A micro-flow-batch analyzer using webcam for spectrophotometric determination of Ortho-phosphate and aluminium(III) in tap water

Abstracts

In this work, a micro-flow-batch analyzer (µFBA) using webcam for spectrophotometric determination of ortho-phosphate and aluminium(III) in tap water is proposed. PO4(3-) and Al3+ determinations were performed by employing molybdenum blue and quecertin methods, respectively. In order to build linear analytical curves, a mathematical model was used on the basis of the RGB (red, green and blue) values. Working ranges were from 25 to 500 µg L-1 for both analytes with estimated limits of detection, relative standard deviation (RSD, n = 5) and sampling rate of 0.82 µg L-1, < 0.8% and 160 h-1 for PO4(3-), and 0.93 µg L-1, < 1.1% and 160 h-1 for Al(III). Comparing with the reference methods, no statistically significant differences were observed when applying the paired t-test at a 95% confidence level. Recovery study shows results between 98.1 and 102.8%. The proposed µFBA presented satisfactory portability, robustness, flexibility, low-cost device and reduced chemicals consumption.

micro-flow-batch analyzer; digital images; RGB color system; webcam; tap water


Neste trabalho é proposto um analisador micro-flow-batch (µFBA) usando uma webcam para determinação espectrofotométrica de ortofosfato e alumínio(III) em água de torneira. Este microssistema foi feito com uretana-acrilato e lâminas de vidro. As determinações de PO4(3-) e Al3+ foram realizadas empregando os métodos do azul de molibdênio e quercertina, respectivamente. Para a construção de curvas analíticas lineares um modelo matemático foi usado com base nos valores RGB (vermelho, verde e azul). As faixas de trabalho foram de 25 a 500 µg L-1 para ambos os analitos com limites de detecção, desvio padrão relativo (RSD, n = 5) e frequência analítica de 0,82 µg L-1, < 0,8% e 160 h-1 para PO4(3-), e 0,93 µg L-1, < 1,1% e 160 h-1 para Al(III). Comparando com os métodos de referência, diferenças estatisticamente significativas não foram observadas quando aplicado o teste-t pareado com 95% de confiança. O teste de recuperação apresentou resultados entre 98,1 e 102,8%. O µFBA proposto apresentou portabilidade satisfatória, robustez, flexibilidade e reduzido consumo de reagentes.


ARTICLE

A micro-flow-batch analyzer using webcam for spectrophotometric determination of Ortho-phosphate and aluminium(III) in tap water

Marcelo B. Lima; Stéfani I. E. Andrade; Maria S. Silva Neta; Inakã S. Barreto; Luciano F. Almeida; Mário César U. Araújo* * e-mail: laqa@quimica.ufpb.br

Departamento de Química, Centro de Ciências Exatas e da Natureza (CCEN), Universidade Federal da Paraíba, CP 5093, 58051-970 João Pessoa-PB, Brazil

ABSTRACT

In this work, a micro-flow-batch analyzer (µFBA) using webcam for spectrophotometric determination of ortho-phosphate and aluminium(III) in tap water is proposed. PO4 and Al3+ determinations were performed by employing molybdenum blue and quecertin methods, respectively. In order to build linear analytical curves, a mathematical model was used on the basis of the RGB (red, green and blue) values. Working ranges were from 25 to 500 µg L–1 for both analytes with estimated limits of detection, relative standard deviation (RSD, n = 5) and sampling rate of 0.82 µg L–1, < 0.8% and 160 h–1 for PO4, and 0.93 µg L–1, < 1.1% and 160 h–1 for Al(III). Comparing with the reference methods, no statistically significant differences were observed when applying the paired t-test at a 95% confidence level. Recovery study shows results between 98.1 and 102.8%. The proposed µFBA presented satisfactory portability, robustness, flexibility, low-cost device and reduced chemicals consumption.

