575MR.1...

29 . Standford RM, Bown MJ, Fishwick G, et al. Duplex ultrasound scanning is reliable in the detection of endoleak following endovascular aneurysm repair. Eur J Vasc Endovasc Surg. 2006;32:537-41. PMid:16875850. http://dx.doi.org/10.1016/j.ejvs.2006.05.013
http://dx.doi.org/10.1016/j.ejvs.2006.05...
-3030 . Ashoke R, Brown LC, Rodway A, et al. Color duplex ultrasonography is insensitive for the detection of endoleak after aortic endografting: a systematic review. J Endovasc Ther. 2005;12:297-305. PMid:15943504. http://dx.doi.org/10.1583/04-1479R.1
http://dx.doi.org/10.1583/04-1479R.1...
Some studies,1111 . Collins JT, Boros MJ, Combs K. Ultrasound surveillance of endovascular aneurysm repair: a safe modality versus computed tomography. Ann Vasc Surg. 2007;21:671-5. PMid:17980791. http://dx.doi.org/10.1016/j.avsg.2007.07.009
http://dx.doi.org/10.1016/j.avsg.2007.07...
,2222 . Fletcher J, Saker K, Batiste P, Dyer S. Color Doppler diagnosis of perigraft flow following endovascular repair of abdominal aortic aneurysm. Int Angiol. 2000;19:326-30. PMid:11305731.

23 . D'Audiffret A, Desgranges P, Kobeiter H, Becquemin JP. Follow-up evaluation of endoluminally treated abdominal aortic aneurysms with duplex ultrasonography: validation with computed tomography. J Vasc Surg. 2001;33:42-50. PMid:11137922. http://dx.doi.org/10.1067/mva.2001.112215
http://dx.doi.org/10.1067/mva.2001.11221...

24 . McLafferty RB, McCrary BS, Mattos MA, et al. The use of color-flow duplex scan for the detection of endoleaks. J Vasc Surg. 2002;36:100-4. PMid:12096265. http://dx.doi.org/10.1067/mva.2002.123089
http://dx.doi.org/10.1067/mva.2002.12308...
-2525 . Bendick PJ, Bove PG, Long GW, Zelenock GB, Brown OW, Shanley CJ. Efficacy of ultrasound scan contrast agents in the non-invasive follow-up of aortic stent grafts. J Vasc Surg. 2003;37:381-5. PMid:12563210. http://dx.doi.org/10.1067/mva.2003.17
http://dx.doi.org/10.1067/mva.2003.17...
such as one conducted by Collins et al.,1111 . Collins JT, Boros MJ, Combs K. Ultrasound surveillance of endovascular aneurysm repair: a safe modality versus computed tomography. Ann Vasc Surg. 2007;21:671-5. PMid:17980791. http://dx.doi.org/10.1016/j.avsg.2007.07.009
http://dx.doi.org/10.1016/j.avsg.2007.07...
have reported 85% positive predictive values, whereas others2626 . Raman KG, Missig-Carroll N, Richardson T, Muluk SC, Makaroun MS. Color-flow duplex ultrasound scan versus computed tomographic scan in the surveillance of endovascular aneurysm repair. J Vasc Surg. 2003;38:645-51. http://dx.doi.org/10.1016/S0741-5214(03)00909-1
http://dx.doi.org/10.1016/S0741-5214(03)...

27 . Elkouri S, Panneton JM, Andrews JC, et al. Computed tomography and ultrasound in follow-up of patients after endovascular repair of abdominal aortic aneurysm. Ann Vasc Surg. 2004;18:271-9. PMid:15354627. http://dx.doi.org/10.1007/s10016-004-0034-5
http://dx.doi.org/10.1007/s10016-004-003...

28 . AbuRahma AF, Welch CA, Mullins BB, Dyer B. Computed tomography versus color duplex ultrasound for surveillance of abdominal aortic stent-grafts. J Endovasc Ther. 2005;12:568-73. PMid:16212456. http://dx.doi.org/10.1583/05-1575MR.1
http://dx.doi.org/10.1583/05-1575MR.1...

