Evaluation of indirect methods of digitization of cephalometric radiographs in comparison with the direct digital method

Abstracts

OBJECTIVE: To evaluate the indirect digitization method of cephalometric radiographs in comparison with the direct digital method. METHODS: The sample was composed of ten cephalometric radiographs acquired by Orthopantomograph OP100/Orthocef OC100 (GE - Instrumentarium), digital direct. In the Adobe™ Photoshop program, five cephalometric landmarks were set in the images and the impression in transparencies was made. The indirect digitization of the images was performed through the Sony™ DSC-W5 and Canon™ Rebel XT/EOS 350D digital photographic cameras-fixed in a copy stand, at the distances of 25 cm and 60 cm-and through the Hewlett Packard™ Scan Jet 4C scanner. The direct digital images and the indirect ones were inserted and gauged in the Radiocef Studio (Radiomemory™, Brazil) software and the center of the previously marked landmarks was set. The cephalometric computerized analysis generated three angular measurements and four linear ones which were submitted to statistical analysis. RESULTS: The images from the scanner demonstrated small statistically significant alterations, without clinical significance. When digitizing the radiographs at 60 cm, both cameras caused distortions which were statistically significant, but clinically acceptable. At 25 cm, the cameras caused the largest distortions, being more expressive and with clinical significance in the images of Canon™ Rebel XT. CONCLUSIONS: The Hewlett Packard™ Scan Jet 4C scanner with transparency reader and the Sony™ DSC-W5 and Canon™ Rebel XT/EOS cameras operating at 60 cm were shown appropriate for the digitization of cephalometric radiographs. In 25 cm, the digital cameras caused distortions in the image which altered the linear measurements with possibilities of jeopardizing the orthodontic diagnosis.

Digital dental radiography; Orthodontics; Radiographic image interpretation; Computer-assisted cephalometrics


OBJETIVO: avaliar métodos de digitalização indireta de radiografias cefalométricas em comparação ao método digital direto. MÉTODOS: a amostra foi composta de dez radiografias cefalométricas adquiridas pelo Orthopantomograph OP100/Orthocef OC100 (GE - Instrumentarium), digital direto. As imagens foram inseridas no programa Adobe Photoshop® e cinco pontos cefalométricos foram marcados. Procedeu-se a impressão em transparência e foi feita a digitalização indireta das imagens impressas, por meio das câmeras fotográficas digitais Sony® DSC-W5 e Canon® Rebel XT/EOS 350D, fixas em estativa, nas distâncias de 25cm e 60cm e com o scanner Scan Jet 4C Hewlett Packard®. As imagens diretas e indiretas foram inseridas e calibradas no software Radiocef Studio (Radiomemory®, Brasil), marcando-se o centro dos pontos previamente assinalados. A análise cefalométrica computadorizada gerou três grandezas angulares e quatro lineares, as quais foram submetidas a tratamento estatístico. RESULTADOS: as imagens do scanner Scan Jet 4C HP® demonstraram pequenas alterações significativas, porém, sem relevância clínica. Quando da digitalização a 60cm da radiografia, as duas câmeras causaram distorções que foram significativas mas clinicamente aceitáveis; e, a 25cm, causaram as maiores distorções, com importância clínica, na Canon® Rebel XT. CONCLUSÃO: o scanner Scan Jet 4C HP® com leitor de transparências foi o melhor método e as câmeras operando a 60cm mostraram-se adequadas para digitalização de radiografias. As câmeras na distância de 25cm causaram distorções na imagem que alteraram as medidas lineares, podendo a Canon® Rebel XT comprometer o diagnóstico ortodôntico.

Radiografia digital; Ortodontia; Interpretação radiográfica; Cefalometria computadorizada


ORIGINAL ARTICLE

IMSc in Dental Radiology, São Leopoldo Mandic Dental Research Center, Campinas/SP, Brazil. Post-graduate student in Health Sciences, Federal University of Goiás, Goiânia, GO, Brazil. Professor of Radiology, Brazilian Dental Association, Brasília, DF, Brazil

IIPhD in Oral Diagnosis, University of São Paulo, Brazil. Professor of Oral Diagnosis, University of Pernambuco, Recife, PE, Brazil

IIIPhD in Oral Diagnosis, University of São Paulo, Brazil. Professor of Radiology, São Leopoldo Mandic Dental Research Center, Campinas/SP, Brazil

IVMSc in Dental Radiology, São Leopoldo Mandic Dental Research Center, Campinas/SP, Brazil

VSpecialist in Orthodontics and Facial Orthopedics, Brazilian Dental Association, Uberlândia/MG, Brazil

VIPhD in Endodontics, University of São Paulo, Brazil. Chairman and Professor of Endodontics, Federal University of Goiás

Contact address

ABSTRACT

OBJECTIVE: To evaluate the indirect digitization method of cephalometric radiographs in comparison with the direct digital method.

