The effects of tether pretension within vertebral body tethering on the biomechanics of the spine: a Finite Element analysis

Abstract

This study investigates the biomechanics of the spine after insertion of vertebral body tethering (VBT) with different cord pretensions. For that purpose, a Finite Element model of the average thoracolumbar spine was stepwise calibrated and validated. The VBT instrumentation was inserted in the left side of the L1-L2 segment with different cord pretensions. As a second test, the L1-L2 segment was submitted to an external pure moment of 6 Nm in left and right lateral bending. The range of motion (ROM) for the spine with VBT was determined with respect to its initial post VBT position. Pretension forces of 100 N and 300 N resulted in a change of scoliotic angle of 2.7° and 5.3° to the left side of the spine, respectively. The ROM of the native spine was 4.5° in right lateral bending and reduced to 1.8° and 1.4° for the cases of the spine with a cord pretension of 100 N and 300 N, respectively. In left lateral bending, the absolute ROM of the native spine was 4.6°. For the cases of a cord pretension of 100 N and 300 N, the spine bent 1.9° and 0.8° to the left side from its initial post VBT position, respectively.

Keywords
Spine; Scoliosis; Vertebral Body Tethering; Growth modulation; Fusionless; Curve correction; Finite Element Method

Graphical Abstract

1 INTRODUCTION

Adolescent idiopathic scoliosis (AIS) is a complex 3D deformity of the spine that worsens during skeletal growth. The spine develops a side-to-side curvature, usually in an elongated “S” or “C” shape in the frontal plane instead of growing straight (Šarčević and Tepavčević, 2019Šarčević, Z., Tepavčević, A., 2019. Association between adolescent idiopathic scoliosis and sacroiliac joint dysfunction in young athletes. Medicine (Baltimore). 98, e15161. https://doi.org/10.1097/MD.0000000000015161
https://doi.org/10.1097/MD.0000000000015...
). Anterior vertebral body tethering (VBT) is a fusionless treatment option that is considered a new promising method for the management of selective cases of AIS (Pehlivanoglu et al., 2020Pehlivanoglu, T., Oltulu, I., Ofluoglu, E., Sarioglu, E., Altun, G., Korkmaz, M., Yildirim, K., Aydogan, M., 2020. Thoracoscopic Vertebral Body Tethering for Adolescent Idiopathic Scoliosis: A Minimum of 2 Years’ Results of 21 Patients. J. Pediatr. Orthop. 40, 575–580. https://doi.org/10.1097/BPO.0000000000001590
https://doi.org/10.1097/BPO.000000000000...
). It is a growth modulating technique that derives benefit from the Heuter-Volkman principle which states that increased mechanical compression load in comparison with normal values slows growth and distraction accelerates it (Stokes et al., 1996Stokes, I.A.F., Spence, H., Aronsson, D.D., Kilmer, N., 1996. Mechanical Modulation of Vertebral Body Growth. Spine (Phila. Pa. 1976). 21, 1162–1167. https://doi.org/10.1097/00007632-199605150-00007
https://doi.org/10.1097/00007632-1996051...
). During surgery, this effect is achieved by tensioning the polyethylene-terephthalate tether (cord) that is connected to screws inserted in the lateral convex side of the vertebral bodies. One of the main advantages of VBT over spinal fusion is the ability to preserve the mobility of the spine while maintaining growth potential.

Multiple clinical reports have shown that VBT is safe and effective in correcting deformity of the spine (Crawford and Lenke, 2010Crawford, C.H., Lenke, L.G., 2010. Growth Modulation by Means of Anterior Tethering Resulting in Progressive Correction of Juvenile Idiopathic Scoliosis. J. Bone Jt. Surgery-American Vol. 92, 202–209. https://doi.org/10.2106/JBJS.H.01728
https://doi.org/10.2106/JBJS.H.01728...
; Miyanji et al., 2020Miyanji, F., Pawelek, J., Nasto, L.A., Simmonds, A., Parent, S., 2020. Safety and efficacy of anterior vertebral body tethering in the treatment of idiopathic scoliosis. Bone Jt. J. 102-B, 1703–1708. https://doi.org/10.1302/0301-620X.102B12.BJJ-2020-0426.R1
https://doi.org/10.1302/0301-620X.102B12...
; Samdani et al., 2015Samdani, A.F., Ames, R.J., Kimball, J.S., Pahys, J.M., Grewal, H., Pelletier, G.J., Betz, R.R., 2015. Anterior vertebral body tethering for immature adolescent idiopathic scoliosis: one-year results on the first 32 patients. Eur. Spine J. 24, 1533–1539. https://doi.org/10.1007/s00586-014-3706-z
https://doi.org/10.1007/s00586-014-3706-...
). Recently, Miyanji et al. (2020)Miyanji, F., Pawelek, J., Nasto, L.A., Simmonds, A., Parent, S., 2020. Safety and efficacy of anterior vertebral body tethering in the treatment of idiopathic scoliosis. Bone Jt. J. 102-B, 1703–1708. https://doi.org/10.1302/0301-620X.102B12.BJJ-2020-0426.R1
https://doi.org/10.1302/0301-620X.102B12...
reported the cases of 57 consecutive patients who underwent VBT mostly at the thoracic region (96.5%). The mean preoperative major curve of 51° was significantly improved to a mean of 24.6° at the first postoperative visit, and 16.3° at one year (Miyanji et al., 2020Miyanji, F., Pawelek, J., Nasto, L.A., Simmonds, A., Parent, S., 2020. Safety and efficacy of anterior vertebral body tethering in the treatment of idiopathic scoliosis. Bone Jt. J. 102-B, 1703–1708. https://doi.org/10.1302/0301-620X.102B12.BJJ-2020-0426.R1
https://doi.org/10.1302/0301-620X.102B12...
). Biomechanical studies on human spines confirmed that VBT can preserve motion of the spine in flexion-extension and axial rotation while reducing motion in lateral bending (Lavelle et al., 2016Lavelle, W.F., Moldavsky, M., Cai, Y., Ordway, N.R., Bucklen, B.S., 2016. An initial biomechanical investigation of fusionless anterior tether constructs for controlled scoliosis correction. Spine J. 16, 408–413. https://doi.org/10.1016/j.spinee.2015.11.004
https://doi.org/10.1016/j.spinee.2015.11...
; Nicolini et al., 2021bNicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
).

Despite the promising results, overcorrection and under-correction after VBT are common causes of revision surgery (Buyuk et al., 2021Buyuk, A.F., Milbrandt, T.A., Mathew, S.E., Potter, D.D., Larson, A.N., 2021. Does preoperative and intraoperative imaging for anterior vertebral body tethering predict postoperative correction? Spine Deform. 9, 743–750. https://doi.org/10.1007/s43390-020-00267-2
https://doi.org/10.1007/s43390-020-00267...
). The intraoperative tensioning of the cord is one key component to achieve appropriate curve correction but its effects are poorly investigated. Lavelle et al. (2016)Lavelle, W.F., Moldavsky, M., Cai, Y., Ordway, N.R., Bucklen, B.S., 2016. An initial biomechanical investigation of fusionless anterior tether constructs for controlled scoliosis correction. Spine J. 16, 408–413. https://doi.org/10.1016/j.spinee.2015.11.004
https://doi.org/10.1016/j.spinee.2015.11...
evaluated the effects of pretension of the tether at the most superior and most inferior vertebrae as opposed to all vertebrae. The thoracic spines were tested by applying only a pretension of 100 N on the cord as also performed in the biomechanical study of Nicolini et al. (2021b)Nicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
with thoracolumbar spines. However, the effects of further tensioning were not investigated although a tensioning of 200 N and 300 N are common in clinical scenarios. Therefore, it remains unknown how much tension to place on the scoliotic spinal segments within the VBT construct although this is a key engineering parameter to control correction of the deformity.

Numerical data that could contribute to the clinical and biomechanical studies as well as to the improvement of the VBT technique remains sparse. Finite Elements (FE) models of the spine have made important contributions in the field of biomechanics but accurate and clinically relevant modeling of the human spine remains challenging (Dreischarf et al., 2014Dreischarf, M., Zander, T., Shirazi-Adl, A., Puttlitz, C.M., Adam, C.J., Chen, C.S., Goel, V.K., Kiapour, A., Kim, Y.H., Labus, K.M., Little, J.P., Park, W.M., Wang, Y.H., Wilke, H.J., Rohlmann, A., Schmidt, H., 2014. Comparison of eight published static finite element models of the intact lumbar spine: Predictive power of models improves when combined together. J. Biomech. 47, 1757–1766. https://doi.org/10.1016/j.jbiomech.2014.04.002
https://doi.org/10.1016/j.jbiomech.2014....
). Few FE models (Ezquerro et al., 2011Ezquerro, F., Vacas, F.G., Postigo, S., Prado, M., Simón, A., 2011. Calibration of the finite element model of a lumbar functional spinal unit using an optimization technique based on differential evolution. Med. Eng. Phys. 33, 89–95. https://doi.org/10.1016/j.medengphy.2010.09.010
https://doi.org/10.1016/j.medengphy.2010...
; Heuer et al., 2007bHeuer, F., Schmidt, H., Klezl, Z., Claes, L., Wilke, H.-J., 2007b. Stepwise reduction of functional spinal structures increase range of motion and change lordosis angle. J. Biomech. 40, 271–280. https://doi.org/10.1016/j.jbiomech.2006.01.007
https://doi.org/10.1016/j.jbiomech.2006....
; Schmidt et al., 2007Schmidt, H., Heuer, F., Drumm, J., Klezl, Z., Claes, L., Wilke, H.-J., 2007. Application of a calibration method provides more realistic results for a finite element model of a lumbar spinal segment. Clin. Biomech. 22, 377–84. https://doi.org/10.1016/j.clinbiomech.2006.11.008
https://doi.org/10.1016/j.clinbiomech.20...
) are calibrated considering experimental data from stepwise resection studies (Li et al., 2017Li, Y., Shen, Z., Huang, M., Wang, X., 2017. Stepwise resection of the posterior ligamentous complex for stability of a thoracolumbar compression fracture. Medicine (Baltimore). 96, e7873. https://doi.org/10.1097/MD.0000000000007873
https://doi.org/10.1097/MD.0000000000007...
; Zhang et al., 2011Zhang, H., Muzumdar, A., Bucklen, B., Hussain, M., Khalil, S., 2011. Effects of stepwise transection of spinal vomponents in lumbar spine: an experimental study of one L1-S1 cadaver sample, in: Poster Session Presented at: 57th Annual Meeting of the Orthopaedic Research Society. Long Beach (CA), p. Jan 13 – 16.). The experimental procedure is conducted by testing the spine in different loading directions and gradually resecting the passive spinal structures to obtain their individual contribution to the biomechanics of the spine. Posteriorly, each spinal structure is sequentially added to the FE models and calibrated by varying and choosing the material properties that best fit the experimental results (Ezquerro et al., 2011Ezquerro, F., Vacas, F.G., Postigo, S., Prado, M., Simón, A., 2011. Calibration of the finite element model of a lumbar functional spinal unit using an optimization technique based on differential evolution. Med. Eng. Phys. 33, 89–95. https://doi.org/10.1016/j.medengphy.2010.09.010
https://doi.org/10.1016/j.medengphy.2010...
; Nicolini et al., 2021aNicolini, L.F., Beckmann, A., Laubach, M., Hildebrand, F., Kobbe, P., Roesler, C.R. de M., Fancello, E.A., Markert, B., Stoffel, M., 2021a. An experimental-numerical method for the calibration of Finite Element models of the spine. Submitt. to Med. Eng. Phys.; Schmidt et al., 2006Schmidt, H., Heuer, F., Simon, U., Kettler, A., Rohlmann, A., Claes, L., Wilk, H.-J., 2006. Application of a new calibration method for a three-dimensional finite element model of a human lumbar annulus fibrosus. Clin. Biomech. 21, 337–344.). This method allows for more precise modeling than obtaining material properties from the literature data and accounts for preferably physiological behavior of the spine (Damm et al., 2020Damm, N., Rockenfeller, R., Gruber, K., 2020. Lumbar spinal ligament characteristics extracted from stepwise reduction experiments allow for preciser modeling than literature data. Biomech. Model. Mechanobiol. 19, 893–910. https://doi.org/10.1007/s10237-019-01259-6
https://doi.org/10.1007/s10237-019-01259...
; Schmidt et al., 2007Schmidt, H., Heuer, F., Drumm, J., Klezl, Z., Claes, L., Wilke, H.-J., 2007. Application of a calibration method provides more realistic results for a finite element model of a lumbar spinal segment. Clin. Biomech. 22, 377–84. https://doi.org/10.1016/j.clinbiomech.2006.11.008
https://doi.org/10.1016/j.clinbiomech.20...
). Moreover, most FE models do not take into consideration the regional variance of geometrical and material properties of the intervertebral disc (IVD) (Cassidy et al., 1989Cassidy, J.J., Hiltner, A., Baer, E., 1989. Hierarchical Structure of the Intervertebral Disc. Connect. Tissue Res. 23, 75–88. https://doi.org/10.3109/03008208909103905
https://doi.org/10.3109/0300820890910390...
; Guerin and Elliott, 2006Guerin, H.A.L., Elliott, D.M., 2006. Degeneration affects the fiber reorientation of human annulus fibrosus under tensile load. J. Biomech. 39, 1410–1418. https://doi.org/10.1016/j.jbiomech.2005.04.007
https://doi.org/10.1016/j.jbiomech.2005....
; Holzapfel et al., 2005Holzapfel, G.A., Schulze-Bauer, C.A.J., Feigl, G., Regitnig, P., 2005. Single lamellar mechanics of the human lumbar anulus fibrosus. Biomech. Model. Mechanobiol. 3, 125–140. https://doi.org/10.1007/s10237-004-0053-8
https://doi.org/10.1007/s10237-004-0053-...
) although they are important characteristics for the accuracy when predicting internal stresses such as the intradiscal pressure.

