Development of Molecularly Imprinted Polymers for Solid Phase Extraction of Parabens in Plasma Samples and Analysis by UHPLC-MS/MS

Abstract

Two molecularly imprinted polymers (MIPs) were synthesized, by sol-gel and radical polymerization procedures. These MIPs were evaluated as selective stationary phases for miniaturized solid phase extraction (MISPE) to determine parabens in plasma samples by ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS). Selective cavities of the MIPs, based on molecular recognition (hydrogen bonds), were revealed after removal of the template (benzylparaben). Both MIPs were characterized by scanning electron microscope (SEM) and Fourier-transform infrared spectroscopy (FTIR). MIP prepared by radical process presented higher selective binding capacity for the target parabens than MIP prepared by sol-gel. Several variables that influence MISPE performance, including sample pH, sorbent washing, elution solvent, and elution solvent volume were also evaluated. According to the analytical validation parameters, the MISPE/UHPLC-MS/MS method is suitable for the analysis of parabens in plasma samples. The effectiveness of this method was demonstrated through the analyses of the umbilical plasma samples from postpartum women.

Keywords:
parabens; molecularly imprinted polymer; miniaturized solid phase extraction; UHPLC-MS/MS; plasma samples


Introduction

Parabens, neutral esters derived from p-hydroxybenzoic acid, are used on large scale as chemical preservatives in cosmetic products. Parabens at low concentrations (0.016 to 0.4%) inhibits the growth of bacteria and fungi. These compounds also have low reactivity and showed low package absorption.11 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.

2 Darbre, P. D.; Harvey, P. W.; J. Appl. Toxicol. 2008, 28, 561.
-33 Darbre, P. D.; Aljarrah, A.; Miller, W. R.; Coldham, N. G.; Sauer, M. J.; Pope, G. S.; J. Appl. Toxicol. 2004, 24, 5. When parabens are present in high concentrations in the dermis, they may accumulate in the human body tissues in a similar way to other lipophilic compounds, which are bioaccumulative, or they may induce allergic dermatitis in sensitive individuals.33 Darbre, P. D.; Aljarrah, A.; Miller, W. R.; Coldham, N. G.; Sauer, M. J.; Pope, G. S.; J. Appl. Toxicol. 2004, 24, 5.,44 Frederiksen, H.; Jorgensen, N.; Andersson, A.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262.

Although the use of parabens is regulated, traces concentrations of unhydrolyzed parabens have been determined in different biological samples, including plasma,55 Zhao, Y.; Liu, G.; Shen, H.; Shen, J.; Aubry, A. F.; Sivaraman; L.; Arnold M. E.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2014, 947, 68. breast milk,66 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538.

7 Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.

8 Alshana, U.; Ertas, N.; Goger, N. G.; Food Chem. 2015, 15181, 1.
-99 Gómez, R. R.; Pijuán, M. R.; Lucena, R.; Cárdenas, S.; Gómez, A. Z.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26. ovarian tissues,1010 Sajid, M.; Basheer, C.; Narasimhan, K.; Choolani, M.; Lee, H. K.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2015, 1000, 192. urine,44 Frederiksen, H.; Jorgensen, N.; Andersson, A.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262.,77 Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.,1111 Azzouz, A.; Rascón, A. J.; Ballesteros, E.; J. Pharm. Biomed. Anal. 2016, 119, 16.

12 Jardim, V. C.; Melo, L. P.; Domingues, D. S.; Queiroz, M. E. C.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2015, 974, 35.

13 Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.
-1414 Navarro, M. V.; Carballo, M. C. M.; Torres, R. F.; Mochón, M. C.; López, M. A. B.; Anal. Bioanal. Chem. 2016, 408, 1615. and serum44 Frederiksen, H.; Jorgensen, N.; Andersson, A.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262.,77 Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.,1313 Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.,1515 Soria, F. V.; Ballesteros, O.; Rodríguez, I.; Gómez, A. Z.; Ballesteros, L.; Cela, R.; Navalón, A.; Anal. Bioanal. Chem. 2013, 405, 7259. by chromatographic or electrophoresis techniques.