29 . Standford RM, Bown MJ, Fishwick G, et al. Duplex ultrasound scanning is reliable in the detection of endoleak following endovascular aneurysm repair. Eur J Vasc Endovasc Surg. 2006;32:537-41. PMid:16875850. http://dx.doi.org/10.1016/j.ejvs.2006.05.013
http://dx.doi.org/10.1016/j.ejvs.2006.05...
-3030 . Ashoke R, Brown LC, Rodway A, et al. Color duplex ultrasonography is insensitive for the detection of endoleak after aortic endografting: a systematic review. J Endovasc Ther. 2005;12:297-305. PMid:15943504. http://dx.doi.org/10.1583/04-1479R.1
http://dx.doi.org/10.1583/04-1479R.1...
found ultrasonography to be less sensitive than tomography, as was the case of Ashoke et al.,3030 . Ashoke R, Brown LC, Rodway A, et al. Color duplex ultrasonography is insensitive for the detection of endoleak after aortic endografting: a systematic review. J Endovasc Ther. 2005;12:297-305. PMid:15943504. http://dx.doi.org/10.1583/04-1479R.1
http://dx.doi.org/10.1583/04-1479R.1...
who observed sensitivity of 70% in their study. In our study, four patients had endoleaks Domingos Moraes Filho, Fernando Barbosa Trevisan et al. that were identified using angiotomography. Three of these were clinically significant type IA leaks and all three were also identified by USV, while the single case of type II endoleak, which was not expanding, was not detected by USV (Table 2). It can therefore be concluded that although USV does not offer a relatively high degree of sensitivity, as has been reported in other studies, it does offer adequate sensitivity for identifying endoleaks and indicating their treatment, without impacting on the quality of care.

Table 2
Correlations for patients with endoleaks: USV × angiotomography.

Another advantage of USV is its ability to detect the direction of endoleak flow, which cannot easily be investigated using angiotomography. Parent et al.3131 . Parent FN, Meier GH, Godziachvili V, et al. The incidence and natural history of type I and II endoleak: a 5-year followup assessment with 1 color duplex ultrasound scan. J Vasc Surg. 2002;35:474-81. PMid:11877694. http://dx.doi.org/10.1067/mva.2002.121848
http://dx.doi.org/10.1067/mva.2002.12184...
related the Doppler waveform to analysis of type II endoleaks as follows: a 'anterograde and retrograde' pattern was associated with spontaneous resolution, whereas the presence of a monophasic or biphasic waveform was associated with persistent leakage. However, we did not observed this in our study, since two of the three endoleaks correctly identified by ultrasonography, both type IA, were erroneously classified as type II and one was repaired using a proximal extension and the other was repaired by embolization (Figure 3). The principal cause of this difference was probably an error of interpretation by the USV examiner.

Figure 3
(a) Transverse and longitudinal vascular ultrasonography (b) showing type II endoleak. (c) Angiotomography showing type IA endoleak.

It is important to point out that for assessment of stenosis or occlusion of the aortoiliac segment, it is arteriography and not angiotomography that is the gold-standard diagnostic method.3232 . Moreira RCP. Estudo comparativo de eco-Doppler cm arteriografia na avaliação da doença oclusiva aortoiliaca. J Vasc Bras. 2009;8:3-13. http://dx.doi.org/10.1590/S1677-54492009000100002
http://dx.doi.org/10.1590/S1677-54492009...
Additionally, Beeman et al.3333 . Beeman BR, Doctor LM, Doerr K, McAfee-Bennett S, Dougherty MJ, Calligaro KD. Duplex ultrasound imaging alone is suficient for midterm endovascular aneurysm repair survaillance: A cost analysis study and prospective comparison with computed tomography scan. J Vasc Surg. 2009;50:1019-24. PMid:19656651. http://dx.doi.org/10.1016/j.jvs.2009.06.019
http://dx.doi.org/10.1016/j.jvs.2009.06....
believe that USV is also more precise than angiotomography for identifying patency and for detecting other problems such as tortuosity and stenoses, since the color flow images combined with analysis of the waveform provide anatomic and hemodynamic data that cannot be acquired using angiotomography. In their study,3333 . Beeman BR, Doctor LM, Doerr K, McAfee-Bennett S, Dougherty MJ, Calligaro KD. Duplex ultrasound imaging alone is suficient for midterm endovascular aneurysm repair survaillance: A cost analysis study and prospective comparison with computed tomography scan. J Vasc Surg. 2009;50:1019-24. PMid:19656651. http://dx.doi.org/10.1016/j.jvs.2009.06.019
http://dx.doi.org/10.1016/j.jvs.2009.06....
USV precisely identified all seven cases in which there were problems with graft permeability requiring treatment, which was also the case in our study, since we correctly identified the only case of branch occlusion using USV.