Another important step in the development of FE models of the spine is the construction of the geometry since it strongly influences the response of numerical results (Niemeyer et al., 2012Niemeyer, F., Wilke, H.J., Schmidt, H., 2012. Geometry strongly influences the response of numerical models of the lumbar spine: a probabilistic finite element analysis. J. Biomech. 45, 1414–1423. https://doi.org/10.1016/j.jbiomech.2012.02.021
https://doi.org/10.1016/j.jbiomech.2012....
). One possible approach to this problem is to choose a geometry obtained via image segmentation of an individual (e.g. computed tomography) and calibrate its material properties from the aforementioned experimental data. The geometry is derived from a real spine and consequently has feasible anthropometric proportions. However, it may carry deviations from an anthropometric mean that may induce numerical results with biases. This may result in difficulties in obtaining a good agreement with the average/median values regarding the biomechanics of the spine reported in the literature. An alternative approach is to construct or controllably modify a geometry to statistically represent the average values of anthropometric data. In this case, the model may carry some geometric relationship that is not verified in a human being since it was "synthesized". On the other hand, with the statistically representative construction of the anthropometric data, accompanied by an observation of the anatomical proportions and proper assembly, it is possible to minimize the risk of incurring the above-mentioned problem. Furthermore, this allows for estimating and reporting average values, which is the preferred method of data publication for most clinical and biomechanical studies.

The main objective of this study is to investigate the biomechanical behavior (change of scoliotic angle and range of motion) of the spine after insertion of VBT with different cord pretensions. For that purpose, a FE model statistically representative of the average spine was developed, stepwise calibrated, and validated. The study is of clinical interest since the cord pretension is an important factor to avoid overcorrection and under-correction of scoliotic spines treated with VBT which are common causes of revision surgery. It also fills a gap in the literature since there is no study that numerically evaluated the effects of different cord tension within VBT on thoracolumbar spines.

2 MATERIALS AND METHODS

The FE model of the thoracolumbar spine was developed in Abaqus® (Dassault Systèmes, Waltham, MA, USA) as shown in Figure 1.

Figure 1
Perspective and lateral view of the intact T10-L3 spine.

2.1 Geometry

The geometry of the vertebrae and sacrum (T10-S1) was derived from CT images of a man who had no history of spinal pathology. It was modified to represent an average spine of a young man following a manual iterative procedure. Firstly, the original geometry of each vertebra and sacrum was modified by translating and rotating the control points that define the shape of the surfaces (Non-Uniform Rational B-Spline - NURBS) using the software Rhinoceros® (Robert McNeel & Associates, Seattle, USA). The dimensions considered during this process (Figure 2) were the superior and inferior vertebral body width (SVBW and IVBW), superior and inferior vertebral body length (SVBL and IVBL), anterior and posterior vertebral body height (AVBH and PVBH), lateral vertebral body height (LVBH), transverse process length (TPL), vertebral canal superior width and length (VCSW and VCSL), superior and inferior interfacet width (SFFW and IFFW), interfacet height (FFH), superior and inferior facet length (SFL and IFL), superior and inferior facet width (SFW and IFW), superior and inferior longitudinal facet angle (SLFA and ILFA), and superior and inferior transverse facet angle (STFA and ITFA) as well as the morphology of the vertebral endplates (Masharawi et al., 2010Masharawi, Y., Dar, G., Peleg, S., Steinberg, N., Medlej, B., May, H., Abbas, J., Hershkovitz, I., 2010. A morphological adaptation of the thoracic and lumbar vertebrae to lumbar hyperlordosis in young and adult females. Eur. Spine J. 19, 768–773. https://doi.org/10.1007/s00586-009-1256-6
https://doi.org/10.1007/s00586-009-1256-...
, 2005Masharawi, Y., Rothschild, B., Salame, K., Dar, G., Peleg, S., Hershkovitz, I., 2005. Facet tropism and interfacet shape in the thoracolumbar vertebrae: characterization and biomechanical interpretation. Spine (Phila. Pa. 1976). 30, E281--E292. https://doi.org/10.1097/01.brs.0000164098.00201.8d
https://doi.org/10.1097/01.brs.000016409...
, 2004Masharawi, Y., Rothschild, B., Dar, G., Peleg, S., Robinson, D., Been, E., Hershkovitz, I., 2004. Facet orientation in the thoracolumbar spine: three-dimensional anatomic and biomechanical analysis. Spine (Phila. Pa. 1976). 29, 1755–1763. https://doi.org/00007632-200408150-00009 [pii]
https://doi.org/00007632-200408150-00009...
; Masharawi and Salame, 2011Masharawi, Y., Salame, K., 2011. Shape variation of the neural arch in the thoracic and lumbar spine: Characterization and relationship with the vertebral body shape. Clin. Anat. 24, 858–867. https://doi.org/10.1002/ca.21175
https://doi.org/10.1002/ca.21175...
; van der Houwen et al., 2010van der Houwen, E.B., Baron, P., Veldhuizen, A.G., Burgerhof, J.G.M., van Ooijen, P.M.A., Verkerke, G.J., 2010. Geometry of the Intervertebral Volume and Vertebral Endplates of the Human Spine. Ann. Biomed. Eng. 38, 33–40. https://doi.org/10.1007/s10439-009-9827-6
https://doi.org/10.1007/s10439-009-9827-...
). Secondly, the vertebrae were positioned in the 3D space considering the intervertebral disc (IVD) height and lordosis angle (Been et al., 2011Been, E., Li, L., Hunter, D.J., Kalichman, L., 2011. Geometry of the vertebral bodies and the intervertebral discs in lumbar segments adjacent to spondylolysis and spondylolisthesis: pilot study. Eur. Spine J. 20, 1159–1165. https://doi.org/10.1007/s00586-010-1660-y
https://doi.org/10.1007/s00586-010-1660-...
, 2010Been, E., Barash, A., Marom, A., Kramer, P.A., 2010. Vertebral Bodies or Discs: Which Contributes More to Human-like Lumbar Lordosis? Clin. Orthop. Relat. Res. 468, 1822–1829.; Boulay et al., 2006Boulay, C., Tardieu, C., Hecquet, J., Benaim, C., Mouilleseaux, B., Marty, C., Prat-Pradal, D., Legaye, J., Duval-Beaupère, G., Pélissier, J., 2006. Sagittal alignment of spine and pelvis regulated by pelvic incidence: standard values and prediction of lordosis. Eur. Spine J. 15, 415–422.; Cheng et al., 1998Cheng, X.G., Sun, Y., Boonen, S., Nicholson, P.H.F., Brys, P., Dequeker, J., Felsenberg, D., 1998. Measurements of vertebral shape by radiographic morphometry: sex differences and relationships with vertebral level and lumbar lordosis. Skeletal Radiol. 27, 380–384. https://doi.org/10.1007/s002560050402
https://doi.org/10.1007/s002560050402...
; Eijkelkamp et al., 2007Eijkelkamp, M.F., Klein, J.P., Veldhuizen, A.G., Horn, J.R. van, Verkerke, G.J., 2007. The geometry and shape of the human intervertebral disc, in: Spine [Submitted for Publication]. The Spine, pp. 75–83.; Labelle et al., 2005Labelle, H., Roussouly, P., Berthonnaud, É., Dimnet, J., O’Brien, M., 2005. The Importance of Spino-Pelvic Balance in L5–S1 Developmental Spondylolisthesis: A Review of Pertinent Radiologic Measurements. Spine (Phila. Pa. 1976). 30, S27–S34. https://doi.org/10.1097/01.brs.0000155560.92580.90
https://doi.org/10.1097/01.brs.000015556...
; Nissan and Issachar, 1986Nissan, M., Issachar, G., 1986. Dimensions of human lumbar vertebrae in the sagittal plane. J. Biomech. 19, 753–758.; Vialle et al., 2005Vialle, R., Levassor, N., Rillardon, L., Templier, A., Skalli, W., Guigui, P., 2005. Radiographic analysis of the sagittal alignment and balance of the spine in asymptomatic subjects. J. bone Jt. Surg. 87, 260–267. https://doi.org/10.2106/JBJS.D.02043
https://doi.org/10.2106/JBJS.D.02043...
; White and Panjabi, 1990White, A.A., Panjabi, M.M., 1990. Clinical Biomechanics of the Spine, 2nd ed. J. B. Lippincott Company, Philadelphia.). The geometries of the IVDs were created by sketching their lateral profile and using the morphology and positions of the vertebrae. The nucleus pulposus was dimensioned with 44% of the total cross-section area of the IVD (White and Panjabi, 1990White, A.A., Panjabi, M.M., 1990. Clinical Biomechanics of the Spine, 2nd ed. J. B. Lippincott Company, Philadelphia.) with its center located approximately at 3.5 mm posterior from the center of the IVD (Noailly et al., 2007Noailly, J., Wilke, H.J., Planell, J.A., Lacroix, D., 2007. How does the geometry affect the internal biomechanics of a lumbar spine bi-segment finite element model? Consequences on the validation process. J. Biomech. 40, 2414–2425. https://doi.org/10.1016/j.jbiomech.2006.11.021
https://doi.org/10.1016/j.jbiomech.2006....
; Weisse et al., 2012Weisse, B., Aiyangar, A.K., Affolter, C., Gander, R., Terrasi, G.P., Ploeg, H., 2012. Determination of the translational and rotational stiffnesses of an L4 – L5 functional spinal unit using a specimen-specific finite element model. J. Mech. Behav. Biomed. Mater. 13, 45–61. https://doi.org/10.1016/j.jmbbm.2012.04.002
https://doi.org/10.1016/j.jmbbm.2012.04....
).

