OBJECTIVE: To introduce a new coupling system between screw driver and interference screw, and biomechanical tests that validate the safety of its application. METHODS: The new system was submitted to biomechanical torsion assays. Two types of analysis were performed: maximum torque of manual insertion of the screws into bovine bone; destructive assays of torsion of the system using an INSTRON 55MT machine. The same tests were also performed on a control group, using a commercially available interference screw coupling system (Acufex®). RESULTS: In the tests on manual insertion of screws in bovine femurs, the average values found with a digital torque meter were 1.958 N/m for Acufex® and 2.563 N/m for FMRP. Considering p>0.05, there were no statistical differences between the two groups (p=0.02) in the values for maximum torque of insertion, in the two systems studied. The average values for maximum torque of torsion resisted by the screw were 15N/m for the Acufex® screw and 13N/m for the FMRP screw, again with no statistical differences between the two groups (p>0.05). In the evaluation of angular deformation, there was also no significant difference between the two screw types (p=0.15). CONCLUSION: The new coupling system for interference screws developed at FMRP-USP revealed a torsion resistance that is comparable with the system already available on the market and regulated for international use.
Knee; Bone screw; Biomechanics