Acessibilidade / Reportar erro

Comparative evaluation of pyrogens tests in pharmaceutical products

Avaliação comparativa de testes de pirogênios em produtos farmacêuticos

A comparison of methodologies for detection of pyrogens in pharmaceutical products was performed. The rabbit pyrogen test was optimized and the dose-response curve was obtained for the 2nd International Standard for bacterial endotoxins, establishing 13.81 EU/mL/kg as the concentration of endotoxin necessary to induce a temperature rise of 0.5ºC. The 0.5ºC cut-off was shown to give results that were more compatible with the pyrogenic doses for humans. The Limulus amoebocyte lysate test (LAL) was standardized with gel-clot and chromogenic endpoints, and used for the comparative evaluation of pharmaceutical products showing good agreement. The use of beta-glucan-reactive and non-reactive LAL reagents identified some products with false-positive results. The interference test was carried out and the specifications validated for some new products as the maximum valid dilution. The results emphasized the importance and limitations of the assays recommended for the evaluation of purity and quality control of parenteral medicinal products, improving the existing methodologies in the context of reduction and replacement in the use of animal models.

bacterial endotoxin; Limulus amoebocyte lysate; pyrogens; rabbit pyrogen test


Sociedade Brasileira de Microbiologia USP - ICB III - Dep. de Microbiologia, Sociedade Brasileira de Microbiologia, Av. Prof. Lineu Prestes, 2415, Cidade Universitária, 05508-900 São Paulo, SP - Brasil, Ramal USP 7979, Tel. / Fax: (55 11) 3813-9647 ou 3037-7095 - São Paulo - SP - Brazil