Figure 2
Dimensions considered during the modification of the geometry of the vertebrae.

The following equation was used to compare the dimension of the model (dm) with the anthropometric data (da) with respect to the anthropometric standard deviation (SD):

Z = 100 ( d m - d a ) / S D (1)

2.2 Mesh

Tetrahedral elements were used to represent the vertebrae, as they can conform to complex geometries (Table 1). The cortical bone of the vertebral body was modeled using shell elements with a wall thickness of 0.29 mm (Ritzel et al., 1997Ritzel, H., Amling, M., Pösl, M., Hahn, M., Delling, G., 1997. The thickness of human vertebral cortical bone and its changes in aging and osteoporosis: a histomorphometric analysis of the complete spinal column from thirty-seven autopsy specimens. J. Bone Miner. Res. 12. https://doi.org/10.1359/jbmr.1997.12.1.89
https://doi.org/10.1359/jbmr.1997.12.1.8...
). The IVD mesh was created using hexahedral elements. Only 1% of the FEs representing the vertebrae and IVD had an aspect ratio greater than 3, which is acceptable according to Burkhart et al. (2013)Burkhart, T.A., Andrews, D.M., Dunning, C.E., 2013. Finite element modeling mesh quality, energy balance and validation methods: A review with recommendations associated with the modeling of bone tissue. J. Biomech. 46, 1477–1488. https://doi.org/10.1016/j.jbiomech.2013.03.022
https://doi.org/10.1016/j.jbiomech.2013....
. Tie constraint was used to guarantee continuity between the superior and inferior surfaces of the IVDs and their adjacent vertebrae since their meshes are different.

Table 1
Element types used for each structure in the FE model.

A mesh quality assessment was performed to evaluate the optimal mesh density that provides accurate results while keeping simulation time tractable. For that purpose, the L1-L2 IVD was tested using the traditional pure moment load protocol with a magnitude moment of 7.5 Nm in flexion. The range of motion (ROM, angular displacement of L1 vertebra with respect to L2 vertebra) and pressure at the center of the IVD were evaluated for different mesh sizes using quadratic shapes functions. A mesh sensitive analysis was not performed with the vertebrae since they have a minor effect on the movements of the spine compared to the soft tissue. Additionally, linear interpolation was used for the FEs representing the vertebrae.

2.3 Constitutive models

The nucleus pulposus was modeled as a hyperelastic Mooney-Rivlin material using the strain energy density function

W n = C 10 n I - 3 + C 01 n I I - 3 + D n J α - 1 2 (2)

where C10n, C01n and Dn are the material constants defining matrix stiffness and compressibility. The elastic volume ratio is represented by Jα while I and II are the first and second invariant of the right Cauchy-Green deformation tensor, respectively. The subscript .n indicates the belonging of the material parameters to the nucleus pulposus.

The annulus fibrosus was modelled using the hyperelastic Holzapfel–Gasser–Ogden model (Holzapfel et al., 2000Holzapfel, G.A., Gasser, T.C., Ogden, R.W., 2000. A new constitutive framework for arterial wall mechanics and a comparative study of material models. J. Elast. Phys. Sci. solids 61, 1–48. https://doi.org/10.1023/A:1010835316564
https://doi.org/10.1023/A:1010835316564...
). The total strain energy density function is composed of the strain energy density function of the isotropic matrix represented by the Neo-Hookean model and the strain energy density function of the fibers. It has the form

W=C10I-3+1DJα2-12-ln(Jα)+K12(K2)α=1NexpK2Eα2-1,(3)

with

EακI-3+1-3κI4αα-1,(4)

where C10 and D are material constants of the annulus ground substance related to its stiffness and compressibility. The material parameters K1 and K2 describes the exponential stress-strain relation for the collagen fibers, and κ represents the level of dispersion in the fiber directions. The constitutive model considers that the fibers provide resistance only to traction. Additionally, I4αα are the pseudo-invariants of the distortion part of the right Cauchy-Green deformation tensor and unit vectors in the fiber directions. The number of fiber families N was considered equal to two. Moreover, the orientation of the fibers at the anterior aspect of the IVD with respect to its transverse plane is defined by the parameter β.

The annulus fibrosus was divided into five radial and five circumferential sections (Figure 3) to simulate the variance of properties in the annulus found in histological anatomy studies and experimental studies (Ebara et al., 1996Ebara, S., AIatridis, J.C., Setton, L.A., Foster, R.J., Van Mow, C., Weidenbaum, M., 1996. Tensile properties of nondegenerate human lumbar anulus fibrosus. Spine (Phila. Pa. 1976). 21, 452–461.; Eberlein et al., 2001Eberlein, R., Holzapfe, G.A., Schulze-Bauer, C.A.J., 2001. An Anisotropic Model for Annulus Tissue and Enhanced Finite Element Analyses of Intact Lumbar Disc Bodies. Comput. Methods Biomech. Biomed. Engin. 4, 209–229. https://doi.org/10.1080/10255840108908005
https://doi.org/10.1080/1025584010890800...
; Holzapfel et al., 2005Holzapfel, G.A., Schulze-Bauer, C.A.J., Feigl, G., Regitnig, P., 2005. Single lamellar mechanics of the human lumbar anulus fibrosus. Biomech. Model. Mechanobiol. 3, 125–140. https://doi.org/10.1007/s10237-004-0053-8
https://doi.org/10.1007/s10237-004-0053-...
; Zhu et al., 2008Zhu, D., Gu, G., Wu, W., Gong, H., Zhu, W., Jiang, T., Cao, Z., 2008. Micro-structure and mechanical properties of annulus fibrous of the L4-5 and L5-S1 intervertebral discs. Clin. Biomech. 23, S74-82.). The coefficients K1 and K2 defining the material properties of the fibres were varied linearly in the circumferential and radial direction. The coefficients k1c,k2c and βc were defined to be responsible for creating a gradient of k1,k2 and β values along the perimeter of the annulus fibrosus, respectively. Similarly, k1r and k2r were set to control the change of k1 and k2 values along the radial direction of the annulus fibrosus, respectively. For these reasons, the subscripts (.)c and (.)r indicate the parameter is responsible for creating a gradient of the material properties values in the circumferential and radial direction of the annulus fibrosus, respectively. Operationally, the properties of each layer of the IVD are calculated based on the values of the outer layer of the most anterior aspect of the IVD (region A1, Figure 3). For instance, a k1c equal to 0.1 means that the k1 value of region B1 and k1 value of region C1 are 10% and 20% higher than the k1 value of region A1, respectively.

Figure 3
Intervertebral disc divided into five radial and five circumferential sections to account for the regional difference in material properties. The figure also indicates the subregion of C4.

The spinal ligaments were modeled as bar elements of nonlinear stiffness connecting the adjacent vertebrae (Figure 4). The gap at each pair of facet joints was modeled using soft contact interactions of the type nonlinear penalization. The model does not consider any tangential resistance at the surface of the facet joints since their coefficient of friction is negligible (Cramer and Darby, 2014Cramer, G.D., Darby, S.A., 2014. Clinical Anatomy of the Spine, Spinal Cord and ANS, 3rd ed. Mosby, St. Louis.). The input data was calculated using the equation

P=aebh-h0-1 h<h0,(5)

where P is the contact pressure and h is the overclosure while a, b and h0 are material constants.

Figure 4
Front view (A) and perspective view (B) showing the location of the modeled ligaments. It contains the anterior longitudinal ligament (ALL), intertransverse ligaments (ITL), posterior longitudinal ligament (PLL), capsular ligaments (CLL), flavum ligament (FVL), interspinous ligament (ISL), and supraspinous ligament (SSL).

2.4 Material properties and calibration

The cortical and cancellous bones of the vertebral body were modeled as an isotropic elastic material with a Young Modulus of 12,000 MPa and 100 MPa, respectively (Guan et al., 2006Guan, Y., Yoganandan, N., Zhang, J., Pintar, F.A., Cusick, J.F., Wolfla, C.E., Maiman, D.J., 2006. Validation of a clinical finite element model of the human lumbosacral spine. Medical and Biological Engineering and Computing. Med. Biol. Eng. Comput. 44, 633–641.; Kulduk et al., 2015Kulduk, A., Altun, N.S., Senkoylu, A., 2015. Biomechanical comparison of effects of the Dynesys and Coflex dynamic stabilization systems on range of motion and loading characteristics in the lumbar spine: a finite element study. Int. J. Med. Robot. Comput. Assist. Surg. 11, 400–405. https://doi.org/10.1002/rcs.1636
https://doi.org/10.1002/rcs.1636...
; C.-L. Liu et al., 2011Liu, C.-L., Zhong, Z.-C., Hsu, H.-W., Shih, S.-L., Wang, S.-T., Hung, C., Chen, C.-S., 2011. Effect of the cord pretension of the Dynesys dynamic stabilisation system on the biomechanics of the lumbar spine: a finite element analysis. Eur. Spine J. 20, 1850–1858. https://doi.org/10.1007/s00586-011-1817-3
https://doi.org/10.1007/s00586-011-1817-...
). The posterior elements of the vertebrae were simplified by considering only one type of bone with intermediate stiffness considering a Young modulus of 3,500 MPa (Ayturka and Puttlitz, 2011Ayturka, U.M., Puttlitz, C.M., 2011. Parametric convergence sensitivity and validation of a finite element model of the human lumbar spine. Comput. Methods Biomech. Biomed. Engin. 14, 695–705.; Guan et al., 2006Guan, Y., Yoganandan, N., Zhang, J., Pintar, F.A., Cusick, J.F., Wolfla, C.E., Maiman, D.J., 2006. Validation of a clinical finite element model of the human lumbosacral spine. Medical and Biological Engineering and Computing. Med. Biol. Eng. Comput. 44, 633–641.; Shirazi-Adl et al., 1986Shirazi-Adl, A., Ahmed, A.M., Shrivastava, S.C., 1986. A finite element study of a lumbar motion segment subjected to pure sagittal plane moments. J. Biomech. 19, 331–350. https://doi.org/10.1016/0021-9290(86)90009-6
https://doi.org/10.1016/0021-9290(86)900...
).