Biological samples may not be introduced in their nature state in chromatographic systems, due to their endogenous compounds, mainly the proteins that can: (i) suppress the ionization of analytes, during the ionization process (liquid chromatography-mass spectrometry (LC-MS) analysis); (ii) co-elute with analytes, during the chromatographic separation; or (iii) adsorb irreversibly in analytical column. Therefore, several sample preparation techniques, including protein precipitation,1616 Mayes, A. G.; Whiticombe, M. J.; Adv. Drug Delivery Rev. 2005, 57, 1742. solid phase extraction (SPE),44 Frederiksen, H.; Jorgensen, N.; Andersson, A.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262.,55 Zhao, Y.; Liu, G.; Shen, H.; Shen, J.; Aubry, A. F.; Sivaraman; L.; Arnold M. E.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2014, 947, 68.,1111 Azzouz, A.; Rascón, A. J.; Ballesteros, E.; J. Pharm. Biomed. Anal. 2016, 119, 16.,1313 Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120. stir-membrane solid-liquid-liquid microextraction (SM-SLLME),99 Gómez, R. R.; Pijuán, M. R.; Lucena, R.; Cárdenas, S.; Gómez, A. Z.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26. dispersive liquid-liquid microextraction (DLLME),88 Alshana, U.; Ertas, N.; Goger, N. G.; Food Chem. 2015, 15181, 1.,1515 Soria, F. V.; Ballesteros, O.; Rodríguez, I.; Gómez, A. Z.; Ballesteros, L.; Cela, R.; Navalón, A.; Anal. Bioanal. Chem. 2013, 405, 7259. and microextraction by packed sorbent (MEPS),1212 Jardim, V. C.; Melo, L. P.; Domingues, D. S.; Queiroz, M. E. C.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2015, 974, 35. have been used in combination with chromatographic or electrophoresis techniques to determine parabens in biological samples.

Due to the complexity of biological samples and the low concentration (trace levels) of parabens in biological samples, molecularly imprinted polymers have been used to increase selectivity and sensitive of the SPE method.66 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. Molecularly imprinted polymers (MIPs), synthetic materials based on biomimetic systems (molecular recognition sites), such as enzyme-substrate or antigen-antibody,1616 Mayes, A. G.; Whiticombe, M. J.; Adv. Drug Delivery Rev. 2005, 57, 1742.,1717 Yan, H.; Ho Row, K.; Int. J. Mol. Sci. 2006, 7, 155. have been used as sorbent for SPE (miniaturized solid phase extraction (MISPE)) to determine different analytes in biological samples.1818 Du, B.; Qu, T.; Chen, Z.; Cao, X.; Han, S.; Shen, G.; Wang, L.; Talanta 2014, 129, 465.

19 Cormack, P. A. G.; Harun, N; Anderson, R. A.; Anal. Bioanal. Chem. 2010, 39, 2449.
-2020 Clausen, D. N; Visentainer, J. V.; Tarley, C. R. T.; Analyst 2014, 139, 19, 5021.

Most polymers of molecular imprinting based on acrylic polymers or organic acrylates are synthesized by radical polymerization. The main advantages of the radical MIP are chemical stability, robustness, reproducibility in synthesis procedure, higher adsorption capacity, and low cost.2121 Ferrer, I.; Lanza, F.; Tolokan, A.; Horvath, V.; Sellergren, B.; Horvai, G.; Barceló, D.; Anal. Chem. 2000, 72, 3934. The sol-gel process is an alternative technique. The organic-inorganic materials are obtained through sequential stages of hydrolysis and condensation reactions of metal alkoxides in acidic or basic condition. MIP via sol-gel process are porous, robust, chemically inert, and exhibit high affinity, selectivity and efficient mass transfer rate.66 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538.

This manuscript describes the development of two molecularly imprinted polymers by distinct process, sol-gel and radical polymerization that were used as stationary phase for MISPE to determine parabens in plasma samples by ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) to assess the human exposure to these compounds.

Experimental

Materials

The chromatographic standards of methyl, ethyl, propyl and butyl-parabens were acquired from Sigma-Aldrich (São Paulo, SP, Brazil). The diluted standard solutions of parabens were prepared in methanol from their respective stock solutions, 10.0 mg mL-1. These solutions were stable for 12 months under storage conditions at -4 °C. Methanol, acetonitrile, tetrahydrofuran, all with chromatographic grade, and sodium acetate, were acquired from J. T. Baker (Phillipsburg, NJ, USA). The ultrapure water was purified by Milli-Q (Millipore, São Paulo, Brazil) system. The mobile phase was filtered through a vacuum filtration system (Millipore, São Paulo, Brazil), and degassed in an ultrasound bath. The 4-vinylpyridine, ethylene glycol dimethacrylate, tetraethyl orthosilicate and (3-aminopropyl) trimethoxysilane reagents were acquired from Sigma-Aldrich (São Paulo, SP, Brazil), and the 2,2'-azobisisobutyronitrile from Merck (Darmstadt, Germany). The ammonium hydroxide was acquired from J. T. Baker (Phillipsburg, NJ, USA).

