Study of Inorganic Profiles of Street Cocaine Samples Using ICP-MS and ICP OES

Abstract

Determining inorganic constituents in illicit drugs can indicate its purity and presence of adulterants. In this study, we analyzed 52 street cocaine samples, seized in three different regions of Espírito Santo, state of Brazil. Inductively coupled plasma optical emission spectrometry was used to determine Al, Ca, Cu, Fe, Mn, Mg, Zn concentrations and inductively coupled plasma mass spectrometry to determine Mo, Co, Pb and P concentrations. From analyte recovery tests, the accuracy was considered acceptable and the proposed method satisfactory. Most of the samples exhibited a relatively homogeneous inorganic profile with similar concentrations of investigated elements, however some samples had very discrepant concentrations. High concentrations of Al, Ca, Fe, Mg, Mn, and P were found, indicating that adulterants such as gypsum, marble powder, limestone, cement, and others must be used to increase the profitability of the illicit drug market.

Keywords:
inorganic elements; cocaine; ICP-MS; ICP OES; multivariate analysis


Introduction

The abuse and use of illicit drugs are one of the biggest challenges facing society in the XXI century.11 Montangne, M. In The Epidemiology of Cocaine Use and Abuse; Schober, S.; Schade, C., eds.; National Institute on Drug Abuse: Rockville, Maryland, 1991. These drugs are generally used for pleasurable effects, since most of them affect the central nervous system, altering the state of consciousness, mood, and behavior of users. These changes can lead to either euphoria or depression; however, can also adversely affect the health of users, their family or community, as well as the social development, stability and safety, creating significant risks and problems.22 Wright, M. G. M.; Gliksman, L.; Khenti, A.; Furegato, A. R. F.; Rev. Latino-Am. Enfermagem 2009, 17, 751.

The United Nations Office on Drugs and Crimes (UNODC) reported that coca and cocaine-related substances were the drugs with the second largest quantities seized in 2017, over the world. The global quantity cocaine production has risen by 50 per cent over the past decade, reaching record levels in 2017, which was mainly due to the production increase in Colombia, which more than quadrupled over the period 2013-2017. It is also estimated that 18.1 million people were past-year users of the drug over the world, in the same year.33 World Drug Report; United Nations Publication, Sales No. E.19.XI.8, 2019, available at https://wdr.unodc.org/wdr2019/, accessed in July 2020.
https://wdr.unodc.org/wdr2019/...

For those reasons, the analysis of abuse drugs in forensic chemistry is a highly valuable research topic, because the reduction of drug traffics and the number of chemical dependents has become current challenges for society.44 Farias, R.; Introdução à Química Forense, 2nd ed.; Editora Átomo: São Paulo, 2008. Cocaine is a drug of plant origin and contains alkaloids, which have a considerable interest in forensic chemistry research.55 Romão, W.; Schwab, N. V.; Bueno, M. I. M. S.; Sparrapan, R.; Eberlin, M. N.; Martiny, A.; Sabino, D.; Maldaner, A. O.; Quim. Nova 2011, 34, 1717. Cocaine (benzoylmethylecgonine) is a tropane alkaloid (8-methyl-8-azabicyclo[3,2,1] octane) with a bicyclic structure, formed from the primary alkaloid present in Erythroxylum coca Lamarck, a native bush, found in some Andean countries such as Peru, Bolivia and Colombia.55 Romão, W.; Schwab, N. V.; Bueno, M. I. M. S.; Sparrapan, R.; Eberlin, M. N.; Martiny, A.; Sabino, D.; Maldaner, A. O.; Quim. Nova 2011, 34, 1717.

6 Conceicão, V. N.; Souza, L. M.; Merlo, B. B.; Filgueiras, P. R.; Poppi, R. J.; Romão, W.; Quim. Nova 2011, 37, 1538.
-77 Casale, J. F.; Klein, R. F. X.; Forensic Sci. Rev. 1993, 5, 95. Cocaine primarily acts stimulating the central nervous system, followed by depression. This explains why cocaine addicts increasingly crave the drug in an attempt to re-establish the initial euphoric state.22 Wright, M. G. M.; Gliksman, L.; Khenti, A.; Furegato, A. R. F.; Rev. Latino-Am. Enfermagem 2009, 17, 751.

Street cocaine is sold illicitly and contains cocaine hydrochloride in addition to many other substances, such as contaminants, adulterants, and diluents, and is rarely found in a 100% pure state.88 Bernardo, N. P.; Siqueira, M. E. P. B.; Paiva, M. N. J.; Maia, P. P.; Int. J. Drug Policy 2003, 14, 331.,99 Broséus, J.; Gentile, N.; Pont, F. B.; Gongora, J. M. G.; Esseiva, L. G. P.; Forensic Sci. Int. 2015, 257, 307. In Brazil, the purity of seized street cocaine ranges up to 85%,1010 Silva, M. J.; Anjos, E. V.; Honorato, R. S.; Pimmentel, M. F.; Paim, A. P. S.; Anal. Chim. Acta 2008, 629, 736. which is similar to France (up to 88%),1111 Dujourdy, L.; Besacier, F.; Forensic Sci. Int . 2008, 179, 111. while the seized cocaine blocks have a higher purity range (80 to 90%).1212 Fucci, N.; Forensic Sci. Int . 2007, 172, 35. Cocaine impurities may originate from coca leaves during the production process. Diluents and adulterants are added to cocaine, to increase the drug amounts and, thus, the sales profit. These substances are cheaper than pure cocaine and, besides diluting the drug, may act synergistically (adulterants) producing similar effects than cocaine.1313 Weiner, A. I.; Bayer, M. J.; Mckay Jr., C. A.; Demeo, M.; Starr, E.; Am. J. Emerg. Med. 1998, 16, 517.

