Acessibilidade / Reportar erro

Phone positioning influence in high-frequency audiometry

Research considers high frequency tonal audiometry as a tool for the early diagnosis of auditory alterations derived from etiological agents. AIM: to investigate possible differences in high frequency audiometry of individuals with normal hearing, based on the person who places the earphone. PATIENTS AND METHOD: clinical and experimental study with 55 undergraduate students from a country side branch of the São Paulo State University, with normal hearing, underwent two tests each; for the first, the evaluator positioned the earphone on the participant; for the second one, the participant did it by him/herself. An AC40 audiometer calibrated to emit pure tone was used in the frequencies of 10, 12.5, and 16 kHz. RESULTS: The kappa(k) coefficient statistical analysis was used to verify the agreement between the two ways of earphone positioning of earphone, bearing a ≥0.70 kappa value as a criterion. Results attained for both ears were below this criterion, with k average of 0.50. DISCUSSION: results indicate a risk of compromising the exam reliability when the patient him/herself adjusts phone to his/her own ear. CONCLUSION: when performing audiometric assessment, this variable must be considered in order to attain reliable results.

pure-tone; audiometry; audiology; hearing


ABORL-CCF Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial Av. Indianápolis, 740, 04062-001 São Paulo SP - Brazil, Tel./Fax: (55 11) 5052-9515 - São Paulo - SP - Brazil
E-mail: revista@aborlccf.org.br