METHODS: The sample was composed of ten cephalometric radiographs acquired by Orthopantomograph OP100/Orthocef OC100 (GE - Instrumentarium), digital direct. In the Adobe™ Photoshop program, five cephalometric landmarks were set in the images and the impression in transparencies was made. The indirect digitization of the images was performed through the Sony™ DSC-W5 and Canon™ Rebel XT/EOS 350D digital photographic cameras-fixed in a copy stand, at the distances of 25 cm and 60 cm-and through the Hewlett Packard™ Scan Jet 4C scanner. The direct digital images and the indirect ones were inserted and gauged in the Radiocef Studio (Radiomemory™, Brazil) software and the center of the previously marked landmarks was set. The cephalometric computerized analysis generated three angular measurements and four linear ones which were submitted to statistical analysis.

RESULTS: The images from the scanner demonstrated small statistically significant alterations, without clinical significance. When digitizing the radiographs at 60 cm, both cameras caused distortions which were statistically significant, but clinically acceptable. At 25 cm, the cameras caused the largest distortions, being more expressive and with clinical significance in the images of Canon™ Rebel XT.

CONCLUSIONS: The Hewlett Packard™ Scan Jet 4C scanner with transparency reader and the Sony™ DSC-W5 and Canon™ Rebel XT/EOS cameras operating at 60 cm were shown appropriate for the digitization of cephalometric radiographs. In 25 cm, the digital cameras caused distortions in the image which altered the linear measurements with possibilities of jeopardizing the orthodontic diagnosis.

Keywords: Digital dental radiography. Orthodontics. Radiographic image interpretation. Computer-assisted cephalometrics.

INTRODUCTION

The direct digital X-ray has become an alternative to the conventional X-ray due to the possibility of image manipulation, radiation dose reduction to the patient, better filing and information access.1,2 However, in many radiological clinics this is not a reality yet and the traditional film continues to be the image receiver. In computerized cephalometry, the information of the radiographic image must be introduced in the software, through the direct acquisition or by the indirect digitization of the radiographs. The indirect digitization was initially made by plotting in digitizer tables3 and later by video cameras.4 Since 1993 it was observed that the conventional radiographic images could be converted into a digital sign by using a high resolution scanner,5 and then it became recommended by the manufacturers of cephalometry softwares. It is very similar to a Xerox machine and is available in three types: laser, rotating drum and flatbed.6 They are all endowed with light source in line shape that scans the image by measuring the amount of reflected or transmitted light in each dot. The captured light is turned into an electric sign, with the aid of photodetector groups which also form a line, and the electric sign is digitized and sent to the computer. The flatbed scanner was shown sensitive to the scanning arrangements, power state and image locations, while such inconsistencies were not observed in the rotating drum scanner (VXR-12),6 what can be explained by its design. However, the space resolution, geometric distortion and CCD (charge coupled device) structure interference of this scanner require further studies.6 When comparing the manual cephalometry to the computerized one, by using the VXR-12 scanner as a digitizer, there were statistically significant amplifications in both linear and angular measurements, although 21 out of the 27 presented differences smaller than 2 degrees or 2 mm, what is within of the norms of most of the cephalometric analyses and therefore without clinical significance.7 Another researchers scanned thirty lateral cephalograms in 300 dpi, and the findings demonstrated that the use of computer software for cephalometric analysis carried out on scanned images does not increase the measurement error when compared with manual tracing.8 High quality image equipment are very expensive and this is an unfavorable factor that added to the time consumption to digitize radiographs into a scanner has been motivating clinicians to use digital photographic cameras, with the aim of replacing the scanner. However, the literature in such field is scarce, leading to lack of standardization and consequent unreliability of the results.