Other limitations of duplex ultrasonography include its operator dependency, inter-patient physical variations and the need for intestinal preparation.4. Kranokpirakasa P, Kaufman JA. Follow-up of endovascular aneurysm repair: plain radiography, ultrasound, CT/CT angiography, MR imaging/MR angiography, or what? J Vasc Interv Radiol. 2008;19:S27-36. PMid:18502384. http://dx.doi.org/10.1016/j.jvir.2008.03.009
http://dx.doi.org/10.1016/j.jvir.2008.03...
However, improvements in ultrasonography technology and techniques can improve and increase its diagnostic power. Contrast mediums, for example, have been developed that can overcome some of the limitations of ultrasonography.3434 . Correas JM, Bridal L, Lesavre A, Mejean A, Claudon M, Helenon O. Ultrasound contrast agents: properties, principles of action, tolerance, and artifacts. Eur Radiol. 2001;11:1316-28. PMid:11519538. http://dx.doi.org/10.1007/s003300100940
http://dx.doi.org/10.1007/s003300100940...
Ultrasonography contrasts are noninvasive, easy to handle, well-tolerated and, apparently, increase the sensitivity of ultrasonography. Use of ultrasonography contrasts significantly improves sensitivity for detection of endoleaks,3535 . Giannoni MF, Palombo G, Sbarigia E, Speziale F, Zaccaria A, Fiorani P. Contrast-enhanced ultrasound imaging for aortic stent-graft surveillance. J Endovasc Ther. 2003;10:208-17. http://dx.doi.org/10.1583/1545-1550(2003)010<0208:CUIFAS>2.0.CO;2
http://dx.doi.org/10.1583/1545-1550(2003...

36 . Napoli V, Bargellini I, Sardella SG, et al. Abdominal aortic aneurysm: contrast-enhanced US for missed endoleaks after endoluminal repair. Radiology. 2004;233:217-25. PMid:15454621. http://dx.doi.org/10.1148/radiol.2331031767
http://dx.doi.org/10.1148/radiol.2331031...
-3737 . Henao EA, Hodge MD, Felkai DD, et al. Contrast-enhanced duplex surveillance after endovascular abdominal aortic aneurysm repair: Improved efficacy using a continuous infusion technique. J Vasc Surg. 2006;43:259-64. PMid:16476596. http://dx.doi.org/10.1016/j.jvs.2005.09.045
http://dx.doi.org/10.1016/j.jvs.2005.09....
particularly in cases of endotension, which is obscured in angiotomography, but can be seen with the aid of contrast.3636 . Napoli V, Bargellini I, Sardella SG, et al. Abdominal aortic aneurysm: contrast-enhanced US for missed endoleaks after endoluminal repair. Radiology. 2004;233:217-25. PMid:15454621. http://dx.doi.org/10.1148/radiol.2331031767
http://dx.doi.org/10.1148/radiol.2331031...
,3737 . Henao EA, Hodge MD, Felkai DD, et al. Contrast-enhanced duplex surveillance after endovascular abdominal aortic aneurysm repair: Improved efficacy using a continuous infusion technique. J Vasc Surg. 2006;43:259-64. PMid:16476596. http://dx.doi.org/10.1016/j.jvs.2005.09.045
http://dx.doi.org/10.1016/j.jvs.2005.09....