The mechanical properties of the soft tissues and the contact of the facet joints were calibrated considering the inverse order of experimental resection studies (Coombs et al., 2013Coombs, D., Bushelow, M., Laz, P., Rao, M., Rullkoetter, P., 2013. Stepwise Validated Finite Element Model of the Human Lumbar Spine, in: ASME 2013 Conference on Frontiers in Medical Devices: Applications of Computer Modeling and Simulation. American Society of Mechanical Engineers, pp. 1–2. https://doi.org/10.1115/FMD2013-16167
https://doi.org/10.1115/FMD2013-16167...
; Damm et al., 2020Damm, N., Rockenfeller, R., Gruber, K., 2020. Lumbar spinal ligament characteristics extracted from stepwise reduction experiments allow for preciser modeling than literature data. Biomech. Model. Mechanobiol. 19, 893–910. https://doi.org/10.1007/s10237-019-01259-6
https://doi.org/10.1007/s10237-019-01259...
; Heuer et al., 2007aHeuer, F., Schmidt, H., Claes, L., Wilke, H.-J., 2007a. Stepwise reduction of functional spinal structures increase vertebral translation and intradiscal pressure. J. Biomech. 40, 795–803.; Jaramillo et al., 2016Jaramillo, H.E., Puttlitz, C.M., McGilvray, K., García, J.J., 2016. Characterization of the L4–L5–S1 motion segment using the stepwise reduction method. J. Biomech. 49, 1248–1254. https://doi.org/10.1016/j.jbiomech.2016.02.050
https://doi.org/10.1016/j.jbiomech.2016....
). The spinal structures were sequentially added to the FE models and calibrated by varying and choosing the set of material properties that best fit the experimental results (Ezquerro et al., 2011Ezquerro, F., Vacas, F.G., Postigo, S., Prado, M., Simón, A., 2011. Calibration of the finite element model of a lumbar functional spinal unit using an optimization technique based on differential evolution. Med. Eng. Phys. 33, 89–95. https://doi.org/10.1016/j.medengphy.2010.09.010
https://doi.org/10.1016/j.medengphy.2010...
; Nicolini et al., 2021aNicolini, L.F., Beckmann, A., Laubach, M., Hildebrand, F., Kobbe, P., Roesler, C.R. de M., Fancello, E.A., Markert, B., Stoffel, M., 2021a. An experimental-numerical method for the calibration of Finite Element models of the spine. Submitt. to Med. Eng. Phys.; Schmidt et al., 2006Schmidt, H., Heuer, F., Simon, U., Kettler, A., Rohlmann, A., Claes, L., Wilk, H.-J., 2006. Application of a new calibration method for a three-dimensional finite element model of a human lumbar annulus fibrosus. Clin. Biomech. 21, 337–344.).

The experimental tests were performed as described in our previous studies (Beckmann et al., 2020Beckmann, A., Nicolini, L.F., Grevenstein, D., Backes, H., Oikonomidis, S., Sobottke, R., Kobbe, P., Hildebrand, F., Stoffel, M., Markert, B., Siewe, J., Herren, C., 2020. Biomechanical in vitro test of a novel dynamic spinal stabilization system incorporating polycarbonate urethane material under physiological conditions. J. Biomech. Eng. 142. https://doi.org/10.1115/1.4044242
https://doi.org/10.1115/1.4044242...
, 2018Beckmann, A., Herren, C., Mundt, M., Siewe, J., Kobbe, P., Sobottke, R., Pape, H.-C., Stoffel, M., Markert, B., 2018. A new in vitro spine test rig to track multiple vertebral motions under physiological conditions. Biomed. Eng. (NY). 63, 341–347. https://doi.org/10.1515/bmt-2016-0173
https://doi.org/10.1515/bmt-2016-0173...
; Nicolini et al., 2021bNicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
). The spine specimens had the vertebra, intervertebral discs, and the seven aforementioned ligaments preserved. They were tested under a pure moment load and a rotational speed of 1°/s in flexion-extension, lateral bending and axial rotation with the custom-built machine (Nicolini et al., 2021bNicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
). The load was recorded by torque sensors that provided a maximum error of 1% relative to the target value while the kinematics of the vertebrae were measured by a electromagnetic tracking system that provides deviations lower than 0.1 mm and 0.1° (Beckmann et al., 2018Beckmann, A., Herren, C., Mundt, M., Siewe, J., Kobbe, P., Sobottke, R., Pape, H.-C., Stoffel, M., Markert, B., 2018. A new in vitro spine test rig to track multiple vertebral motions under physiological conditions. Biomed. Eng. (NY). 63, 341–347. https://doi.org/10.1515/bmt-2016-0173
https://doi.org/10.1515/bmt-2016-0173...
). The data was recorded at a frequency of 100 Hz (Brandes et al., 2021Brandes, L.L., Nicolini, L.F., Greven, J., Lichte, P., Stopinski, T.T., Sattler, M., Hildebrand, F., Pishnamaz, M., 2021. Biomechanical Performance of BoneHelix® Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model. Life 11, 1189. https://doi.org/10.3390/life11111189
https://doi.org/10.3390/life11111189...
). Some of the L1-L2 spine specimens were tested in different resection states to obtain the individual contribution of the IVD, the facet joints and the ligaments to the kinematics of the spine.

For the calibration procedure, the experimental data was prepared in such a way that the FE model could: 1) replicate the median experimental moment-ROM curves of twelve L1-L2 intact segments in the intact state (some already published by Nicolini et al. (2021b)Nicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
, and 2) reproduce the proportional change of ROM after removal of a spinal structure as observed in our resection study with five L1-L2 spines in the resected state. For this purpose, the experimental median ROM of twelve L1-L2 intact segments (Rn¯) were scaled using the equation:

Rd¯=γRn¯ ,(6)

where Rd¯ represents the predicted ROM of the spine with defects while γ is a coefficient defined as

γ = m e d i a n R d R n (7)

The scale factor γ represents a proportional increase in ROM caused by the resection of a spinal structure. It was calculated by dividing the ROM obtained after the resection of a spinal structure (Rd) by the ROM of the native spine (Rn). This process was performed for each of the five spine segments and the median was taken. Moreover, this rule was applied for each loading direction and datapoint of the moment-ROM curve individually (Nicolini et al., 2021aNicolini, L.F., Beckmann, A., Laubach, M., Hildebrand, F., Kobbe, P., Roesler, C.R. de M., Fancello, E.A., Markert, B., Stoffel, M., 2021a. An experimental-numerical method for the calibration of Finite Element models of the spine. Submitt. to Med. Eng. Phys.).

The R-squared function comparing the numerical values with the experimental ones was implemented as a cost function to be minimized. The calibration of the material properties of the IVD were calibrated considering motion within the three anatomical planes while the material properties of the contact of the facet joints were calibrated considering axial rotation. The following design vectors defining the mechanical properties of the IVD and the facet joints were set to be optimized via an in-house developed Genetic Algorithm.

x={C10n,C01n,C10,k1,k2,k,β,k1c,k2c,k1r,k2r,βc}T,(8)
x={a,b,h0}T,(9)

An optimization algorithm was used to unequivocally determine the force-displacement curves that represent the material properties of the ligaments and replicate the experimental moment-ROM data of the spine in flexion-extension. During all simulations, the degrees of freedom of the lower part of the L2 vertebra were constraint and the L1 vertebra was bent similar to the experimental flexibility tests. For the sake of shortness, this article presents only the calibration results of the L1-L2 segment.

2.5 Verification and validation

After the calibration of the material properties of the L1-L2 IVD under a pure moment, the tests of Holzapfel et al. (2005)Holzapfel, G.A., Schulze-Bauer, C.A.J., Feigl, G., Regitnig, P., 2005. Single lamellar mechanics of the human lumbar anulus fibrosus. Biomech. Model. Mechanobiol. 3, 125–140. https://doi.org/10.1007/s10237-004-0053-8
https://doi.org/10.1007/s10237-004-0053-...
were numerically replicated to verify the regional variation of the tensile properties within the IVD. They performed tensile tests with single lamellar annulus specimens whereby the lamellar fiber direction was aligned with the load axis. The specimens were obtained from the ventro-lateral external (VLe), the ventrolateral internal (VLi), the dorsal external (De), and the dorsal internal (Di) of the L1-L2 IVD. Moreover, the forces at the contact of the facet joints were verified with the experimental values of Niosi et al. (2008)Niosi, C.A., Wilson, D.C., Zhu, Q., Keynan, O., Wilson, D.R., Oxland, T.R., 2008. The effect of dynamic posterior stabilization on facet joint contact forces: an in vitro investigation. Spine (Phila. Pa. 1976). 33, 19–26. https://doi.org/10.1097/BRS.0b013e31815e7f76
https://doi.org/10.1097/BRS.0b013e31815e...
by testing the L1-L2 segment under a pure moment of 7.5 Nm in different loading directions.

For validation purposes, the L1-L2 segment was submitted under multiple bending moments in flexion-extension or lateral bending combined with axial rotation. The numerical ROM values were compared to those obtained experimental by Wilmanns et al. (2021)Wilmanns, N., Beckmann, A., Nicolini, L.F., Herren, C., Sobottke, R., Hildebrand, F., Siewe, J., Kobbe, P., Markert, B., Stoffel, M., 2021. Biomechanical in Vitro and Finite Element Study On Different Sagittal Alignment Postures of the Lumbar Spine During Multiaxial Daily Motion. J. Biomech. Eng. 144, 9. https://doi.org/10.1115/1.4053083
https://doi.org/10.1115/1.4053083...
.