Keywords: micro-flow-batch analyzer, digital images, RGB color system, webcam, tap water

RESUMO

Neste trabalho é proposto um analisador micro-flow-batch (µFBA) usando uma webcam para determinação espectrofotométrica de ortofosfato e alumínio(III) em água de torneira. Este microssistema foi feito com uretana-acrilato e lâminas de vidro. As determinações de PO4 e Al3+ foram realizadas empregando os métodos do azul de molibdênio e quercertina, respectivamente. Para a construção de curvas analíticas lineares um modelo matemático foi usado com base nos valores RGB (vermelho, verde e azul). As faixas de trabalho foram de 25 a 500 µg L–1 para ambos os analitos com limites de detecção, desvio padrão relativo (RSD, n = 5) e frequência analítica de 0,82 µg L–1, < 0,8% e 160 h–1 para PO4, e 0,93 µg L–1, < 1,1% e 160 h–1 para Al(III). Comparando com os métodos de referência, diferenças estatisticamente significativas não foram observadas quando aplicado o teste-t pareado com 95% de confiança. O teste de recuperação apresentou resultados entre 98,1 e 102,8%. O µFBA proposto apresentou portabilidade satisfatória, robustez, flexibilidade e reduzido consumo de reagentes.

Introduction

Recently, the micro-flow-batch analyzer (µFBA)1 was proposed in order to improve the features of the traditional flow-batch system,2 such as portability, lower consumption of reagents and sample, and less waste generation. As the flow-batch analyzers, this microsystem also combines favorable characteristics of both flow and batch analysis systems. The transportation of reagents, samples or other solutions into the micro-chamber (µCH) are carried out in a flow mode and, as in the batch system, the sample processing (mixture/homogenization/reaction to generate colorful product), and absorbance measurement is carried out into µCH (micro-beaker).3

Unlike an usual analog to digital converter employed in analytical instruments affords resolution of 12 bit (4096), recent advances in digital image acquisition technology have offered video cameras (webcam) based on charge-coupled devices (CCD), which are capable to capture digital images with up to 24 bits (16.7 million colors). In fact, by using the RGB (red, green and blue) color system, the primary colors are combined in different intensities with values varying in the range 0-255 (8 bits) per color.4 Thus, procedures based on digital images are very sensitive since they included abilities to identify little difference between the colors of images (the analytical signals).4-6

Mathematical models for digital images and RGB data have also been employed to construct analytical curves in quantitative determinations.5 Lyra et al.4 proposed a method using the norm in an RGB three-dimensional space concept for analytical response in digital images obtained from flame emission spectrometry (FES). Maleki et al.7 used digital images for simultaneous determination of Al(III) and Fe(III) in alloys using chrome azurol S (CAS) as the chromogenic reagent. The RGB based artificial neural network (ANN) models were built from digital images of the Al(III)-CAS and Fe(III)-CAS complexes.

Recently, Wongwilai et al.8 employed a webcam for real time detection in the course of an acid-base titration by means of color change in phenolphtalein indicator for increasing acid concentration in the solution of an automatic flow system. Lima et al.9 demonstrated that a webcam with CCD sensor can be coupled to µFBA and used as digital image-based method for quantitative tannin determinations in green tea.

The content of phosphorus in tap water sample is an important index to evaluate the quality of water in environmental monitoring. Molybdenum blue spectrophotometric method10 was a general method to the determining of phosphorus in the ortho-phosphate form (PO4), which had the advantages of high sensitivity, less interference and better stability.10,11 Therefore, the molybdenum blue method seems promising to be adopted as a basis for the development of a miniaturized automatic method using digital images for ortho-phosphate determination.

Aluminium(III) ion occurs into the environment by natural processes and also from anthropogenic sources.12 The toxic effects (it provokes encephalopathy, osteomalicia and microcytic anemia) are well-known also to patients with chronic renal failure.13 This ion is considered to be one of the risk factors to Alzheimer's disease.14 Although it is present in low concentrations in tap water, Al(III) is commonly used as flocculating agent in potable water treatment units, and produces problems due to excess (or insufficient) coagulant, particularly during periods of fast variation in water quality.15,16

Quercetin is a well-known photometric reagent for determination of Al(III) traces in water and in biological samples. As described by Norfun et al.,17 it forms a stable complex with Al(III) and is relatively free from interfering species commonly present in water. Thus, quercetin seems promising to be adopted as a basis for the development of a µFBA method using digital images for determination of Al(III) in water.

In the present study, an automatic µFBA method using digital images with solenoid micro-pumps is proposed for determining of ortho-phosphate and aluminium(III) using a webcam and RGB data in tap water. The phosphorus and aluminium determinations were performed by employing the molybdenum blue and quecertin methods, respectively. The colored complexes formed by these methods can be quantitatively analyzed by digital images from webcam as detector. µFBA is a novel alternative for automatic determination of these ions in tap water, comparing satisfactorily to the recently methods described in the literature.10,11,17,18

Experimental

Reagent solutions

All reagents were of analytical grade and freshly distilled and deionized water (> 18 MΩ cm–1) was used to prepare all solutions.