Additionally, using simple X-rays combined with USV as the primary imaging method for patient follow-up after endovascular repair of abdominal aortic aneurysms could also reduce the radiological risks associated with angiotomography.3838 . Carrafiello G, Recaldini C, Lagana D, Piffaretti G, Fugazzola C. Endoleak detection and classification after endovascular treatment of abdominal aortic aneurysm: value of CEUS over CTA. Abdom Imaging. 2008;33:357-62. PMid:17619925. http://dx.doi.org/10.1007/s00261-007-9268-3
http://dx.doi.org/10.1007/s00261-007-926...
The endovascular procedure and serial lifelong angiotomography involve a substantial ionizing radiation load, in addition to the fact that angiotomography requires administration of iodine-based contrast agents that are associated with nephrotoxic effects.3939 . Rao QA, Newhouse JH. Risk of nephropathy after intravenous administration of contrast material: a critical literature analysis. Radiology. 2006;239:392-7. PMid:16543592. http://dx.doi.org/10.1148/radiol.2392050413
http://dx.doi.org/10.1148/radiol.2392050...
Nevertheless, it is known that ultrasonography can also suffer from a series of limitations. In obese patients with inadequate intestinal preparation, the abdominal examination can be limited by an inappropriate acoustic window.4. Kranokpirakasa P, Kaufman JA. Follow-up of endovascular aneurysm repair: plain radiography, ultrasound, CT/CT angiography, MR imaging/MR angiography, or what? J Vasc Interv Radiol. 2008;19:S27-36. PMid:18502384. http://dx.doi.org/10.1016/j.jvir.2008.03.009
http://dx.doi.org/10.1016/j.jvir.2008.03...
However, the limited sample of 30 cases and the short observation period notwithstanding, this study observed low variability in aneurysm sac dimensions with a strong correlation coefficient, and also a low number of false negative results, just one in fact, which had no clinical significance since the aneurysm did not exhibit significant growth.

CONCLUSIONS

It is therefore understood that vascular ultrasonography combined with abdominal X-ray could be used as a possible substitute for angiotomography as the primary method for follow-up of patients with aneurysms of the abdominal aorta operated using endovascular techniques. Abdominal X-ray proved to be an excellent supplementary examination for assessing the structure of endoprostheses. Ultrasonography had a high correlation coefficient and low variability when compared with angiotomography for measurement of aneurysm diameter. With regard to endoleaks, vascular ultrasonography identified all clinically significant leaks.

  • Financial support: None.
  • The study was carried out at Hospital Universitário de Londrina.