The digitization systems based on cameras, unlike the scanning systems, present lower reproducibility because they require position and zoom adjustment.6 The lens of the camera usually focus the light into a plan behind it, and in conventional cameras such plan contains the photographic film. Nevertheless, in the digital cameras it is replaced by a sensor which captures luminous pulses and transforms them into electric pulses which are converted into digital image; in other words, the sensor generates the pixels.9 Not only the size of the pixel is important, but also the size of the sensor, because the larger the area to absorb light the better the final image. Most of the popular cameras use 1/1, 8-in or 2/3-in sensors.9 The sensor can be CCD (charge coupled device) or CMOS (complementary metal oxide semiconductor). As the lenses of the cameras are not plane, the digitized images may present distortions in barrel or pincushion shapes.10 In the former the images seem to be inflated, and take place where the focal distance is smaller, as for the latter there is a compression of the image in its own center and it is registered in larger focal distances. Such distortions are more visible in the images with perfectly straight lines, mainly when they are close to the edge,10 therefore also visualized in the radiographs digitized by the cameras. When comparing the manual cephalometry to the computerized one, by using a Pulnix TM-760 (512 x 512 pixels resolution) digital camera as a digitizer, it was observed that the calibration of the digital image produces tiny significant errors associated to the angular and linear measurements and to the demarcation of the cephalometric landmarks, which tend to be larger in the digital images than in the conventional ones; and that the space resolution of the digital image is lower than in the conventional X-ray.2 To digitize radiographs with digital photographic cameras a light box is recommended with a high frequency fluorescent lamp and intensity enough for the films to be clearer and sharpen for the sensor to accomplish the acquiring. The further the camera is from the light box using the zoom to frame the X-ray, the better it will be for the sharpness of the focus, the depth of the field and the homogeneity of the lighting.11 For radiographs presenting clear center and dark extremities, a dark mask is used around it to compensate the automatic exposure which is focused in the center of the image.11 When testing an amateur digital camera to digitize images of forty bone trauma films and to transmit them through Telemedicine net to be assessed by specialists, it was observed that there was no significant difference in the diagnostic precision between the conventional film and digital image, as well as the quality of the image, which was classified as excellent.12 The direct digital radiograph obtained by storage phosphor technique has already been recognized as reliable in computerized cephalometry, when compared to the manual cephalometry in conventional radiographic films.13,14

This work aimed to evaluate the reliability of using-in computerized cephalometric studies-the indirect digitization of lateral cephalometric radiograph by means of two models of digital photographic cameras, as well as an flatbed scanner with a transparency reader, in comparison with direct digital radiograph obtained in CCD.

MATERIAL AND METHODS

Ten lateral cephalometric radiographs from the Orthopantomograph OP100/Orthocef OC100 (GE - Instrumentarium) digital direct unit were randomly selected; the files belonged to São Leopoldo Mandic Post-graduations Center. In the Adobe™ Photoshop program, using the Paintbrush tool, five cephalometric landmarks were set: N = nasion, S = sella, Ar = articulare, Go = gonion and Me = menton. The radiographs were printed, in 100% size, in 3M™ transparencies with a HP™ Laser Jet 1320 printer and indirectly digitized by HP™ Scan Jet 4C scanner with transparency reader, in 75 dpi, and by the following digital photographic cameras: Sony™ DSC W5 (5.1 megapixels) and Canon™ Rebel XT/EOS 350D (8.0 megapixels, 55 mm lens). The cameras were fastened in a Incaf™ copy standy (Fig 1), perpendicularly, 25 cm and 60 cm away from the printed radiographs, positioned with a Desetec™ millimeter ruler on a negatoscope (four Osram™ Dulux F 36W/21-840 fluorescent lamps) in an semi-dark room. The cameras operated in the automatic mode, without flash and using optical zoom until the image filled out the entire camera visor. The distance and parallelism of the cameras and radiographs were verified with a Starret™ measure tape and a Tramontina™ level. All the direct and indirect digital images were inserted into the Radiocef Studio (Radiomemory™, Brazil) cephalometry software. For the calibration of the direct digital images, acquired in 350 dpi, the upper and lower borders of the image were marked, previously measured in the Adobe Photoshop™ (16 cm). For the indirect digital images of the cameras the numbers zero and sixteen were marked in the photographed ruler; and the images of the scanner, in 75 dpi, were not gauged because this is the standard resolution of the software. An experienced radiologist, using the zoom tool of the software, marked the center of the landmarks, previously located in the Photoshop. The cephalometric analysis, created by planes with distant landmarks distributed in the center and in the periphery of the image, generated four linear measurements and three angular ones (Fig 2). The effects of variables Camera and Distance were statistically assessed through the variance analysis with repeated measures and the comparisons between mean pairs by Student t test for paired samples. The significance level of p < 0.05 was adopted for all tests.