2.6 Modeling and simulation of the spine with VBT

The geometry of a VBT system was obtained from a manufacturer (Globus Medical Inc, Pennsylvania, USA) and tested using the model of L1-L2 segment (Figure 5). The screw was designed with an inner diameter of 4.75 mm and an outer diameter of 5.5 mm. Oriented by an experienced surgeon, the screws were inserted close to the foramen with bicortical purchase at the left lateral side of the L1 and L2 vertebrae and parallel to the frontal plane. The anchors had an inner and outer diameter of 5.5 and 6.5 mm, respectively. The screws and anchors were modeled as an isotropic material made of titanium with a Young’s Modulus of 110 GPa and a Poisson's ratio of 0.3 (Lo et al., 2011Lo, C.-C., Tsai, K.-J., Zhong, Z.-C., Chen, S.-H., Hung, C., 2011. Biomechanical differences of Coflex-F and pedicle screw fixation combined with TLIF or ALIF – a finite element study. Comput. Methods Biomech. Biomed. Engin. 14, 947–956. https://doi.org/10.1080/10255842.2010.501762
https://doi.org/10.1080/10255842.2010.50...
; Niinomi et al., 2016Niinomi, M., Liu, Y., Nakai, M., Liu, H., Li, H., 2016. Biomedical titanium alloys with Young’s moduli close to that of cortical bone. Regen. Biomater. 3, 173–185. https://doi.org/10.1093/rb/rbw016
https://doi.org/10.1093/rb/rbw016...
). The mesh was created using tetrahedral elements with linear shape functions. Tie constraint was used to attach each anchor and screw to the vertebra. The cord of 4 mm diameter was modeled using a bar element with a Young’s Modulus 1,500 MPa in tension and negligible resistance to compression (Kulduk et al., 2015Kulduk, A., Altun, N.S., Senkoylu, A., 2015. Biomechanical comparison of effects of the Dynesys and Coflex dynamic stabilization systems on range of motion and loading characteristics in the lumbar spine: a finite element study. Int. J. Med. Robot. Comput. Assist. Surg. 11, 400–405. https://doi.org/10.1002/rcs.1636
https://doi.org/10.1002/rcs.1636...
; C. L. Liu et al., 2011Liu, C.-L., Zhong, Z.-C., Hsu, H.-W., Shih, S.-L., Wang, S.-T., Hung, C., Chen, C.-S., 2011. Effect of the cord pretension of the Dynesys dynamic stabilisation system on the biomechanics of the lumbar spine: a finite element analysis. Eur. Spine J. 20, 1850–1858. https://doi.org/10.1007/s00586-011-1817-3
https://doi.org/10.1007/s00586-011-1817-...
).

Figure 5
Perspective view of the L1-L2 segment instrumented with vertebral body tethering.

The first test evaluated the changes in the coronal plane angles (scoliotic angles) due to a cord tension of up to 300 N within VBT. In the second test, the spine was simulated under an external pure moment up to 6 Nm in left and right lateral bending (Lavelle et al., 2016Lavelle, W.F., Moldavsky, M., Cai, Y., Ordway, N.R., Bucklen, B.S., 2016. An initial biomechanical investigation of fusionless anterior tether constructs for controlled scoliosis correction. Spine J. 16, 408–413. https://doi.org/10.1016/j.spinee.2015.11.004
https://doi.org/10.1016/j.spinee.2015.11...
; Nicolini et al., 2021bNicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
) considering an initial cord pretension of 0 N, 50N, 100 N, 200 N and 300 N. Pretension was applied by controlling the initial stretch of the bar element representing the cord. The ROM was determined with respect to the spine at its respective initial post VBT position (no external load independent of cord pretension).

3 RESULTS

3.1 Geometry

The final geometric model of the vertebrae presented an average of the absolute Z values of 18% and a maximum absolute Z value of 74% considering the dimensions SVBW, IVBW, SVBL, IVBL, AVBH, PVBH, LVBH, TPL, VCSW, VCSL, SFFW, IFFW, FFH, SFL, IFL, SFW, IFW, SLFA, ILFA, STFA and ITFA (Table S1 of supplementary material). The morphology of the vertebral endplates was within 10% of the SD of the anthropometric data. The average of the absolute Z values was 29% for the wedging angle of the vertebrae and IVDs, 53% for sagittal dimensions including lumbar lordosis, sacral slope, pelvic tilt, pelvic incidence and thoracic kyphosis, and 20% for the anterior and posterior height of the IVDs. The comparison of the dimensions of the geometric model and the experimental data are presented in Table S1, Table S2, Table S3, and Table S4 of supplementary material. The lordosis angle of the model represents a young male with a type 3 lordosis (Roussouly et al., 2005Roussouly, P., Sohrab, G., Eric, B., Johanes, D., 2005. Classification of the normal variation in the sagittal alignment of the human lumbar spine and pelvis in the standing position. Spine (Phila. Pa. 1976). 30, 346–53.).

3.2 Mesh

The mesh-sensitive analysis showed an asymptotic behavior of the ROM and intradiscal pressure with the refinement of the mesh (Figure 6). An intermediate mesh of 27991 nodes with quadratic shape functions was selected to model the IVD which had a difference of 0.6% of ROM and 4.5% for the intradiscal pressure when compared to the results of the finest tested mesh (68592 nodes).

Figure 6
Effect of mesh refinement on numerical ROM and intradiscal pressure at the center of the disc.

3.3 Calibration and validation

After calibration, the model of L1-L2 IVD was capable of reproducing the experimental moment-ROM curves presenting an average R-squared of 0.9 considering flexion, extension, lateral bending, and axial rotation (Figure 7). The material parameters are shown in Table S5 of supplementary material. The numerical stress-stretch ratio curves of single lamellae obtained from four anatomical regions of the annulus also presented a similar trend compared to the experimental results (Figure 8). All numerical data points were within SD of the experimental data and the average R-squared was 0.95. The tensile moduli for external lamellae were approximately two times higher than those for internal lamellae. The tensile moduli of the ventral layers were only 10% higher compared to the dorsal layers. Both results agree with the values found by Holzapfel et al. (Holzapfel et al., 2005Holzapfel, G.A., Schulze-Bauer, C.A.J., Feigl, G., Regitnig, P., 2005. Single lamellar mechanics of the human lumbar anulus fibrosus. Biomech. Model. Mechanobiol. 3, 125–140. https://doi.org/10.1007/s10237-004-0053-8
https://doi.org/10.1007/s10237-004-0053-...
).

Figure 7
Comparison of the numerical and experimental moment-ROM curves of the L1-L2 under pure moment in different loading directions. For the first subplot, positive and negative values represent flexion and extension, respectively. The average R-squared was 0.9 considering all loading directions.
Figure 8
Comparison of the numerical and experimental engineering stress-stretch ratio curves of single lamellae obtained from four anatomical regions of the annulus fibrosus: the ventro-lateral external (VLe), the ventro-lateral internal (VLi), the dorsal external (De), and the dorsal internal (Di). The average R-squared for all loading directions was 0.95.

The fiber orientation angle with respect to the transverse plane gradually increases from 30° at the anterior aspect to 42° at the posterior aspect of the modeled IVD. This is consistent with the histological values for the anterior and posterior part of the IVD obtained by Holzapfel et al. (2005)Holzapfel, G.A., Schulze-Bauer, C.A.J., Feigl, G., Regitnig, P., 2005. Single lamellar mechanics of the human lumbar anulus fibrosus. Biomech. Model. Mechanobiol. 3, 125–140. https://doi.org/10.1007/s10237-004-0053-8
https://doi.org/10.1007/s10237-004-0053-...
(25.7 ± 5.8° and 49.3 ± 8.0°) and Cassidy et al. (1989)Cassidy, J.J., Hiltner, A., Baer, E., 1989. Hierarchical Structure of the Intervertebral Disc. Connect. Tissue Res. 23, 75–88. https://doi.org/10.3109/03008208909103905
https://doi.org/10.3109/0300820890910390...
(28° and 45°). The total criss-cross angle of the fibers at the outer layer of the model was 60°, which is within standard deviation compared to the experimental results of 51.9 ± 8.4° obtained by Guerin et al. (Guerin and Elliott, 2006Guerin, H.A.L., Elliott, D.M., 2006. Degeneration affects the fiber reorientation of human annulus fibrosus under tensile load. J. Biomech. 39, 1410–1418. https://doi.org/10.1016/j.jbiomech.2005.04.007
https://doi.org/10.1016/j.jbiomech.2005....
).

The model of the L1-L2 segment agreed with the experimental results for different resection stages in flexion, extension, lateral bending, and axial rotation (Figure 9). The average R-squared for all loading directions and resection stages was 0.85. For the native spine, the mentioned variable was equal to 0.83. The model predicted a force at the facet joints of 0, 5, 4, and 79 N whereas Niosi et al. (2008)Niosi, C.A., Wilson, D.C., Zhu, Q., Keynan, O., Wilson, D.R., Oxland, T.R., 2008. The effect of dynamic posterior stabilization on facet joint contact forces: an in vitro investigation. Spine (Phila. Pa. 1976). 33, 19–26. https://doi.org/10.1097/BRS.0b013e31815e7f76
https://doi.org/10.1097/BRS.0b013e31815e...
measured values of 2 ± 5, 13 ± 14, 11 ± 11, and 56 ± 17 N for flexion, extension, lateral bending, and axial rotation, respectively. The comparison of the numerical and experimental data of the spine under combined loading presented an R-squared of 0.98, 0.94, and 0.90 for flexion, extension, and lateral bending combined with axial rotation, respectively.

Figure 9
Comparison of the numerical and experimental results of the L1-L2 segment under pure moment in different loading directions and resection states. The error bars indicate the difference between the numerical and experimental results.

3.4 Effects of VBT

The scoliotic angle of the spine gradually increased with increasing of the cord pretension (Figure 10). Pretension forces of 50 N, 100 N, 200 N, and 300 N resulted in a change of scoliotic angle of 1.5°, 2.7°, 4.3°, and 5.3° to the left side of the spine, respectively.

Figure 10
Change of scoliotic angle as a function of the tension applied on the cord within anterior vertebral body tethering. A tensioning of 300 N resulted in a change of scoliotic angle of 5.3° as shown in the subfigure on the right side.

Herein, the ROM is presented as the angular displacement of the L1-L2 segment from its neutral position or initial post VBT position to the final position caused by an external moment of ±6 Nm. The ROM of the native spine was 4.5° in right lateral bending and reduced to 1.9°, 1.9°, 1.8°, 1.6°, and 1.4° for the cases of the spine with a cord pretension of 0 N, 50 N, 100 N, 200 N and 300 N, respectively (Figure 11). In left lateral bending, the absolute ROM of the native spine and the instrumented spine with no pretension was 4.6°. For the cases of a cord pretension of 50 N, 100 N, 200 N, and 300 N, the spine bent 3.1°, 1.9°, 1.0°, and 0.8° to the left side from its initial post VBT position, respectively. For the case of 50 N and 100 N, the moment-ROM curve in left lateral bending presented an inflection close to the point in which the cord tension reduced to zero due to the applied external moment.

Figure 11
Moment-ROM curves for the spine with different cord pretension in left and right lateral bending. Positive and negative values represent right and left lateral bending, respectively.

3 DISCUSSION

VBT is an interesting technique that has been approved in 2019 by the Food and Drug Administration (FDA) for US patients with AIS. As an emerging surgical technique, there is a paucity of studies on VBT especially regarding the effects of its engineering parameters that can substantially affect the biomechanics of the spine. This paper presented the first numerical study that evaluated the change of scoliotic angle and range of motion of the spine after insertion of VBT with different cord pretensions. It was motivated by the common clinical scenario related to overcorrection and under-correction of the spine treated with VBT that leads to revision surgery and the fact that the cord pretension is a key parameter to achieve appropriate scoliotic curve correction. For that purpose, this paper also presented the development of a FE model of the spine that is statistically representative of the average spine, stepwise calibrated, and validated.

The proposed method successfully generated a geometric model with dimensions close to the anthropometric data and within SDs. The iterative process is justified by the difficulties of guaranteeing geometric values close to the experimental data, smooth surfaces, and a reasonable anatomical assembly while dealing with data from different studies and dimensions that are connected. This includes the alignment of the vertebral bodies, the vertebral canals, and the facet joints of adjacent vertebrae. Based on the obtained results, we consider that the model is statistically representative of the average of the human thoracolumbar spine.