14 Behrman, A. D.; J. Emerg. Nurs. 2008, 34, 80.
-1515 Coomber, R.; Addict. Res. Theory 2001, 5, 195. During the production process, organic solvents and inorganic compounds, such as calcium oxide (quicklime), sodium carbonate, hydrochloric acid, and sulfuric acid are used to extract cocaine from coca leaves. Extraction process produces a cocaine base paste, which is usually treated with hydrochloric acid and potassium permanganate, to produce cocaine hydrochloride, the active cocaine ingredient.1616 Silva Jr., R. C.; Gomes, C. S.; Goulart Jr., S. S.; Almeida, F. V.; Grobério, T. S.; Braga, J. W. B.; Zacca, J. J.; Vieira, M. L.; Botelho, E. D.; Maldaner, A. O.; Forensic Sci. Int . 2012, 221, 113. Inorganic diluents, such as gypsum plaster, limestone, marble dust, calcium sulfate, and phosphate and talcum powder, are used more frequently than organic diluents such as starch, sugar, wheat flour,1717 Carvalho, D. G.; Mídio, A. F.; Braz. J. Pharm. Sci. 2003, 39, 71. making essential to quantify certain elements and identify the presence of inorganic compounds.

Rarely is found 100% purity cocaine, because generally contaminants derived from refining and adulterants are present, sometimes in high concentration. The different components found in samples can contribute to determine its origin. The analytical information obtained through the analysis of a seized cocaine sample becomes very important for use in investigations by the police intelligence, also as an evidence for legal purposes.1515 Coomber, R.; Addict. Res. Theory 2001, 5, 195.

Besides, knowing the type of adulteration by quantifying the concentration of inorganic elements can also generate data to be used in public policies to prevent trafficking. It can help to elucidate the drug route inside the State, from its origin to the end user, through adulteration processes. It can also assist in the treatment of drug users, since several elements quantified in the study are harmful to health. In addition, knowing the presence/type of contaminants is important because diluting substances such as talc, plaster, lime and chalk, used to increase the volume of the drug, are not absorbed and can cause chronic obstructive pulmonary diseases, such as silicosis which causes the hardening of the lung.1717 Carvalho, D. G.; Mídio, A. F.; Braz. J. Pharm. Sci. 2003, 39, 71.

Many analytical techniques have been applied to determine the active ingredients, adulterants, diluents, and impurities in illicit drug samples. For elemental analysis electrothermal atomic absorption spectrometry (ET AAS), flame atomic absorption spectrometry (F AAS), inductively coupled plasma optical emission spectrometry (ICP OES) and inductively coupled plasma mass spectrometry (ICP-MS) can be employed; however, few studies about the application of these techniques to cocaine analysis have been reported.1818 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 77.

19 19. Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Bermejo-Barrera, A. M.; J. Forensic Sci. 1999, 44, 270.

20 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 1783.

21 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom. 1995, 10, 1011.

22 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Anal. Chim. Acta 1995, 310, 355.
-2323 Magalhães, E. J.; Nascentes, C. C.; Pereira, L. S. A.; Guedes, M. L. O.; Lordeiro, R. A.; Auler, L. M. L. A.; Augusti, R.; Queiroz, M. E. L. R.; Sci. Justice 2013, 53, 425.

Bermejo-Barrera et al.1818 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 77.

19 19. Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Bermejo-Barrera, A. M.; J. Forensic Sci. 1999, 44, 270.

20 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 1783.

21 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom. 1995, 10, 1011.
-2222 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Anal. Chim. Acta 1995, 310, 355. performed cocaine pasta base analysis by diluting the sample (20 times) with 3.5% v v-1 nitric acid, to determine Cr, Ag, Mn, Ni and Pb, in cocaine samples with, approximately, 90% of purity, by ET AAS, F AAS and FAES (flame atomic emission spectrometry). Magalhães et al.2323 Magalhães, E. J.; Nascentes, C. C.; Pereira, L. S. A.; Guedes, M. L. O.; Lordeiro, R. A.; Auler, L. M. L. A.; Augusti, R.; Queiroz, M. E. L. R.; Sci. Justice 2013, 53, 425. diluted the sample (100 times) with 10% v v-1 nitric acid to determine Ca, Mg, Na, P, Al, Fe, Mn and Zn in street cocaine samples, by ICP OES. However, the street cocaine samples were not wholly solubilized, and only the soluble fraction was analyzed. However, we emphasize that to investigate the substances used in street cocaine production or dilution, it is essential to decompose and solubilize the samples entirely before elements determination.

Much information can be obtained by determination of certain elements in street cocaine. Statistical tools can be applied to facilitate the interpretation of results.88 Bernardo, N. P.; Siqueira, M. E. P. B.; Paiva, M. N. J.; Maia, P. P.; Int. J. Drug Policy 2003, 14, 331. Statistic treatments of data have been successfully applied in forensic studies to obtain information about the origin and adulteration of illicit substances2020 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 1783.,2121 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom. 1995, 10, 1011. and other forensic evidence.2222 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Anal. Chim. Acta 1995, 310, 355. Multivariate analysis is widely used for chemical data treatment. In multivariate analysis, correlations between variables are used to extract a large amount of information, which is often impossible to achieve by analyzing each variable individually.2424 Bovens, M.; Ahrens, B.; Alberink, I.; Nordgaard, A.; Salonen, T.; Huhtal, S.; Forensic Sci. Int . 2019, 301, 82.

25 Nascimento, J. A.; Dionísio, A. G. G.; Nascimento, G. C. L.; Freitas, S. K. B.; Araújo, M. C. V.; Quim. Nova 2010, 33, 351.
-2626 Materazzi, S.; Gregori, A.; Ripani, L.; Apriceno, A.; Risoluti, R.; Talanta 2017, 166, 328.