REFERENCES

  • 1
    Greenhalgh RM, Powell JT. Endovascular repair of abdominal aortic aneurysm. N Engl J Med. 2008;388:494-501. PMid:18234753. http://dx.doi.org/10.1056/NEJMct0707524
    » http://dx.doi.org/10.1056/NEJMct0707524
  • 2
    Stavropoulos SW, Charagundla SR. Imaging techniques for detection and management of endoleaks after endovascular aortic aneurysm repair. Radiology. 2007;243:641-55. PMid:17517926. http://dx.doi.org/10.1148/radiol.2433051649
    » http://dx.doi.org/10.1148/radiol.2433051649
  • 3
    Chaer RA, Gushchin P, Rhee R, et al. Duplex ultrasound as the sole long-term surveillance method postendovascular aneurysm repair: a safe alternative for stable aneurysm. Vasc Surg. 2009;49:845-50. PMid:19341877. http://dx.doi.org/10.1016/j.jvs.2008.10.073
    » http://dx.doi.org/10.1016/j.jvs.2008.10.073
  • 4
    Kranokpirakasa P, Kaufman JA. Follow-up of endovascular aneurysm repair: plain radiography, ultrasound, CT/CT angiography, MR imaging/MR angiography, or what? J Vasc Interv Radiol. 2008;19:S27-36. PMid:18502384. http://dx.doi.org/10.1016/j.jvir.2008.03.009
    » http://dx.doi.org/10.1016/j.jvir.2008.03.009
  • 5
    EVAR trial participants. Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial. Lancet. 2005;365:2179-86. http://dx.doi.org/10.1016/S0140-6736(05)66627-5
    » http://dx.doi.org/10.1016/S0140-6736(05)66627-5
  • 6
    Geller SC. Imaging guidelines for abdominal aortic aneurysm repair with endovascular stent grafts. J Vasc Interv Radiol. 2003;14:S263-4. PMid:14514830.
  • 7
    Brenner DJ, Hall EJ. Computed tomography an increasing source of radiation exposure. N Engl J Med. 2007;357:2277-84. PMid:18046031. http://dx.doi.org/10.1056/NEJMra072149
    » http://dx.doi.org/10.1056/NEJMra072149
  • 8
    Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298:317-23. PMid:17635892. http://dx.doi.org/10.1001/jama.298.3.317
    » http://dx.doi.org/10.1001/jama.298.3.317
  • 9
    Conrad MF, Adams AB, Guest JM, et al. Secondary intervention after endovascular abdominal aortic aneurysm repair. Ann Surg. 2009;250:383-9. PMid:19652592.
  • 10
    Bendick PJ, Bove PG, Long GW, Zelenock GB, Brown OW, Shanley CJ. Efficacy of ultrasound scan contrast agents in the noninvasive follow-up of aortic stent grafts. J Vasc Surg. 2003;37:381-5. PMid:12563210. http://dx.doi.org/10.1067/mva.2003.17
    » http://dx.doi.org/10.1067/mva.2003.17
  • 11
    Collins JT, Boros MJ, Combs K. Ultrasound surveillance of endovascular aneurysm repair: a safe modality versus computed tomography. Ann Vasc Surg. 2007;21:671-5. PMid:17980791. http://dx.doi.org/10.1016/j.avsg.2007.07.009
    » http://dx.doi.org/10.1016/j.avsg.2007.07.009
  • 12
    Cantisani V, Ricci P, Grazhdani H, et al. Prospective comparative analysis of colour-Doppler ultrasound, contrast-enhanced ultrasound, computed tomography and magnetic resonance in detecting endoleak after endovascular abdominal aortic aneurysm repair. Eur J Endovasc Surg. 2011;41:186-92. PMid:21095141. http://dx.doi.org/10.1016/j.ejvs.2010.10.003
    » http://dx.doi.org/10.1016/j.ejvs.2010.10.003
  • 13
    Carrafiello G, Laganà D, Recaldini C, et al. Comparison of contrast-enhanced ultrasound and computed tomography in classifying endoleaks after endovascular treatment of abdominal aorta aneurysms: preliminary experience. Cardiovasc Intervent Radiol. 2006;29:969-74. PMid:16897267. http://dx.doi.org/10.1007/s00270-005-0267-x
    » http://dx.doi.org/10.1007/s00270-005-0267-x
  • 14
    Ten Bosch JA, Rouwet EV, Peters CTH, et al. Contrast-enhanced ultrasound versus computed tomographic angiography for surveillance of endovascular abdominal aortic aneurysm repair. J Vasc Interv Radiol. 2010;21:638-43. PMid:20363153. http://dx.doi.org/10.1016/j.jvir.2010.01.032
    » http://dx.doi.org/10.1016/j.jvir.2010.01.032
  • 15
    Fearn S, Lawrence-Brown MM, Semmens JB, Hartley D. Follow-up after endovascular aortic aneurysm repair: the plain radiograph has an essential role in surveillance. J Endovasc Ther. 2003;10:894-901. http://dx.doi.org/10.1583/1545-1550(2003)010<0894:FAEAAR>2.0.CO;2