The mesh-sensitive analysis showed an asymptotic behavior of the ROM and intradiscal pressure with the refinement. Based on these results, the selected mesh of approximately 28000 nodes with quadratic shape functions was considered dense enough for practical purposes and could capture important characteristics such as the morphology of the vertebral endplates. It also keeps the simulations numerically tractable since the clock time for the simulations of one spinal functional unit under pure moment within the anatomical planes took less than 5 min using a computer with an i7 processor and 8 GB of RAM.

The results of the developed FE model agreed with experimental results regarding the 1) ROM of the spine under pure moment in different loading directions, 2) stress-stretch ratio curves from tensile tests of single lamellae of four anatomical regions of the annulus, 3) angle of the fibers for different positions within the IVD, 4) forces at facet joints, 5) ROM of the spine under combined loading. We attribute this achievement due to several aspects. First, the geometric model was generated considering the average anthropometric data. This probably reduced the divergence between the numerical values and the experimental median/average values regarding the biomechanics of the spine presented in the literature. Second, a proper mesh size was chosen based on a mesh sensitive analysis. Third, each spinal structure of the FE model was calibrated considering their individual contribution to the motion of the spine using data from cadaveric stepwise resection studies. This allowed to properly dose the stiffness within the spinal structures. For instance, a satisfactory agreement of the numerical forces at facet joints was possible because the soft tissue was calibrated. A too soft IVD, for example, would increase the stress at the facet joints in axial rotation leading to a divergence of the facet joints forces. Furthermore, a powerful Genetic Algorithm was used to tune 12 parameters representing the material properties of the IVDs including characteristics that are not present in most published FE models such as the regional variance of fibers angles and material properties of the IVD. The algorithm is also able to deal with challenging aspects that can be present during the determination of the material properties of the IVD such as multiple local optima and non-smooth objective functions (Nicolini et al., 2021aNicolini, L.F., Beckmann, A., Laubach, M., Hildebrand, F., Kobbe, P., Roesler, C.R. de M., Fancello, E.A., Markert, B., Stoffel, M., 2021a. An experimental-numerical method for the calibration of Finite Element models of the spine. Submitt. to Med. Eng. Phys.). The Genetic Algorithm was also used to determine the parameters that define the contact of the facet joints. Moreover, the force-displacement curves representing the material properties of the ligaments were unequivocally determined. By varying the material properties of the soft tissue and the contact of the facet joints and choosing the set of material properties that produce the best fit between the numerical and experimental results gave a better agreement with the experimental data rather than directly applying the material properties from the literature (Ezquerro et al., 2011Ezquerro, F., Vacas, F.G., Postigo, S., Prado, M., Simón, A., 2011. Calibration of the finite element model of a lumbar functional spinal unit using an optimization technique based on differential evolution. Med. Eng. Phys. 33, 89–95. https://doi.org/10.1016/j.medengphy.2010.09.010
https://doi.org/10.1016/j.medengphy.2010...
; Nicolini et al., 2021aNicolini, L.F., Beckmann, A., Laubach, M., Hildebrand, F., Kobbe, P., Roesler, C.R. de M., Fancello, E.A., Markert, B., Stoffel, M., 2021a. An experimental-numerical method for the calibration of Finite Element models of the spine. Submitt. to Med. Eng. Phys.; Schmidt et al., 2006Schmidt, H., Heuer, F., Simon, U., Kettler, A., Rohlmann, A., Claes, L., Wilk, H.-J., 2006. Application of a new calibration method for a three-dimensional finite element model of a human lumbar annulus fibrosus. Clin. Biomech. 21, 337–344.).

The numerical results confirmed that cord pretension is a key factor to change the scoliotic angle of the spine. A pretension force of 100 N and 300 N resulted in a change of scoliotic angle of 2.7° and 5.3° to the left side of the spine, respectively. This value is associated with the flexibility/stiffness behavior of the spine and may be predicted based on the moment-ROM curve of the native spine under pure moment as an approximation. The cord was attached 26.2 mm from the middle sagittal plane of the spine. Therefore, a pretension of 100 N produces a moment of 2.6 Nm which also provides an angular displacement of 2.7° according to the moment-ROM of the native spine (Figure 11). In the cadaveric study of Lavelle et al. [24], a value of 9.9±5.5° was found for the case of a T4–T12 continuous sequentially tensioned with 100 N, which means 1.24° per spinal segment when assuming a uniform distribution along the spine. The seventeen patients studied by Newton et al. (2018)Newton, P.O., Kluck, D.G., Saito, W., Yaszay, B., Bartley, C.E., Bastrom, T.P., 2018. Anterior Spinal Growth Tethering for Skeletally Immature Patients with Scoliosis. J. Bone Jt. Surg. 100, 1691–1697. https://doi.org/10.2106/JBJS.18.00287
https://doi.org/10.2106/JBJS.18.00287...
presented an average thoracic curve magnitude of 52 ± 10° preoperatively and 31 ± 10° immediately postoperatively. Considering that there was an average of 5.8 segments tethered per patient and assuming uniform distribution, the change of scoliotic angle was 3.6° per segment. However, Newton et al. (2018)Newton, P.O., Kluck, D.G., Saito, W., Yaszay, B., Bartley, C.E., Bastrom, T.P., 2018. Anterior Spinal Growth Tethering for Skeletally Immature Patients with Scoliosis. J. Bone Jt. Surg. 100, 1691–1697. https://doi.org/10.2106/JBJS.18.00287
https://doi.org/10.2106/JBJS.18.00287...
did not report the value of the tension applied to the cord during surgery.

Primary bending stability in the opposite direction to insertion of VBT is an important requirement for VBT (Nicolini et al., 2021bNicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
https://doi.org/10.1007/s00586-021-07035...
). In right lateral bending, the stiffness generated from cord tension resulted in a reduction of ROM of the spine. The instrumentation with no cord pretension and with a pretension of 300 N were capable of reducing 59% and 69% of the ROM of the native spine, respectively. The relatively small change between the case of no pretension and the case with a cord pretension of 300 N suggests that the cord pretension is capable of changing the neutral position of the spine but does not add much stiffness to the spine in right lateral bending.

The moment-ROM curve of the native spine in left lateral bending was not affected after inserting a cord without pretension since it was modeled with no resistance to compression. However, the implant inserted with pretension added stiffness to the spinal system. The ROM was reduced in left lateral bending mainly because the soft tissue with nonlinear stiffness was tensioned at the right side of the spine after insertion of VBT.

This study also presents some limitations. The effects of VBT were investigated using a FE model representing the average of adult spines, while VBT is usually used in adolescents with scoliosis. This limitation could potentially be overcome by scaling the geometric model and creating scoliosis. However, it is not fully understood the biomechanical effects after such modification and it is not guaranteed that it would provide a geometry that represents an adolescent. Moreover, there are no stepwise resection studies on adolescent scoliotic spines which could provide experimental data for the calibration of the FE model. We opted to use the model of a non-scoliotic spine to allow comparison with the biomechanical experiments with cadaveric spines. Furthermore, the bone parts were modelled as isotropic material (Guan et al., 2006Guan, Y., Yoganandan, N., Zhang, J., Pintar, F.A., Cusick, J.F., Wolfla, C.E., Maiman, D.J., 2006. Validation of a clinical finite element model of the human lumbosacral spine. Medical and Biological Engineering and Computing. Med. Biol. Eng. Comput. 44, 633–641.; Kulduk et al., 2015Kulduk, A., Altun, N.S., Senkoylu, A., 2015. Biomechanical comparison of effects of the Dynesys and Coflex dynamic stabilization systems on range of motion and loading characteristics in the lumbar spine: a finite element study. Int. J. Med. Robot. Comput. Assist. Surg. 11, 400–405. https://doi.org/10.1002/rcs.1636
https://doi.org/10.1002/rcs.1636...
; C.-L. Liu et al., 2011Liu, C.-L., Zhong, Z.-C., Hsu, H.-W., Shih, S.-L., Wang, S.-T., Hung, C., Chen, C.-S., 2011. Effect of the cord pretension of the Dynesys dynamic stabilisation system on the biomechanics of the lumbar spine: a finite element analysis. Eur. Spine J. 20, 1850–1858. https://doi.org/10.1007/s00586-011-1817-3
https://doi.org/10.1007/s00586-011-1817-...
), whereas they are supposed to be anisotropic materials. This simplification would have a minor effect on the results of our study since most of the motion of the spine is defined by the properties of the soft tissue due to the differences in stiffness compared to the bone. However, a more sophisticated modelling might be necessary if one intends to perform an analysis involving the stresses acting in the vertebrae.

4 CONCLUSION

This numerical study evaluated the change of scoliotic angle and range of motion of the spine after insertion of VBT with different cord pretensions. We conclude that the cord pretension substantially influences the resulting scoliotic angle while the stiffness of the tether provides primary stability to the spine for lateral movement in the opposite direction of the implant. The increase of cord pretension leads to a reduction of ROM of the spine in lateral bending. Due to the very satisfactory agreement between the numerical and experimental results (geometry, ROM, stresses in the IVD, facet joint contact forces, etc.), we conclude that the model is calibrated, validated and has the potential be used as a tool to assist on design, testing and evaluation of spinal implants.

Supplementary Material

Supplementary material accompanies this paper.

Table S1 Comparison between experimental and numerical results for the superior and inferior vertebral body width (SVBW and IVBW), superior and inferior vertebral body length (SVBL and IVBL), anterior and posterior vertebral body height (AVBH and PVBH), lateral vertebral body height (LVBH), transverse process length (TPL), vertebral canal superior width and length (VCSW and VCSL), superior and inferior interfacet width (SFFW and IFFW), interfacet height (FFH), superior and inferior facet length (SFL and IFL), facet length (IFL), superior and inferior facet width (SFW and IFW), superior and inferior longitudinal facet angle (SLFA and ILFA) and superior and inferior transverse facet angle (STFA and ITFA). Values are in mm followed by standard deviation. Table S2 Comparison between wedging angle of vertebrae and intervertebral discs of the model with the ones obtained in the literature. Values are in degrees followed by the standard deviation. Positive and negative values indicate lordotic and kyphotic wedging, respectively. Table S3 Comparison of the lumbar lordosis, sacral-slope, pelvic tilt, pelvic incidence of the model with the ones obtained in the literature. Values are in degrees followed by the standard deviation. Table S4 Anterior and posterior height of the intervertebral disc (mm) obtained in the literature and the model. Table S5 Material properties obtained after calibration of the L1-L2 intervertebral disc.