For spectrophotometric determination of aluminium(III), a stock solution of 100.0 mg L–1 aluminium(III) was prepared as previously17,19 from pure (> 99.0% w/w, Sigma) product by dissolving the appropriate mass of aluminium chloride hexahydrate in 0.10 mol L–1 hydrochloric acid solution.

Quercetin stock solution (1000.0 mg L–1) was prepared as previously17,19 from the pure product (Sigma) dissolving an appropriate weight and using an ethanol/water solution (60%, v/v). The surfactant solution was prepared by dissolving 0.0127 g of cetyltrimethylammonium bromide (CTAB) (Sigma), in an acetate buffer solution with a concentration of 0.10 mol L–1, at a pH of 5.5.

The stock phosphorus solution (100 mg L–1) was prepared by appropriate dilution of potassium dihydrogen phosphate (KDP) (Sigma). The standard solutions (ranging from 0.10 to 10.0 mg L–1 of phosphorus) were prepared by mixing the stock solution with the acid solutions (H2SO4, HNO3 and H2O2, all 1% v/v) (Sigma) and were used for plotting the analytical curve in µFBA.

A 340 mmol L–1 solution of ascorbic acid (C6H8O6) (Sigma) was prepared by weighing 6 g of ascorbic acid and dissolving it in 100 mL of distilled water. A 5.4 mol L–1 sulfuric acid solution was prepared by transferring an aliquot of 30 mL of concentrated H2SO4 in 100 mL of water.

As described by Norfun et al.,17 aluminium(III) determination required the addition of a masking agent, which consisted of 0.625 g of thiourea (Sigma), 0.1 mol L–1 ascorbic acid (15.0 mL) and 0.1 mol L–1 1,10-phenanthroline (25.0 mL). Then, the pH of the sample was adjusted to 5.5 with 1 mol L–1 sodium hydroxide (Sigma), transferred into a 250 mL volumetric flask and made up to the mark with deionized distilled water. Finally, it was well mixed and subsequently analyzed by the proposed µFBA method.

For the preparation of the stock solution of 32 mmol L–1 ammonium molybdate (Sigma), 20 g of salt were weighed and dissolved in 400 mL of water. After 4 h agitation, the solution was transferred to a 500 mL volumetric flask and the volume was completed with water.

The stock solution of 9 mmol L–1 potassium antimony tartrate (Sigma) was prepared by weighing 3.0 g of potassium antimony tartrate and dissolving them in distilled water. The solution was then transferred to a 1000 mL volumetric flask and the volume completed with distilled water. This solution was stored in amber glass bottles under refrigeration.

The reagent solution was prepared by mixing 35 mL of concentrated sulfuric acid in 500 mL of water, then adding 215 mL of stock solution of 0.032 mol L–1 ammonium molybdate and 72 mL of 9 mmol L–1 potassium tartrate and antimony. The mixture was transferred to a 1000 mL flask and the volume completed with water. A solution of nitric acid (10% v/v) was used to clean the microsystem fabricated.17

Sample preparation

Tap water samples were collected from different sources in the city of João Pessoa, Paraíba, Brazil. The samples were filtered through 0.45 µm cellulose membrane filters, acidified with nitric acid (1% v/v), and stored in polyethylene bottles before analysis.

Apparatus

To fabricate the micro-chamber (µCH) in urethane-acrylate resin with glass slides, it was used a commercial UV light source (Fotolight-MD2-A4, Carimbos Medeiros Ltda, Brazil), with two sets of mercury lamps (BLB-15W-T8, SCT black light).1

For layout design of the µCH, the CorelDraw® X5 program was used. The layout printing was on polyester transparency films for laser printing using an HP LaserJet P2014. After UV exposure, channels on the substrate were revealed by the removal of the non-exposed resin with an ultrasonic bath (model UltraCleaner 800, Unique, Brazil), as previously.3

A spectrophotometer model 8453 Hewlett-Packard diode array UV-Vis, equipped with cuvette (with an inner volume of about 4 mL and an optical path of 1.0 cm), was used for absorbance measurements when employing the reference method.

A Microsoft LifeCam Cinema 720p HD 30FPS was used in conjunction with LabVIEW® 7.1 software (National Instruments®) to control the µFBA system. The images were captured by means of the software written in Delphi (version 3.0).