Some statistic tools, widely used for data treatment, include principal component analysis (PCA), hierarchical cluster analysis (HCA), Pearson’s correlation coefficient (r) and boxplot.2727 Chudzinska, M.; Baralkiewiscz, D.; Food Chem. Toxicol. 2011, 49, 2741.

28 Ferreira, E. C.; Rodrigues, S. H. B. G.; Ferreira, M. M. C.; Nóbrega, J. A.; Nogueira, A. R. A.; Ecletica Quim. 2002, 27, 77.
-2929 Neto, M. M. J.; Moita, C. G.; Quim. Nova 1998, 21, 467. PCA is an unsupervised learning that attempts to identify patterns by reducing the dimensionality of the data for easier visualization. The new variables, called principal components, must explain the variance of data with minimal loss of information. HCA is a hierarchical process in which each step of the hierarchical matrix and the data matrix are decreased to one dimension, by progressively joining similar pairs until all of the points are joined into a single group. The purpose of HCA is to represent the data in a two-dimensional space to determine clustering and natural patterns. The results are presented as dendrograms, where samples are grouped according to their similarities.2929 Neto, M. M. J.; Moita, C. G.; Quim. Nova 1998, 21, 467.

Pearson’s correlation coefficient (r), or the correlation coefficient, is used to measure the degree of linear correlation between two quantitative variables. It is a dimensionless measure, ranging between +1.0 and –1.0.2929 Neto, M. M. J.; Moita, C. G.; Quim. Nova 1998, 21, 467. Boxplot graphically demonstrates the central position of the data (median) and the tendency, as well as any indication of symmetry or asymmetry of the data. Unlike many other statistical tools, boxplot shows outliers and can easily compare the samples.2929 Neto, M. M. J.; Moita, C. G.; Quim. Nova 1998, 21, 467.

This study aimed to develop a method to determine Al, Ca, Cu, Fe, Mn, Mg and Zn by ICP OES and Mo, Co, Pb and P by ICP-MS, in street cocaine samples. Some statistical tools were used to analyze the obtained data to verify correlations and patterns and find similar and discrepant characteristics between the samples.

Experimental

Instrumentation

The Al, Ca, Cu, Fe, Mn, Mg and Zn determination was performed using an ICP OES spectrometer, Optima 7000 Dual View model (PerkinElmer, Waltham, USA), and the Mo, Co, Pb and P determination was performed using an ICP-MS spectrometer, NexIon 300D model (PerkinElmer, Waltham, USA).

Cocaine samples were acid decomposed using a microwave oven equipped with closed PFA vessels (Multiwave 3000) and a rotor (16MF100/HF100) (Anton Paar, Graz, Austria). Cocaine samples were weighed using an analytical balance ED224S (Sartorius Weighing Technology, Goettingen, Germany) with an accuracy of ± 0.0001 g.

Reagents

All solutions were prepared using ultrapure water with 18.2 MΩ cm of resistivity (PURELAB Ultra Mk 2, ELGA, High Wycombe, UK). The 65% m m-1 HNO3, purchase from Synth (São Paulo, Brazil), was purified in an acid distiller (Sub-boiling Distillacid BSB939 IR, BERGHOF, Berchtesgaden, Germany) and used after adequate dilution.

Standard solutions of Ca, Cu, Fe, Mn, Mg, Zn, Mo, Co, and Pb were prepared in 2% v v-1 HNO3, by appropriate dilution of a multielement solution 1000 mg L-1 (PlasmaCAL QC Standard 3 140-102-051, SCP Science, Quebec, Canada). Al and P standard solutions were prepared in 2% v v-1 HNO3, by appropriate dilution of mono-element standard solutions 1000 mg L-1, purchased from SCP Science (Quebec, Canada). The concentration of the standard solutions for Al, Ca, Cu, Fe, Mn, Mg and Zn analytical curves were 0, 50, 100, 500 and 1000 µg L-1. The concentration of the standard solutions for Mo, Co, Pb and P analytical curves were 0, 0.5, 1, 2, 5, 10 and 20 µg L-1. Argon (99.9992%) was used as plasma, nebulizer, and auxiliary gases. Nitrogen (99.99%) was used as shear gas and nitrogen (99.9929%) as purging gas in ICP OES. All gases were purchased from Air Products (São Paulo, Brazil).

Samples

The Civil Police of the State of Espírito Santo-Brazil (in accordance to the cooperation agreement No. 1007/2012) provided fifty-two cocaine samples, seized in March 2014, in twenty cities, over three geographic regions of the state. Region 1, called metropolitan area, location in which the city Vitória, capital of Espírito Santo state, is located, where eighteen samples were seized; region 2, north of the state, where sixteen of the samples were seized; and region 3, the south of the state, where eighteen of the samples were seized.

Procedure

Sample preparation consisted of decomposition of 0.1 g of sample, with 6 mL of 5 % v v-1 HNO3, in closed-vessel in microwave system. The heating program consisted of two stages: at 130 °C for 10 min, followed by 180 °C for 10 min. After decomposition, samples were transferred to polypropylene flasks and diluted to 15.0 mL, using ultrapure water.

Al, Ca, Cu, Fe, Mn, Mg and Zn concentrations were determined by ICP OES, and Mo, Co, Pb and P by ICP-MS. The emission lines were selected according to Bermejo-Barrera et al.2121 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom. 1995, 10, 1011. or by considering those with higher sensitivity and absence of interferences. For ICP-MS determinations, the more abundant isotopes or suffering low interference, were selected.3030 Nicácio, J.; Lara, P. C. P.; Silva, J. C. J.; Ciminelli, V.; Silva, J. B. B.; At. Spectrosc. 2007, 28, 1.