This material is available as part of the online article from https://www.scielo.br/j/lajss

References

  • Ayturka, U.M., Puttlitz, C.M., 2011. Parametric convergence sensitivity and validation of a finite element model of the human lumbar spine. Comput. Methods Biomech. Biomed. Engin. 14, 695–705.
  • Beckmann, A., Herren, C., Mundt, M., Siewe, J., Kobbe, P., Sobottke, R., Pape, H.-C., Stoffel, M., Markert, B., 2018. A new in vitro spine test rig to track multiple vertebral motions under physiological conditions. Biomed. Eng. (NY). 63, 341–347. https://doi.org/10.1515/bmt-2016-0173
    » https://doi.org/10.1515/bmt-2016-0173
  • Beckmann, A., Nicolini, L.F., Grevenstein, D., Backes, H., Oikonomidis, S., Sobottke, R., Kobbe, P., Hildebrand, F., Stoffel, M., Markert, B., Siewe, J., Herren, C., 2020. Biomechanical in vitro test of a novel dynamic spinal stabilization system incorporating polycarbonate urethane material under physiological conditions. J. Biomech. Eng. 142. https://doi.org/10.1115/1.4044242
    » https://doi.org/10.1115/1.4044242
  • Been, E., Barash, A., Marom, A., Kramer, P.A., 2010. Vertebral Bodies or Discs: Which Contributes More to Human-like Lumbar Lordosis? Clin. Orthop. Relat. Res. 468, 1822–1829.
  • Been, E., Li, L., Hunter, D.J., Kalichman, L., 2011. Geometry of the vertebral bodies and the intervertebral discs in lumbar segments adjacent to spondylolysis and spondylolisthesis: pilot study. Eur. Spine J. 20, 1159–1165. https://doi.org/10.1007/s00586-010-1660-y
    » https://doi.org/10.1007/s00586-010-1660-y
  • Boulay, C., Tardieu, C., Hecquet, J., Benaim, C., Mouilleseaux, B., Marty, C., Prat-Pradal, D., Legaye, J., Duval-Beaupère, G., Pélissier, J., 2006. Sagittal alignment of spine and pelvis regulated by pelvic incidence: standard values and prediction of lordosis. Eur. Spine J. 15, 415–422.
  • Brandes, L.L., Nicolini, L.F., Greven, J., Lichte, P., Stopinski, T.T., Sattler, M., Hildebrand, F., Pishnamaz, M., 2021. Biomechanical Performance of BoneHelix® Compared with Elastic Stable Intramedullary Nailing (ESIN) in a Pediatric Tibia Fracture Model. Life 11, 1189. https://doi.org/10.3390/life11111189
    » https://doi.org/10.3390/life11111189
  • Burkhart, T.A., Andrews, D.M., Dunning, C.E., 2013. Finite element modeling mesh quality, energy balance and validation methods: A review with recommendations associated with the modeling of bone tissue. J. Biomech. 46, 1477–1488. https://doi.org/10.1016/j.jbiomech.2013.03.022
    » https://doi.org/10.1016/j.jbiomech.2013.03.022
  • Buyuk, A.F., Milbrandt, T.A., Mathew, S.E., Potter, D.D., Larson, A.N., 2021. Does preoperative and intraoperative imaging for anterior vertebral body tethering predict postoperative correction? Spine Deform. 9, 743–750. https://doi.org/10.1007/s43390-020-00267-2
    » https://doi.org/10.1007/s43390-020-00267-2
  • Cassidy, J.J., Hiltner, A., Baer, E., 1989. Hierarchical Structure of the Intervertebral Disc. Connect. Tissue Res. 23, 75–88. https://doi.org/10.3109/03008208909103905
    » https://doi.org/10.3109/03008208909103905
  • Cheng, X.G., Sun, Y., Boonen, S., Nicholson, P.H.F., Brys, P., Dequeker, J., Felsenberg, D., 1998. Measurements of vertebral shape by radiographic morphometry: sex differences and relationships with vertebral level and lumbar lordosis. Skeletal Radiol. 27, 380–384. https://doi.org/10.1007/s002560050402
    » https://doi.org/10.1007/s002560050402
  • Coombs, D., Bushelow, M., Laz, P., Rao, M., Rullkoetter, P., 2013. Stepwise Validated Finite Element Model of the Human Lumbar Spine, in: ASME 2013 Conference on Frontiers in Medical Devices: Applications of Computer Modeling and Simulation. American Society of Mechanical Engineers, pp. 1–2. https://doi.org/10.1115/FMD2013-16167
    » https://doi.org/10.1115/FMD2013-16167
  • Cramer, G.D., Darby, S.A., 2014. Clinical Anatomy of the Spine, Spinal Cord and ANS, 3rd ed. Mosby, St. Louis.
  • Crawford, C.H., Lenke, L.G., 2010. Growth Modulation by Means of Anterior Tethering Resulting in Progressive Correction of Juvenile Idiopathic Scoliosis. J. Bone Jt. Surgery-American Vol. 92, 202–209. https://doi.org/10.2106/JBJS.H.01728
    » https://doi.org/10.2106/JBJS.H.01728
  • Damm, N., Rockenfeller, R., Gruber, K., 2020. Lumbar spinal ligament characteristics extracted from stepwise reduction experiments allow for preciser modeling than literature data. Biomech. Model. Mechanobiol. 19, 893–910. https://doi.org/10.1007/s10237-019-01259-6
    » https://doi.org/10.1007/s10237-019-01259-6
  • Dreischarf, M., Zander, T., Shirazi-Adl, A., Puttlitz, C.M., Adam, C.J., Chen, C.S., Goel, V.K., Kiapour, A., Kim, Y.H., Labus, K.M., Little, J.P., Park, W.M., Wang, Y.H., Wilke, H.J., Rohlmann, A., Schmidt, H., 2014. Comparison of eight published static finite element models of the intact lumbar spine: Predictive power of models improves when combined together. J. Biomech. 47, 1757–1766. https://doi.org/10.1016/j.jbiomech.2014.04.002
    » https://doi.org/10.1016/j.jbiomech.2014.04.002
  • Ebara, S., AIatridis, J.C., Setton, L.A., Foster, R.J., Van Mow, C., Weidenbaum, M., 1996. Tensile properties of nondegenerate human lumbar anulus fibrosus. Spine (Phila. Pa. 1976). 21, 452–461.
  • Eberlein, R., Holzapfe, G.A., Schulze-Bauer, C.A.J., 2001. An Anisotropic Model for Annulus Tissue and Enhanced Finite Element Analyses of Intact Lumbar Disc Bodies. Comput. Methods Biomech. Biomed. Engin. 4, 209–229. https://doi.org/10.1080/10255840108908005
    » https://doi.org/10.1080/10255840108908005
  • Eijkelkamp, M.F., Klein, J.P., Veldhuizen, A.G., Horn, J.R. van, Verkerke, G.J., 2007. The geometry and shape of the human intervertebral disc, in: Spine [Submitted for Publication]. The Spine, pp. 75–83.
  • Ezquerro, F., Vacas, F.G., Postigo, S., Prado, M., Simón, A., 2011. Calibration of the finite element model of a lumbar functional spinal unit using an optimization technique based on differential evolution. Med. Eng. Phys. 33, 89–95. https://doi.org/10.1016/j.medengphy.2010.09.010
    » https://doi.org/10.1016/j.medengphy.2010.09.010
  • Guan, Y., Yoganandan, N., Zhang, J., Pintar, F.A., Cusick, J.F., Wolfla, C.E., Maiman, D.J., 2006. Validation of a clinical finite element model of the human lumbosacral spine. Medical and Biological Engineering and Computing. Med. Biol. Eng. Comput. 44, 633–641.
  • Guerin, H.A.L., Elliott, D.M., 2006. Degeneration affects the fiber reorientation of human annulus fibrosus under tensile load. J. Biomech. 39, 1410–1418. https://doi.org/10.1016/j.jbiomech.2005.04.007
    » https://doi.org/10.1016/j.jbiomech.2005.04.007
  • Heuer, F., Schmidt, H., Claes, L., Wilke, H.-J., 2007a. Stepwise reduction of functional spinal structures increase vertebral translation and intradiscal pressure. J. Biomech. 40, 795–803.
  • Heuer, F., Schmidt, H., Klezl, Z., Claes, L., Wilke, H.-J., 2007b. Stepwise reduction of functional spinal structures increase range of motion and change lordosis angle. J. Biomech. 40, 271–280. https://doi.org/10.1016/j.jbiomech.2006.01.007
    » https://doi.org/10.1016/j.jbiomech.2006.01.007
  • Holzapfel, G.A., Gasser, T.C., Ogden, R.W., 2000. A new constitutive framework for arterial wall mechanics and a comparative study of material models. J. Elast. Phys. Sci. solids 61, 1–48. https://doi.org/10.1023/A:1010835316564
    » https://doi.org/10.1023/A:1010835316564
  • Holzapfel, G.A., Schulze-Bauer, C.A.J., Feigl, G., Regitnig, P., 2005. Single lamellar mechanics of the human lumbar anulus fibrosus. Biomech. Model. Mechanobiol. 3, 125–140. https://doi.org/10.1007/s10237-004-0053-8
    » https://doi.org/10.1007/s10237-004-0053-8
  • Jaramillo, H.E., Puttlitz, C.M., McGilvray, K., García, J.J., 2016. Characterization of the L4–L5–S1 motion segment using the stepwise reduction method. J. Biomech. 49, 1248–1254. https://doi.org/10.1016/j.jbiomech.2016.02.050
    » https://doi.org/10.1016/j.jbiomech.2016.02.050
  • Kulduk, A., Altun, N.S., Senkoylu, A., 2015. Biomechanical comparison of effects of the Dynesys and Coflex dynamic stabilization systems on range of motion and loading characteristics in the lumbar spine: a finite element study. Int. J. Med. Robot. Comput. Assist. Surg. 11, 400–405. https://doi.org/10.1002/rcs.1636
    » https://doi.org/10.1002/rcs.1636
  • Labelle, H., Roussouly, P., Berthonnaud, É., Dimnet, J., O’Brien, M., 2005. The Importance of Spino-Pelvic Balance in L5–S1 Developmental Spondylolisthesis: A Review of Pertinent Radiologic Measurements. Spine (Phila. Pa. 1976). 30, S27–S34. https://doi.org/10.1097/01.brs.0000155560.92580.90
    » https://doi.org/10.1097/01.brs.0000155560.92580.90
  • Lavelle, W.F., Moldavsky, M., Cai, Y., Ordway, N.R., Bucklen, B.S., 2016. An initial biomechanical investigation of fusionless anterior tether constructs for controlled scoliosis correction. Spine J. 16, 408–413. https://doi.org/10.1016/j.spinee.2015.11.004
    » https://doi.org/10.1016/j.spinee.2015.11.004
  • Li, Y., Shen, Z., Huang, M., Wang, X., 2017. Stepwise resection of the posterior ligamentous complex for stability of a thoracolumbar compression fracture. Medicine (Baltimore). 96, e7873. https://doi.org/10.1097/MD.0000000000007873
    » https://doi.org/10.1097/MD.0000000000007873