The webcam was connected to the USB port of an Intel Core2Duo 2 Gigabytes notebook (PC) and configured to capture 24-bit digital images (16.7 million colors) at a rate of 30 frames s–1 and a 640 pixels × 480 pixels spatial resolution.

Fabrication process and assembly of the µ FBA system

The homemade µCH with two glass slides and total volume about 110 µL was built in urethane-acrylate photoresist using the methodology described elsewhere.9 This microsystem is mounted onto a suitable support in a black (darkroom) box measuring 17 cm × 11 cm × 9 cm to preserve the system from the effects of spurious environmental radiation while in operation (Figure 1).


Six white high intensity LEDs (NPE, Thailand) were arranged in the wall behind the webcam in order to give a constant light intensity throughout the experiment. The interior walls of the box were covered with white paper in order to provide uniform illumination and reduce glare.

The fluids are added with solenoid micro-pumps with nominal values of 8 µL (µP1-µP4) and 20 µL (µP5) per pulse (Biochem Valve Inc., Boonton, NJ, USA). Teflon® tubes with 0.5 mm internal diameter were used for fluid transport. A 0.4 mm nylon wire was used within µCH to ensure an efficient homogenization. The nylon wire was coupled to a CD/DVD-ROM motor drive (model MDN3GT3CPAC, 2000 rpm, 5 V dc).

Analytical procedure

The µFBA manifold is shown in Figure 2. The analyzer was operated as described in Table 1 for the different determinations. The solenoid micro-pumps were actuated at 5 Hz, yielding flow rates from 0.96 mL min–1 (µP1 at µP4) and 2.4 mL min–1 (µP5).


Ortho -phosphate determination

The reagents ammonium molybdate with potassium antimony tartrate (4 pulses, adding 32 µL by µP2), ascorbic acid (4 pulses, adding 32 µL by µP3) and sample or standard solution (4 pulses, adding 32 µL by µP1) were simultaneously added. The homogenization was performed by the drive motor (DM) coupled to the nylon wire (NW) by 2 s. After these steps, digital images are then captured and µCH is emptied (5 pulses, removing 96 µL by µP5).

Afterwards, µCH is cleaned by activation of µP4 (12 pulses), adding 96 µL of cleaning solution (10% v/v HNO3) while activated, and DM is activated for 2 s performing the agitation. Then, µP5 is activated (5 pulses) to discard the contents of the µCH. This cleaning and discard procedure must be done twice to effectively clean the µCH.

The procedure for in line blank preparation is similar to described for the sample analysis. The difference is that 0.1 mol L–1 H2SO4 solution is used instead of the sample or standard solutions.

Aluminium(III) determination

The sample or standard solutions (4 pulses, adding 32 µL by µP1), quercetin solution (4 pulses, adding 32 µL by µP2) and CTAB solution (4 pulses, adding 32 µL by µP3) were simultaneously added. The homogenization was performed by the drive motor (DM) coupled to the nylon wire (NW) by 2 s. After these steps, the digital images are then captured and µCH is emptied (5 pulses, removing 96 µL by µP5).

Afterwards, µCH is cleaned by activation of µP4 (12 pulses), adding 96 µL of water and DM is activated for 2 s performing the agitation. Then, µP5 is activated (5 pulses) to discard the contents of µCH. This cleaning and discard procedure must be done twice to effectively clean µCH.

The procedure for in line blank preparation is similar to described for the sample analysis. The difference is that water of µP4 (12 pulses) is used instead of the sample or standard solutions.

The treatment of digital images and mathematical model

The treatment of the captured digital images was made by means of a second software also written in Delphi (version 3.0). The routine with the working stages of this software is similar to that used elsewhere6,9 and illustrated in Figure 3. Initially the user selects the most homogeneous region in the image which will define the coordinates of the selected region, and will also be used for all other images. Then, the software scans all the pixels, column by column, to extract the RGB component for each pixel and calculate a mean integer value of each RGB component. These mean values are used in the RGB-based value calculation (analytical response) as described below.


The RGB-based values were calculated by means of a mathematical model developed from the concept of vector norm "||ν||",6 calculated as shown in equation 1:

where

s–b, s–b and s–b result from the difference between the s, s and s average values obtained from digital images of the standard solutions and samples and b, b, and b, from the blank.