31 Chan, K. W.; Tan, G. H.; Wong, R. C.; Sci. Justice 2013, 53, 73.

32 Chan, K. W.; Tan, G. H.; Wong, R. C.; J. Anal. At. Spectrom . 2011, 26, 1813.
-3333 Watters, R. L.; Eberhardt, K. R.; Beary, E. S.; Fassett, J. D.; Metrologia 1997, 34, 87. The ICP OES and ICP-MS operations conditions, accessories, spectral lines and isotopes determined are cited in Tables 1 and 2, respectively.

Table 1
ICP OES operational parameters
Table 2
ICP-MS operational parameters

Statistical analysis

Statistical analysis of data obtained was performed using the software MATLAB.3434 MATLAB, v.8.0 and Statistics Toolbox, v.8.1; The MathWorks, Inc., Natick, Massachusetts, United States, 2012. To obtain a separation of the samples by similarity, based on elements concentrations determined, HCA and PCA were performed, after data normalization. The separation between cocaine samples was calculated by the Euclidean distance, and the agglomeration by Ward’s method. Person’s correlation coefficient was used to measure the relationship, or association, between the elements in the samples. Lastly, a boxplot was built to show the distribution of the total inorganic mass in the samples, according to the seizure region.

Results and Discussion

Sample preparation

Initially, sample preparation procedures were performed according to previous studies.1616 Silva Jr., R. C.; Gomes, C. S.; Goulart Jr., S. S.; Almeida, F. V.; Grobério, T. S.; Braga, J. W. B.; Zacca, J. J.; Vieira, M. L.; Botelho, E. D.; Maldaner, A. O.; Forensic Sci. Int . 2012, 221, 113.

17 Carvalho, D. G.; Mídio, A. F.; Braz. J. Pharm. Sci. 2003, 39, 71.

18 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 77.

19 19. Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Bermejo-Barrera, A. M.; J. Forensic Sci. 1999, 44, 270.
-2020 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 1783. A total of 2 mL of 35% v v-1 HNO3 was added to 0.5 g of cocaine sample and the solution diluted to 10 mL. However, this procedure was unsatisfactory since the samples were not wholly solubilized, which is usually necessary for determining inorganic constituents. Sample solubilization with methanol was also investigated, which was also ineffective at complete solubilizing the samples.

In previous researches,1616 Silva Jr., R. C.; Gomes, C. S.; Goulart Jr., S. S.; Almeida, F. V.; Grobério, T. S.; Braga, J. W. B.; Zacca, J. J.; Vieira, M. L.; Botelho, E. D.; Maldaner, A. O.; Forensic Sci. Int . 2012, 221, 113.

17 Carvalho, D. G.; Mídio, A. F.; Braz. J. Pharm. Sci. 2003, 39, 71.

18 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 77.

19 19. Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Bermejo-Barrera, A. M.; J. Forensic Sci. 1999, 44, 270.
-2020 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 1783. it was observed that highly pure cocaine samples were soluble in water and ethanol. However, the drug samples analyzed in the present study were insoluble in the same media, indicating the presence of high quantities of insoluble adulterants or diluents. As a consequence, different sample preparation procedures were investigated using direct solubilization and acid decomposition assisted by microwave radiation. Different sample quantities (0.050, 0.100, 0.200 and 0.500 g) and 6 mL of HNO3 at different concentration (5, 10, 30, 40 and 50% v v-1) were tested.

Most samples treated with HNO3 solution, without submission to microwave decomposition (direct solubilization), were not wholly solubilized, regardless of the acid concentration and sample mass used. Therefore, decomposition of samples was tested using the same sample masses and HNO3 solution concentrations. Six milliliters of 5% v v-1 HNO3 were enough to completely decompose most of the samples, regardless of the mass used, so this acid concentration was chosen. The sample solution was then diluted to 15.0 mL with ultrapure water. Since part of the acid is consumed during decomposition process, the final acidity of the sample solutions is below 2% v v-1, which does not usually cause spectral interferences or instrumentation problem during the analysis. After evaluation and choose of the acid concentration adequate for decomposition, the solutions with different amount of sample mass were analyzed. Using 0.05 g of cocaine, the analytical signal was very low and close to the limit of quantification (LOQ) while using 0.1 g the signals were adequate for analytes quantifications.

Analytical figures of merit

Analytes recovery tests were performed to verify if the sample preparation procedures were adequate for Al, Ca, Cu, Fe, Mn, Mg and Zn determination by ICP OES and Mo, Co, Pb and P determinations by ICP-MS in cocaine samples. Thus, Al, Ca, Cu, Fe, Mn, Mg, and Zn were added to obtain 100 ng mL-1 of these elements in the spiked sample solution in four street cocaine samples (A, B, C, and D) for ICP OES analysis, while Mo, Co, Pb, and P were added to obtain 10 ng mL-1 of these elements in the spiked sample solution in three of them for ICP-MS analysis.

Recovery values (Table 3) approached 100% for most of the studied elements, except for one sample, where Ca exhibited a recovery (186%) much higher than the acceptable range (80-110%).3535 Brito, N. M.; Junior, O. P. A.; Polese, L.; Ribeiro, M. L.; Pesticidas: Rev. Ecotoxicol. Meio Ambiente 2003, 13, 129. ,3636 Official Methods of Analysis of AOAC International, 18th ed.; Horwitz, W.; Latimer Jr., G. W., eds.; AOAC: Rockville, 2011, appendix F. This may have been caused by the high Ca content in D sample that saturated signal.1616 Silva Jr., R. C.; Gomes, C. S.; Goulart Jr., S. S.; Almeida, F. V.; Grobério, T. S.; Braga, J. W. B.; Zacca, J. J.; Vieira, M. L.; Botelho, E. D.; Maldaner, A. O.; Forensic Sci. Int . 2012, 221, 113. Thus, the recovery values (Table 3) indicated that sample preparation procedures and optimized analysis conditions used in this study were adequate for determining the investigated elements in the street cocaine samples.