RESULTS

Table 1 shows the mean values obtained from the measurements performed in the direct digital radiographs as well as in the indirect ones from the scanner and from the combinations of cameras and distances. The mean values of the combinations marked with an asterisk showed statistically significant differences at the level of 5% in relation to the correspondent mean obtained by the direct digital procedure. Figure 3 displays the distortions produced by the evaluated methods: In (A) direct digital image, without distortions; in (B) image digitized by the HP™ Scan Jet 4C scanner, without perceptible distortion in a visual observation; In (C, D, E and F) the images of cameras, in both distances, with the presence of distortions in the borders. It can be observed that within the distance of 60 cm (C, D) the images were less altered; while within 25 cm (E, F) there was larger distortion, mainly in the images digitized by Canon™ Rebel XT (E).


Table 2 shows the results of the variance analysis of repeated measures, indicating the measurements in which the values were significantly influenced by the distance and by the camera type. The angular measurement (Ar-Go).Me and the linear ones S-N and S-Go suffered statistically significant alterations (p <0.05) according to the camera type, as well as the angular (S-N).Ar and the linear ones S-Go and N-Me according to the distance (p <0.05). Table 3 shows the mean values of the measured variables, according to the camera type and distance. Table 4 demonstrates the influence of the camera type and distance on the measurements, confirming the existence of statistically significant alterations (p <5%) for all images when the distance is changed. S-N horizontal line was the most affected (p = 0.006), followed by the vertical one N-Me (p = 0.004). The larger mean difference among the two distances was seen in Canon™ Rebel camera (2.99 mm) in the vertical measurement S-Go, which was located towards the center of the lens.

DISCUSSION

The lateral cephalometric radiographs allow us to quantify facial and dental relationships,15 by the comparison of the cephalometric measurements obtained from populational samples, making it possible to evaluate the extension of morphologic deviations in relation to normality, as well as to scrutinize such measurements in relation to the morphologic characteristics of an individual. Among the several auxiliary instruments used to enhance orthodontic diagnosis, the cephalometric analyses are indeed valuable. Nevertheless, they are subject to erroneous and mistaken interpretations, in function of the necessary registrations to obtain them. In the computerized cephalometry, besides the mistakes committed in the conventional method, there is also the possibility of two other problems: the identification of the cephalometric landmarks in function of the loss of quality of the images2,16 and mistakes caused by calibration.2 To identify these was not the aim of this study, which had just the intention of assess the alterations in the measurements performed in the images digitized by the digital photographic cameras and by scanner, considering that the latter is globally accepted, while the cameras are still little investigated. With the aim of not making location or demarcation mistakes, the cephalometric landmarks were previously marked in the direct digital images and confirmed in the cephalometry software on the images digitized with the scanner and the cameras.

Tradition imposes us to use angular and linear measurements for the evaluation of the structures of the craniofacial frame,17,18 however, individual measures are insignificant if they are not correctly interpreted in a global context. In a general way, linear measurements are more reliable than angular ones, and the latter can be influenced by the former ones. For instance: an increased or reduced length of the cranial base (S-N) may alter the (S-N).A, (S-N).B and (A-N).B angles,18 just the same way that an increased inclination of S-N19 in relation to the Frankfurt plane decreases the angular measurements (S-N).A, (S-N).B and (A-N).B, being able to bring about mistaken interpretations for the individual.

