Objective:
To assess the accuracy in the prediction of both early and late preeclampsia by flow-mediated dilation of the brachial artery (FMD), a biophysical marker for endothelial dysfunction.
Materials and Methods:
A total of 91 patients, considered at high risk for development of preeclampsia were submitted to brachial artery FMD between 24 and 28 weeks of gestation.
Results:
Nineteen out of the selected patients developed preeclampsia, 8 in its early form and 11 in the late form. With a cut-off value of 6.5%, the FMD sensitivity for early preeclampsia prediction was 75.0%, with specificity of 73.3%, positive predictive value (PPV) of 32.4% and negative predictive value (NPV) of 91.9%. For the prediction of late preeclampsia, sensitivity = 83.3%, specificity = 73.2%, PPV = 34.4% and NPV = 96.2% were observed. And for the prediction of all associated forms of preeclampsia, sensitivity = 84.2%, specificity = 73.6%, PPV = 45.7% and NPV = 94.6% were observed.
Conclusion:
FMD of the brachial artery is a test with good accuracy in the prediction of both early and late preeclampsia, which may represent a positive impact on the follow-up of pregnant women at high risk for developing this syndrome.
Preeclampsia; Ultrasonography; Vascular endothelium; Prediction