As a result, a linear relationship was observed between the analyte concentration (C) in the standards solution (or sample) and RGB data, for which the following equation is valid:

As demonstrated in previous work,6 equation 2 provides the basis for building µFBA analytical curves, establishing a linear relationship between ||ν|| (RBG data value adopted as analytical response) and analyte concentration in standards solutions. Furthermore, the vectors associated with the digital images from each analyte should be positioned on the same support line in the RGB three-dimensional space.

Batch reference method

For comparison, the proposed µFBA performance was evaluated against a molybdenum blue method (reference method).20 Standard solutions were prepared from 1.0 to 10.0 mg L–1. The analytical signals (absorbance) were measured at a maximum absorbance of around 882 nm. The analysis of each sample was performed in triplicate and the concentrations were calculated from the analytical curve.

The reference spectrophotometric procedure for aluminium(III) in water analysis was carried out according to standard methods (Standard Methods, 20th ed., Method 3500-Al B, Eriochrome cyanine R method).21 The analyses were performed at a maximum absorbance of around 535 nm.

Results and Discussion

µ FBA parameters

Volumes of reagents and samples to be added inside µCH during the automatic procedure were evaluated in order to improve the sensitivity and reproducibility of the analytical signal. The system value selection (univariate method) was carried out in conformance with studies by Chuan-Xiao et al.10 (for PO4) and Norfun et al.17 (for Al3+).

For both methods, volumes of 32 µL of sample (or standard solution) and reagents were selected as the best compromise between reproducibility and sensitivity considering the volume used in the total mixing chamber (about 96 µL). The range evaluated and values selected for each parameter are shown in Table 2. If need, other sample dilutions may be carried out in line by simply changing the operational parameters in µFBA control software.

The captured digital images

Figure 4 shows the digital images obtained from six standard solutions with different concentrations. The first image of the sequences is the blank solution. All images represent a selected area equal to 35 × 55 pixels. As can be seen in Figure 4, the images present an increase in the intensity for the color dark purple with analytical concentration in each standard solution.


Each sampled image was a matrix with 640 × 480 pixels and the delimited region by the user has been a matrix with about 35 pixels × 55 pixels. The pixel resolution was 87×87 dots in-1. The RGB-based value calculations are based on the product 2R × 2G × 2B, where R, G and B are the red, green and blue color components, respectively. These components may assume integer values in a range from 0 to 7, reaching up to 16,777,216 colors.

Other linear relationships between complex color and the ratios B/R, B/G, B/Itot (Itot = RGB), and log (B) were also studied to possibly maximize precision, as already done in other works.19 However, in all cases, the results were poor when compared to using vector norm for concentration of the complex formed. The statistical quality of the regression was verified by the residuals and analysis of variance (ANOVA). They confirmed the homoscedastic distribution of residuals and a significant linear regression.

Potential interfering ions

Natural water is a complex matrix presenting several chemical species, as ions, which could cause interferences in the analysis of interest. Thus, it is necessary to test the effect of the ion coexistence on the color reaction for the determination of the studied ions. In this sense, a set of assays was performed in order to establish the tolerance limit of the proposed methods. It was considered as a tolerable limit the interferent concentration of foreign ions that produced an error up to a 5% deviation in the determination of ortho-phosphate and aluminium(III) alone on comparing with measurements obtained in the absence of the assayed substance.

For the determination of ortho-phosphate, some ions usually present in water were selected according to the literature17 and results obtained are summarized in Table 3. As can be see, the spectrophotometric procedure has a wide tolerance for the assayed chemical species and severe interferent effects do not take place, at a relatively low concentration level. Compared with the spectrophotometric method, factors that do not influence the determination would not have any effect for the present method. Therefore, the proposed spectrophotometric method using digital images has a satisfactory practicability.

For the determination of Al(III) effects of interfering ions (Table 3) were again evaluated in a similar way as made by Norfun et al.17 Briefly, the maximum w/w ratio of interfering solutions containing 50 µg L–1 Al(III) and different concentrations of some metals which might be present in waters were tested, and the absorbances were measured. All tested cations and anions caused interference < 5% for the determination of the analyte of interest. However, Cu(II), Fe(II) and Fe(III) were again observed as the most serious interferences. Thus, a solution containing thiourea, ascorbic acid and 1,10-phenanthroline was also used as masking for Cu(II), Fe (II) and Fe(III), as adopted by Norfun et al.17

The µ FBA applications

For the ortho-phosphate determination in tap water, the regression equation was A = 15.4796 + 0.0214 C, where A is the analytical response and C is the analytic concentration in µg L–1 of PO4. The linear correlation coefficient (r) was 0.9995 (n = 5) in the range between 25.0 to 500 µg L–1 (25, 50, 100, 250 and 500 µg L–1). For the determination of Al3+ in tap water using the µFBA system, the regression equation was A = 21.1871 + 0.0298 C, where C is the aluminium(III) concentration in µg L–1 in the measuring solution. The squared linear correlation coefficient was 0.996 (n = 5) in the range between 25.0 to 500.0 µg L–1 (25, 50, 100, 250 and 500 µg L–1).

