ANDRADES, A.O. et al. Heating produced by therapeutic ultrasound in the presence of a metal plate in the femur of canine cadavers. Arq. Bras. Med. Vet. Zootec. [online]. 2014, vol.66, n.5, pp.1343-1350. ISSN 0102-0935.  http://dx.doi.org/10.1590/1678-6676.

The present study aimed to assess the heat generated by a therapeutic ultrasound (TUS) in a metal bone plate and adjacent structures after fixation to the femur of canine cadavers. Ten pairs of hind limbs were used, and they were equally distributed between groups that were subjected to 1- and 3-MHz frequencies, with each frequency testing 1- and 2-W/cm² intensities. The right hind limb was defined as the control group (absence of the metal plate), and the left hind limb was the test group (presence of the metal plate). Therefore, the control groups (CG) were denominated CGI, using TUS with 1-MHz frequency and 1-W/cm² intensity; CGII, using 1-MHz frequency and 2-W/cm² intensity; CGIII, using 3-MHz frequency and 1-W/cm² intensity; and CGIV, using 3-MHz frequency and 2-W/cm² intensity. For each control group, its respective test group (TG) was denominated TGI, TGII, TGIII and TGIV. The TUS was applied to the lateral aspect of the thigh using the continuous mode and a 3.5-cm² transducer in a 6.25-cm² area for 2 minutes. Sensors were coupled to digital thermometers that measured the temperature in different sites before (t0) and after (t1) of the TUS application. The temperatures in t1were higher in all tested groups. The intramuscular temperature was significantly higher (P<0.05) in the groups used to test the 3-MHz frequency in the presence of the metal plate. The therapeutic ultrasound in the continuous mode using frequencies of 1 and 3 MHz and intensities of 1 and 2 W/cm2 for 2 minutes caused heating of the metal plate and adjacent structures after fixation to the femur of canine cadavers.

Keywords : dog; physical therapy; deep heat therapy; heating; contracture.