    » http://dx.doi.org/10.1583/1545-1550(2003)010<0894:FAEAAR>2.0.CO;2
  • 16
    Hodgson R, McWilliams RG, Simpson A, et al. Migration versus apparent migration: Importance of error due to positioning variation in plain radiographic follow-up of aortic stentgrafts. J Endovasc Ther. 2003;10:902-10. http://dx.doi.org/10.1583/1545-1550(2003)010<0902:MVAMIO>2.0.CO;2
    » http://dx.doi.org/10.1583/1545-1550(2003)010<0902:MVAMIO>2.0.CO;2
  • 17
    Murphy M, Hodgson R, Harris PL, McWilliams RG, Hartley DE, Lawrence-Brown MM. Plain radiographic surveillance of abdominal aortic stent-grafts: the Liverpool/Perth protocol. J Endovasc Ther. 2003;10:911-2. http://dx.doi.org/10.1583/1545-1550(2003)010<0911:PRSOAA>2.0.CO;2
    » http://dx.doi.org/10.1583/1545-1550(2003)010<0911:PRSOAA>2.0.CO;2
  • 18
    Ederle FA, Wilson SE, Johnson GR, et al. Variability in measurement of abdominal aortic aneurysms. J Vasc Surg. 1995;21:945-52. http://dx.doi.org/10.1016/S0741-5214(95)70222-9
    » http://dx.doi.org/10.1016/S0741-5214(95)70222-9
  • 19
    Singh K, Bønaa KH, Solberg S, Sørlie DG, Bjørk L. Intra- and interobserver variability in ultrasound measurements of abdominal aortic diameter. The Tromsø Study. Eur J Vasc Endovasc Surg. 1998;15:497-504. http://dx.doi.org/10.1016/S1078-5884(98)80109-3
    » http://dx.doi.org/10.1016/S1078-5884(98)80109-3
  • 20
    Sprouse LR, Meier GH, Parent FN, DeMasi RJ, Glickman MH, Barber GA. Is ultrasound more accurate than axial computed tomography for determination of maximal abdominal aortic aneurysm diameter? Eur J Vasc Endovasc Surg. 2004;28:28-35. PMid:15177228. http://dx.doi.org/10.1016/j.ejvs.2004.03.022
    » http://dx.doi.org/10.1016/j.ejvs.2004.03.022
  • 21
    Sun Z. Diagnostic value of color duplex ultrasonography in the follow-up of endovascular repair of abdominal aortic aneurysm. J Vasc Interv Radiol. 2006;17:759-64. PMid:16687740. http://dx.doi.org/10.1097/01.RVI.0000217944.36738.02
    » http://dx.doi.org/10.1097/01.RVI.0000217944.36738.02
  • 22
    Fletcher J, Saker K, Batiste P, Dyer S. Color Doppler diagnosis of perigraft flow following endovascular repair of abdominal aortic aneurysm. Int Angiol. 2000;19:326-30. PMid:11305731.
  • 23
    D'Audiffret A, Desgranges P, Kobeiter H, Becquemin JP. Follow-up evaluation of endoluminally treated abdominal aortic aneurysms with duplex ultrasonography: validation with computed tomography. J Vasc Surg. 2001;33:42-50. PMid:11137922. http://dx.doi.org/10.1067/mva.2001.112215
    » http://dx.doi.org/10.1067/mva.2001.112215
  • 24
    McLafferty RB, McCrary BS, Mattos MA, et al. The use of color-flow duplex scan for the detection of endoleaks. J Vasc Surg. 2002;36:100-4. PMid:12096265. http://dx.doi.org/10.1067/mva.2002.123089
    » http://dx.doi.org/10.1067/mva.2002.123089
  • 25
    Bendick PJ, Bove PG, Long GW, Zelenock GB, Brown OW, Shanley CJ. Efficacy of ultrasound scan contrast agents in the non-invasive follow-up of aortic stent grafts. J Vasc Surg. 2003;37:381-5. PMid:12563210. http://dx.doi.org/10.1067/mva.2003.17
    » http://dx.doi.org/10.1067/mva.2003.17
  • 26
    Raman KG, Missig-Carroll N, Richardson T, Muluk SC, Makaroun MS. Color-flow duplex ultrasound scan versus computed tomographic scan in the surveillance of endovascular aneurysm repair. J Vasc Surg. 2003;38:645-51. http://dx.doi.org/10.1016/S0741-5214(03)00909-1
    » http://dx.doi.org/10.1016/S0741-5214(03)00909-1
  • 27
    Elkouri S, Panneton JM, Andrews JC, et al. Computed tomography and ultrasound in follow-up of patients after endovascular repair of abdominal aortic aneurysm. Ann Vasc Surg. 2004;18:271-9. PMid:15354627. http://dx.doi.org/10.1007/s10016-004-0034-5
    » http://dx.doi.org/10.1007/s10016-004-0034-5
  • 28
    AbuRahma AF, Welch CA, Mullins BB, Dyer B. Computed tomography versus color duplex ultrasound for surveillance of abdominal aortic stent-grafts. J Endovasc Ther. 2005;12:568-73. PMid:16212456. http://dx.doi.org/10.1583/05-1575MR.1