When comparing the image digitized by the HP™ Scan Jet 4C scanner in 75 dpi to the direct digital X-ray (Table 1), it was observed that all angular and the linear measurement N-Me did not suffer statistically significant alterations. Significant amplification was verified in the linear measurements S-N (0.82%, 0.55 mm), Go-Me (0.89%, 0.60 mm) and S-Go (0.9%, 0.72 mm), however, those are clinically acceptable because the differences for the measurements in the direct digital X-ray were lower than 1 mm and 1 degree-which are below the norm of the most used cephalometric analyses. Therefore, as other scanner types studied,6,8 we can also consider the flatbed scanner with transparency reader reliable for digitization of cephalometric radiographs. The differences between the scanner and the direct digital X-ray can be explained because of the fact that the scanner is sensitive to the "scanning" arrangements, including the location and orientation of the image and power state.6 However, we believe that there may be a minimum difference in the distance between the two landmarks set in the digital image and the real distance between them, informed to the "software" in the moment of the calibration of the images, being this hypothesis corroborated by authors who have claimed that the calibration of the digital image produces few but significant errors.2

The Sony™ W5 camera did not show statistically or clinically significant alterations in the angular measurements at the distances of 25 cm and 60 cm (Table 1), because the differences of the means for the direct digital X-ray were all lower than or equal to 1 degree. All the linear measurements, in both distances, suffered statistically significant amplifications, except for Go-Me at 60 cm. In the cephalometric analyses, we can say that at 60 cm there was no clinical significance, because the largest alteration in comparison with the direct digital radiograph was lower than 2 mm (N-Me = 1.84 mm). At 25 cm three linear measurements showed differences a little higher than 2 mm (S-Go = 2.43 mm; N-Me = 2.31 mm; SN = 2.28 mm), alterations that are very close to the norm of most of the analyses and therefore without clinical importance.

The Canon™ camera showed statistically significant alterations in the angular measurements (S-N).Ar and (Ar-Go).Me, at distances of 25 cm and 60 cm (Table 1), however, all lower than 2 degrees. The largest difference was for (S-N).Ar (1.57 degrees at 25 cm and 0.86 degrees at 60 cm), which also meant no clinical significance. All the linear measurements of Canon™, at 25 and 60 cm, underwent statistically significant amplifications, except N-Me and Go-Me at 60 cm. At 60 cm, S-N and S-Go presented with statistically significant alterations, although considered clinically insignificant in agreement with the cephalometric analyses, because the highest difference was of 2 mm for S-Go. However, when digitized at 25 cm all linear measurements presented statistically significant alterations, being S-Go the most enlarged one (5 mm), followed by S-N (2.97 mm). S-Go is used to evaluate the subsequent vertical proportion of the face in comparison with the previous facial height (N-Me),20 while S-N evaluates the length of the cranial base in relation to the mandibular plane (true horizontal:Go-Me), in the description of facial patterns (long face versus short face), and in the description of the vertical growth (low and high angles). Therefore the alterations caused by Canon™ at 25 cm could cause mistaken interpretations and might consequently interfere in the diagnosis as well as in the individualized orthodontic treatment plan.

When photographing, the cameras were with the upper part turned to the lower board of the radiographs (Fig 1), pointing to the less amplified horizontal measurement (Go-Me), in both cameras at 25 and 60 cm. The most amplified in the Canon™ at 25 and 60 cm, and in Sony™ at 25 cm was the vertical S-Go, located in the center of the lens, while in Sony™ at 60 cm was the vertical N-Me, in the left periphery of the lens. This fact confirms that the form and the convexity of the lens cause different distortions in different parts of the images10 and that there are construction differences between the lenses of both cameras, because they belong to different manufacturers.

Cameras can be useful to digitize cephalometric radiographs, but, before its clinical use, it is recommended to compare them to other well-known and reliable methods, observing the choice of the lens, lens-object distance, use of a copy stand, correct positioning and image calibration.

CONCLUSION

The Hewlett Packard™ Scan Jet 4C scanner with transparency reader was the best method and the Sony™ DSC-W5 as well as the Canon™ Rebel XT/EOS 350D, fixed in copy standy, operating at 60 cm were considered adequate for radiographic digitization. The cameras positioned at 25 cm caused distortions in the images, altering the linear measurements, and the Canon™ Rebel XT/EOS 350D may jeopardize the orthodontic diagnosis.