All analytical curves were statistically validated by analysis of variance (ANOVA), showing no significant lack of fit in the proposed models at a 95% confidence level. The limit of detection (LOD) and the limit of quantification (LOQ) for both methods were estimated based on the criteria established by International Union of Pure and Applied Chemistry (IUPAC).22 LOD was evaluated as 3 times the standard deviation of the blank measurement and LOQ was evaluated as being 10 times the standard deviation of the blank measure. For the PO4 determination, LOD and LOQ were 0.82 and 2.75 µg L–1, respectively; and for the Al(III) determination, LOD and LOQ were 0.92 and 3.09 µg L–1, respectively.

Table 4 presents the results for the proposed µFBA, and the reference spectrophotometric method for PO4 and Al(III) determinations in tap water.20,21 No statistically significant differences were observed between the results at a confidence level of 95% when applying the paired t-test. The relative standard deviation (RSD) was less than 1.1 and 0.8%, for the ortho-phosphate and aluminium(III), respectively, from three replicates.

Recovery tests were also performed and, in this case, three samples were used for each analyte. 1.0 mL standard solution with known concentrations of 10.0, 25.0 and 50.0 µg L–1 for each analyte was added to 9.0 mL of the tap water (21.4, 85.4 and 109.2 µg L–1 of PO4, and 45.8, 69.1 and 112.1 µg L–1 Al(III)), for measurement using the proposed µFBA. The recovery values obtained are shown in Table 5. As can be seen, the recoveries obtained for each of the samples were within the 98.1-102.8% range.

It is worth noting that a solution containing thiourea, ascorbic acid and 1,10-phenanthroline was used as masking for Cu (II), Fe (II) and Fe(III). Nevertheless, the question concerning to the use of 1,10-phenanthroline as a masking reagent for iron needs to be commented. Ascorbic acid is an efficient reducing reagent for Fe (III), while 1,10-phenanthroline reacts with Fe(II) and forms a compound that absorb radiation around 530 nm, so that it should cause interference. The results shown that no significant interfering effect was observed, thus for analytical proposal, the procedure was ready to use. According to Meyer et al.,23 thiourea is a versatile ligand for iron, thus this feature would inhibit the formation of the complex with 1,10-phenatroline. In this sense, the use of 1,10-phenanthroline would be not necessary. However, such finding was not investigated in this work.

Table 6 presents selected analytical features of the proposed µFBA, and other automated systems as described in the literature for ortho-phosphate10,11 and aluminium(III)17,18 determinations in water. In general, µFBA (compared to recently methods) presents satisfactory parameters such as limit of detection, relative standard deviation, working range, sampling rate, in line preparation of calibration solutions, elimination of the carrier fluid and no associated fluid dispersion problems, such as loss of sensitivity.

Conclusion

This study proposes a portable µFBA for photometric determination of water quality parameters. For the illustration, the analyzer was applied for determination of two important quality parameters in tap water: PO4 and Al3+. This system presents a high sampling throughput, thanks to optimizations in the addition times and volumes of the fluids in the microsystem that allowed homogenization efficient and discrete measurements, with lower waste generation, contributing to the basic principles of green chemistry and the advancement of microanalysis.

By the use of an inexpensive webcam for analytical detection, the proposed µFBA offers an alternative to traditional spectrophotometric analysis. The use of the webcam dispenses the wavelength selection, which allow reduces costs and simplifies the instrumentation. With the webcam it is also possible to implement chemometric treatments due to the trivariate nature of the detection (when used RGB data), as well as the spatial-resolution characteristics inherent in digital images.

Acknowledgements

The authors would like to thank the Brazilian agencies (CNPq and CAPES) for the research fellowships and scholarships.