In order to obtain plasma samples for analysis, immediately after the partum, the blood collection in venipuncture was performed, kept in the antecubital fossa, by the right arm from ten women in the Hospital das Clínicas da Faculdade de Medicina de Ribeirão Preto, University of São Paulo (Ribeirão Preto, SP, Brazil). The venipuncture of fetal blood was performed in the umbilical cord immediately after handling the placenta. Then the blood samples were centrifuged at 3000 rpm for 10 min, and the supernatant (plasma) was stored at -20 °C, until the time of analysis.

Methods

The employed LC system was a Varian 230 ProStar (Varian, California, USA) equipment. Signals were monitored at 250 nm by a Varian 310 ProStar UV detector. Separation was performed in an RP 18 LichroCART(r) column (125 mm × 4 mm × 5 µm particle size-Merck, Darmstadt, Germany), at room temperature (25 °C), using methanol and water (62:38 v/v) as the mobile phase, in the isocratic mode, at a flow rate of 1.0 mL min-1.

The UHPLC-MS/MS system consisted of a Waters ACQUITY UPLC (Waters Corporation, Milford, MA, USA) using an ACQUITY UPLC binary solvent manager and ACQUITY UPLC sample manager, coupled with a Waters XEVO TQD sequential mass spectrometer (tandem quadrupole) with electrospray ionization (ESI). Chromatographic separations were carried out using a Kinetex C18 (100 mm × 2.1 mm × 1.7 µm) (Phenomenex, CA, USA) column at 40 °C, under isocratic conditions with mobile phase composed of acetonitrile:water (40:60 v/v) at a flow rate of 0.3 mL min-1. Data were acquired in the MRM (multiple reaction monitoring) mode, with negative ionization. The analytical parameters were as follows: capillary voltage: 3.30 kV, ion source temperature: 150 °C, extractor: 3.0 V, and RF lens: 2.5 V. Nitrogen was used as desolvation gas at a flow-rate of 1000 L h-1, and at the temperature of 450 °C. Cone voltage (CV) was set up individually for each analyte (Table 1). The product ions were selected according to the fragmentation pathways in the daughter scan mode. Argon was the collision gas; collision energy (CE) was optimized for each analyte individually (Table 1). Instrument control, peak detection and integration were carried out using MassLynx 4.1 Data System. QuanLynx software was used for data processing and quantitation: regression analysis of standard, matrix calibration curves and calculation of concentrations.

Table 1
MS/MS transition (MRM), optimal declustering potential (DP), optimal collision energy (CE) and retention time (tR) for the parabens studied. Dwell time = 0.027. *IS: Internal standard (deuterated propylparaben)

Two specific transitions were optimized for each molecule, to increase the selectivity and reliability of the method. The first transition was used to quantify the analytes; while the second MRM transition was used for qualitative identification.

The MS/MS detection method was optimized by direct individual infusion of each compound, to obtain better response of the precursor ion. ESI interface in the positive and negative modes were evaluated for all the compounds. ESI negative mode was selected, because it provided higher sensitivity for all of target analytes.

Syntheses of the molecularly imprinted polymers

For radical polymerization, in a conic vial, following the stoichiometric ratio (1:4:20 v/v/v) among template, functional monomer, and crosslinking agent; 22 mg of benzylparaben (template molecule) was dissolved in acetonitrile (porogen), followed by the addition of 49 µL of the 4-vinylpyridine (functional monomer), and 430 µL of ethylene glycol dimethacrylate (crosslinking agent). This vial was mixed for 10 s and sonicated for 10 min. After this procedure, 16 mg of 2,2'-azobisisobutyronitrile (initiator) was added to the reaction vial, and incubated at 60 °C for 24 h for polymerization. The acetonitrile volume (15, 30, 60 and 90 mL) was also evaluated.

