Abstract
(1) Background:
The Commercial Kit SIRE Nitratase® PlastLabor, is a drug susceptibility test kit used to detect Mycobacterium tuberculosis resistance to first-line TB treatment drugs. The present study aimed at evaluating its performance in a multicenter study.
(2) Methods:
To determine its accuracy, the proportion methods in Lowenstein Jensen medium or the BACTECTMMGITTM960 system was used as a gold standard.
(3) Results:
The study revealed that the respective accuracies of the kit with 190 M. tuberculosis clinical isolates, using the proportion methods in Lowenstein Jensen medium or BACTECTMMGITTM960 system as a gold standard, were 93.9% and 94.6%, 96.9% and 94.6%, 98.0% and 97.8%, and 98.0% and 98.9%, for streptomycin, isoniazid, rifampicin, and ethambutol, respectively.
(4) Conclusion:
Thus, the kit can rapidly screen resistance to streptomycin, isoniazid, rifampicin, and ethambutol. Additionally, it does not require sophisticated equipment; hence, it can be easily used in the laboratories of low and middle income countries.
Keywords:
tuberculosis; MDR; nitrate reductase; technology