Table 3
Recovery values of Al, Ca, Cu, Fe, Mn, Mg, Zn, P, Co, Pb and Mo spiked in street cocaine samples

As expected, the determination coefficient (R2) of calibration curves were typically greater than 0.999. The limit of detection (LODm) and the limit of quantification (LOQm) of the analytical method were calculated according to Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO),3737 Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO); DOQ-CGCRE-008 Orientação sobre Validação de Métodos Analíticos; INMETRO: Rio de Janeiro, 2016, available at http://inmetro.gov.br/Sidoq/Arquivos/Cgcre/DOQ/DOQ-Cgcre-8_05.pdf, accessed in July 2020.
http://inmetro.gov.br/Sidoq/Arquivos/Cgc...
using equations 1 and 2, respectively, and considering the curve sensibility, the sample mass and the final volume used. The obtained LODm and LOQm, presented in Table 4, were inferior to values reported1818 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 77.,2121 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom. 1995, 10, 1011. and were adequate for the proposed determinations.

(1) LODm = X ¯ + 3 s
(2) LOQm = X ¯ + 10 s

where X is the average value of the blank samples and s is the sample standard deviation of the blank samples.

Table 4
Limits of detection (LODm) and quantification (LOQm) of the proposed method

Analytes determination in the analyzed samples

Using the proposed sample preparation procedure, 52 street cocaine samples were prepared in triplicate and analyzed. The average of the obtained results and the relative standard deviation (RSD) are presented in Table 5. The RSD is defined as the ratio between the standard deviation and the mean, expressing the extent of variability in relation to the mean. As can be seen in Table 5, most measurements had satisfactory repeatability (RSD < 5%).3737 Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO); DOQ-CGCRE-008 Orientação sobre Validação de Métodos Analíticos; INMETRO: Rio de Janeiro, 2016, available at http://inmetro.gov.br/Sidoq/Arquivos/Cgcre/DOQ/DOQ-Cgcre-8_05.pdf, accessed in July 2020.
http://inmetro.gov.br/Sidoq/Arquivos/Cgc...
Only a few elements showed high RSD for certain samples. Cu was the element that most presented results with RSD above 5%, but due to its low concentration a greater range of RSD is accepted.3737 Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO); DOQ-CGCRE-008 Orientação sobre Validação de Métodos Analíticos; INMETRO: Rio de Janeiro, 2016, available at http://inmetro.gov.br/Sidoq/Arquivos/Cgcre/DOQ/DOQ-Cgcre-8_05.pdf, accessed in July 2020.
http://inmetro.gov.br/Sidoq/Arquivos/Cgc...

Table 5
Average concentrations of Al, Ca, Cu, Fe, Mn, Mg, Zn, Co, Mo, Pb and P and relative standard deviation (RSD)

Using Pearson’s correlation coefficient, concentrations of the investigated elements were correlated (Table 6) to determine relations between them and suggest chemicals used in production or dilution of cocaine. As can be seen in Table 6, most of the elements are not correlated to each other, since r values were close to zero. Only the pairs Al-Cu, Fe-Cu, Mn-Cu, and Fe-Mn, showed high correlations, with r > 0.7. The pairs Al-Mn, Al-Fe, and Ca-P, showed moderate correlations, with r higher than 0.6.

Table 6
Pearson's correlations for Al, Ca, Cu, Fe, Mn, Mg, Zn, Mo, Co, Pb and P concentrations in cocaine samples

Ca, Mg, P, Al and Fe showed higher concentrations, comparing to the other elements (Table 5), indicating that they may be part of the chemical composition of the adulterants added to the drug. Chan et al.3838 Chan, K. W.; Tan, G. H.; Wong, R. C. S.; Anal. Lett. 2012, 45, 1122. analyzed inorganic compounds in heroin samples and also found Ca, Mg and Fe present at relatively higher levels, besides Na and K elements.

Ca concentrations ranged from around 20 µg g-1 to higher than 50000 µg g-1 (saturated signal), presenting the most significant concentrations among all elements. Six samples from region 1 (R1-d, R1-f, R1-g, R1-j, R1-p and R1-q), three from region 2 (R2-a, R2-i and R2-n) and six from region 3 (R3-a, R3-b, R3-c, R3-g, R3-i and R3-q) exhibited high concentrations of Ca, over 1000 µg g-1, which could be related to the use of calcium oxide (quicklime) in cocaine refining or to the use of some diluents as gypsum, limestone and marble powder.

Mg concentrations ranged from values below the detection limit to 45870 µg g-1. Two samples from region 1 (R1-f and R1-g), one from region 2 (R2-n) and three from regions 3 (R3-b, R3-h and R3-j) exhibited high concentrations of Mg, over 1000 µg g-1, which could be related to addition of talcum powder (Mg3Si4O10(OH)2) as cocaine diluent. The high Mg concentrations in regions R1-f and R1-g, R2-n and R3-b can also be related to the addition of limestone as a diluent, since those samples also presented high Ca concentrations.

P concentrations ranged from values below the limit of detection to 105000 µg g-1. Only eight samples (R1-g and R1-p, R2-a and R2-n, R3-a, R3-b, R3-c and R3-i) exhibited P concentration higher than 1000 µg g-1. All those samples also had high Ca concentration, indicating the use of calcium phosphate as diluent. This behavior may explain the significant Person correlation observed between Ca and P concentrations (r = 0.6).

Concentrations of Al ranged from 0.45 to 3934 µg g-1 and nine samples had Al concentration higher than 200 µg g-1. The high amount of Al in those samples could be related to the use of white clay as a drug diluent. White clay is a natural material, composed by hydrated silicates of aluminum and iron and can contain specific contaminants like Mn, Cu, and Pb. This may explain the high correlation between Al and Cu (r = 0.817), Al and Fe (r = 0.618), Al and Mn (r = 0.658), Cu and Fe (0.70), Cu and Mn (0.851) and Fe and Mn (0.740), as can be seen in Table 6.

