Quality control and dosimetry in computed tomography units

OBJECTIVE: Evaluation of equipment conditions and dosimetry in computed tomography services utilizing protocols for head, abdomen, and lumbar spine in adult patients (in three different units) and pediatric patients up to 18 months of age (in one of the units evaluated). MATERIALS AND METHODS: Computed tomography dose index and multiple-scan average dose were estimated in studies of adult patients with three different units. Additionally, entrance surface doses as well as absorbed dose were estimated in head studies for both adult and pediatric patients in a single computed tomography unit. RESULTS: Mechanical quality control tests were performed, demonstrating that computed tomography units comply with the equipment-use specifications established by the current standards. Dosimetry results have demonstrated that the multiple-scan average dose values were in excess of up to 109.0% the reference levels, presenting considerable variation amongst the computed tomography units evaluated in the present study. Absorbed doses obtained with pediatric protocols are lower than those with adult patients, presenting a reduction of up to 51.0% in the thyroid gland. CONCLUSION: The present study has analyzed the operational conditions of three computed tomography units, establishing which parameters should be set for the deployment of a quality control program in the institutions where this study was developed.

Computed tomography equipment; Quality control; Dosimetry; Dose reduction


Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem Av. Paulista, 37 - 7º andar - conjunto 71, 01311-902 - São Paulo - SP, Tel.: +55 11 3372-4541, Fax: 3285-1690, Fax: +55 11 3285-1690 - São Paulo - SP - Brazil
E-mail: radiologiabrasileira@cbr.org.br