vol.44 issue4The Brazilian profile of intellectual property in radiology and imaging diagnosis in an international context in the years 2000 to 2009Radiological propedeutics of femoroacetabular impingement in times of computed tomography and magnetic resonance imaging: what a radiologist needs to know author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Radiologia Brasileira

Print version ISSN 0100-3984

Abstract

MENDES, Adriana Carla Rodrigues; RAMOS, Christiane Lucena  and  ABREU, Danilo Wanderley Matos de. Evaluation of medical X-ray machines in Paraíba state radiology centers between 2008 and 2009. Radiol Bras [online]. 2011, vol.44, n.4, pp. 244-248. ISSN 0100-3984.  http://dx.doi.org/10.1590/S0100-39842011000400010.

OBJECTIVE: To evaluate the light and radiation fields congruence and radiation beam alignment in medical X-ray equipment in Paraíba state radiology centers by means of two quality control tests. MATERIALS AND METHODS: A loaded cassette, a measuring tape and a bubble level were utilized in the field size and alignment testing. The evaluation of collimation systems accuracy and X-ray beam alignment was undertaken during health inspections performed in radiology centers between 2008 and 2009. RESULTS: In 2008, 121 X-ray machines were evaluated in the Paraíba state. In 2009, 117 machines were tested. From this universe, 86 machines were selected for comparison, since they were evaluated both in 2008 and 2009, with 18.60% (n = 16) showing test results improvement from one year to another. CONCLUSION: The percentage of problematic X-ray machines decreased between 2008 and 2009, notwithstanding no quality assurance program has been observed in Paraíba state radiology centers.

Keywords : Equipment; X-rays; Quality control.

        · abstract in Portuguese     · text in English | Portuguese     · pdf in English | Portuguese