Submitted: November 12, 2014

Published: March 28, 2014

  • 1. Lima, M. B.; Barreto, I. S.; Andrade, S. I. E.; Almeida, L. F.; Araújo, M. C. U.; Talanta 2012, 100, 308.
  • 2. Diniz, P. H. G. D.; Almeida, L. F.; Harding, D. P.; Araújo, M. C. U.; TrAC, Trends Anal. Chem 2012, 35, 39.
  • 3. Lima, M. B.; Barreto, I. S.; Andrade, S. I. E.; Neta, M. S. S.; Almeida, L. F.; Araújo, M. C. U.; Talanta 2012, 98, 118.
  • 4. Lyra, W. S.; Santos, V. B.; Dionízio, A. G. G.; Martins, V. L.; Almeida, L. F.; Gaião, E. N.; Diniz, P. H. G. D.; Silva, E. C.; Araújo, M. C. U.; Talanta 2009, 77, 1584.
  • 5. Gaião, E. N.; Martins, V. L.; Lyra, W. S.; Almeida, L. F.; Silva, E. C.; Araújo, M. C. U.; Anal. Chim. Acta 2006, 570, 283.
  • 6. Tôrres, A. R.; Lyra W. S.; Andrade, S. I. E.; Andrade, R. A. N.; Silva, E. C.; Araújo, M. C. U.; Gaião, E. N.; Talanta 2011, 84, 601.
  • 7. Maleki, N.; Safavi, A.; Sedaghatpour, F.; Talanta 2004, 64, 830.
  • 8. Wongwilai, W.; Lapanantnoppakhun, S.; Grudpan, S.; Grudpan, K.; Talanta 2010, 81, 1137.
  • 9. Lima, M. B.; Andrade, S. I. E.; Barreto, I. S.; Almeida, L. F.; Araújo, M. C. U.; Microchem. J. 2013, 106, 238.
  • 10. Chuan-Xiao, Y.; Xiang-Ying, S.; Bin, L.; Hui-Ting, L.; Chin. J. Anal. Chem. 2007, 35, 850.
  • 11. Gentle, B. S.; Ellis, P. S.; Faber, P. A.; Grace, M. R.; McKelvie, I. D.; Anal. Chim. Acta 2010, 674, 117.
  • 12. čančar, J.; Milačič, R.; Anal. Bioanal. Chem. 2006, 386, 999.
  • 13. Savory, J.; Bertholf, R. L.; Wills, M. R.; Clin. Endocrinol. Metab 1985, 14, 681.
  • 14. Yokel, R. A. In Elements and Their Compounds in the Environment, vol. 2.; Merian, E.; Anke, M.; Ihnat, M.; Stoeppler, M., eds., Wiley-VCH: Weinheim, Germany, 2004, p. 635.
  • 15. Tria, J.; Butler, E. C. V.; Haddad, P. R.; Bowie, A. R.; Anal. Chem. Acta 2007, 588, 153.
  • 16. Vanloot, P.; Branger, C.; Margaillan, A.; Brach-Papa, C.; Boudenne, J. L.; Coulomb, B.; Anal. Bioanal. Chem. 2007, 389, 1595.
  • 17. Norfun, P.; Pojanakaroon, T.; Liawraungrath, S.; Talanta 2010, 82, 202.
  • 18. Mesquita, R. B. R.; Rangel, A. O. S. S.; J. Braz. Chem. Soc. 2008, 19, 1171.
  • 19. Andrade, S. I. E.; Lima, M. B.; Barreto, I. S.; Lyra, W. S.; Almeida, L. F.; Araújo, M. C. U.; Silva, E. C.; Microchem. J. 2013, 109, 106.
  • 20. American Public Health Association (APHA), American Water Works Association (AWWA), Water Environment Federation (WEF); Standard Methods for the Examination of Water and Wastewater, 19th ed.; American Public Health Association: Washington, DC, 1998.
  • 21. Clesceri, L. S.; Greenberg, A. E.; Eaton, A. D.; Standard Methods for the Examination of Water and Wastewater, 20th ed.; American Public Health Association: Washington, DC, USA, 1998.
  • 22. McNaught, A. D.; Andrew, W.; IUPAC Compendium of Chemical Terminology, 2nd ed.; Royal Society of Chemistry: Cambridge, UK, 1997.
  • 23. Meyer, S.; Demeshko, S.; Dechert, S.; Meyer, F.; Inorg. Chim. Acta 2010, 363, 3088.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      30 May 2014
    • Date of issue
      May 2014

    History

    • Received
      12 Nov 2014
    • Accepted
      28 Mar 2014
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br