For sol-gel procedure, 12.2 µL of the (3-aminopropyl)trimethoxysilane (APTMS) (precursor), and 500 µL of the benzylparaben solution (5 mg mL-1) were added to the reaction vial, and mixed for 10 s. After this step, 88.4 µL of tetraethyl orthosilicate (TEOS) and 25 µL of NH4OH were also added to the reaction vial. This vial was incubated for 12 h at room temperature (25 °C).

Monoliths obtained in both syntheses procedures were crushed, ground and sieved to obtain particles in the 50 and 100 µm size range. Subsequently, the template was removed from the MIP selective cavities by soxhlet extraction for approximately 72 h, using methanol:ethanol solution (90:10 v/v). The template removal was monitored by liquid chromatography-ultraviolet detector (LC-UV) analysis, until the benzylparaben presence was not observed in the chromatograms.

To evaluate the selectivity of each MIP synthesized, non-imprinted polymers (NIP) were prepared using the same procedure, but omitting the template molecule.

Physical and chemical characterization of the MIPs

The surface morphology analysis of the MIP and NIP polymers were performed on a scanning electron microscope (Zeiss EVO5O) at the Department of Chemistry, FFCLRP, University of São Paulo. Measurements by scanning electron microscopy were carried out with magnitude of 50.00 k× and nominal resolution of 200 nm; and magnitude of 5.00 k× and nominal resolution of 2 µm. The measurements were made under 20.00 and 5.00 kV.

Infrared analyses were performed on a Fourier transform infrared spectrophotometer (ABB Bomem MB100 series) at Department of Chemistry, FFCLRP, University of São Paulo. The spectra were analyzed in KBr pellet in a range from 4000 to 400 cm-1 and 200 cumulative scans.

Miniaturized MISPE process

Sterile CA syringe filters RC of 0.20 µm pore size and 4 mm diameter were used as extraction devices for SPE. These SPE devices were packed with 10 mg of the synthesized polymers. Initially, the SPE devices containing the MIP phase were conditioned with 200 µL of methanol and 200 µL of Milli-Q water. Prior to the MISPE procedure, the proteins of plasma samples (200 µL) were precipitated with acetonitrile (400 µL). After the centrifugation, the supernatant was evaporated under N2 flow, and the dried extract was reconstituted with 200 µL of the ultrapure water. This extract was percolate through the MIP phase. For the removal of endogenous compounds, 100 µL of the ultrapure water were also percolated through this phase. After this step, parabens were eluted from the MIP with methanol (300 µL). The average flow-rate of the solvents through the cartridges was 500 µL min-1 (10 mmHg). The extract was evaporated under N2 flow, and the dried extract was reconstituted with mobile phase (methanol:water solution, 80:20 v/v). 5 µL of this solution were injected into UHPLC-MS/MS system.

Analytical validation

The MISPE (radical polymerization)/UHPLC-MS/MS method was validated using paraben-free plasma samples (blank plasma samples) spiked with the internal standard (45 ng mL-1) and parabens standard solutions at the plasma concentrations ranging from 1 to 50 ng mL-1.

The linearity was evaluated by calibration curves constructed by linear regression of the ratio between the parabens and the internal standard (Y) peak areas versus the nominal parabens concentrations (X, ng mL-1). The linearity of the calibration curve was determined using linear regression analysis. The lower limit of quantification (LLOQ) was defined as the lowest concentration in the calibration curve that could be quantitatively measured with acceptable precision and accuracy (within 20%).

The selectivity of the method was evaluated by comparison of MRM chromatograms of the spiked plasma samples at the LLOQ level with those of the human blank plasma samples. The response of co-eluting interferences should be less than 20% of the response of the analytes and less than 5% of the peak area of the internal standard (IS). Inter-assay accuracy and precision values were determined by replicate (n = 5) analyses (MISPE/UHPLC-MS/MS assays) of the blank samples spiked with the analytes at six different concentrations throughout the linear range. The accuracy was calculated by the relative standard error (RSE), values outside the range of ± 15% of the nominal value were not accepted, except for the limit of quantitation for which values outside the range of ± 20% of the nominal value was not accepted.

During validation, carry-over was assessed by injecting three aliquots of the same blank plasma sample: one before and two after analysis of the plasma sample spiked with analytes at the concentration that corresponded to the upper limit of quantification (ULOQ: 50 ng mL-1). Carry-over in the blank sample that evaluated immediately after the ULOQ sample should not be greater than 20% of the analytes signals at LLOQ chromatogram, and not greater than 5% of the internal standard signal (IS).