Mn, Cu, and Zn were present at intermediate concentrations in comparison to the other elements, ranging from below LODm to 30 µg g-1. All samples had Mn concentration below 20 µg g-1, except for R1-f, R3-j, and R3-q. The higher Mn concentration in those three samples suggest the use of potassium permanganate during the production process to obtain cocaine in its base form. It could also be explained by the use of talcum powder as diluent since, in R1-f and R3-j samples, the Mg concentration was also very high; however, no significant correlation was observed for Mg and Mn (r = 0.362) in this study.

Co and Pb were found in very low concentrations, reaching a maximum of 17.4 and 3.7 μg g-1, respectively. These low concentrations indicate the presence of Co and Pb in reagents and adulterants used in the drug production process. Mo also exhibited low concentrations, reaching a maximum of 12.1 μg g-1. This has occurred because Mo is not presented at significant concentration in drug adulterants usually. Instead, Mo is a plant micronutrient so that it may be related to the purity of the drug. In other words, the higher the Mo concentration, the less dilution the drug may have suffered.2121 Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom. 1995, 10, 1011.

Magalhães et al.2323 Magalhães, E. J.; Nascentes, C. C.; Pereira, L. S. A.; Guedes, M. L. O.; Lordeiro, R. A.; Auler, L. M. L. A.; Augusti, R.; Queiroz, M. E. L. R.; Sci. Justice 2013, 53, 425. quantified inorganic elements in street cocaine samples seized in Minas Gerais and Amazonas, states of Brazil. As well as our results, Ca, P and Mg maximum concentrations were higher than the other elements investigated, as can be seen on Table 7. Except for Al, P, Co and Cu, the maximum concentration of the inorganic elements was lower than the maximum concentration found in this study. For example, the Ca, Mg, Pb, and Zn maximum concentrations in such samples were, respectively, 3.6, 2.3, 5.1, and 39.4 times higher than those found in this study.

Table 7
Comparison of results

Liu et al.3939 Liu, C.; Hua, Z.; Meng, X.; Forensic Sci. Int . 2017, 276, 77. also determined inorganic elements in cocaine samples seized in China. As well as our results, Ca and Mg concentrations were higher than those of the other elements investigated. For Al, Mn, Mo and Co, the authors found, respectively, a maximum concentration of almost 7, 9, 61 and 25 times lower than of this study and for Pb and Zn, almost 55 and 22 times higher, respectively.

Frequency histograms of the number of samples that presented concentration of the studied elements in different ranges of values are shown in Figure 1. The histogram allows a better understanding about the concentration profiles in cocaine samples. It can be seen that, for P element, for example, most of the samples presented similar and low concentration of this element and just a few of them presented higher concentrations. In general, most of the samples exhibited concentrations of elements investigated in a low value range, however some samples showed very discrepant concentrations. This pattern can be observed for all analyzed elements, except for Pb, which concentration profile was slightly more heterogeneous, since many samples showed concentrations in different ranges.

Figure 1
Histogram of elements concentrations in street cocaine samples.

This pattern was also observed by Magalhães et al.2323 Magalhães, E. J.; Nascentes, C. C.; Pereira, L. S. A.; Guedes, M. L. O.; Lordeiro, R. A.; Auler, L. M. L. A.; Augusti, R.; Queiroz, M. E. L. R.; Sci. Justice 2013, 53, 425. and Liu et al.3939 Liu, C.; Hua, Z.; Meng, X.; Forensic Sci. Int . 2017, 276, 77. who reported that some cocaine samples had discrepant concentrations of investigated elements. From 180 analyzed samples, according to Liu et al.,3939 Liu, C.; Hua, Z.; Meng, X.; Forensic Sci. Int . 2017, 276, 77. 45 of them were considered as outliers, because of the discrepant element concentrations found.

Chemometrics analysis

Multivariate analysis, HCA and PCA, of the concentration of investigated elements was carried out in order to classify the cocaine samples according to their similarities and differences. Since variables (elements concentrations) has different orders of magnitude, the normalization procedure, before multivariate analysis, was essential to prevent one variable from overlapping another.

For PCA results, the first three principal components explained only 61.2% of the total data variance. This low capture of the variance occurred because most of the variables were not correlated, as can be seen in Table 6. In this case, variables reduction in principal components can be inefficient. From the loadings of features in PC1, PC2 and PC3, the more correlated variables Al, Cu, Fe, and Mn were the dominating features in the first principal component; Ca and P dominate in the second and Co, Mg and Pb dominate in the third principal component.

The 3D plot of PC1, PC2, and PC3 is shown in Figure 2. The first three components cannot discriminate the cocaine samples in different regions, through the concentration of elements. The plot shows that almost all samples are located in the center of the box, and some few samples, especially from region 3, are located far from the center. Samples close to the center are those in which the concentration of the elements is close to the average. On the other hand, samples far from center, showed greater variation in concentrations, differing from the others. Chan et al.3131 Chan, K. W.; Tan, G. H.; Wong, R. C.; Sci. Justice 2013, 53, 73. also carried out PCA in 309 heroin samples, using 16 inorganic elements concentration, in order to classify them according to the distribution lines. The score plot presented the samples clustered into two dense areas, indicating that samples were likely from at least two major distribution lines.

Figure 2
Scores plots of first three principal components (PC1, PC2, and PC3) for inorganic elements concentrations in cocaine samples.

In HCA analysis, the dendrogram obtained (Figure 3a) reveled the similarity or difference between the samples, as well as the distance levels between the clusters. Four groups were formed at a linkage distance of 8 (dashed line). This groups contained 36, 3, 10 and 1 samples, respectively. As can be seen in the dendrogram, the largest group contains samples from the three seizure regions, showing no significant differences in the drugs from these regions.

