Influence of the instrument used for cervical preflaring on the precision of 2 Eletronic Apex Locators

Influência do instrumento utilizado no preparo cervical na precisão de 2 localizadores apicais eletrônicos

ABSTRACT

Objective:

This study was aimed at evaluating the influence of cervical preflaring using LA-Axxes No. 1 bur (SybronEndo, Glendora, USA) or S1 and SX ProTaper files (Dentsply Maillefer, Ballaigues, Switzerland) on the accuracy of the Electronic Apex Locators (EALs): Root ZX mini (J. Morita Corporation, Tokyo, Japan), and Joypex 5 (Denjoy, Changsha, China).

Methods:

Thirty mandibular incisors were accessed, and the root canal length (RCL) was determined with a K-file #15, with the aid of a stereo microscope. Afterwards, the specimens were divided into 2 groups (n = 15) referring to the cervical preflaring with LA-Axxess or ProTaper files. The teeth were embedded in alginate and the RCL was determined by the EALs before and after preflaring. Data were classified into: accurate, if the difference in RCL measurement were ≤0.05 mm; and inaccurate, if the difference were >0.5 mm or beyond the RCL.

Results:

McNemar's test (α<0.05) was used to detect differences in the accuracy of the EALs before and after each preflaring with different instruments, and to detect difference in accuracy among devices. No differences were found concerning the accuracy of the EALs (P > 0.05) after the cervical preflaring, regardless of the used instrument.

Conclusion:

The preflaring procedure increased the number of accurate measurements for both EALs, with statistical difference for Joypex 5 when the preflaring was performed with LA-Axxess. However, after the cervical preflaring, the EALs showed similar accuracy, regardless of the used instrument.

Indexing terms:
Dental instruments; Endodontics; Odontometry