Acessibilidade / Reportar erro

Residual solvent determination by head space gas chromatography with flame ionization detector in omeprazole API

Residual solvents in pharmaceutical samples are monitored using gas chromatography with head space. Based on good manufacturing practices, measuring residual solvents is mandatory for the release testing of all active pharmaceutical ingredients (API). The analysis of residual organic solvents (methanol, acetone, cyclohexane, dichloromethane, toluene) in Omeprazole, an active pharmaceutical ingredient was investigated. Omeprazole is a potent reversible inhibitor of the gastric proton pump H+/K+-ATPase. The Head space gas chromatography (HSGC) method described in this investigation utilized a SPB TM-624, Supelco, 30 m long x 0.25 mm internal diameter, 1.4µm-thick column. Since Omeprazole is a thermally labile compound, the selection of the proper injector temperature is critical to the success of the analysis. The injector temperature was set at 170ºC to prevent degradation. The initial oven temperature was set at 40ºC for 12 min and programmed at a rate of 10ºC min-1 to a final temperature of 220ºC for 5 min. Nitrogen was used as a carrier gas. The sample solvent selected was N,N-dimethylacetamide. The method was validated to be specific, linear, precise, sensitive, rugged and showed excellent recovery.

Headspace-gas chromatography; Omeprazole; Method validation; Residual solvents release testing


Universidade de São Paulo, Faculdade de Ciências Farmacêuticas Av. Prof. Lineu Prestes, n. 580, 05508-000 S. Paulo/SP Brasil, Tel.: (55 11) 3091-3824 - São Paulo - SP - Brazil
E-mail: bjps@usp.br