Figure 3
Dendrogram of samples (a) and variables (b) obtained by the Euclidean distance and the Ward’s method.

According to PCA and HCA results, the samples presented a very similar inorganic profile, so it was not possible to separate them, with respect to the three seizure regions. It became clear that no patterns of adulteration were detected in the drug samples. For this reason, it was not possible to classify them in different groups in this illicit market. Only similarities were found in the impurity content, which was not enough to classify the samples. The small extension of the state of Espírito Santo can be a reason for the similarity between samples. It can also be considered that, although we have worked with a relatively high number of samples for this type of work (52 street cocaine samples seized), perhaps a larger number of samples could lead to the observation of some trend.

Different from the above, Myors et al.4040 Myors, R.; Wells, R. J.; Skopec, S. V.; Crisp, P.; Iavetz, R.; Skopec, Z.; Ekangakic, A.; Robertson, J.; Anal. Commun. 1998, 35, 403. built HCA models capable of indicate the heroin sample origins by only considering the inorganic elements distribution, analyzed by coupled plasma mass spectrometry. The dendogram placed together the samples from one specific region and separated from the samples from other parts of the world. However, the authors used data of 73 inorganic compounds and a total of 188 heroin samples, numbers much higher than we used in this research. Liu et al.4141 Liu, C.; Hua, Z.; Bai, Y.; Liu, Y.; Forensic Sci. Int . 2014, 239, 37. also could discriminate 150 heroin samples according to two geography origins, using measurements of 19 inorganic trace elements and a two-dimensional score scatter plot based on PCA analysis.

For future researches, we recommend to develop a methodology to purify the cocaine samples and analyze impurities and purified cocaine, separately. Maybe that could show by chemometrics analysis the traces of the different cultivars, indicating their origin, as well as those coming from one or more drug producing and/or supplying regions. Besides, this data could also be combined with measurements of organic compounds to provide a more powerful discriminating technique.

Cluster analysis was also performed on the 11 measured elements. According to the dendogram shown in Figure 3b, Al, Cu, Fe and Mn, which were highly correlated (Table 6), were grouped in a single cluster while, P and Ca that were significantly correlated were in another group. The other elements, which did not have significant correlation between them, became separated. The elements concentrations (Table 5) were summed to calculate the percentage (m m-1) of them in each cocaine sample. Figure 4 shows a boxplot of the total concentration in samples, separated by the seizure regions.

Figure 4
Boxplot of total concentration (m m-1) of investigated elements in the street cocaine samples, by seizure regions.

The total concentration in samples from region 3 reveals more dispersion, followed by region 1. Besides, the data are asymmetric, due to the distortion caused by the extremely high concentrations of some elements in some samples. These results evidenced that, in general, samples from region 3 suffered more dilution, whereas samples from region 2 did lower dilution.

Conclusions

Fifty-two samples of street cocaine, seized in different cities of State of Espírito Santo, were analyzed, and concentration results were used to determine inorganic profiles through chemometrics treatments. The method used to quantify Al, Ca, Cu, Fe, Mn, Mg, and Zn by ICP OES and Mo, Co, Pb and P by ICP-MS, after microwave assisted acid decomposition, proved to be appropriated, since the results from recovery tests were within an acceptable range. Multivariate analysis allowed to correlate the inorganic analytes and propose some substances that may have been used during cocaine production or its dilution.

The profile of the samples with respect to the investigated elements was very similar, according to HCA and PCA and they could not be discriminated by region. This result is not surprising since Espírito Santo is a small extension state. Some samples exhibited high and discrepant concentrations of some studied elements indicating a high level of dilution. Most of those samples were from region 3.

Acknowledgments

The authors are grateful to Civil Police of Espírito Santo State, FAPES (37372.591.18380.05102018, Edital CNPq/FAPES No 23/2018-Programa de Apoio a Núcleos Emergentes (PRONEM) Outorga 597/2018), and CNPq (305359/2017-7 and 422555/2018-5) for financial support. Also, the authors would like thanks LabPetro/UFES-Research and Methodology Development Laboratory for Crude Oil Analysis for supporting this study. This study was financed in part by the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior-Brazil (CAPES)-finance code 001.

