vol.29 número3Effect of laser vs LED in the near infrared region on the skeletal muscle activity: clinical studyCurrent perception threshold and reaction time in the assessment of sensory peripheral nerve fibers through sinusoidal electrical stimulation at different frequencies índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

Compartir


Revista Brasileira de Engenharia Biomédica

versión ISSN 1517-3151

Resumen

NEGRAO, João Francisco Ribeiro; ARAUJO, Guilherme Augusto Limeira; COSTA JUNIOR, Carlos Tavares da  y  SOUZA, Daniel Cardoso de. Electromagnetic interference reduction by dynamic impedance balancing applied to biosensors. Rev. Bras. Eng. Bioméd. [online]. 2013, vol.29, n.3, pp.269-277. ISSN 1517-3151.  http://dx.doi.org/10.4322/rbeb.2013.030.

INTRODUCTION: Electromagnetic interference caused by electric power lines adversely affects the signals of electronic instruments, especially those with low amplitude levels. This type of interference is known as common-mode interference. There are many methods and architectures used to minimize the influence of this kind of interference on electronic instruments, the most common of which is the use of band-reject filters. This paper presents the analysis, development, prototype and test of a new reconfigurable filter architecture for biomedical instruments, aiming to reduce the common-mode interference and preserve the useful signal components in the same frequency range as that of the noise, using the technique of dynamic impedance balancing. METHODS: The circuit blocks were mathematically modeled and the overall closed-loop transfer function was derived. Then the project was described and simulated in the VHDL_AMS language and also in an electronics simulation software, using discrete component blocks, with and without feedback. After theoretical analysis and simulation results, a prototype circuit was built and tested using as input a signal obtained from ECG electrodes. RESULTS: The results from the experimental circuit matched those from simulation: a 97.6% noise reduction was obtained in simulations using a sinusoidal signal, and an 86.66% reduction was achieved using ECG electrodes in experimental tests. In both cases, the useful signal was preserved. CONCLUSION: The method and its architecture can be applied to attenuate interferences which occur in the same frequency band as that of the useful signal components, while preserving these signals.

Palabras clave : Biomedical instruments; Electromagnetic interference; Common-Mode Rejection; Instrumentation amplifier; Dynamic impedance balancing.