Matrix effects were investigated using at least five lots of blank matrix from individual donors. Considering each analyte and the IS, the matrix factor (MF) was determined for each lot of matrix, by calculating the ratio between the peak area of the analytes in the presence of matrix (measured by analyzing spiked blank matrix after extraction of the analyte), and the peak area of the analytes in the absence of matrix (pure solution of the analyte). The IS normalized MF was also calculated by dividing the MF of the analyte by the MF of the IS. The CV of the IS-normalized MF calculated from the five lots of matrix should not be greater than 15%. This determination (n = 5) was conducted at low and high concentrations of analytes (3.0 and 40 ng mL-1), respectively.

Results and Discussion

Optimization of the syntheses

For radical polymerization, the functional monomer, 4-vinylpyridine, was selected due to its alkalinity. Acetonitrile (porogenic solvent), which is a polar protic solvent, was evaluated in different volumes (15, 30, 60 and 90 mL), data not shown. Increasing the acetonitrile volume in reactional medium, the MIP porosity reduces, and consequently, the extraction efficiency, and parabens diffusion decrease.

For sol-gel process, the synthesis of MIP consisted of the hydrolysis and condensation reactions of the alkoxide precursors (APTMS and TEOS) in a basic medium in the presence of benzylparaben, template molecule. The NH4OH was used as a catalyst for the sol-gel reaction. Under basic conditions, the condensation of the silanol groups occurs preferably among highly branched oligomers, which generate the gel particles, and then the polymeric materials with high porosity, after drying.2222 Pilau, E. J.; Silva, R. G.; Jardim, I. C. S. F.; Augusto, F.; J. Braz. Chem. Soc. 2008, 19, 1136. The addition of TEOS results in higher network intersection. The choice of APTMS was due to its specific properties, such as hydrophobicity, and reactivity.

For both syntheses, benzylparaben was chosen as a template, because among the evaluated parabens, benzylparaben presents selective cavities with suitable size, shape, and functionality for selective extraction of analogues molecules. Furthermore, benzylparaben has not been used as a preservative in cosmetic products and should not be found in human plasma samples.

Characterization of MIPs

MIP obtained by radical polymerization presented higher porosity and more spherical and homogeneous particles than MIP obtained by sol-gel process (Figure 1). The MIP obtained by sol-gel process resulted in a cluster of non-regular shape of the particles.

The rougher surface of the MIP synthesized by radical polymerization (Figure 1) should be resulted in higher area surface area, and consequently, higher extraction efficiency.

Figure 1
Scanning electron microscopy of the polymers obtained by: (a) radical polymerization with magnitude 5.00 k×; (b) sol-gel process with magnitude 5.00 k×; (c) radical polymerization with magnitude 50.00 k× and (d) sol-gel process with magnitude 50.00 k×.

Based on the analysis of the IR spectra, the spectral band at 3439 cm-1 (radical MIP and sol-gel MIP) is related to the N-H vibrations of symmetric and asymmetric stretch of the primary amine.

Figure 2 illustrates the absorption bands of C-H bonds at 2936 (Figure 2b) and at 2953 cm-1 (Figure 2a) (relating to the asymmetric stretching of the CH2 groups). Figure 2b illustrates the symmetrical angular deformation out of the plane of the grouping NH amine in the region of 795 cm-1. This band is very sensitive to the formation of hydrogen bonds. The bands present at 1633 (Figure 2b) and at 1602 cm-1 (Figure 2a) are typical binding generated by C-H functional groups (amino) present in both polymers. It is also possible to allocate the bands in the spectrum radical MIP (Figure 2a) in 1728 (C=O), 1618 (C=C), 1160 and 1255 cm-1 (symmetric and asymmetric ester C-O stretch bands) which are related to the presence of ethylene glycol dimethacrylate (EGDMA) in the MIP.

Figure 2
FTIR (KBr) spectrum of compound (a) radical MIP; and (b) via sol-gel process.

For MIP via sol-gel, the confirmation of the presence of silicon atoms bonded to the polymer chain can be attributed to the bands in the region of 970 cm-1, corresponding to the deformation of free silanol groups (Si-OH). It can also be observed the presence of bands in the region of 1094 cm-1, related to the connection stretching silicates groups (Si-O). The characteristic bands of siloxane bonds are in the region of 469 cm-1.