    » http://dx.doi.org/10.1583/05-1575MR.1
  • 29
    Standford RM, Bown MJ, Fishwick G, et al. Duplex ultrasound scanning is reliable in the detection of endoleak following endovascular aneurysm repair. Eur J Vasc Endovasc Surg. 2006;32:537-41. PMid:16875850. http://dx.doi.org/10.1016/j.ejvs.2006.05.013
    » http://dx.doi.org/10.1016/j.ejvs.2006.05.013
  • 30
    Ashoke R, Brown LC, Rodway A, et al. Color duplex ultrasonography is insensitive for the detection of endoleak after aortic endografting: a systematic review. J Endovasc Ther. 2005;12:297-305. PMid:15943504. http://dx.doi.org/10.1583/04-1479R.1
    » http://dx.doi.org/10.1583/04-1479R.1
  • 31
    Parent FN, Meier GH, Godziachvili V, et al. The incidence and natural history of type I and II endoleak: a 5-year followup assessment with 1 color duplex ultrasound scan. J Vasc Surg. 2002;35:474-81. PMid:11877694. http://dx.doi.org/10.1067/mva.2002.121848
    » http://dx.doi.org/10.1067/mva.2002.121848
  • 32
    Moreira RCP. Estudo comparativo de eco-Doppler cm arteriografia na avaliação da doença oclusiva aortoiliaca. J Vasc Bras. 2009;8:3-13. http://dx.doi.org/10.1590/S1677-54492009000100002
    » http://dx.doi.org/10.1590/S1677-54492009000100002
  • 33
    Beeman BR, Doctor LM, Doerr K, McAfee-Bennett S, Dougherty MJ, Calligaro KD. Duplex ultrasound imaging alone is suficient for midterm endovascular aneurysm repair survaillance: A cost analysis study and prospective comparison with computed tomography scan. J Vasc Surg. 2009;50:1019-24. PMid:19656651. http://dx.doi.org/10.1016/j.jvs.2009.06.019
    » http://dx.doi.org/10.1016/j.jvs.2009.06.019
  • 34
    Correas JM, Bridal L, Lesavre A, Mejean A, Claudon M, Helenon O. Ultrasound contrast agents: properties, principles of action, tolerance, and artifacts. Eur Radiol. 2001;11:1316-28. PMid:11519538. http://dx.doi.org/10.1007/s003300100940
    » http://dx.doi.org/10.1007/s003300100940
  • 35
    Giannoni MF, Palombo G, Sbarigia E, Speziale F, Zaccaria A, Fiorani P. Contrast-enhanced ultrasound imaging for aortic stent-graft surveillance. J Endovasc Ther. 2003;10:208-17. http://dx.doi.org/10.1583/1545-1550(2003)010<0208:CUIFAS>2.0.CO;2
    » http://dx.doi.org/10.1583/1545-1550(2003)010<0208:CUIFAS>2.0.CO;2
  • 36
    Napoli V, Bargellini I, Sardella SG, et al. Abdominal aortic aneurysm: contrast-enhanced US for missed endoleaks after endoluminal repair. Radiology. 2004;233:217-25. PMid:15454621. http://dx.doi.org/10.1148/radiol.2331031767
    » http://dx.doi.org/10.1148/radiol.2331031767
  • 37
    Henao EA, Hodge MD, Felkai DD, et al. Contrast-enhanced duplex surveillance after endovascular abdominal aortic aneurysm repair: Improved efficacy using a continuous infusion technique. J Vasc Surg. 2006;43:259-64. PMid:16476596. http://dx.doi.org/10.1016/j.jvs.2005.09.045
    » http://dx.doi.org/10.1016/j.jvs.2005.09.045
  • 38
    Carrafiello G, Recaldini C, Lagana D, Piffaretti G, Fugazzola C. Endoleak detection and classification after endovascular treatment of abdominal aortic aneurysm: value of CEUS over CTA. Abdom Imaging. 2008;33:357-62. PMid:17619925. http://dx.doi.org/10.1007/s00261-007-9268-3
    » http://dx.doi.org/10.1007/s00261-007-9268-3
  • 39
    Rao QA, Newhouse JH. Risk of nephropathy after intravenous administration of contrast material: a critical literature analysis. Radiology. 2006;239:392-7. PMid:16543592. http://dx.doi.org/10.1148/radiol.2392050413
    » http://dx.doi.org/10.1148/radiol.2392050413

Publication Dates

  • Publication in this collection
    Jul-Sep 2014

History

  • Received
    04 Oct 2013
  • Accepted
    19 Mar 2014
Sociedade Brasileira de Angiologia e de Cirurgia Vascular (SBACV) Rua Estela, 515, bloco E, conj. 21, Vila Mariana, CEP04011-002 - São Paulo, SP, Tel.: (11) 5084.3482 / 5084.2853 - Porto Alegre - RS - Brazil
E-mail: secretaria@sbacv.org.br