SciELO - Scientific Electronic Library Online

 
vol.88 issue3Ventricular electrical activation in cardiac resynchronization as characterized by body surface potential mappingRadiofrequency catheter ablation of premature ventricular contractions originating in the right ventricular outflow tract author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Arquivos Brasileiros de Cardiologia

Print version ISSN 0066-782X

Abstract

COSTA, Roberto et al. Incidence of shock and quality of life in young patients with implantable cardioverter-defibrillator. Arq. Bras. Cardiol. [online]. 2007, vol.88, n.3, pp. 258-264. ISSN 0066-782X.  http://dx.doi.org/10.1590/S0066-782X2007000300002.

OBJECTIVES: To analyze the incidence and causes of ICD therapies in children and young adults and verify their impact on the quality of life (QoL). METHODS: From March/1977 to February/2006, 29 patients (15.7±5.4 years old) were submitted to ICD implants. Aborted cardiac arrest (41.5%), sustained ventricular tachycardia (27.6%) and primary prophylaxis of sudden cardiac death (30.9%) indicated device therapy. The number of therapies was evaluated by interviewing patients and by ICD diagnostic data. The SF-36 questionnaire was used to measure the QoL and the results were compared to healthy population. The expectative of freedom from ICD therapies were estimated by the Kaplan-Meier method. RESULTS: After 2.6±1.8 years follow-up, 8 (27.6%) patients received 141 appropriate ICD shocks due to ventricular tachycardia (6) or ventricular fibrillation (2), and 11 (37.9%) patients received 152 inappropriate ICD shocks due to supraventricular tachyarrhythmias (8) or oversensing (3). Expectative of freedom from appropriate shocks was 74.2±9.0% and 66.7±10.7% after one and three years, respectively. Compared to healthy population, QoL decreased in physical function (61.7±28.7), vitality (64.7±19.1), mental health (65.9±22.7) and role-emotional domains (66.7±38.5). All patients referred fear and concern related to ICD use. CONCLUSION: Despite the efficacy of ICD therapies, the high incidence of appropriate and inappropriate shocks interfered in patients’ QoL and adaptation to the device.

Keywords : Defibrillators; implantable; quality of life; adolescent.

        · abstract in Portuguese     · text in English | Portuguese     · pdf in English | Portuguese