Comparison of extraction efficiency of MIPs obtained

The extraction efficiency (MISPE) of the imprinted polymers (10 mg) obtained by sol-gel process and radical polymerization were evaluated using blank plasma samples spiked with parabens at concentration of 500 ng mL-1. The average value of peak areas obtained from the MISPE/LC-UV analysis is illustrated in Figure 3a.

Figure 3
(a) Extraction efficiency (MISPE/LC-UV, n = 3) from polymers obtained by radical polymerization and sol-gel process and (b) extraction efficiency (MISPE/LC-UV, n = 3) from MIP and NIP obtained by radical polymerization.

The MIP obtained by radical polymerization presented lower standards deviation, and higher extraction efficiency for ethyl and propyl-parabens than the MIP obtained by sol-gel process (Figure 3a). Furthermore, the MIP synthesized via radical polymerization presented higher imprinting factors (Table 2). The polymeric non-imprinted material (radical NIP) presented non-specific interactions, and much lower extraction efficiency than radical MIP. These results confirm the selectivity (complementary shape and size) of radical MIP for parabens analysis. Due to the presence of hydroxyl groups on the polymer surface (not encapsulated groups), the MIP synthesized via sol-gel can result in greater number of non-specific interactions.

Table 2
Imprinting factor of radical MIP (MISPE/LC-UV analysis), and linearity of the MISPE/LC MS/MS method

Selectivity of MIP by radical polymerization

Figure 3b and Table 2 illustrate the selectivity (imprinting factor) of radical MIP that was evaluated using blank plasma samples spiked with parabens at concentration of 500 ng mL-1 by MISPE/LC-UV analysis.

Analytical validation

The selectivity of the method can be verified by representative MISPE/UHPLC-MS/MS (MRM mode) chromatograms of a blank plasma sample (Figure 4a) and of a blank plasma sample spiked with the target parabens at a concentration of 1 ng mL-1 (Figure 4b).

Figure 4
Representative MISPE/UHPLC-MS/MS (MRM) chromatograms of the (a) blank plasma spiked with parabens at LLOQ level (1.0 ng mL-1; and (b) blank plasma sample.

The chromatograms evidenced that the method was able to measure the analytes in the presence of endogenous components.

The MISPE/UHPLC-MS/MS method presented a linear range from the 1 ng mL-1 (LLOQ) up to 50 ng mL-1 with correlation coefficients greater than 0.993. The calibration standards (1.0, 3.0, 10, 25, 40, and 50 ng mL-1) presented CV lower than 15%. Table 2 also illustrates the linear regression equations and the corresponding correlation coefficient for all the target parabens.

Accuracy (expressed as relative standard error) of the method ranged from -11 to 14.5%. The inter-assay precision presented coefficient of variation ranging from 1.4 to 12.2% for all the analytes (Table 3). The MISPE procedure was performed manually, and this fact justified these results. Carry-over in the blank sample, after the analysis of ULOQ sample, was not greater than 20% of the analytes signals at LLOQ chromatogram, and not greater than 5% of the IS signal.

Table 3
LOD, LLOQ, inter-assay precision (coefficient of variation, CV), inter-assay accuracy (relative standard error, RSE), and matrix factor (MF) of the MISPE(MIP)/UHPLC-MS/MS method for the determination of parabens in plasma samples

The CV of the IS-normalized MF (matrix effects) was not greater than 15% (Table 3).

The developed MISPE/UHPLC-MS/MS method was compared with other recent reported methods for the determination of parabens in biological samples, Table 4. The main advantages of this method (MISPE/UHPLC-MS/MS) are the selectivity of the MISPE stationary phase, the reuse of this sorbent (over 50 times without significant extraction efficiency loss), and reduced volume of the biological sample and the organic solvents. However, due to the lower amount of the MIP phase used (10 mg), the developed method did not presented very low LLOQ.

Table 4
Comparison of present method with other methods for the determination of parabens in biological samples

Analyses of human umbilical plasma samples from postpartum women

The effectiveness of the MISPE-UHPLC-MS/MS method for parabens determination in plasma samples was evaluated by analyzing umbilical plasma samples from postpartum women (n = 10). According to the results from the MISPE/UHPLC-MS/MS analyses, parabens were not present at concentrations higher than the LLOQ value.