REFERENCES

  • 1. Wenzel A. Influence of computerized information technologies on image quality in dental radiographs. Tandlaegebladet. 1991 Sep;95(12):527-9.
  • 2. Forsyth DB, Shaw WC, Richmond S, Roberts CT. Digital imaging of cephalometric radiographs. Part 2: image quality. Angle Orthod. 1996;66(1):43-50.
  • 3. Faber RD, Burstone CJ, Solonche DJ. Computerized interactive orthodontic treatment planning. Am J Orthod. 1978 Jan;73(1):36-46.
  • 4. Lowey MN. The development of a new method of cephalometric and study cast mensuration with a computer controlled, video image capture system; part I: video image capture system. Br J Orthod. 1993 Aug;20(3):203-14.
  • 5. Brooks SL, Miles DA. Advances in diagnostic imaging in dentistry. Dent Clin North Am. 1993 Jan;37(1):91-111.
  • 6. Chen SK, Chiang TC. Digitizing of radiographs with a roller-type CCD scanner. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1997 Jun;83(6):719-24.
  • 7. Chen YJ, Chen SK, Yao JC, Chang HF. The effects of differences in landmark identification on the cephalometric measurements in the traditional versus digitized cephalometry. Angle Orthod. 2004 Apr;74(2):155-61.
  • 8. Sayinsu K, Isik F, Trakyali G, Arun T. An evaluation of the errors in cephalometric measurements on scanned cephalometric images and conventional tracings. Eur J Orthod. 2007 Feb;29(1):105-8.
  • 9. Nilce K, Gurevich GJ. How digital cameras work. HowStuffWorks. [cited 2004 Oct 6]. Available from: http://eletronics.howstuffworks.com/digital-camera4.htm
  • 10. Bockaert V. The 123 of digital Imaging. USA: Asimex; 2003.
  • 11. Whitehouse R, Moulding F. Latitude and noise comparisons between digital cameras and radiographic film scanner. J Telemed Telecare. 2000;6 Suppl 1:S41-2.
  • 12. Krupinski E, Gonzales M, Gonzales C, Weinstein RS. Evaluation of a digital camera for acquiring radiographic images for telemedicine applications. Telemed J E Health. 2000 Fall;6(3):297-302.
  • 13. Geelen W, Wenzel A, Gotfredsen E, Kruger M, Hansson LG. Reproducibility of cephalometric landmarks on conventional film, hardcopy, and monitor-displayed images obtained by the storage phosphor technique. Eur J Orthod. 1998 Jun;20(3):331-40.
  • 14. Chen YJ, Chen SK, Huang HW, Yao CC, Chang HF. Reliability of landmark identification in cephalometric radiography acquired by a storage phosphor imaging system. Dentomaxillofac Radiol. 2004 Sep;33(5):301-6.
  • 15. Krogman W, Sassouni V. A syllabus in roentgenographic cephalometry. Philadelphia: Center for Research in Child Growth;1957.
  • 16. Houston WJ, Maher RE, McElroy D, Sherriff M. Sources of error in measurements from cephalometric radiographs. Eur J Orthod. 1986 Aug;8(3):149-51.
  • 17. Downs WB. Variations in facial relationships: their significance in treatment and prognosis. Am J Orthod. 1948 Oct;34(10):812-40.
  • 18. Steiner C. The use of cephalometrics as an aid to planning and assessing orthodontic treatment. Am J Orthod. 1960;46:721-35.
  • 19. Moorrees C. Natural head position: the key to cephalometry. In: Jacobsen A. Radiographic cephalometry. Chicago: Quintessence; 1995. p. 175-84.
  • 20. Horn A. Facial height index. Am J Orthod Dentofacial Orthop. 1992 Aug;102(2):180-6.
  • Evaluation of indirect methods of digitization of cephalometric radiographs in comparison with the direct digital method

    Cleomar Donizeth RodriguesI; Márcia Maria Fonseca da SilveiraII; Orivaldo TavanoIII; Ronaldo Henrique ShibuyaIV; Giovanni ModestoV; Carlos EstrelaVI
  • Publication Dates

    • Publication in this collection
      18 Aug 2010
    • Date of issue
      Aug 2010

    History

    • Accepted
      May 2010
    • Received
      Feb 2010
    Dental Press International Av. Luís Teixeira Mendes, 2712 , 87015-001 - Maringá - PR, Tel: (55 44) 3033-9818 - Maringá - PR - Brazil
    E-mail: artigos@dentalpress.com.br