  • Liu, C.-L., Zhong, Z.-C., Hsu, H.-W., Shih, S.-L., Wang, S.-T., Hung, C., Chen, C.-S., 2011. Effect of the cord pretension of the Dynesys dynamic stabilisation system on the biomechanics of the lumbar spine: a finite element analysis. Eur. Spine J. 20, 1850–1858. https://doi.org/10.1007/s00586-011-1817-3
    » https://doi.org/10.1007/s00586-011-1817-3
  • Lo, C.-C., Tsai, K.-J., Zhong, Z.-C., Chen, S.-H., Hung, C., 2011. Biomechanical differences of Coflex-F and pedicle screw fixation combined with TLIF or ALIF – a finite element study. Comput. Methods Biomech. Biomed. Engin. 14, 947–956. https://doi.org/10.1080/10255842.2010.501762
    » https://doi.org/10.1080/10255842.2010.501762
  • Masharawi, Y., Dar, G., Peleg, S., Steinberg, N., Medlej, B., May, H., Abbas, J., Hershkovitz, I., 2010. A morphological adaptation of the thoracic and lumbar vertebrae to lumbar hyperlordosis in young and adult females. Eur. Spine J. 19, 768–773. https://doi.org/10.1007/s00586-009-1256-6
    » https://doi.org/10.1007/s00586-009-1256-6
  • Masharawi, Y., Rothschild, B., Dar, G., Peleg, S., Robinson, D., Been, E., Hershkovitz, I., 2004. Facet orientation in the thoracolumbar spine: three-dimensional anatomic and biomechanical analysis. Spine (Phila. Pa. 1976). 29, 1755–1763. https://doi.org/00007632-200408150-00009 [pii]
    » https://doi.org/00007632-200408150-00009
  • Masharawi, Y., Rothschild, B., Salame, K., Dar, G., Peleg, S., Hershkovitz, I., 2005. Facet tropism and interfacet shape in the thoracolumbar vertebrae: characterization and biomechanical interpretation. Spine (Phila. Pa. 1976). 30, E281--E292. https://doi.org/10.1097/01.brs.0000164098.00201.8d
    » https://doi.org/10.1097/01.brs.0000164098.00201.8d
  • Masharawi, Y., Salame, K., 2011. Shape variation of the neural arch in the thoracic and lumbar spine: Characterization and relationship with the vertebral body shape. Clin. Anat. 24, 858–867. https://doi.org/10.1002/ca.21175
    » https://doi.org/10.1002/ca.21175
  • Miyanji, F., Pawelek, J., Nasto, L.A., Simmonds, A., Parent, S., 2020. Safety and efficacy of anterior vertebral body tethering in the treatment of idiopathic scoliosis. Bone Jt. J. 102-B, 1703–1708. https://doi.org/10.1302/0301-620X.102B12.BJJ-2020-0426.R1
    » https://doi.org/10.1302/0301-620X.102B12.BJJ-2020-0426.R1
  • Newton, P.O., Kluck, D.G., Saito, W., Yaszay, B., Bartley, C.E., Bastrom, T.P., 2018. Anterior Spinal Growth Tethering for Skeletally Immature Patients with Scoliosis. J. Bone Jt. Surg. 100, 1691–1697. https://doi.org/10.2106/JBJS.18.00287
    » https://doi.org/10.2106/JBJS.18.00287
  • Nicolini, L.F., Beckmann, A., Laubach, M., Hildebrand, F., Kobbe, P., Roesler, C.R. de M., Fancello, E.A., Markert, B., Stoffel, M., 2021a. An experimental-numerical method for the calibration of Finite Element models of the spine. Submitt. to Med. Eng. Phys.
  • Nicolini, L.F., Kobbe, P., Seggewiß, J., Greven, J., Ribeiro, M., Beckmann, A., Da Paz, S., Eschweiler, J., Prescher, A., Markert, B., Stoffel, M., Hildebrand, F., Trobisch, P.D., 2021b. Motion preservation surgery for scoliosis with a vertebral body tethering system: a biomechanical study. Eur. Spine J. https://doi.org/10.1007/s00586-021-07035-4
    » https://doi.org/10.1007/s00586-021-07035-4
  • Niemeyer, F., Wilke, H.J., Schmidt, H., 2012. Geometry strongly influences the response of numerical models of the lumbar spine: a probabilistic finite element analysis. J. Biomech. 45, 1414–1423. https://doi.org/10.1016/j.jbiomech.2012.02.021
    » https://doi.org/10.1016/j.jbiomech.2012.02.021
  • Niinomi, M., Liu, Y., Nakai, M., Liu, H., Li, H., 2016. Biomedical titanium alloys with Young’s moduli close to that of cortical bone. Regen. Biomater. 3, 173–185. https://doi.org/10.1093/rb/rbw016
    » https://doi.org/10.1093/rb/rbw016
  • Niosi, C.A., Wilson, D.C., Zhu, Q., Keynan, O., Wilson, D.R., Oxland, T.R., 2008. The effect of dynamic posterior stabilization on facet joint contact forces: an in vitro investigation. Spine (Phila. Pa. 1976). 33, 19–26. https://doi.org/10.1097/BRS.0b013e31815e7f76
    » https://doi.org/10.1097/BRS.0b013e31815e7f76
  • Nissan, M., Issachar, G., 1986. Dimensions of human lumbar vertebrae in the sagittal plane. J. Biomech. 19, 753–758.
  • Noailly, J., Wilke, H.J., Planell, J.A., Lacroix, D., 2007. How does the geometry affect the internal biomechanics of a lumbar spine bi-segment finite element model? Consequences on the validation process. J. Biomech. 40, 2414–2425. https://doi.org/10.1016/j.jbiomech.2006.11.021
    » https://doi.org/10.1016/j.jbiomech.2006.11.021
  • Pehlivanoglu, T., Oltulu, I., Ofluoglu, E., Sarioglu, E., Altun, G., Korkmaz, M., Yildirim, K., Aydogan, M., 2020. Thoracoscopic Vertebral Body Tethering for Adolescent Idiopathic Scoliosis: A Minimum of 2 Years’ Results of 21 Patients. J. Pediatr. Orthop. 40, 575–580. https://doi.org/10.1097/BPO.0000000000001590
    » https://doi.org/10.1097/BPO.0000000000001590
  • Ritzel, H., Amling, M., Pösl, M., Hahn, M., Delling, G., 1997. The thickness of human vertebral cortical bone and its changes in aging and osteoporosis: a histomorphometric analysis of the complete spinal column from thirty-seven autopsy specimens. J. Bone Miner. Res. 12. https://doi.org/10.1359/jbmr.1997.12.1.89
    » https://doi.org/10.1359/jbmr.1997.12.1.89
  • Roussouly, P., Sohrab, G., Eric, B., Johanes, D., 2005. Classification of the normal variation in the sagittal alignment of the human lumbar spine and pelvis in the standing position. Spine (Phila. Pa. 1976). 30, 346–53.
  • Samdani, A.F., Ames, R.J., Kimball, J.S., Pahys, J.M., Grewal, H., Pelletier, G.J., Betz, R.R., 2015. Anterior vertebral body tethering for immature adolescent idiopathic scoliosis: one-year results on the first 32 patients. Eur. Spine J. 24, 1533–1539. https://doi.org/10.1007/s00586-014-3706-z
    » https://doi.org/10.1007/s00586-014-3706-z
  • Šarčević, Z., Tepavčević, A., 2019. Association between adolescent idiopathic scoliosis and sacroiliac joint dysfunction in young athletes. Medicine (Baltimore). 98, e15161. https://doi.org/10.1097/MD.0000000000015161
    » https://doi.org/10.1097/MD.0000000000015161
  • Schmidt, H., Heuer, F., Drumm, J., Klezl, Z., Claes, L., Wilke, H.-J., 2007. Application of a calibration method provides more realistic results for a finite element model of a lumbar spinal segment. Clin. Biomech. 22, 377–84. https://doi.org/10.1016/j.clinbiomech.2006.11.008
    » https://doi.org/10.1016/j.clinbiomech.2006.11.008
  • Schmidt, H., Heuer, F., Simon, U., Kettler, A., Rohlmann, A., Claes, L., Wilk, H.-J., 2006. Application of a new calibration method for a three-dimensional finite element model of a human lumbar annulus fibrosus. Clin. Biomech. 21, 337–344.
  • Shirazi-Adl, A., Ahmed, A.M., Shrivastava, S.C., 1986. A finite element study of a lumbar motion segment subjected to pure sagittal plane moments. J. Biomech. 19, 331–350. https://doi.org/10.1016/0021-9290(86)90009-6
    » https://doi.org/10.1016/0021-9290(86)90009-6
  • Stokes, I.A.F., Spence, H., Aronsson, D.D., Kilmer, N., 1996. Mechanical Modulation of Vertebral Body Growth. Spine (Phila. Pa. 1976). 21, 1162–1167. https://doi.org/10.1097/00007632-199605150-00007
    » https://doi.org/10.1097/00007632-199605150-00007
  • van der Houwen, E.B., Baron, P., Veldhuizen, A.G., Burgerhof, J.G.M., van Ooijen, P.M.A., Verkerke, G.J., 2010. Geometry of the Intervertebral Volume and Vertebral Endplates of the Human Spine. Ann. Biomed. Eng. 38, 33–40. https://doi.org/10.1007/s10439-009-9827-6
    » https://doi.org/10.1007/s10439-009-9827-6
  • Vialle, R., Levassor, N., Rillardon, L., Templier, A., Skalli, W., Guigui, P., 2005. Radiographic analysis of the sagittal alignment and balance of the spine in asymptomatic subjects. J. bone Jt. Surg. 87, 260–267. https://doi.org/10.2106/JBJS.D.02043
    » https://doi.org/10.2106/JBJS.D.02043
  • Weisse, B., Aiyangar, A.K., Affolter, C., Gander, R., Terrasi, G.P., Ploeg, H., 2012. Determination of the translational and rotational stiffnesses of an L4 – L5 functional spinal unit using a specimen-specific finite element model. J. Mech. Behav. Biomed. Mater. 13, 45–61. https://doi.org/10.1016/j.jmbbm.2012.04.002
    » https://doi.org/10.1016/j.jmbbm.2012.04.002
  • White, A.A., Panjabi, M.M., 1990. Clinical Biomechanics of the Spine, 2nd ed. J. B. Lippincott Company, Philadelphia.
  • Wilmanns, N., Beckmann, A., Nicolini, L.F., Herren, C., Sobottke, R., Hildebrand, F., Siewe, J., Kobbe, P., Markert, B., Stoffel, M., 2021. Biomechanical in Vitro and Finite Element Study On Different Sagittal Alignment Postures of the Lumbar Spine During Multiaxial Daily Motion. J. Biomech. Eng. 144, 9. https://doi.org/10.1115/1.4053083
    » https://doi.org/10.1115/1.4053083
  • Zhang, H., Muzumdar, A., Bucklen, B., Hussain, M., Khalil, S., 2011. Effects of stepwise transection of spinal vomponents in lumbar spine: an experimental study of one L1-S1 cadaver sample, in: Poster Session Presented at: 57th Annual Meeting of the Orthopaedic Research Society. Long Beach (CA), p. Jan 13 – 16.
  • Zhu, D., Gu, G., Wu, W., Gong, H., Zhu, W., Jiang, T., Cao, Z., 2008. Micro-structure and mechanical properties of annulus fibrous of the L4-5 and L5-S1 intervertebral discs. Clin. Biomech. 23, S74-82.