Conclusion

In this work, new MIPs phases were synthesized by radical polymerization and by sol-gel process for the analysis of the target parabens in plasma samples by MISPE/UHPLC-MS/MS.

The MIP obtained by radical polymerization presented lower standards deviation, and higher extraction efficiency for ethyl and propyl-parabens than the MIP obtained by sol-gel process. Furthermore, the MIP synthesized via radical polymerization presented higher imprinting factor. The polymeric non-imprinted material (radical NIP) presented non-specific interactions, and much lower extraction efficiency than radical MIP. These results confirm the selectivity (complementary shape and size) of radical MIP for parabens analysis.

The optimization of the MISPE process favored not only the analytical method sensitivity, but also reduced the number of non-specific interaction with the printed material. The miniaturized MISPE, compared with SPE conventional, reduced the amount of MIP phase (10 mg), minimized the organic solvent, and allowed the reusese of the extraction phase.

According to the parameters of the analytical validation, the miniaturized MISPE/UHPLC-MS/MS method is suitable for analysis of parabens in human plasma samples. The applicability of this method was demonstrated by the umbilical plasma sample analyzes from postpartum women that did not presented the target parabens in concentrations above the LLOQ values established.

  • FAPESP has sponsored the publication of this article.

Acknowledgments

The authors acknowledge financial support from FAPESP (No. 12/21178-0) and Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq).

References

  • 1
    Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.
  • 2
    Darbre, P. D.; Harvey, P. W.; J. Appl. Toxicol 2008, 28, 561.
  • 3
    Darbre, P. D.; Aljarrah, A.; Miller, W. R.; Coldham, N. G.; Sauer, M. J.; Pope, G. S.; J. Appl. Toxicol. 2004, 24, 5.
  • 4
    Frederiksen, H.; Jorgensen, N.; Andersson, A.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262.
  • 5
    Zhao, Y.; Liu, G.; Shen, H.; Shen, J.; Aubry, A. F.; Sivaraman; L.; Arnold M. E.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci 2014, 947, 68.
  • 6
    Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538.
  • 7
    Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.
  • 8
    Alshana, U.; Ertas, N.; Goger, N. G.; Food Chem. 2015, 15181, 1.
  • 9
    Gómez, R. R.; Pijuán, M. R.; Lucena, R.; Cárdenas, S.; Gómez, A. Z.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26.
  • 10
    Sajid, M.; Basheer, C.; Narasimhan, K.; Choolani, M.; Lee, H. K.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci 2015, 1000, 192.
  • 11
    Azzouz, A.; Rascón, A. J.; Ballesteros, E.; J. Pharm. Biomed. Anal. 2016, 119, 16.
  • 12
    Jardim, V. C.; Melo, L. P.; Domingues, D. S.; Queiroz, M. E. C.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci 2015, 974, 35.
  • 13
    Hines, E. P.; Mendola, P.; Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.
  • 14
    Navarro, M. V.; Carballo, M. C. M.; Torres, R. F.; Mochón, M. C.; López, M. A. B.; Anal. Bioanal. Chem 2016, 408, 1615.
  • 15
    Soria, F. V.; Ballesteros, O.; Rodríguez, I.; Gómez, A. Z.; Ballesteros, L.; Cela, R.; Navalón, A.; Anal. Bioanal. Chem. 2013, 405, 7259.
  • 16
    Mayes, A. G.; Whiticombe, M. J.; Adv. Drug Delivery Rev. 2005, 57, 1742.
  • 17
    Yan, H.; Ho Row, K.; Int. J. Mol. Sci. 2006, 7, 155.
  • 18
    Du, B.; Qu, T.; Chen, Z.; Cao, X.; Han, S.; Shen, G.; Wang, L.; Talanta 2014, 129, 465.
  • 19
    Cormack, P. A. G.; Harun, N; Anderson, R. A.; Anal. Bioanal. Chem. 2010, 39, 2449.
  • 20
    Clausen, D. N; Visentainer, J. V.; Tarley, C. R. T.; Analyst 2014, 139, 19, 5021.
  • 21
    Ferrer, I.; Lanza, F.; Tolokan, A.; Horvath, V.; Sellergren, B.; Horvai, G.; Barceló, D.; Anal. Chem. 2000, 72, 3934.
  • 22
    Pilau, E. J.; Silva, R. G.; Jardim, I. C. S. F.; Augusto, F.; J. Braz. Chem. Soc. 2008, 19, 1136.