SciELO - Scientific Electronic Library Online

 
vol.19 issue5Hydroxylation of a hederagenin derived saponin by a Xylareaceous fungus found in fruits of Sapindus saponariaInfraspecific variability in the essential oil composition of Lychnophora ericoides author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Journal of the Brazilian Chemical Society

Print version ISSN 0103-5053

Abstract

CERVINI, Priscila  and  CAVALHEIRO, Éder Tadeu Gomes. Determination of paracetamol at a graphite-polyurethane composite electrode as an amperometric flow detector. J. Braz. Chem. Soc. [online]. 2008, vol.19, n.5, pp. 836-841. ISSN 0103-5053.  http://dx.doi.org/10.1590/S0103-50532008000500005.

A bare graphite-polyurethane composite was evaluated as an amperometric flow injection detector in the determination of paracetamol (APAP) in pharmaceutical formulations. A linear analytical curve was observed in the 5.00 x 10-5 to 5.00 x 10-3 mol L-1 range with a minimum detectable net concentration of 18.9 µmol L-1 and 180 determinations h-1, after optimization of parameters such as the detection potential, sample loop volume, and carrier solution flow rate. Interference of ascorbic acid was observed, however, it was possible overcome the interference, reaching results that agreed with HPLC within 95% confidence level. These results showed that the graphite-polyurethane composite can be used as an amperometric detector for flow analysis in the determination of APAP.

Keywords : paracetamol; graphite-PU composite electrode; FIA.

        · abstract in Portuguese     · text in English     · pdf in English