References

  • 1
    Montangne, M. In The Epidemiology of Cocaine Use and Abuse; Schober, S.; Schade, C., eds.; National Institute on Drug Abuse: Rockville, Maryland, 1991.
  • 2
    Wright, M. G. M.; Gliksman, L.; Khenti, A.; Furegato, A. R. F.; Rev. Latino-Am. Enfermagem 2009, 17, 751.
  • 3
    World Drug Report; United Nations Publication, Sales No. E.19.XI.8, 2019, available at https://wdr.unodc.org/wdr2019/, accessed in July 2020.
    » https://wdr.unodc.org/wdr2019/
  • 4
    Farias, R.; Introdução à Química Forense, 2nd ed.; Editora Átomo: São Paulo, 2008.
  • 5
    Romão, W.; Schwab, N. V.; Bueno, M. I. M. S.; Sparrapan, R.; Eberlin, M. N.; Martiny, A.; Sabino, D.; Maldaner, A. O.; Quim. Nova 2011, 34, 1717.
  • 6
    Conceicão, V. N.; Souza, L. M.; Merlo, B. B.; Filgueiras, P. R.; Poppi, R. J.; Romão, W.; Quim. Nova 2011, 37, 1538.
  • 7
    Casale, J. F.; Klein, R. F. X.; Forensic Sci. Rev 1993, 5, 95.
  • 8
    Bernardo, N. P.; Siqueira, M. E. P. B.; Paiva, M. N. J.; Maia, P. P.; Int. J. Drug Policy 2003, 14, 331.
  • 9
    Broséus, J.; Gentile, N.; Pont, F. B.; Gongora, J. M. G.; Esseiva, L. G. P.; Forensic Sci. Int 2015, 257, 307.
  • 10
    Silva, M. J.; Anjos, E. V.; Honorato, R. S.; Pimmentel, M. F.; Paim, A. P. S.; Anal. Chim. Acta 2008, 629, 736.
  • 11
    Dujourdy, L.; Besacier, F.; Forensic Sci. Int . 2008, 179, 111.
  • 12
    Fucci, N.; Forensic Sci. Int . 2007, 172, 35.
  • 13
    Weiner, A. I.; Bayer, M. J.; Mckay Jr., C. A.; Demeo, M.; Starr, E.; Am. J. Emerg. Med 1998, 16, 517.
  • 14
    Behrman, A. D.; J. Emerg. Nurs 2008, 34, 80.
  • 15
    Coomber, R.; Addict. Res. Theory 2001, 5, 195.
  • 16
    Silva Jr., R. C.; Gomes, C. S.; Goulart Jr., S. S.; Almeida, F. V.; Grobério, T. S.; Braga, J. W. B.; Zacca, J. J.; Vieira, M. L.; Botelho, E. D.; Maldaner, A. O.; Forensic Sci. Int . 2012, 221, 113.
  • 17
    Carvalho, D. G.; Mídio, A. F.; Braz. J. Pharm. Sci 2003, 39, 71.
  • 18
    Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 77.
  • 19
    19. Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Bermejo-Barrera, A. M.; J. Forensic Sci 1999, 44, 270.
  • 20
    Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Talanta 1996, 43, 1783.
  • 21
    Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; J. Anal. At. Spectrom 1995, 10, 1011.
  • 22
    Bermejo-Barrera, P.; Moreda-Piñeiro, A.; Moreda-Piñeiro, J.; Bermejo-Barrera, A.; Anal. Chim. Acta 1995, 310, 355.
  • 23
    Magalhães, E. J.; Nascentes, C. C.; Pereira, L. S. A.; Guedes, M. L. O.; Lordeiro, R. A.; Auler, L. M. L. A.; Augusti, R.; Queiroz, M. E. L. R.; Sci. Justice 2013, 53, 425.
  • 24
    Bovens, M.; Ahrens, B.; Alberink, I.; Nordgaard, A.; Salonen, T.; Huhtal, S.; Forensic Sci. Int . 2019, 301, 82.
  • 25
    Nascimento, J. A.; Dionísio, A. G. G.; Nascimento, G. C. L.; Freitas, S. K. B.; Araújo, M. C. V.; Quim. Nova 2010, 33, 351.
  • 26
    Materazzi, S.; Gregori, A.; Ripani, L.; Apriceno, A.; Risoluti, R.; Talanta 2017, 166, 328.
  • 27
    Chudzinska, M.; Baralkiewiscz, D.; Food Chem. Toxicol 2011, 49, 2741.
  • 28
    Ferreira, E. C.; Rodrigues, S. H. B. G.; Ferreira, M. M. C.; Nóbrega, J. A.; Nogueira, A. R. A.; Ecletica Quim 2002, 27, 77.
  • 29
    Neto, M. M. J.; Moita, C. G.; Quim. Nova 1998, 21, 467.
  • 30
    Nicácio, J.; Lara, P. C. P.; Silva, J. C. J.; Ciminelli, V.; Silva, J. B. B.; At. Spectrosc 2007, 28, 1.
  • 31
    Chan, K. W.; Tan, G. H.; Wong, R. C.; Sci. Justice 2013, 53, 73.
  • 32
    Chan, K. W.; Tan, G. H.; Wong, R. C.; J. Anal. At. Spectrom . 2011, 26, 1813.
  • 33
    Watters, R. L.; Eberhardt, K. R.; Beary, E. S.; Fassett, J. D.; Metrologia 1997, 34, 87.
  • 34
    MATLAB, v.8.0 and Statistics Toolbox, v.8.1; The MathWorks, Inc., Natick, Massachusetts, United States, 2012.
  • 35
    Brito, N. M.; Junior, O. P. A.; Polese, L.; Ribeiro, M. L.; Pesticidas: Rev. Ecotoxicol. Meio Ambiente 2003, 13, 129.
  • 36
    Official Methods of Analysis of AOAC International, 18th ed.; Horwitz, W.; Latimer Jr., G. W., eds.; AOAC: Rockville, 2011, appendix F.
  • 37
    Instituto Nacional de Metrologia, Qualidade e Tecnologia (INMETRO); DOQ-CGCRE-008 Orientação sobre Validação de Métodos Analíticos; INMETRO: Rio de Janeiro, 2016, available at http://inmetro.gov.br/Sidoq/Arquivos/Cgcre/DOQ/DOQ-Cgcre-8_05.pdf, accessed in July 2020.
    » http://inmetro.gov.br/Sidoq/Arquivos/Cgcre/DOQ/DOQ-Cgcre-8_05.pdf
  • 38
    Chan, K. W.; Tan, G. H.; Wong, R. C. S.; Anal. Lett 2012, 45, 1122.
  • 39
    Liu, C.; Hua, Z.; Meng, X.; Forensic Sci. Int . 2017, 276, 77.
  • 40
    Myors, R.; Wells, R. J.; Skopec, S. V.; Crisp, P.; Iavetz, R.; Skopec, Z.; Ekangakic, A.; Robertson, J.; Anal. Commun 1998, 35, 403.
  • 41
    Liu, C.; Hua, Z.; Bai, Y.; Liu, Y.; Forensic Sci. Int . 2014, 239, 37.

Publication Dates

  • Publication in this collection
    20 Jan 2021
  • Date of issue
    Jan 2021

History

  • Received
    18 Apr 2020
  • Accepted
    30 July 2020
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br